Abstract
purpose. To quantify the reflectance of the swine fundus as a function of illumination angle and wavelength using a novel technique of intravitreal illumination.
methods. Enucleated swine eyes were illuminated with a scanning monochromator coupled to a fiber-optic probe placed in the vitreous at several locations. Intravitreal illumination was used to reduce the effects of extraneous reflections and scatter from the anterior structures of the eye, including glints from the internal limiting membrane and blood vessels. Intravitreal illumination provided different light paths and additional information to apply to retinal reflectance modeling. A 25-mm2 region of the illuminated fundus was imaged while the angle of illumination was varied over a maximum range of 22°. Multispectral images from areas free of large blood vessels were acquired. The diffusely reflected intensity was integrated over the pupil with a solid angle of 0.028 steradians. The spectral reflectance function was calculated for multiple illumination angles.
results. Multispectral fundus image sets were obtained for two enucleated swine eyes by using intravitreal illumination. The fundus spectral reflectance function showed decreasing reflectance with increasing illumination angle, rapid changes in the 430- to 480-nm range, and a fairly consistent reflectance decrease for 480 to 700 nm.
conclusions. Variations in fundus spectral reflectance with change in the illumination angle were found to deviate from Lambertian behavior, varying from Lambertian by 5% across the spectrum in one sample and 20% in a second sample. Intravitreal illumination resulted in markedly decreased extraneous reflections.
Optical methods for diagnosing and treating ocular disease rely on a fundamental understanding of the interaction between light and ocular tissue. Data obtained from noninvasive optical techniques such as fundus photography, optical coherence tomography, and laser polarimetry may benefit from improved characterization of the reflectance and scattering properties of the fundus.
Several studies of the optical properties of the fundus have been performed.
1 2 3 4 5 6 7 8 9 Van Norren and Tiemeijer
2 measured reflectance in the foveal and peripheral fundus and mathematically modeled the fundus. Their reflectance model used two reflective and four absorbing layers to describe the fundus. Delori and Pflibsen
3 performed in vivo spectral reflectance measurements in human subjects with a wide range of fundus pigmentation and applied the model of van Norren and Tiemeijer
2 to the resultant data. Delori and Pflibsen
3 also applied the model of Kubelak and Munk
10 to their data and achieved a better fit to measured spectra. Kubelak and Munk introduced a two-constant theory for the description of the reflectance properties of a material: the absorption and the scattering coefficients. The adaptation of this model by Delori and Pflibsen
3 treats the choroid as a diffuse absorbing scatterer backed by a reflecting sclera. Hammer and Schweitzer
4 performed high-spectral-resolution studies with imaging spectroscopy at high spatial resolutions along a bar-shaped field in the retina.
The earlier works in fundus reflectometry used transcorneal illumination. The illuminating light entered the eye through the cornea, passing through the anterior chamber, lens, and vitreous to illuminate the fundus.
1 2 3 4 5 6 7 8 9 Transcorneal fundus reflectance measurements are impeded by extraneous reflections, or glints, from the cornea, retinal vessels, and the fundus, as well as backscattering from the anterior chamber, lens, and vitreous. The illumination and the measurement directions are nearly the same for transcorneal illumination. In the previous studies, the fundus was generally assumed to be a perfectly diffuse or Lambertian structure. A primary objective of this research was to assess the validity of this assumption.
We introduce the novel technique of intravitreal illumination for the measurement of fundus reflectance in enucleated swine eyes. The use of intravitreal illumination allows for variation of illumination angle, eliminates the corneal and retinal vessel glints, and reduces the errors in quantitative spectroradiometry arising from multiple passes through the anterior ocular structures. Our technique results in light reflected from the fundus undergoing a single pass through the anterior chamber. The illumination direction is varied and measurements of fundus reflectance are made from a fixed viewing angle. Reflectance measurements for various illumination angles are compared with each other and with the predictions of the Lambertian model.
The fiber-optic intraocular illuminator was placed in the first sclerotomy, the closest one to the pupil. This sclerotomy provided the smallest angle of illumination (θ) between the fiber-optic intraocular illuminator and the optic axis for the target area. The illuminated fundus was focused on the camera. A suitable vessel-free target area was identified by using the retinal vessels as landmarks. The depth of the fiber-optic intraocular illuminator was adjusted to maximize the dynamic range of the camera (i.e., the brightest pixel in the image is nearly saturated). An image was acquired at each of the 36 wavelengths listed in
Table 1 . The wavelengths were referenced with respect to the peak of the in-band spectral power distribution. The reflected light was collected over a solid angle of 0.028 steradians, the solid angle subtended by the eye’s pupil, which was the aperture stop of this system.
