Abstract
Purpose.:
To determine the effects of silicone hydrogel lens wear and lens–solution interactions on ocular surface sensitivity.
Methods.:
Forty-eight adapted lens wearers completed the study, which comprised two phases. Phase 1 included habitual lens wear, no lens wear (7 ± 3 days), and balafilcon A lenses (PV; PureVision; Bausch & Lomb, Rochester, NY) with a hydrogen peroxide-based regimen for 2 weeks; phase 2 included wear of PV with the use of a multipurpose solution containing either polyhexamethylene-biguanide (PHMB) or Polyquad/Aldox (Alcon Laboratories, Fort Worth, TX) preservative, each for 1 week, with a 2-week washout period between solutions. Tactile and pneumatic (mechanical and chemical) stimuli were delivered, and thresholds were determined by Cochet-Bonnet (Luneau Ophthalmologie, Chartres, France) and Belmonte (Cooperative Research Centre for Eye Research and Technology, Sydney, NSW, Australia) pneumatic esthesiometers, respectively. Corneal and conjunctival thresholds and staining scores were assessed at baseline, after 2 and 8 hours of lens wear on day 1 and at the end of each wearing cycle (2 hours).
Results.:
In phase 1, compared to the no-lens baseline, corneal tactile thresholds increased at the 1-day, 8-hour and the 2-week visits (P < 0.05), whereas conjunctival mechanical thresholds decreased at the 1-day, 2-hour and the 2-week visits (P < 0.05). In phase 2, the chemical thresholds were lower with PHMB-preserved solution compared with the Polyquad/Aldox system at the 1-day, 2-hour and the 1-week visits (P < 0.05). Staining scores correlated inversely with conjunctival chemical thresholds (all P < 0.05).
Conclusions.:
Ocular surface sensitivity changed in adapted lens wearers, when lenses were refit after a no-lens interval and during lens wear with different care regimens. The corneal staining that was observed with certain lens–solution combinations was accompanied by sensory alteration of the ocular surface—that is, higher levels of staining correlated with increased conjunctival chemical sensitivity. (ClinicalTrials.gov number, NCT00455455.)
Highly oxygen-permeable silicone hydrogel (SH) lenses offer certain advantages over traditional hydrogel lenses by eliminating lens-induced hypoxia and producing less detrimental effects on corneal homeostasis.
1 Despite these ocular health benefits, SH lens wear is generally similar to hydrogel lens wear in its mechanical interaction with the ocular surface and the effects on the structure and physiology of the tear film.
2
The contact lens interacts with the cornea and conjunctiva, each being innervated by sensory nerve endings
3 that are functionally heterogeneous
4 : Mechanonociceptors respond to mechanical stimuli only, cold receptors signal downward temperature changes in the non-noxious range, and polymodal nociceptors respond to mechanical, chemical, and thermal stimuli.
4 The cornea and conjunctiva provide sensory input to the functional unit comprising the ocular surface (cornea, conjunctiva, and meibomian glands), the lacrimal glands, and the sensory and motor nerves that connect them.
5 Through a complex network, the afferent and efferent nerves link the components of the integrated unit into a homeostatic loop, with the primary function of protecting and maintaining the health of the ocular surface and the tear film.
6 In addition, corneal sensory nerves exert various trophic effects on the cornea, which may play a role in the modulation of wound healing after corneal injuries.
7
The measurement of ocular surface sensitivity is one way to assess the functioning of the sensory nerves and has been a useful clinical indicator of corneal health, in contact lens wear and corneal disease and during the healing process after various corneal injuries and refractive surgery.
8 –10 In the past, sensitivity has been measured mainly by Cochet-Bonnet–type esthesiometers (Luneau Ophthalmologie, Chartres, France),
11,12 in which hair or nylon filaments of variable diameters and lengths are used to deliver tactile stimuli to the ocular surface. The more recently developed pneumatic esthesiometers
13 –15 deliver controlled air pulses at various temperature and air-CO
2 mixtures to the ocular surface and allow measurements of ocular surface sensitivity over a range of thermal, mechanical, and chemical stimuli.
Reduced corneal sensitivity has been reported in contact lens wear.
16 –24 The reduction in sensitivity seems to be eliminated after the cessation of lens wear.
