Standard Microlithographic Mosaics to Assess Endothelial Cell Counting Methods by Light Microscopy in Eye Banks Using Organ Culture
Author Affiliations
  • Nilanjana Deb-Joardar
    From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France;
  • Gilles Thuret
    From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France;
  • Georges-André Racine
    Center of MicroNanoTechnology, Ecole Polytechnique Fédérale de Lausanne, Lausanne, Switzerland;
  • David Pons
    Laboratory of Signal Treatment and Instrumentation, Université Jean Monnet, Saint-Etienne, France; and
  • Gerald Brun
    Laboratory of Signal Treatment and Instrumentation, Université Jean Monnet, Saint-Etienne, France; and
  • Olivier Parriaux
    Laboratory of Signal Treatment and Instrumentation, Université Jean Monnet, Saint-Etienne, France; and
  • Michel Peoc’h
    From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France;
  • Sophie Acquart
    From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France;
    French Blood Center/Eye Bank of Saint Etienne, Saint Etienne, France.
  • Philippe Gain
    From the Laboratory Cell Adherence and Survival in Cancers and Grafts EA 3063, Université Jean Monnet, Saint Etienne, France;
Investigative Ophthalmology & Visual Science October 2006, Vol.47, 4373-4377. doi:https://doi.org/10.1167/iovs.06-0536
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Nilanjana Deb-Joardar, Gilles Thuret, Georges-André Racine, David Pons, Gerald Brun, Olivier Parriaux, Michel Peoc’h, Sophie Acquart, Philippe Gain; Standard Microlithographic Mosaics to Assess Endothelial Cell Counting Methods by Light Microscopy in Eye Banks Using Organ Culture. Invest. Ophthalmol. Vis. Sci. 2006;47(10):4373-4377. https://doi.org/10.1167/iovs.06-0536.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

purpose. To develop standard microscopic hexagonal mosaics mimicking the human corneal endothelium for quality control of endothelial cell density (ECD) measurement and verification of cell counting strategy by light microscopy in eye banks using organ culture.

methods. A standard slide, the Keratotest, was developed with 10 laser-engraved mosaics and different predetermined “cell” densities representing the range of ECDs observed routinely. Horizontal and vertical micrometric scales were etched adjacently to each mosaic, and a standard microscopy resolution test pattern was included. The Keratotest was applied to assess the reliability of a computer-assisted analyzer developed for corneal endothelial evaluation based on light microscopy images.

results. The Keratotest consisted of 10 microlithographic homogeneous mosaics of 1-mm2 printed area and 1.2-μm cell boundary thickness. The micrometric scale associated with each mosaic aided in simultaneous verification of microscope calibration, and the test pattern aided in checking the microscope resolution. The design was unalterable and reproducible, and the glass slide incorporated in a carbon fiber support ensured easy handling and safe transport. Evaluation of the Keratotest mosaics by the computer-assisted analyzer found a high level of agreement (error margin between +0.12 and –0.46%) with the laser-engraved cell density.

conclusions. This prototype device enabled assessment of reliability of ECD measurement in eye banks. It also allowed verification of the calibration and resolution of light microscopes. Periodic validation of counting procedure in eye banks with mosaics of known “cell” densities should be useful for standardization of donor corneal tissue quality control.

