Abstract
Purpose.:
We compared the subfoveal choroidal thickness (SFCT) measured on the images obtained by two spectral-domain optical coherence tomographic (SD-OCT) instruments and one swept-source OCT (SS-OCT) instrument.
Methods.:
A cross-sectional, prospective noninterventional study was done in which SFCT was measured in the images obtained by two SD-OCT instruments; Heidelberg Spectralis-OCT (Spectralis-SD-OCT) and Topcon 3D OCT-1000 Mark II (Topcon-SD-OCT). Images also were obtained with SS-OCT Atlantis DRI OCT-1 (DRI-SS-OCT). After manual segmentation, the measurements were made using the calipers embedded in each instrument. The intrarater, interrater, and intermachine agreements were assessed.
Results.:
We studied 35 subjects. The intrarater correlation coefficient (95% confidence interval) was 0.994 (0.988–0.994) for Spectralis-SD-OCT, 0.996 (0.993–0.998) for Topcon-SD-OCT, and 0.997 (0.991–0.998) for DRI-SS-OCT (P < 0.001). The interrater correlation coefficient was 0.995 (0.991–0.998) for Spectralis-SD-OCT, 0.995 (0.990–0.998) for Topcon-SD-OCT, and 0.996 (0.992–0.998) for DRI-SS-OCT (P < 0.001). The average SFCT was 273.2 μm with Spectralis-SD-OCT, 269.1 μm with the Topcon-SD-OCT, and 280.5 μm with DRI-SS-OCT. The intermachine correlation coefficient was 0.982 (0.964–0.991) for Spectralis-SD-OCT versus Topcon-SD-OCT, 0.907 (0.815–0.953) for Topcon-SD-OCT versus DRI-SS-OCT, and 0.911 (0.832–0.954) for DRI-SS-OCT versus Spectralis-SD-OCT (P < 0.001). The SFCT measured with DRI-SS-OCT was significantly thicker than that with Topcon-SD-OCT, with a mean difference of 11.41 ± 30.27 μm (P = 0.032).
Conclusions.:
In normal adult eyes, there was good reproducibility and repeatability of SFCT measurements obtained by the SD-OCT and SS-OCT instruments. However, the choroid measured with DRI-SS-OCT was thicker than that measured with both SD-OCT instruments, and, thus, the choroidal thickness should not be compared between the SD-OCT and SS-OCT instruments. (www.umin.ac.jp/ctr number, UMIN000011259.)
This was a prospective, cross-sectional study that was performed at a single institution from April 30 to May 30, 2013. Approximately 40 volunteers were projected to enroll in the study, and all subjects had a complete ocular examination, including the measurement of the refractive power (spherical equivalent) with an autorefractor keratometer (RM8900; Topcon, Tokyo, Japan), best-corrected visual acuity (BCVA), IOP with a pneumotonometer (CT-80; Topcon), slit-lamp biomicroscopy, and dilated funduscopy. The axial length was measured with optical interferometry (OA-1000; Tomey, Tokyo, Japan). Only the right eye was measured with the three OCT instruments.
The inclusion criteria were an age of ≥20 years and <50 years, BCVA of ≥20/20, and normal fundus by ophthalmoscopy and OCT. The exclusion criteria were history of ocular and systemic diseases, prior ocular surgery or intraocular injections, and high myopia ≥6.0 diopters (D).
All eyes were examined without mydriasis and two separate measurements were made on each eye with each of the three instruments by a single trained examiner. The order of the OCT instrument measurements was random for the patients, and the order for the second OCT examinations also was random. Because of the significant diurnal fluctuations of the choroidal thickness, all OCT recordings on a single patient were made within 1 hour on the same day.
