Abstract
purpose. To develop and test a short and reliable visual field threshold program for the early detection of glaucomatous visual field loss, by adapting the Swedish interactive test algorithm (SITA) to short-wavelength-automated-perimetry (SWAP).
methods. Computer simulations were performed to test the accuracy of several versions of SITA SWAP, and to optimize speed versus reliability. The selected SITA SWAP version was evaluated and compared with the older Full Threshold SWAP and Fastpac SWAP programs in 41 patients with glaucoma and normal subjects.
results. Average test time was 3.6 minutes for SITA SWAP, 11.8 minutes for Full Threshold SWAP, and 7.7 minutes for Fastpac SWAP, differences were significant at P < 0.0001. Mean threshold reproducibility, calculated as absolute difference between two tests, did not differ significantly between programs and was 2.4 dB for SITA, 2.3 dB for Full Threshold, and 2.4 dB for Fastpac SWAP. Simultaneous comparison showed significant differences in threshold sensitivity, P = 0.023: SITA SWAP showed highest sensitivity, 21.6 dB on average, compared with both Full Threshold SWAP and Fastpac SWAP with a mean sensitivity of 17.3 and 17.8 dB, respectively.
conclusions. SITA SWAP was much faster than the older SWAP strategies, and reproducibility did not differ. This implies that SITA SWAP could become a clinically useful method for the detection of early glaucoma. SWAP tests may also be applicable in larger groups of patients because of the increased dynamic range.
Short-wavelength perimetry (SWAP) applies a blue stimulus, Goldmann size V, on an intense yellow background. The currently accepted clinical method, as incorporated in the Humphrey Field Analyzer (HFA; Zeiss-Humphrey Systems, Dublin, CA), utilizes a Goldmann size V narrow-band blue stimulus with a peak transmission of 440 nm presented on a 100-cd/m
2 yellow background.
1
SWAP can detect glaucomatous visual field loss before conventional white-on-white (WW) perimetry.
2 3 4 5 6 Whether this advantage is based on selective losses of a subpopulation of ganglion cells in glaucoma or on reduced redundancy is debatable—that is, whether SWAP responses are simply derived from a subset of ganglion cells that are selectively damaged by glaucoma, or, alternatively, whether because SWAP tests only one type of ganglion cells, other cell types that may still be functional in a given test location therefore cannot contribute to detection of the blue-on-yellow target.
7 There are limitations associated with the SWAP technique, however.
8 SWAP results are affected by age-related yellowing of the lens
9 10 and by cataract.
11 12
SWAP has been reported to require 15% to 17% more time than WW perimetry using the same threshold strategies.
13 The longer test time with SWAP compared with conventional WW perimetry is a practical disadvantage.
Variability in threshold estimates is greater with SWAP than with WW.
13 14 15 16 Intersubject variability in normal subjects, which forms the base for normal limits of threshold deviations, is also larger with SWAP
13 17 than with conventional WW perimetry. Hence, the normal limits for SWAP thresholds are wider than those for WW thresholds, resulting in less significance for the same decibel depression with SWAP perimetry and hence less sensitive probability maps
18 (Fig. 1) .
The SITA Standard and Fast programs were originally developed for WW perimetry with the goal of considerably shortening test times without losing accuracy, compared with the older test programs they were designed to replace: the Full Threshold and Fastpac strategies, respectively (Zeiss-Humphrey Systems).
19 20 SITA Standard tests are approximately 50% shorter than the older Full Threshold test, and SITA Fast is approximately 50% shorter than Fastpac tests, while showing similar reproducibility.
21 22 23 24 25 Threshold sensitivity is higher with the SITA programs (i.e., the dynamic range of stimulus intensity is increased). Intersubject variability is smaller and threshold normal limits are narrower than with the older threshold programs.
26 27 In glaucomatous eyes, slightly more test points show significantly depressed threshold deviations with SITA, and sensitivity to glaucomatous visual field damage is at least as good as with the older standard programs.