17,18 Metabolic impairment (i.e., hypoxia) has been considered to be the principal cause of reduced sensory nerve function of traditional low-oxygen-permeability (Dk) hydrogel lenses.
21,25 Other factors contributing to this sensitivity loss include sensory adaptation
26,27 and acidosis-suppressed corneal nerve function.
8 Most published reports have been based on wearing earlier generation lenses of low-Dk materials, and the sensitivity measurements have been limited to tactile stimulation. It is largely unknown whether SH lens wear affects ocular surface sensitivity differently, particularly when lens-induced hypoxia has been reduced or eliminated. Also, there have been no reports on how tactile, mechanical, and chemical sensitivity changes with SH lens wear over time. Moreover, perhaps due to the unique nature of SH lens materials, there is accumulating evidence that the interaction between certain SH lenses and preserved multipurpose solutions (MPSs) affects the ocular surface, manifesting in corneal fluorescein staining.
28,29 It is unclear whether the interaction between the lens material and care regimens would have an impact on ocular surface sensitivity, although one study with a small sample has intimated that this may be the case.
24
The present study investigated the effects of SH lens wear and lens–solution interactions on ocular surface sensitivity. We used different stimuli delivered by Cochet-Bonnet and Belmonte (Cooperative Research Centre for Eye Research and Technology, Sydney, NSW, Australia) esthesiometers and measured corneal and conjunctival thresholds in a group of adapted contact lens wearers, before and during a short period of cessation of lens wear and after refitting with SH lenses disinfected with a variety of care regimens.
Sample size calculations to detect significant differences were derived from unpublished data; the sample size necessary to achieve a power of 0.80 with an effect size of 0.4 at a 5% significance level was 50. The 50 subjects, consisting of 35 women and 15 men (mean age 25.2 ± 7.4 years; range 18–45), were adapted contact lens wearers and were asymptomatic during at least 8 hours of lens wear. Each subject had no history of eye surgery or systemic or ocular disease and was not using any systemic or topical medication that would affect ocular health. Two subjects did not complete the study: one due to a burning sensation during lens insertion and the other one for personal reasons.
Before entering the study, 27 of 48 subjects who completed the study were wearing conventional hydrogel lenses, and 21 were SH lens wearers. The median length of contact lens wear was 7 years (10th, 90th percentiles: 2, 15 years).
Balafilcon A SH lenses (PureVision; Bausch & Lomb, Rochester, NY) and three lens care regimens were used in the study. A hydrogen peroxide-based regimen (Aosept; CIBA Vision, Duluth, GA) was used in phase I and in a washout period in phase II, to eliminate potential confounding effects of surfactants and preservatives. Two multipurpose solutions (MPSs), a polyhexamethylene biguanide (PHMB)-based product (ReNu MultiPlus; Bausch & Lomb) and a solution preserved with Polyquad (polyquaternium-1) and Aldox (myristamidopropyl dimethylamine; Opti-Free RepleniSH; Alcon Laboratories, Fort Worth, TX), were used in phase II. The subjects wore balafilcon A lenses throughout the study, fitted according to the manufacturer's guidelines on a daily wear basis. Each subject received a new pair of lenses on commencing each lens care regimen period. The lenses were cleaned and disinfected after each wearing period (i.e., after each day of lens wear), using the lens care regimen assigned to each particular phase. The subjects were not permitted to use rewetting drops while wearing the study lenses.
Tactile Stimulation Thresholds.
Pneumatic Mechanical and Chemical Stimulation Thresholds.
Corneal and conjunctival thresholds for tactile, pneumatic mechanical, and chemical stimulation for phase I are presented in
Table 1.
