Video recordings were analyzed with a custom program that detected the occurrence and duration of blinks (recorded at lower left of images). To display weak fluorescence, the range of recorded intensity on the displayed image could be varied, and this was recorded at the lower right of the images (e.g.,
Fig. 2). To improve signal-to-noise ratio, images could be averaged after alignment of a selected area (typically 50 × 100 pixels) of conjunctiva based on a cross-correlation algorithm.
18 The conjunctiva was used for this alignment, partly because the fluorescence image over the conjunctiva has a higher contrast initially than that over the cornea, particularly for low fluorescein concentration (e.g.,
Fig. 3A). The image pattern over the conjunctiva is probably largely due to surface roughness of the conjunctival surface (this would explain the high contrast over the conjunctiva in
Fig. 3A, compared to the low contrast over the relatively smooth cornea). As expected from this, the conjunctival image was also found to show less change than the corneal image (compare
Figs. 2A and
2B). A pixel-by-pixel ratio of two images (or two averages of images, e.g.,
Fig. 2C) could be displayed, again after alignment of the images or averages with the cross-correlation algorithm. Thus the ratio image,
R(
x,
y), was given by
where
x and
y are horizontal and vertical position in pixels,
I 1 and
I 2 are intensities of the first and second images, and
dx and
dy are the horizontal and vertical shifts needed to bring the first image into alignment with the second image. The ratio image thus shows the fractional dimming of fluorescence at every position of the exposed corneal surface. For the low concentration condition, in which quenching is unimportant, any dimming should be largely due to divergent tangential flow. For the high concentration condition, in addition to the effect of any tangential flow, evaporation should cause dimming due to quenching from increasing fluorescein concentration.