24 28
The first purpose of the present study was to modify the SITA algorithm to SWAP, to obtain a fast and reliable test for early detection of glaucomatous field defects. A second purpose was to conduct a pilot clinical test with the new algorithm.
The SITA algorithm applies Bayesian statistics for maximum a priori (MAP) likelihood calculation of thresholds. MAP calculation is an established mathematical statistical method, and was described by Watson and Pelli
35 for threshold estimation in experimental psychophysical tests. King-Smith et al.
36 later developed their method for perimetry, which was adapted later to SWAP by Turpin et al.
37 The SITA algorithm, however, is based on the work in studies by Olsson et al.,
38 Olsson,
39 and Olsson and Rootzén
40 and was designed not only to estimate thresholds but also to save test time by stopping testing at locations when the variability in the threshold estimate is small enough according to predetermined rules.
19 SITA also includes two other time-saving features: a new way to estimate false-positive and false-negative answers
41 42 and improved pacing of the test.
19
The selected version of SITA SWAP showed much shorter test times than both older SWAP programs. This was one of the main goals when modifying the SITA algorithm for SWAP testing. Recent development of threshold strategies has focused on shorter test times.
19 20 43 44 The long test duration using older SWAP methods has been one of the major obstacles for using SWAP in clinical settings. In clinical work, SWAP has therefore been used only in a minority of patients with glaucoma or suspected glaucoma.
Reproducibility of threshold estimates did not differ between the three strategies. In real-time testing, the clinician has to rely on reproducibility instead of accuracy but reliable threshold algorithms are expected to produce acceptable reproducibility. It should be remembered, however, that crude tests, such as screening programs with only two outcomes, seen or not seen, show higher reproducibility than threshold tests with a wide range of outcomes. Our comparison, however, included different, but similar threshold programs.
Accuracy can only be calculated when a comparison with true thresholds is possible, as in simulations. This is one important reason to use simulations when developing new threshold programs. Another advantage is that single test parameters, such as step size or ERF, can be varied and evaluated one at the time in a large number of fields much more rapidly than with real-time measurements.
With SITA SWAP average threshold sensitivity increased by approximately 4 dB compared with the older SWAP programs. This is in agreement with our earlier results comparing WW SITA with WW Full Threshold,
20 21 22 and also with the results of the Turpin et al.,
37 who found approximately a 3-dB difference between the SWAP Full Threshold and the method of King-Smith et al.
36 adapted for SWAP. The higher sensitivity is positive, thus increasing dynamic range. To some extent, the higher sensitivity is due to differences in the definition of thresholds. Bayesian MAP algorithms define the threshold as the stimulus intensity that has a 50% probability to be seen, whereas the older strategies define the threshold as the stimulus intensity last seen at the end of the staircases. The absence of visual fatigue
45 46 also contributes to the higher sensitivity of the shorter SITA tests compared with more time-consuming older tests. When developing the SITA programs for WW perimetry, we found a positive correlation between differences in mean sensitivity and differences in test length,
20 with shorter tests being associated with higher sensitivity.
The difference in threshold sensitivity between test programs has a large influence on the gray-scale representations of such values. This makes a comparison of sensitivity to glaucomatous field loss very difficult when only raw threshold data and resultant gray-scale representations are available. More accurate evaluations should take age-corrected normal thresholds and normal limits for such values into consideration. Such normal thresholds are not available at the present time. A subjective evaluation of the appearance of threshold gray scales, however, suggests that SITA SWAP would show glaucomatous field loss similar to that found with the older SWAP programs. One example is shown in
Figure 6 . It could be expected that narrower normal threshold limits and more sensitive probability maps as found with WW SITA compared with WW Full Threshold
26 also will be found with SITA SWAP, but this remains to be tested. However, an initial analysis was performed to calculate preliminary age-corrected thresholds and empirically derived normal limits at the
P < 0.10 and
P < 0.05 levels for comparison between strategies using the 20 tests obtained with each of the three SWAP strategies in the 10 normal subjects. These crudely calculated limits were considerably wider for Fastpac SWAP and Full Threshold SWAP than for SITA SWAP. It therefore seems likely that SITA SWAP probability maps may become more sensitive to early glaucoma loss than the older SWAP programs.