Table 1. Corneal and Conjunctival Tactile, Pneumatic Mechanical, and Chemical Thresholds in Phase I
Table 1. Corneal and Conjunctival Tactile, Pneumatic Mechanical, and Chemical Thresholds in Phase I
Lens Wear | Tactile (mm/g/s) | Pneumatic Mechanical (Air Flow mL/min) | Chemical (%CO2 Added) |
Cornea | Conjunctiva | Cornea | Conjunctiva | Cornea | Conjunctiva |
Habitual | 31.0 ± 18.2 | 116.0 ± 45.1 | 56.8 ± 30.3 | 60.3 ± 33.2 | 24.1 ± 9.2 | 47.8 ± 16.7 |
None | 28.7 ± 16.0 | 117.1 ± 37.5 | 53.4 ± 26.2 | 64.1 ± 30.6 | 27.0 ± 11.7 | 47.0 ± 15.9 |
PV lens and AOSEPT solution | | | | | | |
Day 1, 2 h | 32.4 ± 15.9 | 122.5 ± 33.5 | 47.9 ± 23.5 | 50.7 ± 24.0 | 25.3 ± 11.0 | 49.2 ± 18.3 |
Day 1, 8 h | 35.1 ± 20.5 | 126.0 ± 42.0 | 47.4 ± 23.8 | 54.3 ± 29.7 | 25.8 ± 11.2 | 49.9 ± 17.9 |
Week 2, 2 h | 34.5 ± 19.0 | 121.4 ± 39.6 | 49.6 ± 24.6 | 53.6 ± 33.5 | 26.2 ± 13.3 | 49.0 ± 18.8 |
There was a significant difference in corneal tactile threshold between visits (P = 0.003; ANOVA). At the 8-hour examination on day 1 and the 2-week visit after the study lens was dispensed, the tactile threshold was higher than that at the end of the no-lens-wear period (Tukey HSD, P = 0.003 and 0.012 for the 1-day, 8-hour and the 2-week visits, respectively). There was no significant change in conjunctival tactile thresholds between visits (P > 0.05, ANOVA).
For pneumatic mechanical thresholds, there were significant differences in both corneal and conjunctival thresholds between visits (P = 0.027 and 0.008 for cornea and conjunctiva, respectively, ANOVA). Corneal thresholds at the 2- and 8-hour examinations on day 1 were lower than those with habitual lens wear (Tukey HSD, P = 0.036 and 0.022 for the 2- and 8-hour examinations, respectively). The conjunctival threshold at the end of the no-lens-wear period was higher than measurements taken after 2 hours of PV lens wear at the 1-day and 2-week visits (Tukey HSD, P = 0.003 and 0.038 for 1-day, 2-hour and 2-week, respectively).
For corneal and conjunctival chemical thresholds, there were no significant differences between visits in phase I (both P > 0.05, ANOVA).
Corneal and conjunctival thresholds to tactile, pneumatic mechanical, and chemical stimulation for phase II are summarized in
Table 2.
Table 2. Corneal and Conjunctival Tactile, Pneumatic Mechanical, and Chemical Thresholds in Phase II
Table 2. Corneal and Conjunctival Tactile, Pneumatic Mechanical, and Chemical Thresholds in Phase II
Lens/Preservative | Tactile (mm/g/s) | Pneumatic Mechanical (Air Flow mL/min) | Chemical (%CO2 Added) |
Cornea | Conjunctiva | Cornea | Conjunctiva | Cornea | Conjunctiva |
PV+Polyquad/Aldox-preserved system | | | | | | |
Baseline (Aosept) | 35.5 ± 19.0 | 123.9 ± 33.3 | 50.4 ± 25.9 | 53.1 ± 32.3 | 26.9 ± 13.0 | 48.7 ± 20.2 |
Day 1, 2 h | 35.1 ± 17.2 | 125.6 ± 31.2 | 46.6 ± 25.7 | 51.4 ± 28.7 | 26.0 ± 12.9 | 51.2 ± 20.1 |
Day 1, 8 h | 34.2 ± 20.3 | 126.6 ± 32.6 | 45.0 ± 24.8 | 49.4 ± 28.6 | 25.6 ± 14.1 | 50.6 ± 21.3 |
Week 1, 2 h | 33.9 ± 18.1 | 120.7 ± 27.5 | 44.3 ± 21.7 | 50.1 ± 32.0 | 25.5 ± 13.4 | 50.5 ± 20.2 |
PV+PHMB-preserved system | | | | | | |
Baseline (Aosept) | 34.2 ± 16.5 | 121.9 ± 36.7 | 46.2 ± 26.5 | 48.5 ± 30.0 | 25.5 ± 12.9 | 49.6 ± 18.7 |
Day 1, 2 h | 34.6 ± 16.3 | 122.6 ± 35.1 | 44.6 ± 28.5 | 46.5 ± 27.5 | 22.9 ± 12.8 | 46.8 ± 21.2 |
Day 1, 8 h | 36.9 ± 18.6 | 121.8 ± 33.7 | 43.5 ± 24.8 | 45.5 ± 27.7 | 24.8 ± 12.8 | 49.0 ± 21.3 |
Week 1, 2 h | 36.0 ± 18.7 | 124.2 ± 33.4 | 45.6 ± 25.8 | 50.5 ± 35.5 | 22.1 ± 13.8 | 46.2 ± 22.4 |
Corneal chemical thresholds were significantly different between the MPSs, regardless of visit (P = 0.049, ANOVA). The threshold with the PHMB-preserved solution was lower than that of the Polyquad/Aldox system. There was a significant difference in corneal chemical thresholds between visits averaged across solutions (P = 0.043, ANOVA). The threshold at 1 week was lower than that at baseline (with Aosept; Tukey HSD, P = 0.025). However, the interaction between solution and visit was not significant (P > 0.05, ANOVA).