We screened 43 Japanese volunteers, but six eyes were excluded because of high myopia, one eye because of a history of intraocular surgery, and one because of poor OCT quality. In the end, we studied 35 eyes in 18 men and 17 women. Clear choroidal images were obtained from all of the eyes from each of the three OCT instruments. The mean ± SD age of all of the volunteers was 32.5 ± 6.5 years, with a range of 22 to 47 years. The mean refractive error (spherical equivalent) was −2.6 ± 1.8 D, with a range from −5.75 to −0.75 D. The mean IOP was 13.9 ± 2.0 mm Hg, with a range from 10 to 19 mm Hg, and the mean axial length was 24.7 ± 1.3 mm, with a range from 22.51 to 27.68 mm.
Representative images obtained from each of the three OCT instruments are shown in
Figure 1. The subfoveal choroidal thickness measured with the Spectralis SD-OCT was 273.20 ± 70.03 μm, compared to 269.1 ± 70.54 μm with the Topcon SD-OCT, and 280.5 ± 77.01 μm with the DRI SS-OCT. The differences in the mean choroidal thickness were not significant. There was no significant difference of either the nasal or temporal choroidal thicknesses (
Table 2).
Table 2 Choroidal Thickness Measurements by Three Different OCT Machines
Table 2 Choroidal Thickness Measurements by Three Different OCT Machines
Location | Choroidal Thickness Measurements, Average ± SD, μm | 1-Way ANOVA With Tukey's Post-Test |
Spectralis SD-OCT | Topcon SD-OCT | DRI SS-OCT | P |
Nasal | 256.3 ± 70.49 | 255.5 ± 70.67 | 262.8 ± 74.47 | >0.05 |
Subfoveal | 273.2 ± 70.03 | 269.1 ± 70.54 | 280.5 ± 77.01 | >0.05 |
Temporal | 258.6 ± 70.38 | 256.1 ± 63.86 | 272.3 ± 75.31 | >0.05 |
The choroidal thicknesses were measured on the images obtained by the commonly used SD-OCT instruments, Spectralis SD-OCT and Topcon SD-OCT, and the newly developed DRI SS-OCT in normal subjects. Our results showed that the two separate measurements had good reproducibility and repeatability with each instrument.
We found that the coefficient of correlation between any two of the three OCT instruments was high, and significant for the subfoveal and parafoveal choroidal thickness measurements. The choroid measured with the DRI SS-OCT was significantly thicker than that with the Topcon SD-OCT (280.5 ± 77.01 μm with DRI SS-OCT versus 269.1 ± 70.54 μm with Topcon SD-OCT, P = 0.032).
The Bland-Altman analyses also showed that the mean choroidal thickness measured by SS-OCT was greater than that by either of the SD-OCT instruments. We did not determine the exact reason for this observation, but the raters noted that the outer border of the choroid was more distinct, probably because SS-OCT can record choroidal images deeper than the SD-OCT (
Fig. 3). In many cases, the sclera-choroidal border in the SD-OCT images may not necessarily be the true border.
The largest difference in the choroidal thickness was 43.5 μm (−28.5–43.5 μm) for Spectralis SD-OCT versus Topcon SD-OCT, 148 μm (−148.0–24.0 μm) for Topcon SD-OCT versus DRI SS-OCT, and 140 μm (−19.0–140.0 μm) for DRI SS-OCT versus Spectralis SD-OCT. By eliminating the one outlier from the DRI SS-OCT measurements, these differences were reduced to 43.5, 75, and 91 μm, respectively. In an earlier analyses of choroidal thickness with different SD-OCT instruments, the largest difference was approximately 90 μm.
18,20 Therefore, it is possible to say that the difference in the choroidal thickness measured by SD-OCT or SS-OCT was not significant. The slight differences in the choroidal thickness determined by different SD-OCT or SS-OCT instruments probably are not important for clinical trials at present. However, when detailed analysis is necessary, this difference might be more important.
Of importance was the presence of an outlier in the SS-OCT data. In our earlier study comparing three different SD-OCT instruments in 43 normal subjects, the choroidal thickness measurements were not significantly different among all of the SD-OCT instruments, and no outliers were observed.
