February 2002
Volume 43, Issue 2
Free
Cornea  |   February 2002
Estimation of Human Corneal Oxygen Consumption by Noninvasive Measurement of Tear Oxygen Tension While Wearing Hydrogel Lenses
Author Affiliations
  • Joseph A. Bonanno
    From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
  • Thomas Stickel
    From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
  • Tracy Nguyen
    From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
  • Trina Biehl
    From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
  • Donna Carter
    From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
  • William J. Benjamin
    Eye Physiology and Ocular Prosthetics Laboratory, School of Optometry, University of Alabama at Birmingham, Birmingham, Alabama.
  • P. Sarita Soni
    From the Borish Center for Ophthalmic Research, Indiana University, School of Optometry, Bloomington, Indiana; and the
Investigative Ophthalmology & Visual Science February 2002, Vol.43, 371-376. doi:
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Joseph A. Bonanno, Thomas Stickel, Tracy Nguyen, Trina Biehl, Donna Carter, William J. Benjamin, P. Sarita Soni; Estimation of Human Corneal Oxygen Consumption by Noninvasive Measurement of Tear Oxygen Tension While Wearing Hydrogel Lenses. Invest. Ophthalmol. Vis. Sci. 2002;43(2):371-376.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

purpose. To devise a procedure for direct estimation of corneal oxygen consumption in human subjects.

methods. Tear oxygen tension (Po 2) was measured at the posterior surface of two standard hydrogel contact lenses (38% water, 0.2 and 0.06 mm thick, oxygen transmissibility [Dk/t] = 4.2 and 14 × 10−9 cm · mL O2/mL · sec · torr) and one newly available hydrogel-silicone polymer lens (Dk/t = 99 × 10−9). The oxygen-sensitive dye, Pd-meso-tetra (4-carboxyphenyl) porphine, bound to bovine serum albumin, was incubated with the lenses overnight. The lenses, coated with the protein–dye complex, were placed on four subjects’ eyes, and tear Po 2was measured in the open eye and after 5 minutes of eye closure, using a time–domain phosphorescence measurement system. Given the tear po 2, lens Dk/t, and corneal thickness, oxygen consumption (Q C, in mL O2/cm3 · sec) could be calculated from established oxygen diffusion models.

results. Protein-dye complex bound to the lens surface enabled reporting of tear po 2 for long periods. As expected, estimated tear po2 was higher in subjects wearing lenses with higher Dk/t: mean open-eye Po 2 = 30.6 ± 3.1 and 8.1 ± 1.3 torr for the thin and thick hydrogel lenses, respectively, and 97.6 ± 22.9 torr for the hydrogel-silicone lens. After 5 minutes of eye closure, tear Po 2 was significantly reduced and reached a new steady state in approximately 20 seconds after eye opening. Fitting a single exponential model to the data and extrapolating to t = 0 provided an estimate of po 2 under the closed lid for the thin hydrogel (Po 2 = 7 ± 2.3 torr) and the hydrogel-silicone lens (Po 2 = 22.6 ± 4 torr). After 5 minutes of eye closure with the thick hydrogel lens, tear Po 2 remained constant for ∼10 seconds after eye opening (mean Po 2 = 3.9 ± 0.7) before increasing to a new steady state. This delay could be accounted for by the time needed for oxygen to diffuse to the posterior surface of the lens. Calculated Q C ranged from 2.2 × 10−4 to 3.7 × 10−6 mL O2/cm3 · sec) at the highest and lowest po 2s, respectively, and is comparable to previous in vitro and in vivo estimates.

conclusions. Tear Po 2 behind hydrogel lenses can be measured in human subjects using the phosphorescence of the porphyrin-protein complex bound to the lens surface. The method is simple, fast, reliable, and noninvasive, allowing quick and direct estimates of Q C. In addition to contact lens wear, this method should be useful for examining the effects of disease, surgery, or topical drugs on the corneal oxygen consumption rate.

