The slit lamp OCT was initially designed at the University of LuΜbeck by Hoerauf et al.
7 The initial SLOCT operated at a wavelength of 830 nm. A modified version with a superluminescent diode at 1310 nm was subsequently described
8 and is now commercially available as the SLOCT (Heidelberg Engineering, GmbH, Dossenheim, Germany) and Visante OCT (Carl Zeiss Meditec, Dublin, CA). Anterior chamber angle measurements with anterior segment OCT have recently been demonstrated to be reproducible. In the study by Li et al.
2 using Visante OCT, the intrasession/intersession intraclass correlation for AOD and TIA measured in the dark were 0.98/0.95 and 0.97/0.92, respectively. With the prototype anterior segment OCT developed by Carl Zeiss Meditec, Radhakrishnan et al.
1 showed the intraclass correlations were between 0.67 and 0.90 for short term measurements (in a single visit) and 0.56 and 0.93 for long-term measurements (median of 3 weeks). Muller et al.
3 also reported high intraclass correlations (0.93β0.97 for AOD, 0.91β0.94 for TIA) for SLOCT angle measurements. Although high reproducibility for anterior chamber angle measurements was observed, we found that SLOCT and Visante OCT had poor agreement
(Figs. 3 4 5) . For example, the difference between SLOCT and Visante OCT measurements of TIA would be within 28Β° in 95% of observations. The poor agreement is not likely to be related to the difference in pupil size, because there was no correlation between differences in angle measurements and difference in pupil size measured by the two instruments (all with
P β₯ 0.118). Although it is unavoidable to have minimal illumination from the screen display, the finding that the pupil size measured by SLOCT and Visante OCT was comparable indicates that the difference of illumination during imaging was minimal. It is plausible, however, that the poor agreement could be attributable to the differences in the choice of refractive indexes in the calculation of anterior segment dimensions, algorithms for image dewarping, the state of accommodation, and the exact scan locations. The corneal refractive index adopted by SLOCT is 1.376, which is derived from the model of Gullstrand of the human eye for visible light, whereas a group index of 1.389 was used in Visante OCT.
9 The different refractive indexes may lead to different corneal thickness and anterior chamber angle measurements. Because of the scan geometry of the scan probe and the refraction at smooth surfaces of the eye, algorithms for dewarping are incorporated in the analysis software of Visante OCT and SLOCT to correct for image misalignment. It is uncertain whether these algorithms have equivalent adjustment. The use of internal fixation in Visante OCT and external fixation in SLOCT probably would have resulted in a different state of accommodation during the imaging. For each diopter increase in accommodation, the anterior pole of the lens moves forward by an average of 30 ΞΌm.
10 Change in lens position is probably linked with change in the angle width. Finally, since the scan position was based on subjective alignment of a scan line bisecting the pupil, there may be differences in the exact scan location. Because we lack a standard reference for comparison, it is not yet certain which anterior segment OCT provides more accurate estimation of the angle width.Β