Endothelial cell density (ECD) is the principal criterion for quality evaluation of donor corneal tissue in eye banks. The ECD of the grafted cornea determines its survival with time in terms of maintenance of transparency. 1 2 An ECD of 2000 to 2400 cells/mm2 is considered by most eye banks in Europe to be the standard cutoff for corneal delivery. 3 ECD is estimated on a small sample of few hundred cells out of an estimated 250,000 to 350,000 cells present in the cornea. It is therefore imperative to have a precise and reproducible method of ECD evaluation that would help standardize donor tissue evaluation methods and exchange between eye banks. 
In Europe, where organ culture is the storage method of choice, the endothelium is viewed directly under a light microscope after osmotic preparation consisting of incubation of the endothelial face of the cornea for 1 to 4 minutes in 0.9% sodium chloride or 1.8% sucrose. 4 This dilates the intercellular spaces and enables the visualization of cell contours that are otherwise invisible. From this preparation, cells are most often counted manually in real time by observation through a calibrated reticule fitted in the microscope eyepiece or by indication of the cells in a known surface on endothelial photographs. Recent years have seen the emergence of prototypes and commercially available computer-assisted counting systems, currently used in some European eye banks, 3 in which analysis of digitized optical microscopy images allows rapid and effective measurement of ECD of organ-cultured corneas. 5 6 7 8  
Previous studies analyzing ECD measurement methods in 22 eye banks in France have emphasized the need for standardization of counting methods. 9 10 Lack of microscope calibration had been identified as a major factor responsible for gross variation in ECD estimates between eye banks. Differences in manual counting techniques were also shown to influence the ECD values. Until now, the only method of calibration available was a certified graduated micrometer, a traditional method in microscopy. Essentially, this measures only the magnification ratio between the tissue observed and its image projected in a calibrated reticule or by image analysis software. Whatever the method used, manual or computerized, periodical calibration of the measuring equipment and standardization of counting procedures are essential to ensure reliability of intrabank and interbank cell counts. To address this issue, we present a prototype standard slide, called Keratotest, carrying a series of hexagonal mosaics of different densities mimicking a range of human corneal endothelia. 
Methods
Specifications
The design of the Keratotest had to meet certain essential requirements. It had to correspond to microscopic hexagonal mosaics and mimic the human corneal endothelium, particularly “cell” area and “intercellular” spaces, as close to the reality as possible when observed under standard light microscopes. Cell density (CD) had to be perfectly mathematically predetermined to avoid any bias and had to represent the range of ECDs observed commonly in eye banks (we previously showed that errors in manual counting could partly depend on ECD 10 ). Possibility of simultaneous calibration of the microscope by a classic graduated micrometer would facilitate a complete system check with a single device. Verification of the resolving power of the optical components (eyepiece and objective) of the microscope would be an additional advantage. To withstand mechanical transportation without structural damage, the material had to be durable and unalterable. The manufacturing process had to be reliable and reproducible to facilitate production of a series. 
Manufacturing Process
The Keratotest consisted of a series of laser-engraved regular microlithographic mosaics with predetermined CDs. The steps of design and manufacture are described briefly here. 
Calculation of CD
Given that a hexagon is made of six equilateral triangles of side length (L) and height (H), from Pythagorean theory it is known that H = √3/2;L. Therefore, the area of each triangle is given by ½LH = √¾L2, and the area of a hexagon is 6 × √¾L2 = 3√3/2L2. From this, we determined mathematically the number of the hexagons that could be engraved within a given surface measuring 1 mm2 so as to obtain a pattern of 10 mosaics of different CDs. The cell border area was calculated according to the Bourne hypothesis, by which the intercellular space is assumed to be negligible and the true cell border is located at the center of each border line. 11 Therefore, the total surface area of each cell was given by its internal surface area plus half the cell border area. Next, from the CD, we calculated the best-fit circle diameter of the hexagon, which enabled us to obtain its vectorial coordinates and to generate the holographic pattern. 
Creation of Holographic Pattern