Assessment of metabolic activity in vivo by noninvasive techniques is a desirable approach for studying the normal physiology of ocular tissues and how it is altered by disease and drugs, surgery, or other interventions. Fluorescence- and phosphorescence-based techniques offer high sensitivity and are applicable to studying the physiology of many ocular structures that are optically accessible. Autofluorescence of naturally occurring substances (e.g., the reduced nicotinamide adenine dinucleotide-to-nicotinamide adenine dinucleotide [NADH/NAD] ratio) can detect the tissue’s metabolic state 1 and is advantageous, because addition of exogenous agents is not required. Unfortunately, the autofluorescence signal is typically very weak or may require ultraviolet illumination, but two-photon techniques could circumvent the radiation hazard. 2 Fluorescent and phosphorescent dyes are more sensitive, but must be delivered in usable concentrations to the site of interest and/or may have toxic interactions, which could limit clinical applicability. 
Previously, we have reported the use of a phosphorescence-quenching technique to measure the tear po 2 beneath contact lenses in rabbits. 3 This method has also been used to determine anterior chamber 4 and retinal vasculature 5 po 2. Although it is of interest to contact lens researchers and manufacturers to use tear po 2 to assess lens performance, this approach can also be used to measure corneal oxygen consumption (Q C) in vivo. Clinical response to contact lens wear and laboratory hypoxia-induced corneal swelling studies have hinted that there is a wide variability in corneal oxygen demand in the normal young population. 6 Furthermore, it has been shown that oxygen uptake into corneas of diabetic rabbits 7 and humans 8 is suppressed. Thus a sensitive, easily administered, and quick measurement of Q C could be useful in studying the effects of disease, surgery, topical drug use, or contact lens wear on the metabolic status of the cornea. 
The steady state tear Po 2 under a contact lens is determined primarily by Q C. 9 10 Thus, given the tear Po 2 under a contact lens of known oxygen transmissibility (Dk/t), it is possible to estimate Q C. A previous attempt to directly measure in vivo human Q C required the use of tight-fitting, fluid-filled goggles and oxygen-consuming Clark-type electrodes. 11 A less invasive, but indirect and time-consuming approach, relied on corneal swelling responses to estimate the po 2 in subjects wearing contact lenses of known transmissibility. 12 In contrast, the phosphorescence technique that we report in the current study, is a direct measurement of tear Po 2 beneath contact lenses that requires only a few minutes of hydrogel contact lens wear. 
Methods
Subjects
Four subjects (two men, two women; mean age, 25 ± 2 years) who were free of ocular and systemic disease and had not worn contact lenses for at least 6 months participated in this study. The research adhered to the tenets of the Declaration of Helsinki and was approved by the Indiana University Human Subjects Committee. 
Instrumentation
The principles of oxygen measurement by phosphorescence quenching have been previously described in detail. 3 Briefly, a 10-μsec excitation flash excites a probe whose phosphorescence is quenched by oxygen. The relationship between phosphorescence decay lifetime and oxygen concentration follows a linear relation described by the Stern-Volmer equation. We used a commercially available system (Oxyspot; Medical Instruments, Inc.; now available through Harvard Apparatus, Holliston, MA). The flash excitation was coupled to the illumination optics of a slit lamp (model FS-1; Nikon, Melville, NY) through a fiber optic cable. The slit was adjusted to provide a 2 × 2-mm2 illumination. Phosphorescence was collected by a gated photomultiplier tube mounted on the slit lamp’s camera port. Gating delay, sample number, and sampling rate were controlled by computer (Oxyspot software; Harvard Apparatus, running on a Windows 95 [Microsoft, Redmond, WA]–compatible computer). The phosphorescence decay constant (τ) was computed after each flash, and the average τ of 10 successive flashes was computed for each po 2 data point. Also determined was the correlation coefficient for the fit of the data to a first-order exponential decay. At a po 2 less than 50 torr, data with correlation coefficients less than 0.9 were rejected. At higher po 2 levels, a correlation coefficient of 0.8 was used as the cutoff. Typically, poor correlations occurred during blinks or eye movements. 
