From a clinical perspective, the reproducibility of any new technology must be verified, as it influences the diagnostic accuracy and the ability to monitor disease progression.
22 The reproducibility values (i.e., ICC) for peripapillary (global and per region) RNFL thickness measurements using some of the commercially available Fd-OCT instruments (Cirrus HD-OCT; Carl Zeiss Meditec, Inc.; 3D OCT-1000; Topcon, Tokyo, Japan; and RTVue OCT; Optovue Corporation, Fremont, CA) ranged from 0.74 to 0.99.
22–26 Two studies have also investigated the reproducibility of Spectralis for peripapillary RNFL thickness measurements performed by a single examiner.
27,28 In the study by Langenegger et al.,
27 the reproducibility values (i.e., ICC) for peripapillary RNFL thickness measurements in healthy subjects using the Spectralis device
without the EBF function were 0.98 (global), 0.86 (TS), 0.83 (T), 0.91 (TI), 0.91 (NI), 0.83 (N), and 0.91 (NS). With the EBF function activated, ICC values increase to 0.99 in all but TI (0.97) and NI (0.98) regions.
27 Wu et al.
28 also obtained similar results about the reproducibility of Spectralis with the EBF function activated for peripapillary RNFL measurements, with ICC values ranging from 0.97 (T) to 0.99 (global, NS, and NI). The singularity of the present study relies on the evaluation of the Spectralis reproducibility for peripapillary RNFL thickness measurements obtained by two different examiners with the use of two distinct acquisition modes (HS and HR), thus providing supplementary information about the use of this Fd-OCT instrument for this purpose. Our results showed that interobserver reproducibility (with and without the EBF function) of the peripapillary RNFL thickness measurements on HS mode was very good for all (ICC > 0.80) but the temporal region, which was good. Interestingly, the ICC values for interobserver reproducibility on HR mode were generally lower than those on HS mode. The intraobserver ICC values for peripapillary RNFL thickness measurements on HS mode with the EBF function activated were greater than or equal to 0.98 in all but TS (0.97) and T (0.93) regions. Serbecic et al.
29 compared the detection of RNFL measurements using either HS or HR mode and found the highest coefficient of variation in the temporal sector with no significant differences between the modes of image acquisition. The high reproducibility of peripapillary RNFL thickness measurements with the Spectralis can be attributable to different technical factors, such as improved image resolution, imaging speed (40,000 A-scans/s) or retinal segmentation algorithms. However, given the results of the current study, it is likely that its improved reproducibility has also been positively influenced by the EBF function. Moreover, the overall greater ICC values obtained for HS scanning may be related, at least in part, to the fact that the ART tool was not used during HR scanning.