High-quality, two-dimensional drawings of 10 mosaic patterns with adjacent horizontal and vertical micrometer were obtained in DXF format (Drawing eXchange Format; AutoCAD software; Autodesk SA, Neuchatel, Switzerland) containing the holographic data. A simplified microscopy resolution test pattern conforming to the National Bureau of Standards (NBS 1010a Microscopy Test Chart) consisting of three bars of equal spatial frequency in the horizontal and vertical directions was also designed. These were next converted to the machine-specific CIF (Caltech Intermediate File format) format using a translator (CAD; LinkCAD, Burlington, IA). 
Photolithography
Photolithographic mosaics were made at the Swiss Federal Institute of MicroNanoTechnology in Lausanne, Switzerland. A glass wafer was coated with a 32-nm–thick chromium layer in a vacuum chamber on the principle of evaporation. The chromium layer was covered with a 1.5-μm–thick layer of S1813 photoresist (Rohm and Haas Electronic Materials, Coventry, UK) with the use of a spin coater. The holographic pattern was generated by a high-resolution, laser beam direct writing lithography machine (DWL200 Laser Writer; Heidelberg Instruments Mikrotechnik GmbH, Heidelberg, Germany) that uses a Kr laser light source (413-nm wavelength). The target substrate was placed on an XY table with Mach-Zender interferometer position feedback, which permitted high-precision movement with a 50-nm encoding resolution and a 2-μm depth of focus. 
Generation of Test Slide
The exposed pattern was transferred to the glass wafer by standard photoresist development and selective etching of exposed chromium. Etched samples were characterized by optical microscopy and scanning electron microscopy. The photoresist, developed by analogy to a photographic film, was baked at 120°C on a hotplate to make it hard and resilient to the etching solution. The substrate was then immersed in an etching solution containing ceric ammonium nitrate. At the end of the process, resin residues were eliminated with a stripping solution. The engraved surface was shielded in a coverslip mounted with resin (Eukitt; ProSciTech, Kirwan, Australia) to ensure mechanical protection. The mount was then cut and incorporated into an unalterable carbon-fenestrated support of dimensions similar to those of a microscope slide (76 × 26 mm). 
Example of Application to Computer-Assisted Counting System
We used the Keratotest to assess a tri-image endothelial analyzer (Sambacornea; Samba Technologies, Meylan, France) developed in our laboratory specifically for the evaluation of ECD and morphometry based on light microscopy images. 7 The analyzer itself was calibrated with a graduated micrometer of 1-mm certified length (Leica Microsystems, Rueil-Malmaison, France) different from that included in the Keratotest. The 10 mosaics were counted with the tri-image endothelial analyzer five times each by two independent observers. Each time, three random images of a mosaic were captured after displacement and rotation of the Keratotest to ensure that each view was unique. Images were analyzed in a masked fashion to obtain a mean CD for that observation. Each observer selected a different counting zone on each mosaic each time to enable analysis of widely varying CDs. Selections were subjected to automated analysis, as described elsewhere. 7 Analysis was performed only by threshold adjustment, and no manual touch-up of contour segments was made. For each mosaic, the (mean ± SD) CD was calculated, and the percentage error was compared with the actual CD of the laser-engraved mosaic. Results of the two analyses were compared. 
Results
Presentation of the Keratotest
The Keratotest consisted of homogeneous hexagons of 10 different CDs identified by the letters A to J engraved in a central window of a single glass slide (Fig. 1) . High-precision laser engraving allowed drawing of cells with boundaries of 1.2 μm, as observed for human endothelial cells after osmotic dilation (Fig. 2) . The exact CD of each of the 10 mosaics was recalculated after taking into consideration the necessary approximations (for design of hexagons, calculations were rounded off to four decimal places), precision of engraving, and tolerance of the laser writer; the difference between the theoretical CD and the final CD was negligible (Table 1) . The graduated micrometer, placed adjacently to each mosaic, enabled simultaneous vertical and horizontal calibration of the microscope. The test pattern helped to verify the optical resolution of the microscope. Resolution was determined by observing the line pair pattern, where all three lines and spaces in both the horizontal and the vertical direction could be discerned. Spatial frequency, expressed in terms of cycles per millimeter, was indicated by the numeric value shown with each test element. The glass slide incorporated in the carbon fiber support ensured easy maintenance, handling, and safe transport. 
Application to Computer-Assisted Counting System
The CD estimation of each mosaic (mean ± SD of five measures) of the Keratotest performed by the tri-image endothelial analyzer (Fig. 3)found a level of agreement ranging from +0.12% to –0.46% compared with the laser-engraved CD (Table 2)