Lenses
Three contact lenses were used: (1) a newly available hydrogel-silicone lens with Dk/t = 99 × 10−9 (Balafilcon; Bausch & Lomb, Rochester, NY; Dk = 99 × 10−11 cm2 · mL O2/mL · sec· torr; thickness = 0.1 mm), a 38% water lens with Dk/t = 14 × 10−9 (Polymacon, Dk = 8.4 × 10−11, thickness = 0.06 mm; Metroptics, Glendora, CA), and a 0.2-mm thick lens (Dk/t = 4.2 × 10−9; Polymacon; Metroptics). All lenses had back surface radii of 8.6 mm, diameter of 13.5 mm and −0.50 D power to achieve uniform thickness. 
Preparation of Oxygen-Sensitive Dye and Lens
A 1:9 part mixture of the oxygen-sensitive phosphorescent dye Pd meso-tetra (4-carboxyphenyl) porphine and bovine serum albumin (BSA) was obtained from Harvard Apparatus. The powder was dissolved into Ringer’s solution with final concentrations (in mM) of 140 NaCl, 2 K2HPO4, 0.61 MgCl2, 1.4 Ca+-gluconate, and 28.5 Na+-gluconate (pH 7.5). Osmolarity was adjusted to 300 ± 5 mOsm. The solution was forced through a 0.2-μm filter and placed in a sterile container. A new sterile contact lens was placed in the dye solution and incubated overnight at room temperature. The next day, the lens was rinsed with sterile saline and placed on the subject’s right eye. 
It is conceivable that a protein coating on the surface of a contact lens could provide another layer of resistance to the passage of oxygen across the lens. This would potentially decrease the Dk/t of the lens. Indeed, very thick, denatured albumin coatings placed on contact lenses have been found to lower the amount of oxygen reaching the cornea, but coatings of such thickness are not encountered in practice. 13 Protein coatings from normal wear or mild protein applications in the laboratory do not significantly alter oxygen transmissibility 14 15 and are more representative of the slight coatings that were applied in this study. Nevertheless, we measured the Dk/t of 5 to 7 lenses from each of the three lens types that were incubated with protein and dye as described earlier (18 coated lenses in all) and compared the Dk/t with that obtained with an identical number of uncoated lenses. The polarographic method for determining hydrogel Dk/t has been described in detail previously. 16 An electronic thickness gauge (ET-1; Rehder Development Co., Castro Valley, CA) was used to verify that the mean thickness of coated and uncoated lenses were the same. We found no significant difference of Dk/t between coated lenses and uncoated lenses of the same material and thickness. 
Procedure
Once the stained lens was inserted, it was allowed to settle on the eye for at least 10 minutes. The subject was seated at the slit lamp and asked to fixate, using the left eye, on an LED placed across the room. The subject was allowed to blink at will. The operator aligned the flash illumination on the center of the cornea and adjusted the PMT voltage to bring the signal on scale with the 12-bit analog-to-digital (A/D) converter of the system (Oxyspot; Harvard Apparatus). Open-eye measurements were then made at 0.5 Hz over 60 seconds and repeated 1 to 2 times to assure that the lens was completely settled on the eye, which was judged by the successive data sets’ being within ±5 torr at a po 2 less than 50 torr and ±10 torr at a higher po 2. Phosphorescence from the anterior surface of the lens was avoided by including a delay between the flash and data collection. Because the anterior surface is exposed to air (155 torr), its phosphorescence decays very rapidly (half-life, 20 μsec). Therefore, we used a delay of at least 40 μsec. 
To estimate closed-eye po 2, we asked that subjects close their eyes for 5 minutes while continuing to hold position in the slit lamp after an open-eye measurement. When directed to open their eyes, they immediately took up original fixation. They were instructed to try not to blink for the first 10 seconds and then to blink at will thereafter. Concomitant with eye opening, data collection was started and continued for at least 40 seconds at 1 Hz. Alignment of the flash illumination area with the central cornea was generally preserved, but, occasionally, small adjustments were necessary. In the first 10 seconds after eye opening, most of the individual data point correlation coefficients had to be acceptable to reconstruct the Po 2 change between the closed-eye and the open-eye condition. If more than two of the data points in this period had correlations that were not acceptable, the procedure was repeated until acceptable measures were obtained. The data collected after eye opening was fit to a first-order exponential model  
