Our study demonstrated that with the Cirrus SD-OCT, SL, SS, TM, and SC were identified in 77%, 82%, 34%, and 10% of quadrants, respectively, whereas the corresponding figures for iVue were 74%, 66%, 68%, and 13% of quadrants, respectively. Both devices identified SL equally well, and SC equally poorly, but SS was more frequently identified in Cirrus compared with iVue images, whereas the TM was more frequently identified in iVue images compared with Cirrus. The variability in these figures could be explained by the inherent differences in quality of image acquisition and processing between the different machines, and the subjectivity of image interpretation between observers. In addition, the magnifications of the images are different. The magnification of the Cirrus machine is 2-fold that of the iVue. This results in the SC in the iVue image appearing compressed and lower in position compared with that in the Cirrus image.
Currently available software analysis programs require the manual localization of the scleral spur, which can at times be difficult, especially in closed angles or where there is a smooth transition from cornea to sclera. With AS-OCT, Sakata et al.
16 found that the sclera spur could not be detected in approximately 30% of angle quadrants, a problem that was worse in the superior and inferior quadrants. In our study, we defined angle closure as contact between the iris and angle wall anterior to the scleral spur. We had difficulty determining angle status due to an inability to clearly discern the SS in some images. Although higher-resolution OCT devices improved the resolution of angle structures, in our study, angle status was not determined in 10% of quadrants and 13% of eyes using Cirrus SD-OCT, and 17% of quadrants and 25% of eyes using iVue SD-OCT, mainly due to an inability to identify the SS in these images. Inherent limitations of SD-OCT imaging technology could be possible reasons. The iVue uses markers for standardizing the image acquisition point, resulting in less penetration depth, attenuation of details at the scleral spur area, and motion artifacts that preclude distinct identification of landmarks. Possible solutions to these limitations include averaging of images and increasing the width to include both angles, and postimage acquisition processing similar to those of high-resolution retinal images.
More recently, Cheung et al.
17 proposed the SL as a new anatomical landmark, independent of the SS location, for assessing ACA width quantitatively with Cirrus SD-OCT. They developed a computer-aided program to define two new quantitative parameters for assessing ACA width: Schwalbe's line–angle opening distance (SL-AOD) measured at the SL, and Schwalbe's line–trabecular-iris space area (SL-TISA) measured 500 μm from the SL. In their study, 20% of images had to be excluded because of poor image quality. In the remaining images, SL and SS could be identified in 95% and 85% of quadrants, respectively. SL-AOD and SL-TISA were significantly correlated with SS parameters (all
r > 0.85) and gonioscopic grading (all
r > 0.69). In eyes with closed angles (
n = 36), SL parameters showed strong correlations with gonioscopic grading (
r ranged from 0.43 to 0.44). These findings suggest that novel angle parameters, based on SL as a landmark, may be useful to quantify ACA width and to assess for risk of angle closure.
17 Both the devices used in our study identified SL in about 75% of images. A further study on these SL-derived quantitative parameters derived from images from both the Cirrus and iVue devices, and their correlation with gonioscopy could be performed to assess suitability of these devices for quantification of angle closure.
We found that the agreement of Cirrus and iVue SD-OCT with gonioscopy for identification of angle closure in eyes was fair (AC1 = 0.35 and 0.50, respectively), with both the Cirrus and iVue OCT detecting angle closure at a lower rate compared with gonioscopy. On SD-OCT, a radiolucent gap is sometimes present between the cornea and sclera such that the iris does not seem to be in apposition with the angle, rendering the angle “open” on SD-OCT but “closed” on gonioscopy (
Figure 3). In addition, although gonioscopy allows concurrent visualization of the entire angle quadrant, AS-OCT images are meriodional, and relate only to the particular cross-section of the angle scanned. Thus, they may not be representative of the whole quadrant.
In addition, scan acquisition with the patient looking at the side of the instrument may potentially change the image profile of the angle. Unfortunately, imaging of the angle with the current SD-OCT is technically not possible with the eye in the primary position, and we attempted to standardize this by using specified targets for each patient to fixate on. This may be an additional reason for the poor agreement between gonioscopy and these devices.
Although intraobserver agreement on identification of angle structures was good for both devices (AC1 = 0.6–1 for Cirrus and 0.88–0.94 for iVue), the interobserver agreement on identification of angle closure was poor to fair on the Cirrus (kappa = 0.20–0.40) and fair on iVue (kappa = 0.35–0.47). In this study, we found that angle structure identification was generally easier in horizontal scans compared with vertical scans. As detailed in
Table 2, angle structures were most frequently identified in either nasal or temporal quadrants, in both machines. Localization of angle structures was difficult in the superior and inferior quadrants, a finding consistent with that previously reported by Sakata et al.
16 As a result, interobserver agreement was generally poorer for superior and inferior quadrants than nasal and temporal quadrants. Significant interobserver variability exists for both machines, and angle closure status between observers may not be consistently reproducible.
In conclusion, it was more difficult to determine angle closure status with the iVue SD-OCT compared with Cirrus SD-OCT. Although these new SD-OCT devices provide us with higher resolution and detailed imaging of angle structures, they are unable to determine angle status in all cases, have significant interobserver variability, and have only fair correlation with gonioscopy.