Discussion
We developed a prototype standard slide to assess the reliability of ECD measurement with light microscopy in eye banks using organ culture. Additionally, the slide allowed verification of microscope calibration and optical resolution. Reliability of the counting procedure could be assessed by a periodic validation of eye bank counting methods using the range of mosaics with known CDs. 
Quality assessment of donor tissue and accreditation of medical personnel are important considerations in eye banks worldwide, and the reliability and precision of endothelial cell counting and its periodic standardization are essential prerequisites. In contrast to standardization methods prevalent in other medical disciplines, in which a sample of predetermined value is used for comparison or calibration, no similar tool is available for application to endothelial cell count. A corneal donor tissue sample cannot be used for calibration because no two corneas are alike. We have previously demonstrated that ECD estimation was neither precise nor reproducible in French eye banks. 9 10 Inadequate microscope calibration and nonuniform counting procedures (difference in number of cells assessed, number of reticule units counted, and count strategy—“border” or “zone”) had been identified as the major factors for the high intrabank and interbank variations of ECD. Human error might also have contributed to this variation because a tendency to underestimate was noted in corneas with high ECDs. Based on these findings, we advocated the urgent need for standardization of cell counting methods to prevent the delivery of corneas with low ECD at certain centers and wastage in others. 
As a first step in this direction, we designed Alizarin red–stained test corneas to evaluate the reproducibility of cell counts between eye banks. 10 Endothelial cells were rendered perfectly visible because of membrane staining, thereby eliminating visualization problems encountered during routine light microscopy resulting from nonuniform response of the endothelial monolayer to the osmotic challenge. 5 Flat mounting of the dissected endothelium enabled visualization of the entire endothelium in a single plane, thus eliminating field depth–related measurement errors caused by the combined effects of physiologic corneal curvature and posterior corneal folding. Stained endothelial flat mounts, though representative of the real human endothelium, had certain limitations: (1) the ECD could not be predetermined mathematically, and (2) the slides were not reproducible because each human endothelium is unique. Hence, manufacture of a series to facilitate interbank exchange was not possible. A different standard that would represent the real endothelium but would be mathematically accurate and reproducible and would be useful for training new technicians was required to assess the reliability of eye bank counting methods. 
The importance of calibrating specular microscopes for in vivo ECD counts has been emphasized in previous studies. 12 13 Different analysis techniques of specular microscopy images—i.e., use of a fixed-frame or a center method—led to important interbank ECD variation in the Cornea Donor Study. 14 Light microscopic evaluation used for organ-cultured corneas also has similar prerequisites. Computer-assisted analyzers also require calibration of the optical and the computing devices. In the tri-image endothelial analyzer, this is achieved by use of a calibration factor derived from a standard micrometric slide. However, verification of counting procedures or detection of human errors is not spontaneously possible. Our photolithographic device was designed for simultaneous calibration of the microscope and the counting device (reticule, printed grid, or digital image analyzer) and could be adapted to manual and computer-assisted cell counts. The extremely narrow limits of difference between the mathematically determined Keratotest ECD and that obtained after repeated counts by different observers with the computer-assisted analyzer permitted assessment of the reliability of counting methods. 
Patterning of flat surfaces by use of photolithography is a high-precision technology widely used in industrial applications, such as in the manufacture of integrated circuits and microsystems. 15 Choice of this particular method guaranteed the etching precision and hence the accuracy of the CDs obtained. Mathematical predetermination allowed creation of the exact CD desired, thus permitting a large range of ECDs (800–3200 cells/mm2) likely to be encountered in eye banks. The manufacturing process made it possible to create reproducible slides on a semi-industrial scale to meet the requirements of potential users. 
The current prototype allows only validation of the ECD count. Further modifications with mosaics having “cells” of different surface area and shape are ongoing to enable assessment of morphometric parameters increasingly recognized as valuable indicators of donor tissue quality. 16 17  
 