\[\mathrm{P\mbox{\textsc{o}}}_{\mathrm{2}}{=}SS-(SS-I){\cdot}\mathrm{e}^{-kt}\]
where Po 2 is oxygen tension at any time (in torr), SS is the steady state oxygen tension, I is the initial (t = 0) po 2, k is the rate constant, and t is time. Regressions were performed using PSI-Plot software (Poly Software International, Sandy, UT). 
Estimation of Q C
The steady state tear Po 2 under a contact lens is primarily determined by the corneal oxygen consumption rate (Q C). Thus, given the tear Po 2 under a contact lens of known Dk/t, it is possible to determine the oxygen flux into the cornea, j C, which leads to an estimate of Q C. Q C (mL O2/mL · sec) is calculated from the following equations published by Fatt et al. 9 10 17 18  
\[j_{\mathrm{C}}{=}\mathrm{-}\left(\left[\frac{Q_{\mathrm{C}}L_{\mathrm{C}}}{2}\right]\ {+}\left[(P_{\mathrm{t}}-P_{\mathrm{a}})\ \frac{Dk_{\mathrm{C}}}{L_{\mathrm{C}}}\right]\right)\]
and rearranging:  
\[Q_{\mathrm{C}}{=}\left[2(P_{\mathrm{t}}-P_{\mathrm{a}})\ \frac{Dk_{\mathrm{C}}}{L_{\mathrm{C}}^{2}}\right]\ -2\ \frac{j_{\mathrm{C}}}{L_{\mathrm{C}}}\]
where j C is oxygen flux (in mL O2/cm2 · sec) into the cornea, P t is the po 2 in the tears, P a is the po 2 at the endothelial surface (30 torr), L C is corneal thickness, and Dk C is the oxygen permeability (in mL O2 cm2/mL · sec · torr) of the cornea (2.4 × 10−10). 19 20 At the tears–cornea interface, the flux into the cornea must be equal to the flux of oxygen that is leaving the contact lens, given that the thickness of the tears is relatively small (<10 μm). Therefore, j C = j CL, and j CL is calculated by  
\[j_{\mathrm{CL}}{=}\mathrm{-}\ \frac{Dk_{\mathrm{CL}}}{L_{\mathrm{CL}}}\ (P_{\mathrm{ant}}-P_{\mathrm{t}})\]
where Dk CL/L CL is the oxygen transmissibility of the contact lens (Dk/t in the newer notation). P ant is the po 2 at the anterior surface of the lens and is generally assumed to be 155 torr in the open eye and 55 torr in the closed eye. 18 Corneal thickness (L C) integrated over a central 3-mm diameter, was measured with a pachometer (Orbscan; Orbtek, Inc., Salt Lake City, UT). 
Results
In previous studies of tear po 2 behind rigid contact lenses in rabbits, simple instillation of dye–protein complex solution was made directly into the tears, followed by lens insertion. 3 This worked well in sedated rabbits, because an adequate amount of dye was retained behind the lens for 10 to 15 minutes. Using either rigid lenses or hydrogels, this approach did not work in human subjects, presumably because of more frequent blinking and rapid washout relative to the sedated rabbit. Protein binding to hydrogels is a well-known clinical problem, and the porphyrin dye is completely bound to BSA. We took advantage of this property and bound the dye-protein complex to the lens surface by incubation overnight. Large proteins are not expected to penetrate 38% water hydrogels, and we verified this by sectioning lenses and viewing at ×200 magnification (data not shown). 
The quenching constant, q k, and the lifetime in the absence of oxygen (τ0) of this porphyrin–protein complex are well established. 21 22 23 For the eye–contact lens system we used q k = 304 (in torr per second) andτ 0 = 581 μsec, which are the parameters for this dye at 35°C and pH 7.2. 22 We verified that these parameters were appropriate for our instrumentation. A dye-coated thin hydrogel was placed in saline in a cuvette and kept at 35°C. The saline was bubbled with 100% nitrogen gas or air. The estimated Po 2 in nitrogen was less than 0.1 torr and ranged from 140 to 165 torr in air. To verify that the phosphorescence parameters were appropriate for the lens on the eye, a tight-fitting goggle was placed over the eyes of a subject who was wearing a thin hydrogel dye-coated lens. Humidified nitrogen gas or air was passed through the goggle and the phosphorescence decay measured. Again, for air the estimated Po 2 was 140 to 165 torr. Under nitrogen gas the estimated Po 2 was 0.8 torr. This is a reasonable level, because oxygen diffusion from the anterior chamber and occasional oxygen fluxes from the palpebral conjunctiva during blinks make it difficult to obtain absolute anoxia at the corneal surface. 