SITA SWAP was developed to become a fast and reliable clinical method for the detection of early glaucomatous visual field loss in patients without cataract. Considerably reduced test time, increased dynamic range and perhaps also smaller intersubject variability imply that SITA SWAP could become such a test, but further testing is necessary. Definite comparisons cannot be made until normal age-corrected thresholds and normal threshold limits have been determined.
Supported by Zeiss-Humphrey Systems, Dublin, California.
Submitted for publication February 19, 2002; revised June 20, 2002; accepted July 9, 2002.
Commercial relationships policy: F.
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked “
advertisement” in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Corresponding author: Boel Bengtsson, Department of Ophthalmology, Malmö University Hospital, SE-20502 Malmö, Sweden;
[email protected].
Table 1. Accuracy and Test Time of the Four SITA SWAP Versions
Table 1. Accuracy and Test Time of the Four SITA SWAP Versions
SITA Version | Stimulus Step (dB) | ERF | Threshold Reproducibility (|difference| in dB) | Average Test Time (min) | Efficiency (Reproducibility × Test Time) | Number of Test Point Locations with Staircases Interrupted by ERF (%) |
A | 4 | 1.49 | 3.24 | 4.06 | 13.54 | 28 |
B | 4 | 1.29 | 3.16 | 4.24 | 13.72 | 18 |
C | 4 | 0.89 | 2.55 | 5.94 | 15.11 | 0 |
D | 4 and 4-2 | 1.29 | 2.73 | 3.58 | 9.97 | 16 |
The author thanks Buck Cunningham, Software Engineer, Zeiss-Humphrey Systems, Dublin, CA, for programming the experimental SITA SWAP, and Jonny Olsson, PhD, for modifying the SITA calculation model and the computer simulator.
Sample, PA, Johnson, CA, Haegerstrom-Portnoy, G, Adams, AJ. (1996) Optimum parameters for short-wavelength automated perimetry J Glaucoma 5,375-393
[PubMed]Hart, WM, Silverman, SE, Trick, GL, Nesher, R, Gordon, ME. (1990) Glaucomatous visual field damage: luminance and color-contrast sensitivities Invest Ophthalmol Vis Sci 31,359-367
[PubMed]De Jong, LAMS, Snepvangers, CEJ, van den Berg, TJTP, Langerhorst, CT. (1990) Blue-yellow perimetry in the detection of early glaucomatous damage Doc Ophthalmol 75,303-314
[CrossRef] [PubMed]Sample, PA, Weinreb, RN. (1992) Progressive color visual field loss in glaucoma Invest Ophthalmol Vis Sci 33,2068-2071
[PubMed]Johnson, CA, Adams, AJ, Casson, EJ, Brandt, JD. (1993) Blue-on-yellow perimetry can predict the development of glaucomatous field loss Arch Ophthalmol 111,645-650
[CrossRef] [PubMed]Sample, PA, Taylor, JDN, Martinez, GA, Lusky, M, Weinreb, RN. (1993) Short-wavelength color visual fields in glaucoma suspects at risk Am J Ophthalmol 115,225-233
[CrossRef] [PubMed]Johnson, CA. (1994) Selective versus nonselective losses in glaucoma J Glaucoma 1(suppl. 1),S32-S44
Wild, JM. (2001) Short wavelength automated perimetry Acta Ophthalmol Scand 79,546-559
[CrossRef] [PubMed]Sample, PA, Esterson, FD, Weinreb, RN, Boynton, RM. (1988) The aging lens: in vivo assessment of light absorption in 84 human eye Invest Ophthalmol Vis Sci 29,1306-1311
[PubMed]Johnson, CA, Adamas, AJ, Twelker, JD, Quigg, JM. (1988) Age-related changes in the central visual field for short-wave-length-sensitive pathways J Opt Soc Am A 5,2131-2139
[CrossRef] [PubMed]Moss, ID, Wild, JM, Whitaker, DJ. (1995) The influence of age-related cataract on blue-on-yellow perimetry Invest Ophthalmol Vis Sci 36,764-773
[PubMed]Kim, YY, Kim, JS, Shin, DH, Kim, C, Jung, HR. (2001) Effect of cataract extraction on blue-on-yellow visual field Am J Ophthalmol 132,217-220
