Investigative Ophthalmology & Visual Science December 1986, Vol.27, 1776-1780. doi:
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      J G Feghali, D T Azar, P L Kaufman; Comparative aqueous outflow facility measurements by pneumatonography and Schiotz tonography.. Invest. Ophthalmol. Vis. Sci. 1986;27(12):1776-1780.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
This content is PDF only. Please click on the PDF icon to access.
Abstract

Tonography was performed on 36 eyes of 15 normal and 3 primary open angle glaucoma patients using pneumatonography and classical Schiotz tonography. The average values of the coefficient of outflow facility (C) for the whole sample were virtually identical with both methods. However, both intersubject and interobserver variability were significantly higher with penumatonography. Although both methods provide comparable aggregate estimates of aqueous outflow facility, we think that Schiotz tonography is more reliable than pneumatonography because of the greater mechanical stability of the Schiotz instrument on the eye. On the other hand, pneumatonography offers the advantage of a shorter test period (2 min instead of 4).

×
×