Figure 1.
 
Global view of the Keratotest. Small gray circles: ten laser-engraved mosaics (AJ) of various cell densities. Cell densities (in cells/mm2) are: (A) 800; (B) 2400; (C) 3200; (D) 1800; (E) 2000; (F) 2200; (G) 1200; (H) 2800; (I) 1600; and (J) 3600. Large circle: position of the microscopy resolution test pattern. Original magnification, ×4.
Figure 1.
 
Global view of the Keratotest. Small gray circles: ten laser-engraved mosaics (AJ) of various cell densities. Cell densities (in cells/mm2) are: (A) 800; (B) 2400; (C) 3200; (D) 1800; (E) 2000; (F) 2200; (G) 1200; (H) 2800; (I) 1600; and (J) 3600. Large circle: position of the microscopy resolution test pattern. Original magnification, ×4.
Figure 2.
 
(A) Magnified view of the mosaic, G, of the Keratotest showing the regular hexagonal pattern with a cell density of 1200/mm2 and the horizontal and vertical micrometer axes with 100-μm primary, 50-μm secondary, and 10-μm tertiary graduations. Original magnification, ×4. (B) Magnified view of inset showing the intercellular margins (thickness, 1.2 μm) and micrometer details. Original magnification, ×10.
Figure 2.
 
(A) Magnified view of the mosaic, G, of the Keratotest showing the regular hexagonal pattern with a cell density of 1200/mm2 and the horizontal and vertical micrometer axes with 100-μm primary, 50-μm secondary, and 10-μm tertiary graduations. Original magnification, ×4. (B) Magnified view of inset showing the intercellular margins (thickness, 1.2 μm) and micrometer details. Original magnification, ×10.
Table 1.
 
Comparison between the Desired Mathematically Calculated Cell Density and the Final Cell Density Engraved on Keratotest
Table 1.
 
Comparison between the Desired Mathematically Calculated Cell Density and the Final Cell Density Engraved on Keratotest
Desired Cell Density (cells/mm2) Cell Surface Area (μm2) Best-fit Circle Diameter of Hexagon (μm) Approximation to 4 Decimal Places Recalculated Surface Area after Approximation (μm2) Final Cell Density Engraved (cells/mm2)
800 1250 43.8691337650831 43.8691 1249.9981 800.0012
1200 833.333333 35.8189977274514 35.8190 833.3334 1199.9998
1600 625 31.0201619700700 31.0202 625.0015 1599.9961
1800 555.555556 29.2460891767221 29.2461 555.5560 1799.9987
2000 500 27.7452763352521 27.7453 500.0009 1999.9966
2200 454.545455 26.4540830139439 26.4541 454.5460 2199.9972
2400 416.666667 25.3278561883864 25.3279 416.6681 2399.9917
2800 357.142857 23.4490383433568 23.4490 357.1417 2800.0092
3200 312.5 21.9345668825415 21.9346 312.5009 3199.9903
3600 277.777778 20.6801079800467 20.6801 277.7776 3600.0028
Figure 3.
 
Printout of the final analysis results of Keratotest mosaic A by the tri-image endothelial analyzer. Blue line, arrow: selected analysis zone. Green: contour recognition pattern as obtained after appropriate thresholding. Orange: cells represent automatic detection of probable errors in contour recognition. The section on the right shows the cell density (including number of cells counted) and morphometric data (coefficient of variation [CV] of cell area, minimum/maximum/mean cell area, SD, and percentage of hexagonal cells). Note that for mosaic A containing 800 cells, the calculated cell density was 797 cells/mm2, and CV was 1% based on a total of 1859 cells counted on three images. Hexagonality is 100% based on a total of 1578 cells counted on three images (cells intersected by the blue line were excluded for hexagonality measurement).
Figure 3.
 