Figure 1 shows representative open-eye oxygen measurements for the three lenses 10 minutes after lens insertion. Figure 1A shows data from one subject for the high-Dk/t lens taken at 2-second intervals for 60 seconds. The mean ± SD for this data set was 105.6 ± 2.0 torr. Figure 1B shows representative data for the thin hydrogel (mean ± SD; 31.1 ± 1.2 torr). Figure 1C shows data for the thick hydrogel for which the mean ± SD over 60 seconds was 6.7 ± 0.15 torr. These data illustrate that the variability in Po 2 estimates increased with increasing Po 2 and in the four subjects, the average of the SDs of each 60-second data set (30 readings) was 3.85, 2.78, and 0.21 torr, for the Balafilcon (Bausch & Lomb) and thin and thick hydrogel lenses (Metroptics), respectively. Figure 1D summarizes the open-eye data in the four subjects for each lens. 
In preliminary experiments we had subjects who were wearing test lenses close their eyes for 1, 3, 5, and 10 minutes before opening the eyes, to measure the po 2 to the open-eye steady state value. These experiments indicated that approximately 3 minutes of eye closure was sufficient to establish the closed-eye value—that is, 5 or 10 minutes of eye closure did not produce lower estimates. Therefore, for all closed-eye experiments we used 5 minutes of eye closure to assure closed-eye equilibrium. Figure 2 shows representative data from the same subject shown in Figure 1 for the three lenses, after eye closure. Figures 2A and 2B indicate the initial estimate (I) of Po 2 for the Balafilcon (Bausch & Lomb) and the thin hydrogel lenses (Metroptics) at t = 0, by fitting the data to a simple exponential increase to a new steady state. Of interest, Figure 2C shows that with the thick hydrogel lens there was no change in Po 2 for approximately 10 seconds after eye opening. The average of the first six readings was taken as the closed-eye Po 2. Figure 2D summarizes the closed-eye data in the four subjects. 
We suspected that the observed delayed increase in Po 2 just after eye opening with the thick hydrogel lens was due to the time needed for oxygen to diffuse across the 200-μm thickness of the lens. The diffusion coefficient (D) for oxygen in water at 35°C is approximately 3 × 10−5 cm2/sec. 24 From the relation, t = x 2/D, where x is distance and t is time, we can calculate that it would take approximately 13 seconds for a change in oxygen to appear at the posterior lens surface, which is within a few seconds of what we observed. 
Figure 3 shows the mean Q C calculated from the estimated tear Po 2 for the six conditions (three lenses, open and closed eye). The data indicate that Q C varied significantly with tear po 2. Q C ranged from 2.2 × 10−4 at approximately 100 torr surface po 2 to 3.7 × 10−6 mL O2/cm3 · sec at 4 torr). 
Discussion
Our goal in this study was to show that human corneal oxygen consumption could be estimated by a direct, noninvasive measure of tear po 2 beneath contact lenses. Given the oxygen transmissibility of a contact lens, the oxygen flux through the lens can be calculated, because the boundary conditions (an assumed front surface Po 2 and the measured back surface Po 2) are known. In the steady state, oxygen flux out of the lens must be equal to flux into the cornea. From this relation and the corneal thickness, an estimate of Q C can be obtained. 10 At high surface po 2 we estimated Q C to be 2.2 × 10−4 mL O2/cm3 · sec. This is approximately two to three times that reported in dissected rabbit corneas. 25 This difference could be due to species differences or more likely because rabbit corneal oxygen consumption was suppressed by the trauma of explantation to an in vitro measurement apparatus. Weissman and Fazio 26 estimated in vivo human corneal oxygen fluxes based on the known lens Dk/t and estimated surface po 2 from corneal swelling experiments. At 25 torr, Weissman 12 estimates human in vivo Q C to be 4.85 × 10−5. From our data, the calculated Q C was 5.8 to 6.2 × 10−5 mL O2/cm3 · sec at a Po 2 of 25 to 30 torr, which is reasonably close to Weissman’s estimate. That Q C decreases with decreasing surface Po 2 is not unexpected. Early in vivo studies 11 indicated that at approximately 20 torr, Q C began to decrease. Further, recent mathematical modeling of oxygen distribution from the front to the back surface of the cornea has shown that even at an open-eye surface Po 2 of 70 torr, a small portion of the central stroma is anoxic and at surface Po 2 between 30 and 40 torr, basal epithelial cells are hypoxic, 27 which would significantly suppress O2 consumption. 