Conjunctival chemical threshold in phase II appeared to be lower with the PHMB-preserved solution than with the Polyquad/Aldox system. However, the difference was not statistically significant (P > 0.05, ANOVA). There was a significant interaction between solution and visit (P = 0.039, ANOVA). At the 2-hour visit on day 1 and the 1-week visit, there were significant differences between the two MPSs (Tukey HSD, both P = 0.049): lower threshold with the use of PHMB-preserved solution.
For tactile and pneumatic mechanical thresholds, there were no significant differences between solutions and visits and no solution and visit interactions in phase II (all P > 0.05, ANOVA).
The staining scores in phase I with the use of Aosept are shown in
Table 3. Grading of corneal and conjunctival staining for phase II (stratified by MPS) is presented in
Table 4.
Table 3. Grading of Corneal and Conjunctival Staining in Phase I
Table 3. Grading of Corneal and Conjunctival Staining in Phase I
Lens Wear | Cornea (Mean % Area ± SD) | Conjunctiva (Median of Sum; 10th, 90th Percentiles) |
Habitual | 7.5 ± 18.6 | 4.0 (1, 9) |
None | 1.6 ± 4.4 | 1.0 (0, 4) |
PV + Aosept | | |
Day 1, 2 h | 1.5 ± 2.2 | 2.5 (0, 7) |
Day 1, 8 h | 2.3 ± 3.6 | 3.0 (1, 7) |
Week 2, 2 h | 1.8 ± 2.9 | 3.0 (0, 6) |
Table 4. Grading of Corneal and Conjunctival Staining in Phase II
Table 4. Grading of Corneal and Conjunctival Staining in Phase II
Visit | Cornea (Mean % Area ± SD) | Conjunctiva (Median; 10th, 90th Percentiles) |
Polyquad/Aldox System | PHMB-Preserved System | Polyquad/Aldox System | PHMB-Preserved System |
Baseline | 1.3 ± 2.0 | 1.8 ± 3.0 | 2.0 (0, 6) | 3.0 (0, 6) |
Day 1, 2 h | 1.6 ± 2.4 | 29.5 ± 32.5 | 2.0 (0, 5) | 5.0 (1, 9) |
Day 1, 8 h | 5.7 ± 9.9 | 20.3 ± 25.1 | 3.0 (1, 5) | 4.5 (1, 8) |
Week 1, 2 h | 4.7 ± 7.5 | 49.9 ± 39.9 | 3.0 (1, 6) | 6.0 (2, 10) |
There were significant differences in corneal staining between solutions and visits (both P < 0.0001, ANOVA), and the differences between visits were dependent on the solution type (P < 0.0001, ANOVA). With the Polyquad/Aldox system, corneal staining at all follow-up visits was similar to baseline (with Aosept; Tukey HSD all P > 0.05), whereas corneal staining with the PHMB-preserved solution increased at all follow-up visits compared with baseline. The highest score was at the 1-week visit after 2 hours of lens wear (Tukey HSD all P < 0.003).
With the PHMB-preserved solution, conjunctival staining increased over time (Friedman P < 0.001) and was higher than that with the Polyquad/Aldox system at all visits, except baseline (all P < 0.001, Wilcoxon with Bonferroni correction). In addition, corneal staining correlated inversely with the conjunctival chemical thresholds at the 2- and 8-hour visits on day 1 (Pearson r= −0.34 and −0.37 for the 2- and 8-hour visits, respectively, both P < 0.05).