[CrossRef] [PubMed]Wild, JM, Cubbidge, RP, Pacey, IE, Robinson, R. (1998) Statistical aspects on the normal visual field in short-wavelength automated perimetry Invest Ophthalmol Vis Sci 39,54-63
[PubMed]Known, YH, Park, HJ, Jap, A, Ugurlu, S, Caprioli, J. (1998) Test-retest variability in blue-on-yellow perimetry is greater than white-on-white perimetry in normal subjects Am J Ophthalmol 126,29-36
[CrossRef] [PubMed]Olsson, J, Bengtsson, B, Heijl, A. (1998) Frequency-of-seeing curves in short-wavelength perimetry (SWAP) [ARVO abstract] Invest Ophthalmol Vis Sci 39(4),S494Abstract nr 2268
Briggs, JL, Hudson, C, Silvestri, G, Jackson, AJ. (2000) The impact of stimulus condition frequency-of-seeing in automated static perimetry [ARVO abstract] Invest Ophthalmol Vis Sci 41(4),S282Abstract nr 1496
Wild, JM, Moss, ID, Whitaker, D, O’Neill, EC. (1995) The statistical interpretation of blue-on-yellow visual field loss Invest Ophthalmol Vis Sci 36,1398-1410
[PubMed]Heijl, A, Lindgren, G, Olsson, J, Åsman, P. (1989) Visual field interpretation with empirical probability maps Arch Ophthalmol 107,204-208
[CrossRef] [PubMed]Bengtsson, B, Olsson, J, Heijl, A, Rootzén, H. (1997) A new generation of algorithms for computerized threshold perimetry, SITA Acta Ophthalmol Scand 75,368-375
[PubMed]Bengtsson, B, Heijl, A. (1998) SITA Fast, a new rapid perimetric threshold test: description of methods and evaluation in patients with manifest and suspect glaucoma Acta Ophthalmol Scand 786,431-437
Bengtsson, B, Heijl, A, Olsson, J. (1998) Evaluation of a new threshold visual field strategy, SITA, in normal subjects Acta Ophthalmol Scand 76,165-169
[CrossRef] [PubMed]Bengtsson, B, Heijl, A. (1998) Evaluation of a new perimetric threshold strategy in patients with manifest and suspect glaucoma Acta Ophthalmol Scand 76,268-272
[CrossRef] [PubMed]Shirato, S, Inoue, R, Fukushima, K, Suzuki, Y. (1999) Clinical evaluation of SITA: a new family of perimetric testing strategies Graefes Arch Clin Exp Ophthalmol 237,29-34
[CrossRef] [PubMed]Wild, JM, Pacey, IE, O’Neill, EC, Cunliffe, IA. (1999) The SITA perimetric threshold algorithms in glaucoma Invest Ophthalmol Vis Sci 40,1998-2000
[PubMed]Sekhar, GC, Naduvilath, TJ, Lakkai, M, et al (2000) Sensitivity of Swedish interactive threshold algorithm compared with standard full threshold algorithm in Humphrey visual field testing Ophthalmology 107,1303-1308
[CrossRef] [PubMed]Bengtsson, B, Heijl, A. (1999) Inter-subject variability and normal limits of the SITA Standard, SITA Fast, and the Humphrey Full Threshold computerized perimetry strategies, SITA STATPAC Acta Ophthalmol Scand 77,125-129
[CrossRef] [PubMed]Wild, JM, Pacey, IE, Hancock, SA, Cunliffe, IA. (1999) Between-algorithm, between-individual differences in normal perimetric sensitivity: full threshold, Fastpac, and SITA, Swedish Interactive Threshold algorithm Invest Ophthalmol Vis Sci 40,1152-1161
[PubMed]Bengtsson, B, Heijl, A. (1999) Comparing significance and magnitude of glaucomatous visual field defects using the SITA and Full Threshold strategies Acta Ophthalmol Scand 77,143-146
[CrossRef] [PubMed]Heijl, A, Lindgren, A, Lindgren, G. (1989) Test-retest variability in glaucomatous visual fields Am J Ophthalmol 108,130-135
[CrossRef] [PubMed]Weber, J, Rau, S. (1992) The properties of perimetric thresholds in normal and glaucomatous eyes Ger J Ophthalmol 1,79-85
[PubMed]Olsson, J, Heijl, A, Bengtsson, B, Rootzén, H. (1993) Frequency-of-seeing in computerized perimetry Mills, eds. Perimetry Update ,551-556 Kugler Amsterdam.
