The protocol was as follows. A custom-made circular PMMA ring
(Fig. 4)was bonded onto the sclera in the region of the ciliary body approximately 2 mm posterior to the limbus with cyanoacrylate adhesive (Duro Quick Gel super glue; Loctite Corp., Rocky Hill, CT) after the conjunctiva, adipose, and muscle tissues were removed. The ring’s inner radius of curvature was machined to fit the globe (12 mm for human, 9.5 mm for monkey, and 8 mm for rabbit), enabling dissection of the globe with minimal deformation while keeping the ciliary body–zonule–lens framework intact. Dissection was not initiated until the glue had dried, to ensure that the fumes from the glue did not cause any dehydration of the lens. The posterior pole was removed by making a circumferential incision through the sclera approximately 7 mm posterior to the limbus. Excess vitreous was removed, and the eye was placed on a Teflon slide. The cornea and iris were then sectioned. The clinical appearance of all lenses and lens capsules was examined under an operation microscope. All monkey and rabbit lenses were noted to be intact and clear. All human lenses were intact, but some of the older human lenses (>70 years old) had signs of cataract. All lens capsules were intact and transparent, with no signs of dehydration. The mounted tissue specimen was then placed in a well filled with balanced saline solution (
Fig. 5 , left) under the focusing lens of the optical system (
Fig. 5 , right). The well was 33 mm in diameter and 6 mm deep, to enable placement of the PMMA ring. A circular groove 14 mm in diameter and 2 mm deep was made in the center of the well to ensure that the lens did not touch the bottom. The well was filled so that the entire capsule except for a region approximately 3 mm in diameter at the central pole of the lens was covered in balanced saline during the measurements. The lens was adjusted in the
x–
y plane until the incident laser beam was located in the center of the lens. The determination of the center of the lens was subjective. However, as a test of the precision of the
x–
y alignment technique, in one separate experiment optical measurements were repeated five times on one anterior lens capsule. In this experiment the sample was removed from the system between each successive measurement and placed under the focusing objective again, and the position was adjusted until the beam was in the center of the lens. The measurement difference was found to be only 0.5 μm in these successive measurements. The sample was also on a platform with tilt adjustment. The tilt was adjusted until maximum reflected light was sent to the photodetector. This confirmed perpendicularity of the sample, since the maximum light detected at the photodetector signifies that the sample is aligned perpendicularly to the coupling system. The lens objective was translated toward the sample at 0.1 mm/s as the detected signal was acquired at a rate of 1000 Hz (1 sample every 0.1 μm). Except for the alignment experiment (described earlier) in which five measurements were taken, optical measurements were performed three times on each sample, with the sample alignment adjusted before each measurement. The mounted specimen was flipped, and the same procedure was used to measure the posterior lens capsule. Repeatability for the optical technique was defined as the difference between the largest and smallest measurement obtained for the same eye
(Table 1) . The dissection process took an average of 15 minutes per eye, and measurements took an average of 5 minutes per eye. Both dissection and measurements were performed at room temperature. The lens capsule was kept hydrated during the dissection with the saline. In a time-related study on the saline solution (BSS; Alcon Laboratories) as a preservative medium for the crystalline lens, it caused swelling of <3% over a testing period of 4 to 5 hours.
17 Because the surgical preparation and measurements took only ∼20 minutes, the saline solution (BSS; Alcon Laboratories) physiologically maintained the shape of the entire lens. Only the central pole of the capsule was exposed to air during the measurements (∼5 minutes), and dehydration was therefore unlikely. This was demonstrated by the repeatability of the measurements. Successive measurements did not show a trend toward decreasing thickness. The clinical appearance of the lens capsules was normal throughout the experiments, with no clouding of the lens capsule observed. Intensity maxima obtained during the measurements corresponded to the anterior and posterior surfaces of the lens capsule. The intensity maxima were detected with a peak detection algorithm from the graphing software (Origin; Microcal, Northampton, MA), and the validity of the peaks was verified manually. The physical thickness of the lens capsule was calculated by multiplying the distance between successive maxima by the refractive index (1.4 for the human lens capsule
2 ). Because there are no published data on the refractive index of the rabbit or monkey lens capsule, the value was assumed to be 1.4 as well. Slight variations in refractive index of the capsule were found to have a negligible effect on the calculated thickness, assuming that a refractive index in the range of 1.37 to 1.47 produces only a 1.5 μm measurement difference for a 20-μm thick sample.
Figure 6shows typical graphs obtained during the experiments.