The efficiency of CXL decreased significantly as UV-A light irradiances increased from 3 to 18 mW/cm
2. Indeed, corneas treated with the highest tested irradiance (18 mW/cm
2 for 5 minutes) had stiffness that was indistinguishable from untreated controls (
Table 2). Higher light irradiances were associated with lower Young's modulus at each percentage strain tested.
Wernli et al.
20 evaluated Young's modulus using the same total energy fluence and riboflavin concentration as in our study. They also observed a decrease in Young's modulus for high irradiances, but only at irradiances exceeding 50 mW/cm
2. These differences might be explained by several factors. First, the groups had different sizes (10 vs. 50 eyes/group), second, the biomechanical measurements were performed at different times; Wernli and colleagues
20 took measurements at 30 minutes after starting irradiation, regardless of irradiation time. By contrast, we consistently performed measurements at 30 minutes after the end of irradiation. Another difference is that Wernli and colleagues
20 kept corneas immersed in the riboflavin solution. This extended exposure to riboflavin likely increased the amount of riboflavin penetration and subsequent different cross-linking activity.
Also, we observed a Young's modulus that was approximately a factor 2 larger than in the Wernli study. Several factors might be responsible for these differences. First, the machines for biomechanical measurements were not the same (Zwick Z 0.5 versus MINIMAT; Stretton Shropshire) and second, the methods were slightly different (time before biomechanical testing, length of the corneal strips 10 vs. 7 mm). Other, yet unidentified aspects might have further influenced the differences observed.
Lastly, the Wernli study
20 was performed using a beam-optimized device (UV-X 2000; IROC Innocross, Zurich, Switzerland). This device tends to deliver a more homogeneous energy profile to the cornea.
21 In our experiments, a device delivering a less homogeneous distribution of energy with respect to corneal curvature was used (CXL 365 Vario; SCHWIND eye-tech-solutions GmbH & Co., Kleinostheim, Germany). One might speculate that the differences between the studies might be due to this variation in energy distribution. We do not believe that this is the case: the main interest in both studies was to assess relative differences in the cross-linking effect between the current gold standard (3 mW/cm
2 for 30 minutes) and accelerated settings.
In a recent study, Beshtawi et al.
22 analyzed ex vivo human corneas using Scanning Acoustic Microscopy (SAM) to determine stiffness following irradiation at 3 and 9 mW/cm
2. Similar to our results, they found a significant increase in stiffness at both settings when compared with controls. In contrast to our findings, they did not see significant differences between both settings. Several factors might explain this discrepancy: the tissues were different between the Beshtawi study (human corneas) and our experiments (porcine corneas). Also, we performed stress-strain measurements, whereas Beshtawi and colleagues
22 used SAM. Without a doubt, the 9 mW/cm
2 for 10 minutes setting provides cross-links to the cornea and clinical validation is needed to better understand the results of both studies.
Oxygen levels in the cornea are related to the oxygen diffusion flux and local oxygen uptake.
23 Corneal oxygen levels decrease during CXL, presumably due to the transformation of oxygen into reactive oxygen species.
24 The reactive species are thought to catalyze the creation of covalent bonds between collagen and proteoglycan molecules, stiffening the cornea.
5 Oxygen seems to be essential to this process and is probably the rate-limiting substrate in the photochemical reaction. We have previously shown that corneas treated in a low-oxygen state using an irradiance of 9 mW/cm
2 for 10 minutes exhibit a Young's modulus similar to that of untreated controls.
19 High UV-A irradiances would be expected to have higher oxygen usage rates. If oxygen conversion to free radicals outpaces oxygen replenishment by diffusion, the local oxygen levels would fall and collagen cross-linking would be compromised.
24 This would result in lower measured Young's modulus. Our findings support this hypothesis and are in agreement with previously reported data.
24 Alternatively, other yet unknown mechanisms might also contribute to the biomechanical results observed.
In conclusion, we report a steady and significant decline in the biomechanical response (stiffening) of ex vivo corneas with increasing irradiance and decreased treatment times. This may indicate that the Bunsen-Roscoe law knows limitations in an in vivo setup: and cannot be simply applied to the cornea. Whether or not the decline in biomechanical stiffness will be clinically relevant remains to be validated in clinical trials using high-irradiance CXL.