Clinical and Epidemiologic Research  |   November 2007
Optical Coherence Tomography versus Stereoscopic Fundus Photography or Biomicroscopy for Diagnosing Diabetic Macular Edema: A Systematic Review
Author Affiliations
  • Gianni Virgili
    From the Department of Ophthalmology, University of Florence, Florence, Italy; the
  • Francesca Menchini
    Department of Ophthalmology University of Udine, Udine, Italy; and the
  • Andrea F. Dimastrogiovanni
    Department of Ophthalmology University of Udine, Udine, Italy; and the
  • Emilio Rapizzi
    From the Department of Ophthalmology, University of Florence, Florence, Italy; the
  • Ugo Menchini
    From the Department of Ophthalmology, University of Florence, Florence, Italy; the
  • Francesco Bandello
    Department of Ophthalmology University of Udine, Udine, Italy; and the
  • Raffaella Gortana Chiodini
    Fondazione G. B. Bietti per lo Studio e la Ricerca in Oftalmologia–IRCCS (Istituto Ricerca e Cura a Carattere Scientifico), Rome, Italy.
Investigative Ophthalmology & Visual Science November 2007, Vol.48, 4963-4973. doi:https://doi.org/10.1167/iovs.06-1472
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Gianni Virgili, Francesca Menchini, Andrea F. Dimastrogiovanni, Emilio Rapizzi, Ugo Menchini, Francesco Bandello, Raffaella Gortana Chiodini; Optical Coherence Tomography versus Stereoscopic Fundus Photography or Biomicroscopy for Diagnosing Diabetic Macular Edema: A Systematic Review. Invest. Ophthalmol. Vis. Sci. 2007;48(11):4963-4973. https://doi.org/10.1167/iovs.06-1472.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

purpose. To review systematically the sensitivity and specificity of optical coherence tomography (OCT) for diagnosing macular edema attributable to diabetic retinopathy compared with well-established gold standard tests such as fundus stereophotography or contact and noncontact fundus biomicroscopy.

methods. Medline and Embase were searched electronically and six major ophthalmic journals from 1998 to 2006 were hand searched. Two reviewers independently assessed trial searches, studied quality with the QUADAS (Quality Assessment of Diagnostic Accuracy Studies) checklist, and extracted data. The target disease was clinically significant macular edema (CSME) according to Early Treatment of Diabetic Retinopathy Study (ETDRS) criteria. A bivariate model was used to obtain summary estimates of sensitivity and specificity and fit a summary receiver operating characteristic (ROC) curve.

results. Fifteen studies were considered eligible. These studies were of good quality for most items of the QUADAS checklist, but most studies did not report masking of examiners and did not describe how withdrawals and undetermined results were treated. Seven studies included healthy control subjects, which could have artificially enhanced OCT diagnostic performance. All but one study included both eyes of the patients without taking into account the within-subject correlation in statistical analyses. Sensitivity and specificity data could be extracted from only 6 of 15 studies, because appropriate cross tabulations of index and reference tests were not reported by the others. In five of these studies, central retinal thickness cutoffs between 230 and 300 μm were adopted to define abnormal OCT results and considered the central type of CSME only, whereas in one study a complex algorithm accounting for extrafoveal CSME was used. The design of one study was case–control and was excluded from the meta-analysis. The expected operating point on the summary ROC, a pooled estimate of all studies, corresponded to a sensitivity of 0.79 (95% CI: 0.71–0.86), a specificity of 0.88 (95% CI: 0.80–0.93), a positive likelihood ratio of 6.5 (95% CI: 4.0–10.7), and a negative likelihood ratio of 0.24 (95% CI: 0.17–0.32). These values suggest a good overall performance of OCT for diagnosing CSME.

conclusions. OCT performs well compared with fundus stereophotography or biomicroscopy to diagnose diabetic macular edema. The quality of reporting of such studies should be improved, and authors should present cross tabulations of index and reference test results. Data adjusted for within-subject correlation should also be provided, although this issue represents a challenge for systematic reviewers.