To determine the angle of the fiber-optic illuminator, an image of the experimental apparatus was taken with a digital camera (Cool Pix 4300; Nikon Corp., Tokyo, Japan) aligned orthogonally to the plane defined by the sclerotomies. Orthogonality was insured by using an optical table equipped with a railed carrier system. The digital image was analyzed to determine the angle of the fiber-optic illuminator in the eye. We estimated the systematic error in angular measurement at ±2°. This systematic error is indicative of the error in the absolute value of the measured angles, but the relative differences between measured angles are affected only slightly.
The fiber-optic illuminator was successively inserted into each sclerotomy and illuminated the target area of the fundus at increasing angles. The multispectral image-acquisition process was repeated for each sclerotomy. The appropriate depths of the fiber-optic illuminator for the remaining sclerotomies were determined based on illumination geometry, so that the end of the illuminator was kept at a constant distance from the fundus. The depth of the fiber-optic illuminator was verified by measuring the length of the fiber-optic illuminator exterior to the sclera. The only variable between spectral image sets was the angle of illumination in the plane of the sclerotomies. For all sclerotomies, the brightest portion of the illuminating light beam was placed on the target area of the fundus by using retinal blood vessels as landmarks. Four angles of illumination were used for the first enucleated swine eye and three for the second eye.
Table 2lists the angles of illumination used for the two eyes.
A calibration image set was acquired on completion of data collection. A lined integrating sphere (Spectralon; Labsphere, North Sutton, NH) was illuminated by the fiber-optic intraocular illuminator, and the exit pupil of the sphere was imaged by the camera system. This calibration set was used to correct for the spectral intensity variation in the illuminating source and the spectral response of the CCD.
The software package (Mathematica; Wolfram Research, Inc., Champaign, IL) was used to perform image calibration, subtraction, averaging, and plotting. Reflected spectra were constructed for each angle of illumination for each eye. The reflectance spectra were averaged over 25 pixels in the target area and were normalized to the value obtained in the brightest image for each eye. The brightest image corresponded to the smallest illumination angle for each eye.
The results indicate that the spectral reflectance of the swine fundus deviates from the Lambertian model that is customarily used to model fundus reflectance. A Lambertian reflector has three defining reflectance properties: (1) the angular distribution of reflected flux is independent of illumination angle; (2) for a given illumination angle, reflected intensity is independent of viewing angle; and (3) the total reflected intensity decreases according to Lambert’s law (also called the cosine law),
11 \[I_{\mathrm{diffuse}}{=}I_{0}\ {\cdot}\ \mathrm{cos}({\theta})\]
where
I diffuse is the total reflected intensity,
I 0 is the incident intensity, and θ is the angle between the surface normal and the incident light, as illustrated in
Figure 8 .
Optical density is a parameter for describing absorbance, which is defined as OD = log(1/T) where T is the transmittance associated with a medium.
This experiment tested only the third Lambertian property. The reflectance decrease predicted by Lambert’s law was compared with the measured reflectance decrease. The measured reflectance from the smallest illumination angle (26°) was used to generate the three Lambertian reflectance curves for the remaining illumination angles. The third property of the Lambertian model specifically states that the total reflected intensity varies with the cosine of the respective illumination angles, so the measurements in eye 1 should vary as the ratio cos(27°):cos(33°):cos(37°):cos(48°). The percent difference between this prediction and the measurement is plotted in
Figure 9A . For a Lambertian reflector these percent errors would be zero.
Figure 9Bis the corresponding plot for eye 2.
The results of this research indicate that the fundus is not a perfectly diffuse reflector, because the measured intensity at larger illumination angles is less than that predicted by Lambert’s law. In addition, the variation from Lambertian is spectrally dependent. The characteristic spectra of fundus pigments and retinal blood may have a significant dependence on illumination angle. These preliminary findings imply that an accurate model of fundus reflectance requires consideration of illumination angle. Further studies are necessary to determine the analytical form of this term.