Contact lens wear affects the ocular surface in various ways,
2,34,35 and one of the effects has been to change the functioning of the sensory nerves of the ocular surface, as reflected in the response to corneal and conjunctival stimulations. Sensory nerves not only provide information about the relationship between the body and the external environment (e.g., potential dangers or injuries), but also signal the physiological condition of the body, such as local metabolism (e.g., acidic pH, hypoxia) and immune and hormonal activity.
36 The sensory impulses arising from the ocular surface are one of the important aspects of the integrated functional unit and play a part in maintaining the homeostatic environment of the ocular surface.
6
In the present study, conjunctival sensitivity to pneumatic mechanical stimulation increased from the no-lens baseline after the refitting of the SH lenses and a similar trend but smaller magnitude was observed in the cornea, suggesting an altered sensory processing of the ocular surface. This increase in sensitivity appeared to be transient (i.e., greater at the 2-hour and day-1 visits for conjunctiva and cornea, respectively). Stapleton et al.
31 compared the effects of short-term SH and hydrogel lens wear on ocular surface sensitivity in neophytes and found that conjunctival sensitivity increased after the wear of SH lenses. The mechanism that leads to this increase in sensitivity to mechanical stimuli is unclear. Contact lenses applied to the eye can introduce stimulation due to friction on the ocular surface, particularly during the initial phase of wear.
37 This mechanical effect may be greater with SH lens materials because of their relative high elastic modulus.
37 In addition, the structure and physiology of the tear film can be altered during lens wear.
2,34,35,38 The combination of these factors may lead to an increase in sensory input from the ocular surface, signaling the temporary disequilibrium between the components of the functional unit. Moreover, subclinical conjunctival inflammation that has been detected in asymptomatic contact lens wearers
39 is likely to be another contributing factor, as the activities of the sensory nerve could be modified by injury and inflammatory mediators.
4
The change in mechanical sensitivity with SH lens wear in the present study was more pronounced in the conjunctiva. The conjunctiva is not only supplied by free nerve endings to provide sensory input to the functional unit, as in the cornea, but is also a highly reactive tissue.
40 It is richly supplied by blood vessels, connected to the lymphatic system, and filled with immunocompetent cells.
40 In addition, the conjunctiva is directly involved in regulating the secretion of components of the tear film, such as electrolytes, water, and mucin,
41,42 and has a large area covering the ocular surface. It is conceivable that the conjunctiva would be sensitive to any changes occurring on the ocular surface, including the effects of contact lens wear.
In comparison, corneal tactile sensitivity decreased after SH lens wear from the no-lens-wear baseline, which is similar to previous reports with hydrogel lens wear.
16,21,23 The reduction in corneal sensitivity with low-Dk hydrogel lens wear has been attributed to metabolic impairment resulting from lens-induced hypoxia.
21,25 SH lens materials have diminished the lens-induced hypoxic impact on corneal physiology, with resulting improvement of a number of physiological markers, especially when lenses are worn overnight.
43 –45 If hypoxia and mechanical stimulation are responsible for lens-induced alteration in sensitivity but hypoxia effects have been minimized, the mechanism reducing tactile sensitivity with SH lens wear may be due to mechanical stimulation. Adaptation, the reduction in sensitivity after repeated suprathreshold stimulation, is a phenomenon that occurs in a variety of forms in sensory systems.
46 –50 Functionally, adaptation may help to optimize the dynamic range of encoding in a neural system by shifting the sensitivity,
51 –53 and it develops and recovers depending on the time course of the stimulation.
54,55 Although subjects in the present study were adapted daily contact lens wearers, temporary cessation of lens wear and therefore the withdrawal of the mechanical stimulus may have allowed some recovery and therefore no adaptation to the lenses, as shown in
Table 1. After the subject was refitted with SH lenses, the close interaction between the lens and the cornea produced a sustained stimulation of the surface and adaptation, resulting in a shift in corneal sensitivity to the tactile stimuli.
In addition to the changes in pneumatic mechanical and tactile sensitivity with SH lens wear, ocular surface chemical sensitivity seemed to reflect the interaction between SH lens materials and MPS in this study. The higher levels of ocular surface staining induced by the solution preserved with PHMB were accompanied by an increased chemical sensitivity. This result is different from the decreased tactile sensitivity in solution-induced staining reported by Epstein.
