June 2013
Volume 54, Issue 15
Free
ARVO Annual Meeting Abstract  |   June 2013
Predicted changes in the cut surface and weight of colored acrylic intraocular lenses in 20 years
Author Affiliations & Notes
  • Kenji Kawai
    Ophthalmology, Tokai Univ School of Medicine, Isehara, Japan
  • Footnotes
    Commercial Relationships Kenji Kawai, Alcon (F), KOWA (F), HOYA (F), Santen (F), NIDEK (F)
  • Footnotes
    Support None
Investigative Ophthalmology & Visual Science June 2013, Vol.54, 844. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Kenji Kawai; Predicted changes in the cut surface and weight of colored acrylic intraocular lenses in 20 years. Invest. Ophthalmol. Vis. Sci. 2013;54(15):844.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: We previously reported that long-term immersion of intraocular lenses (IOLs) in water at high temperature leads to their opacification and deterioration, depending on differences in materials. This study aimed to examine predicted changes in the cut surface and weight of colored acrylic IOLs (colored IOLs) in 20 years.

Methods: Colored IOLs manufactured by various companies were immersed in 50-mL screw-tube bottles containing ultrapure water at 100 degrees for 115 days (equivalent to 20 years). External appearance and cut surfaces were observed before and after drying of the IOLs, and then percent changes in weight before and after treatment were determined. IOLs used: We used 3 colored IOLs of each of the following 5 types manufactured by different companies - SN60WF (Alcon) and AN6K (KOWA) by the cast-molding method and NY-60 (HOYA), NX-60 (Santen), and NS-60YG (NIDEK) by the lathe-cut method.

Results: External appearance: Opacification was observed to a varying degree in all IOLs except AN6K and NX-60. After drying at a room temperature of 25 degrees for 48 hours, opacification disappeared in all IOLs except SN60WF. Cut surfaces: Opacification was observed to a varying degree for all IOLs. After drying, opacification was observed in all except NS-60YG. Percent changes in weight: Percent changes in weight before drying were unmeasurable for SN60WF, -3% for AN6K, -0.9% for NY-60, +2.3% for NX-60, and +0.6% for NS-60YG. The changes after drying were -1.5% for SN60WF, -9.8% for AN6K, -2.7% for NY-60, -3.7% for NX-60, and -1.5% for NS-60YG.

Conclusions: There were differences in opacification among cut surfaces of IOLs. Percent changes in weight before and after drying differed by approximately 6% for AN6K and NX-60. These differences may be attributable to different methods of manufacturing acrylic resin.

Keywords: 567 intraocular lens • 467 clinical laboratory testing • 684 refractive surgery: optical quality  
×
×