This study shows that tear po 2 can be measured in human subjects using the oxygen quenching of the phosphorescence probe Pd-meso tetra (4-carboxyphenyl) porphyrin. The measurement is not completely noninvasive, because it requires hydrogel lens wear. Lens wear itself can have mechanical effects on the surface epithelium that could suppress Q C. This could be tested by determining whether other mild forms of trauma (e.g., surface drying, brief wear of a rigid contact lens, or brief touch with an applanation tonometer) affects Q C. Because of the nature of measuring phosphorescence decay accelerated by oxygen quenching, the measurement is most sensitive at a low po 2. This is exemplified by the greater variability of individual open-eye data sets at a high po 2. The dye–protein complex immobilized to the surface of contact lenses could act as an oxygen sensor for many hours; however, the individual lengths of the experiments in this study were no more than 20 minutes. Binding the dye complex to the lens was needed, because direct instillation into the tears led to rapid loss of signal (<1 minute) due to washout. Protein–dye binding to the lens surface had no effect on lens Dk/t. Also, binding to the lens did not alter the dye’s quenching parameters, presumably because the primary interaction of the porphyrin is with BSA. We assume that the dye complex is indicating po 2 at the lens surface and not from inside the lens. Because BSA has a molecular weight of ∼66 kDa it is not expected to penetrate beyond the lens surface. Light microscopy of lens sections indicated that only the surface was stained. Furthermore, the delay between eye opening and a change in measured po 2 when subjects wore the thick hydrogel would have been significantly shorter if dye complex had penetrated the lens. Last, because both front and back lens surfaces are stained, the front surface phosphorescence had to be removed. The front surface of the lens is exposed to air (155 torr) and the decay constant at 155 torr is 20 μsec. Thus significant phosphorescence from the front surface could be avoided by delaying data acquisition after the flash by approximately 40 μsec. 
Estimates of closed-eye po 2 have been of interest to contact lens researchers for many years. In the current study, we used closed-eye Po 2 estimates to extend the range of surface po 2 to examine its effect on Q C. After eye opening, the fit of the oxygen measurement data to a simple exponential model was very good. Because of the fitting process, the data, especially in the first 10 seconds, had to be of high quality to get an accurate estimate of the initial Po 2. In practice, data are often rejected because of poor fixation after eye opening or, more commonly, excessive blinking. However, because of the relatively short time of eye closure, it is convenient to repeat the closed-eye data collection until an acceptable data set is obtained. 
Previously, in vivo measurements of corneal oxygen consumption were very cumbersome and somewhat invasive (use of fluid-filled tight fitting goggles) 11 or they were lengthy procedures that relied on the relationship between surface oxygen and corneal swelling. 26 In contrast, the current procedure is relatively simple and quick and has broad applicability. We anticipate that very few human subjects would be unable to tolerate wearing a hydrogel lens for 20 minutes. Within this time, Q C can be determined at two po 2s (open and closed eye). With this technique, the effects of disease, drugs, or surgical interventions on Q C could be determined. For example, it could be used as a measure of the depth of the effects of a disease on the cornea (e.g., diabetes) and of the effects of topical drug use on metabolic activity, recovery of Q C after photoablative surgery, as a determinant in the wound-healing process, or to study the relation between sensory innervation and corneal metabolism (e.g., neurotrophic keratopathy, cataract surgery, or penetrating keratoplasty). 
 
