Corneal cells in vivo are exposed to biophysical cues intrinsic to the extracellular matrix, including topography and compliance.
9,19,20 Increasing numbers of publications from our laboratory and others have documented several cell types to be highly influenced by biophysical cues resulting in changes in their cellular behavior, including cell shape, proliferation, adhesion, and migration, as well as gene expression.
11,17,21–24 Previous studies
8,25 from our laboratory detailed the impact of substratum topographic cues on the transition of corneal fibroblasts to myofibroblasts. We have recently profiled the elastic moduli of human corneal stromal elements
9 informing the subsequent design and fabrication of substrates possessing biologically relevant values for substratum compliance. The mean ± SD elastic modulus of the anterior stroma of the human cornea was measured using AFM as 33.1 ± 6.1 kPa
9 ; thus, the 28-kPa gels used in the present study approximated the compliance of the native human corneal stroma. The 4-kPa gels most closely mimic the rabbit corneal stroma, which is markedly more compliant than the human corneal stroma, with a mean ± SD elastic modulus of 1.1 ± 0.58 kPa as reported by Thomasy et al.
13 It should be noted that other research groups have documented notably different values for the compliance of the corneal stroma. For instance, Winkler et al.
26 measured elastic moduli ranging from 892 to 1861 Pa for the anterior cornea, while Lombardo et al.
27 reported values ranging from 1.14 to 2.63 megapascal. As previously detailed,
9,28 several factors likely account for these differences, including sample preparation, the specific location within the corneal tissue (e.g., anterior versus posterior stroma), and, most importantly, differences in the techniques used for determining the compliance values. For a more detailed presentation of the differences in techniques utilized to obtain elastic modulus values in tissues, we refer the reader to a review article by McKee et al.
28 In the present study, the compliance measurements were internally consistent in that the same technique, AFM, was used to determine the elastic modulus of the cornea
9,13 and of the compliant substrates.
11,12