RT Journal Article A1 Fogagnolo, Paolo A1 Frezzotti, Paolo A1 Mittica, Pietro A1 Digiuni, Maurizio A1 Paderni, Ruggiero A1 Vallenzasca, Elena A1 Rossetti, Luca T1 Repeatibility And Reproducibility Of Applanation Resonance Tonometry JF Investigative Ophthalmology & Visual Science JO Invest. Ophthalmol. Vis. Sci. YR 2012 VO 53 IS 14 SP 5042 OP 5042 SN 1552-5783 AB To assess repeatability (intra-observer variability) and reproducibility (inter-observer variability) of intraocular pressure (IOP) measurements with Applanation Resonance Tonometry (ART) and to evaluate possible influential factors. The study included 178 patients (101 glaucoma, 63 healthy and 14 with OHT; one eye per subject) from two Italian centers (Eye Clinic of San Paolo Hospital of Milan and University of Siena). IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by ART (ART1, ART2) in a randomized sequence, by a single operator to assess intra-operator variability. To assess inter-operator variability 2 operators, chosen randomly (A and B, B and C, A and C), performed 2 measurements each (in random order; i.e. A B A B; B A B A) on the same patient. Repeatability and reproducibility were assessed by the coefficient of variation (CoV) and intraclass correlation coefficient (ICC). ART1 was 0.4±2.8 mmHg higher than ART2 (p<0.001). Results were not influenced by randomization test order. Intra-operator CoV was 7.6%±7.6%, and ICC was 0.88 (95%CI: 0.85-0.91). Inter-operator CoV ranged between 6.0%±10.8% and 8.0%±6.8%, and ICC between 0.906 and 0.917. ART1 and 2 were respectively 1.8±3.8 and 1.4±4.3 mmHg higher than GAT (p<0.001). Results were unaffected by diagnosis, age, central corneal thickness, keratometry, operator, randomization sequence. ART repeatability and reproducibility were almost perfect. Tonometry measurements with ART tended to be overestimated compared with GAT. RD 3/4/2021