24 Comparison between the two studies is difficult to make, because of differences in study design, lens materials, and stimulus modality used. Even though the exact mechanism causing lens–solution interaction–induced staining remains unclear, the presence of this type of staining indicates an alteration in the ocular surface (Muya L, et al.
IOVS 2008;49:ARVO E-Abstract 4869) that may have an impact on the sensory input (or vice versa) to the functional unit. This increased chemical sensitivity suggests sensitization of the polymodal nociceptors (or chemoreceptors
56 ) of the ocular surface. Sensitization, a reduction in threshold to one or more stimulus modalities and/or the development of lower-frequency spontaneous activity,
57 can be caused by repeated noxious stimuli, such as inflammatory mediators on the ocular surface.
4,58 This sensitization makes the polymodal receptors excitable by non-noxious stimuli and enhances the magnitude of their response to noxious stimuli.
57 Many components in ophthalmic solutions have the potential to induce inflammatory changes on the ocular surface,
59 and in the present study, the combination of lens and solution may have acted as exogenous chemical stimuli of the ocular surface, triggering a cascade of events and resulting in enhanced responsiveness of the polymodal receptors. More studies are warranted to explore the pathway and mechanism of sensitivity changes induced by this lens–solution interaction.
In the present study, we found different profiles of the changes in ocular surface sensitivity with lens wear and lens–solution combinations, when measuring with different stimuli from the two esthesiometers used. The basis for the discrepancies between the two types of sensitivity measurement is not fully understood. It has been suggested that the stimulus delivered by the Cochet-Bonnet esthesiometer selectively activates mechanonociceptors that are responsible for the acute, sharp pain induced by direct mechanical contact with the ocular surface.
4,60 On the other hand, the pneumatic stimulation may activate both mechano- and polymodal nociceptors. More recently, it has been proposed that a series of labeled lines consisting of neurons in the spinothalamic tract (the pain-signaling pathway to the brain) signal the homeostatic state of the body (interoception), whereas the pathways processing exteroceptive information (i.e., potentially dangerous stimuli impinging on our body) associated with pain may be different.
61 In addition, studies have suggested that cornea-responsive neurons in the spinal trigeminal nucleus, the subnucleus interpolaris/caudalis transition (Vi/Vc) and the subnucleus caudalis/upper cervical cord transition (Vc/C1), process corneal input differently, and perhaps serve different ocular nociception functions.
62 –64 Therefore, in addition to being related to differences between the two esthesiometers such as stimulus modality, this dissimilarity between tactile and pneumatic mechanical sensitivity may reflect the contribution of different nociceptive signaling pathways; the tactile sensitivity may represent the alarm-alerting functions to exteroceptive stimuli, whereas pneumatic mechanical and chemical sensitivity may provide additional interoceptive information such as changes in the ocular surface and tear film. In addition, since Cochet-Bonnet esthesiometry was always performed after Belmonte esthesiometry, because of its invasiveness, it is possible, although it seems unlikely, that the order of the testing influenced the different profiles measured with the two instruments.
In summary, despite the advances of SH lens materials in eliminating lens-induced hypoxia, this study has demonstrated that SH lens wear and lens–solution interactions affect ocular surface sensitivity. The effects of lens wear on the sensory nerves function appears to be more complex than previously thought. Given that the sensory input arising from the ocular surface plays a critical role in maintaining the optimally balanced state of the lacrimal functional unit, sensitivity measures may be considered as an indicator of the subclinical changes induced by SH lens wear. However, the present study has some limitations, partially because the lens wear period was relatively brief. Questions, such as how ocular surface sensitivity, particularly conjunctival sensitivity, varies after longer-term SH lens wear; whether sensory aspects of the ocular surface contribute to the end-of-day discomfort that is commonly reported by lens wearers, including SH lenses wearers; and the clinical implications of the changes in sensitivity, remain unanswered. The Belmonte pneumatic esthesiometer does provide uniquely useful information, however, when used in assessing different aspects of sensory functioning in contact lens wearers.
Supported by a grant from Alcon Laboratories, Inc., and operating and equipment grants from the National Sciences and Engineering Research Council (NSERC) Canada and the Canada Foundation for Innovation (CFI), respectively.
Disclosure:
P. Situ, Alcon Laboratories (F);
T.L. Simpson, Alcon Laboratories (F);
L.W. Jones, Alcon Laboratories (F);
D. Fonn, Alcon Laboratories (F)