Printout of the final analysis results of Keratotest mosaic A by the tri-image endothelial analyzer. Blue line, arrow: selected analysis zone. Green: contour recognition pattern as obtained after appropriate thresholding. Orange: cells represent automatic detection of probable errors in contour recognition. The section on the right shows the cell density (including number of cells counted) and morphometric data (coefficient of variation [CV] of cell area, minimum/maximum/mean cell area, SD, and percentage of hexagonal cells). Note that for mosaic A containing 800 cells, the calculated cell density was 797 cells/mm2, and CV was 1% based on a total of 1859 cells counted on three images. Hexagonality is 100% based on a total of 1578 cells counted on three images (cells intersected by the blue line were excluded for hexagonality measurement).
Table 2.
 
Comparison of the Cell Density of Keratotest and That Calculated by the Sambacornea Analyzer
Table 2.
 
Comparison of the Cell Density of Keratotest and That Calculated by the Sambacornea Analyzer
Actual Cell Density Mean ± SD Mean Accuracy (%)
1st Observer (n) 2nd Observer (n) 1st Observer 2nd Observer
800 797 ± 3 (797) 798 ± 0 (1749) −0.35 −0.23
2400 2391 ± 4 (2731) 2391 ± 2 (5545) −0.38 −0.37
3200 3188 ± 4 (3697) 3185 ± 3 (7514) −0.36 −0.46
1800 1799 ± 4 (2021) 1796 ± 3 (4092) −0.04 −0.21
2000 2000 ± 5 (2253) 1996 ± 2 (4664) −0.02 −0.19
2200 2197 ± 8 (2414) 2191 ± 1 (5021) −0.15 −0.41
1200 1201 ± 5 (1322) 1197 ± 1 (2705) 0.12 −0.22
2800 2799 ± 8 (3222) 2789 ± 2 (6510) −0.03 −0.41
1600 1603 ± 5 (1792) 1597 ± 1 (3736) 0.20 −0.21
3600 3591 ± 8 (4160) 3585 ± 4 (8649) −0.24 −0.41
The authors thank Samuel Proton for assistance with vectorization and Souad Herrag for technical help. 
NishimuraJK, HodgeDO, BourneWM. Initial endothelial cell density and chronic endothelial cell loss rate in corneal transplants with late endothelial failure. Ophthalmology. 1999;106:1962–1965. [CrossRef] [PubMed]
BourneWM. Cellular changes in transplanted human corneas. Cornea. 2001;20:560–569. [CrossRef] [PubMed]
European Eye Bank Association Directory. 2006; 14th ed.European Eye Bank Association Amsterdam.
PelsL. Organ culture: the method of choice for preservation of human donor corneas. Br J Ophthalmol. 1997;81:523–525. [CrossRef] [PubMed]
Barisani-AsenbauerT, BaumgartnerI, GrabnerG, SturM. Automated digital image analysis of organ culture preserved donor corneas. Ophthalmic Res. 1993;25:94–99. [CrossRef] [PubMed]
ReinhardT, SpelsbergH, HolzwarthD, DahmenN, GodehardtE, SundmacherR. Endothelial evaluation of corneal transplants by digital imaging. Klin Monatsbl Augenheilkd. 1999;214:407–411. [CrossRef] [PubMed]
GainP, ThuretG, KodjikianL, et al. Automated tri-image analysis of stored corneal endothelium. Br J Ophthalmol. 2002;86:801–808. [CrossRef] [PubMed]
RuggeriA, GrisanE, JaroszewskiJ. A new system for the automatic estimation of endothelial cell density in donor corneas. Br J Ophthalmol. 2005;89:306–311. [CrossRef] [PubMed]
ThuretG, ManissolleC, AcquartS, et al. Is manual counting of corneal endothelial cell density in eye banks still acceptable? The French experience. Br J Ophthalmol. 2003;87:1481–1486. [CrossRef] [PubMed]
ThuretG, ManissolleC, AcquartS, et al. Urgent need for normalization of corneal graft quality controls in French eye banks. Transplantation. 2004;78:1299–1302. [CrossRef] [PubMed]
BourneWM. Morphologic and functional evaluation of the endothelium of transplanted human corneas. Trans Am Ophthalmol Soc. 1983;81:403–450. [PubMed]
van SchaickW, van DoorenBT, MulderPG, Volker-DiebenHJ. Validity of endothelial cell analysis methods and recommendations for calibration in Topcon SP-2000P specular microscopy. Cornea. 2005;24:538–544. [CrossRef] [PubMed]
OhnoK, NelsonLR, McLarenJW, HodgeDO, BourneWM. Comparison of recording systems and analysis methods in specular microscopy. Cornea. 1999;18:416–423. [CrossRef] [PubMed]
LassJ, GalR, RuedyK, et al. An evaluation of image quality and accuracy of eye bank measurement of donor cornea endothelial cell density in the Specular Microscopy Ancillary Study. Ophthalmology. 2005;112:431–440. [CrossRef] [PubMed]
MadouM. Fundamentals of Microfabrication. 1997;CRC Press Boca Raton, FL.
YeeRW, MatsudaM, SchultzRO, EdelhauserHF. Changes in the normal corneal endothelial cellular pattern as a function of age. Curr Eye Res. 1985;4:671–678. [CrossRef] [PubMed]
WiffenSJ, NelsonLR, AliAF, BourneWM. Morphologic assessment of corneal endothelium by specular microscopy in evaluation of donor corneas for transplantation. Cornea. 1995;14:554–561. [PubMed]
Figure 1.
 