Olsson, J. (1991) Statistics in Perimetry. Doctoral thesis
Department of Mathematical Statistics, Lund University Lund, Sweden.
Bengtsson, B. (1999) Improved Computerized Perimetric Threshold Strategies. Doctoral thesis
,17 Department. of Ophthalmology, Malmö University Hospital, Lund University Lund, Sweden.
Johnson, CA, Chauhan, BC, Shapiro, LR. (1992) Properties of staircase procedures for estimating thresholds in automated perimetry Invest Ophthalmol Vis Sci 33,2966-2974
[PubMed]Watson, AB, Pelli, DG. (1983) Quest: a Bayesian adaptive psychometric method Percept Psychophys 33,113-120
[CrossRef] [PubMed]King-Smith, PE, Grigsby, SS, Vingrys, AJ, Benes, SC, Supowit, A. (1994) Efficient and unbiased modifications of the QUEST threshold method: theory, simulations, experimental evaluation and practical implication Vision Res 34,885-912
[CrossRef] [PubMed]Turpin, A, Johnson, CA, Spry, PGD. (2001) Development of maximum likelihood procedure for short-wavelength automated perimetry Wall, M Mills, RP eds. Perimetry Update 2000/2001 ,139-147 Kugler The Hague, The Netherlands.
Olsson, J, Åsman, P, Rootzén, H, Heijl, A. (1991) Improved threshold estimates using full staircase data Mills, RP Heijl, A eds. Perimetry Update 1990/1991 ,245-249 Kugler & Ghedini Amstelveen, The Netherlands.
Olsson, J. (1991) Spatial dependence of stimulus-response curves in perimetry with image restoration Statistics in Perimetry. Doctoral thesis
Department of Mathematical Statistics, Lund University Lund, Sweden.
Olsson, J, Rootzén, H. (1994) An image model for quantal response analysis in perimetry Scand J Stat 21,375-387
Olsson, J, Rootzén, H, Heijl, A. (1989) Maximum likelihood estimation of the frequency of false positive and false negative answers from up-and-down staircases of computerized threshold perimetry Heijl, A eds. Perimetry Update 1988/89 ,245-251 Kugler & Ghedini Amstelveen, The Netherlands.
Olsson, J, Bengtsson, B, Heijl, A, Rootzén, H. (1997) An improved method to estimate frequency of false positive answers in computerized perimetry Acta Ophthalmol Scand 75,181-183
[PubMed]Weber, J, Klimaschka, T. (1995) Test time and efficiency of the dynamic strategy in glaucoma perimetry Ger J Ophthalmol 4,25-31
[PubMed]Gonzales de la Rosa, M, Martinez, A, Sanches, MN, et al (1997) Accuracy of tendency-oriented perimetry with the Octopus 1-2-3 perimeter Wall, RP Heijl, A eds. Perimetry Update ,119-123 Kugler Amsterdam.
Heijl, A. (1977) Time changes of contrast threshold during automatic perimetry Acta Ophthalmol 55,512-516
Heijl, A, Drance, SM. (1983) Changes in differential threshold in patients with glaucoma during prolonged perimetry Br J Ophthalmol 67,1847-1854