Figure 1.
 
Tear po 2 in the open eye. (AC) Representative data sets of tear po 2 taken every 2 seconds over 1 minute for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (D) Mean open-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 1.
 
Tear po 2 in the open eye. (AC) Representative data sets of tear po 2 taken every 2 seconds over 1 minute for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (D) Mean open-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 2.
 
Tear po 2 in the closed eye. (AC) Representative data sets of tear po 2 after 5 minutes of closed-eye lens wear for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (A, B, dashed line) The fit to an exponential model. I is initial Po 2 at t = 0, SS is the steady state Po 2, and r 2 is the coefficient of determination. For the thick hydrogel lens (C), the first six data points were averaged to obtain the closed-eye Po 2 estimate. (D) Mean closed-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 2.
 
Tear po 2 in the closed eye. (AC) Representative data sets of tear po 2 after 5 minutes of closed-eye lens wear for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (A, B, dashed line) The fit to an exponential model. I is initial Po 2 at t = 0, SS is the steady state Po 2, and r 2 is the coefficient of determination. For the thick hydrogel lens (C), the first six data points were averaged to obtain the closed-eye Po 2 estimate. (D) Mean closed-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 3.
 
Q C as a function of Po 2. Q C was estimated for each of the six conditions (three lenses, open and closed eye). Error bars, SD.
Figure 3.
 
Q C as a function of Po 2. Q C was estimated for each of the six conditions (three lenses, open and closed eye). Error bars, SD.
Masters B. A noninvasive optical measurement of tear oxygen tension at the corneal epithelium. Curr Eye Res. 1985;4:725–727. [CrossRef] [PubMed]
Piston DW, Masters BR, Webb WW. Three-dimensionally resolved NAD(P)H cellular metabolic redox imaging of the in situ cornea with two-photon excitation laser scanning microscopy. J Microsc. 1995;178:20–27. [CrossRef] [PubMed]
Harvitt D, Bonanno J. Direct noninvasive measurement of tear oxygen tension beneath gas-permeable contact lenses in rabbits. Invest Ophthalmol Vis Sci. 1996;37:1026–1036. [PubMed]
McLaren JW, Dinslage S, Dillon JP, Roberts JE, Brubaker RF. Measuring oxygen tension in the anterior chamber of rabbits. Invest Ophthalmol Vis Sci. 1998;39:1899–1909. [PubMed]
Shonat RD, Wilson DF, Riva CE, Pawlowski M. Oxygen distribution in the retinal and choroidal vessels of the cat as measured by a new phosphorescence imaging method. Appl Opt. 1992;31:3711–3718. [CrossRef] [PubMed]
Sarver M, Polse K, Baggett D. Intersubject difference in corneal edema response to hypoxia. Am J Optom Physiol Opt. 1983;60:128–131. [CrossRef] [PubMed]
Herse P, Petchell M. Oxygen consumption and ATPase activity in the corneal epithelium of rabbits with alloxan induced hyperglycemia. Acta Ophthalmol Scand. 1998;76:528–532. [CrossRef] [PubMed]
Rubinstein MP, Parrish ST, Vernon SA. Corneal epithelial oxygen uptake rate in diabetes mellitus. Eye. 1990;4:757–759. [CrossRef] [PubMed]
Fatt I, Freeman R, Lin D. Oxygen tension distributions in the cornea: a re-examination. Exp Eye Res. 1974;18:357–365. [CrossRef] [PubMed]
Fatt I, Bieber M, Pye S. Steady state distribution of oxygen and carbon dioxide in the in vivo cornea of an eye covered by a gas-permeable contact lens. Am J Optom. 1969;46:3–14. [CrossRef]
Jauregui MJ, Fatt I. Estimation of the in vivo oxygen consumption rate of the human corneal epithelium. Am J Optom Arch Acad Optom. 1972;49:507–511. [CrossRef]
Weissman BA. Oxygen consumption of whole human corneas. Am J Optom Physiol Opt. 1984;61:291–292. [CrossRef] [PubMed]
Benjamin W, Hill R. Ultrathins: the case for continuous care. J Am Optom Assoc. 1980;54:277–279.
Huth SW, Lannom CS, Lannom S. Effect in in vivo and in vitro surface protein deposits on the oxygen permeability of polyhydroxyethylmethacrylate gel contact lenses. Am J Optom Physiol Opt. 1984;61:232–238. [CrossRef] [PubMed]
Refojo MF, Leong FL, Ueno N, Herman J. Oxygen transmissibility and hydration of new and used hydrogel contact lenses. J Am Optom Assoc. 1982;53:215–218. [PubMed]
American National Standards Institute. Merrifield, VA: Optical Laboratories Association, Publication no. ANSI Z80.20-1998:9,16–17,34,43–57.
Fatt I, Bieber M. The steady-state distribution of oxygen and carbon dioxide in the in vivo cornea. Exp Eye Res. 1968;7:103–112. [CrossRef] [PubMed]
Fatt I. Steady-state distribution of oxygen and carbon dioxide in the in vivo cornea: part II. Exp Eye Res. 1968;7:413–430. [CrossRef] [PubMed]
Freeman R, Fatt I. Oxygen permeability of the limiting layers of the cornea. Biophys J. 1972;1972:237–247.
Weisman B, Selzer K, Duffin R, Petit T. Oxygen permeability of rabbit and human corneal stroma. Invest Ophthalmol Vis Sci. 1983;24:645–647. [PubMed]
Vanderkooi JM, Wilson DF. A new method for measuring oxygen concentration in biological systems. Adv Exp Med Biol. 1982;200:189–193.
Vanderkooi J, Maniara G, Green T, Wilson D. An optical method for measurement of dioxygen concentration based upon quenching of phosphorescence. J Bio Chem. 1987;262:5476–5482.
Wilson DF, Vanderkooi JM, Green TJ, Maniara G, DeFeo SP, Bloomgarden DC. A versatile and sensitive method for measuring oxygen. Adv Exp Med Biol. 1987;215:71–77. [PubMed]
Weast RC. CRC Handbook of Chemistry and Physics. 1990; 70th ed. CRC Press Boca Raton, FL.
Freeman R. Oxygen consumption by the component layers of the cornea. J Physiol. 1972;225:15–32. [CrossRef] [PubMed]
Weissman BA, Fazio DT. Human corneal oxygen flux under soft contact lenses. Am J Optom Physiol Opt. 1982;59:635–638. [CrossRef] [PubMed]
Harvitt D, Bonanno J. Re-assessment of corneal oxygen distribution during contact lens wear. Optom Vis Sci. 1999;76:712–719. [CrossRef] [PubMed]
Figure 1.
 