Global view of the Keratotest. Small gray circles: ten laser-engraved mosaics (AJ) of various cell densities. Cell densities (in cells/mm2) are: (A) 800; (B) 2400; (C) 3200; (D) 1800; (E) 2000; (F) 2200; (G) 1200; (H) 2800; (I) 1600; and (J) 3600. Large circle: position of the microscopy resolution test pattern. Original magnification, ×4.
Figure 1.
 
Global view of the Keratotest. Small gray circles: ten laser-engraved mosaics (AJ) of various cell densities. Cell densities (in cells/mm2) are: (A) 800; (B) 2400; (C) 3200; (D) 1800; (E) 2000; (F) 2200; (G) 1200; (H) 2800; (I) 1600; and (J) 3600. Large circle: position of the microscopy resolution test pattern. Original magnification, ×4.
Figure 2.
 
(A) Magnified view of the mosaic, G, of the Keratotest showing the regular hexagonal pattern with a cell density of 1200/mm2 and the horizontal and vertical micrometer axes with 100-μm primary, 50-μm secondary, and 10-μm tertiary graduations. Original magnification, ×4. (B) Magnified view of inset showing the intercellular margins (thickness, 1.2 μm) and micrometer details. Original magnification, ×10.
Figure 2.
 
(A) Magnified view of the mosaic, G, of the Keratotest showing the regular hexagonal pattern with a cell density of 1200/mm2 and the horizontal and vertical micrometer axes with 100-μm primary, 50-μm secondary, and 10-μm tertiary graduations. Original magnification, ×4. (B) Magnified view of inset showing the intercellular margins (thickness, 1.2 μm) and micrometer details. Original magnification, ×10.
Figure 3.
 
