Cup shaped electrodes of Ag/AgCl were fixed with collodion in the following positions: active electrode in Oz,
58 reference electrode in Fpz,
58 and ground in the left arm. Interelectrode resistance was kept below 3000 Ω. The bioelectric signal was amplified (gain 20,000), filtered (band-pass 1–100 Hz) and averaged (200 events free from artifacts were averaged for every trial) by the electro-oculography system (BM 6000; Biomedica Mangoni). Analysis time was 250 ms. The transient VEP response was characterized by several waves with three subsequent peaks of negative, positive, and negative polarity. In visually normal subjects, these peaks have the following implicit times: 75, 100, and 145 ms (N75, P100, N145).
During a recording session, simultaneous VEPs and PERGs were recorded at least twice (2–6 times) and the resulting waveforms were superimposed to check the repeatability of results. For all PERGs and VEPs, implicit times and peak-to-peak amplitudes of each of the averaged waves were directly measured on the displayed records by means of a pair of cursors.
On the basis of previous studies (i.e., Parisi et al.
49 ), we know that intraindividual variability (evaluated by test–retest) is approximately ±2 ms for PERG P50 and VEP P100 implicit times and approximately ±0.25 μV for PERG P50 to N95 and VEP N75 to P100 amplitudes. During the recording session, we considered as superimposable and therefore repeatable two successive waveforms with a difference in milliseconds (for PERG P50 and VEP P100 implicit times) and in microvolts (for PERG P50-N95 and VEP N75–P100 amplitudes), less than the values for intraindividual variability. Sometimes the first two recordings were sufficient to obtain repeatable waveforms; other times, however, further recordings were required (but never >6 in the cohort of patients or control subjects). For statistical analyses, we included PERG and VEP values measured in the recording with the lowest PERG P50 to N95 amplitude.
In each subject or patient, the signal-to-noise ratio (SNR) of PERG and VEP responses was assessed by measuring a noise response while the subject fixated an unmodulated field of the same mean luminance as the stimulus. At least two noise records of 200 events each were obtained, and the resulting grand average was used for measurement. The peak-to-peak amplitude of this final waveform (i.e., average of at least two replications) was measured in a temporal window corresponding to that at which the response component of interest (i.e., VEP N75-P100, PERG P50-N95) was expected to peak. SNRs for this component were determined by dividing the peak amplitude of the component by the noise in the corresponding temporal window. An electroretinographic noise <0.1 μV (mean, 0.082 μV; range, 0.064–0.092 μV; resulting from the grand average of 400–1200 events), and an evoked potential noise <0.15 μV (mean, 0.094 μV; range, 0.075–0.118 μV, resulting from the grand average of 400–1200 events) was observed in all subjects tested. In all subjects and patients, we accepted VEP and PERG signals with a signal-to-noise ratio >2.