Tear po 2 in the open eye. (AC) Representative data sets of tear po 2 taken every 2 seconds over 1 minute for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (D) Mean open-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 1.
 
Tear po 2 in the open eye. (AC) Representative data sets of tear po 2 taken every 2 seconds over 1 minute for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (D) Mean open-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 2.
 
Tear po 2 in the closed eye. (AC) Representative data sets of tear po 2 after 5 minutes of closed-eye lens wear for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (A, B, dashed line) The fit to an exponential model. I is initial Po 2 at t = 0, SS is the steady state Po 2, and r 2 is the coefficient of determination. For the thick hydrogel lens (C), the first six data points were averaged to obtain the closed-eye Po 2 estimate. (D) Mean closed-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 2.
 
Tear po 2 in the closed eye. (AC) Representative data sets of tear po 2 after 5 minutes of closed-eye lens wear for the Balafilcon (Bausch & Lomb) and the thin and thick hydrogel (Metroptics) lenses, respectively. (A, B, dashed line) The fit to an exponential model. I is initial Po 2 at t = 0, SS is the steady state Po 2, and r 2 is the coefficient of determination. For the thick hydrogel lens (C), the first six data points were averaged to obtain the closed-eye Po 2 estimate. (D) Mean closed-eye po 2 in the four subjects for each lens. Error bars, SD.
Figure 3.
 
Q C as a function of Po 2. Q C was estimated for each of the six conditions (three lenses, open and closed eye). Error bars, SD.
Figure 3.
 
Q C as a function of Po 2. Q C was estimated for each of the six conditions (three lenses, open and closed eye). Error bars, SD.
×
×

This PDF is available to Subscribers Only

Sign in or purchase a subscription to access this content. ×

You must be signed into an individual account to use this feature.

×