Printout of the final analysis results of Keratotest mosaic A by the tri-image endothelial analyzer. Blue line, arrow: selected analysis zone. Green: contour recognition pattern as obtained after appropriate thresholding. Orange: cells represent automatic detection of probable errors in contour recognition. The section on the right shows the cell density (including number of cells counted) and morphometric data (coefficient of variation [CV] of cell area, minimum/maximum/mean cell area, SD, and percentage of hexagonal cells). Note that for mosaic A containing 800 cells, the calculated cell density was 797 cells/mm2, and CV was 1% based on a total of 1859 cells counted on three images. Hexagonality is 100% based on a total of 1578 cells counted on three images (cells intersected by the blue line were excluded for hexagonality measurement).
Figure 3.
 
Printout of the final analysis results of Keratotest mosaic A by the tri-image endothelial analyzer. Blue line, arrow: selected analysis zone. Green: contour recognition pattern as obtained after appropriate thresholding. Orange: cells represent automatic detection of probable errors in contour recognition. The section on the right shows the cell density (including number of cells counted) and morphometric data (coefficient of variation [CV] of cell area, minimum/maximum/mean cell area, SD, and percentage of hexagonal cells). Note that for mosaic A containing 800 cells, the calculated cell density was 797 cells/mm2, and CV was 1% based on a total of 1859 cells counted on three images. Hexagonality is 100% based on a total of 1578 cells counted on three images (cells intersected by the blue line were excluded for hexagonality measurement).
Table 1.
 
Comparison between the Desired Mathematically Calculated Cell Density and the Final Cell Density Engraved on Keratotest
Table 1.
 
Comparison between the Desired Mathematically Calculated Cell Density and the Final Cell Density Engraved on Keratotest
Desired Cell Density (cells/mm2) Cell Surface Area (μm2) Best-fit Circle Diameter of Hexagon (μm) Approximation to 4 Decimal Places Recalculated Surface Area after Approximation (μm2) Final Cell Density Engraved (cells/mm2)
800 1250 43.8691337650831 43.8691 1249.9981 800.0012
1200 833.333333 35.8189977274514 35.8190 833.3334 1199.9998
1600 625 31.0201619700700 31.0202 625.0015 1599.9961
1800 555.555556 29.2460891767221 29.2461 555.5560 1799.9987
2000 500 27.7452763352521 27.7453 500.0009 1999.9966
2200 454.545455 26.4540830139439 26.4541 454.5460 2199.9972
2400 416.666667 25.3278561883864 25.3279 416.6681 2399.9917
2800 357.142857 23.4490383433568 23.4490 357.1417 2800.0092
3200 312.5 21.9345668825415 21.9346 312.5009 3199.9903
3600 277.777778 20.6801079800467 20.6801 277.7776 3600.0028
Table 2.
 
Comparison of the Cell Density of Keratotest and That Calculated by the Sambacornea Analyzer
Table 2.
 
Comparison of the Cell Density of Keratotest and That Calculated by the Sambacornea Analyzer
Actual Cell Density Mean ± SD Mean Accuracy (%)
1st Observer (n) 2nd Observer (n) 1st Observer 2nd Observer
800 797 ± 3 (797) 798 ± 0 (1749) −0.35 −0.23
2400 2391 ± 4 (2731) 2391 ± 2 (5545) −0.38 −0.37
3200 3188 ± 4 (3697) 3185 ± 3 (7514) −0.36 −0.46
1800 1799 ± 4 (2021) 1796 ± 3 (4092) −0.04 −0.21
2000 2000 ± 5 (2253) 1996 ± 2 (4664) −0.02 −0.19
2200 2197 ± 8 (2414) 2191 ± 1 (5021) −0.15 −0.41
1200 1201 ± 5 (1322) 1197 ± 1 (2705) 0.12 −0.22
2800 2799 ± 8 (3222) 2789 ± 2 (6510) −0.03 −0.41
1600 1603 ± 5 (1792) 1597 ± 1 (3736) 0.20 −0.21
3600 3591 ± 8 (4160) 3585 ± 4 (8649) −0.24 −0.41
×
×

This PDF is available to Subscribers Only

Sign in or purchase a subscription to access this content. ×

You must be signed into an individual account to use this feature.

×