18 In the SD-OCT images, the choroidal border appeared as a linear structure (as indicated by the arrows in
Figs. 3A,
3B), but in the SS-OCT images, another layer of tissue was observed outside of that border. This additional tissue made the choroid thicker. We were not able to determine why this eye showed the different images with the two types of OCT instruments. Because the images of choroids more than 500 μm did not have a clear choroidal boundary with any SD-OCT instrument, the choroidal thickness might be the factor.
18 However, the choroidal thickness in this case was less than 500 μm; thus, unidentifiable factors, such as pigmentation or vascular structures, might have caused the inconsistent results between the SD-OCT and SS-OCT instruments. Considering the image quality, the image obtained by the SS-OCT likely was to be the true choroidal boundary more than the SD-OCT. Thus, the SS-OCT instrument might be better for evaluating choroidal thickness than the SD-OCT instruments.
The design of our experiments did not allow us to determine the sensitivity of each instrument. In an earlier report, the subfoveal choroidal thickness could be measured in all the subjects even from different ethnic groups,
22 and clear images of the choroid could be obtained from 39 of 43 eyes with all of the SD-OCT instruments.
18 We measured a similar population in this study, and we were able to record clear images from all eyes. To compare the sensitivity of different instruments correctly, interinstrument comparisons should be done on a larger number of eyes and by more graders.
There are strengths of the study, one of which was its prospective design, which allowed us to recruit a number of healthy subjects with a wide range of choroidal thicknesses. Scans of each eye were performed within a limited time, which minimized the possibility of choroidal thickness changes caused by circulating catecholamines, diurnal variations, or fluctuations of the intraocular pressure.
23,24
There also are several limitations of this study. Scans were acquired from comparatively young subjects with no ocular pathology, and the values may not reflect those of patients seen in a routine outpatient setting. For example, the clarity of images of patients could be hindered by an ocular pathology, such as significant media opacity, or masking of choroidal reflectance by intraocular tissue. These issues, especially the age differences, should be remembered in interpreting our results. Second, we used manual segmentation, and this always is a concern because of uncontrollable bias among examiners. We performed interrater comparisons which were found to be high. However, the raters were well-trained and experienced. If the agreement ratio of many different raters is proved to be high on each device, the present results should be more generalized and the present instruments could be used interchangeably with confidence. Third, we only evaluated the specific scanning protocol of two SD-OCT instruments commonly used and one SS-OCT. It is not certain whether our observations can be generalized to other SD-OCT instruments or other scanning protocols. In addition, our results apply only to the choroidal thickness of the subfoveal or parafoveal areas. We do not have any evidence of the choroidal thickness of other areas. This also should be remembered when interpreting the present data.
In conclusion, the subfoveal choroidal thickness measurements obtained with the Spectralis SD-OCT and with the Topcon SD-OCT were significantly correlated with values obtained by the DRI SS-OCT in most of the cases. Although the difference might be clinically insignificant, it should be noted that the choroidal thickness measured with the DRI SS-OCT was thicker than that obtained with the SD-OCT instruments and some of them reached the statistical significance. Because OCT is an easy, reproducible, and noninvasive examination, it will be used more often in a clinical setting. Our results should be indispensable to interpreting the data of each SD-OCT or SS-OCT instruments and a better understanding of ocular diseases, which will further increase the value of this examination.
Supported by a grant from the Research Committee on Chorioretinal Degeneration and Optic Atrophy, Ministry of Health, Labor, and Welfare, Tokyo, Japan; and by a Grant-in-Aid for Scientific Research from the Ministry of Education, Science, and Culture of the Japanese Government. The authors alone are responsible for the content and writing of the paper.
Disclosure: Y. Matsuo, None; T. Sakamoto, None; T. Yamashita, None; M. Tomita, None; M. Shirasawa, None; H. Terasaki, None