Diabetes mellitus results in considerable morbidity and mortality, affecting approximately 180 million people worldwide. 1 The total number of people with diabetes is expected to rise to an estimated 300 million cases by 2025, with the most significant increases in developing countries, thought to be the result of population growth, ageing, obesity, and sedentary lifestyle. 2  
Approximately 25% of people with diabetes have at least some form of diabetic retinopathy, and the incidence increases with the duration of the diabetes. At 10 years, the prevalence of retinopathy in diabetic patients is 7%, after 25 years it is more than 90%. 3 In developed countries, diabetic eye disease represents the leading cause of blindness in adults under 75 years of age. 4 Diabetic macular edema increases with the duration of diabetes, and the prevalence is 5% within the first 5 years after diagnosis and 15% at 15 years. 3  
According to the Diabetic Retinopathy Study Group, the risk of severe visual loss at 2 years was 3.2% for eyes with nonproliferative diabetic retinopathy. 5 The presence of clinically significant macular edema (CSME) increases the risk of moderate visual loss to approximately 30% to 50%, depending on the level of baseline visual acuity. 6  
Diabetic macular edema was defined on the basis of stereoscopic fundus photography in ETDRS studies. 7 This technique is complicated and difficult to use in a clinical setting and was replaced with contact fundus biomicroscopy, which was found to be in close agreement with stereophotography, particularly for CSME. 8 Noncontact fundus biomicroscopy is more commonly used, since sophisticated fundus lenses have been proposed for binocular fundus observation during the past two decades, yet it has been shown to be slightly less sensitive than contact fundus biomicroscopy. 9  
Optical coherence tomography (OCT) has gained increasing popularity as an objective tool to measure retinal thickness and other aspects associated with macular edema. 10 11 12 Standard OCT assessment of diabetic macular edema has been adopted in multicenter trials in patients with diabetic retinopathy by the Diabetic Retinopathy Clinical Research network (DRCR.net). 13 An advantage of using OCT is its quantitative assessment, rather than the qualitative evaluation performed with photography or biomicroscopy. In fact, less than optimal diagnostic performances can be expected in a clinical practice when such qualitative methods are used by untrained physicians. It is worth summarizing the large amount of literature published on this topic in a systematic review to investigate whether OCT may perform as well as fundus photography or its surrogates to diagnose CSME, allowing it to become an objective and quantitative alternative to this gold standard. 
Methods
Purpose
The objective of this review was to evaluate the sensitivity and specificity of OCT for diagnosing macular edema attributable to diabetic retinopathy as assessed by well-established gold standard tests such as fundus stereophotography or fundus biomicroscopy. 
A secondary purpose of this review was to investigate potential sources of heterogeneity of the estimates of sensitivity and specificity among diagnostic studies. 
Definitions and Inclusion Criteria
Diabetes was defined as taking any glucose-lowering medication. The ETDRS definition of CSME was adopted as the definition of the target disease in this review, because it represents the main indication of focal or grid laser photocoagulation. 7 The most common type of CSME is central diabetic macular edema (CDME), defined as retinal thickening within 500 μm of the center of the macula or, alternatively, hard exudates within 500 μm from the center of the macula with thickening of the adjacent retina. The noncentral type of CSME is less common and is defined as a zone of retinal thickening, 1 disc area or larger, any portion of which is located within 1 disc diameter from the center of the macula. 
The reference tests considered as a valid gold-standard for this review were stereoscopic fundus photography and contact lens or noncontact lens biomicroscopy of the fundus. The index test is be OCT, either low-resolution (OCT 2000, the second-generation model) or high-resolution (Stratus OCT, the third-generation model). Studies were included if these definitions of target disease, reference test, and index test were met. 
Search Strategy
Since predefined searches may not retrieve all studies in diagnostic reviews 14 we searched Medline (PubMed 1966–September 2006) and Embase (embase.com 2002–October 2006) using the strategy displayed in Table 1 . We also hand searched the index of the following journals from 1998 to 2006: Ophthalmology, Archives of Ophthalmology, American Journal of Ophthalmology, Investigative Ophthalmology and Visual Science, British Journal of Ophthalmology, and Retina. We also hand searched the references of the articles obtained in full. 
Study Selection
The assessment of the titles and abstracts for eligibility was conducted by two independent reviewers (GV, AFD). Articles of potential interest were retrieved and their inclusion was reassessed. Disagreement at each step was resolved with discussion between the two reviewers. 
The methodologic quality of the included studies was performed independently by two reviewers (FM, AFD) based on the QUADAS (Quality Assessment of Diagnostic Accuracy Studies) checklist. 15 16 Disagreement on study quality was resolved by a third author (GV). 
Assessment of Heterogeneity, Subgroup Analyses, and Sensitivity Analyses
We planned to evaluate the following sources of heterogeneity if data were available: clinical setting (primary ophthalmic care versus retina specialist), diabetic retinopathy severity, OCT threshold or algorithm used, type of OCT, type of gold standard. A further source of heterogeneity assessed in this review was also the type of study, especially cohort studies versus case–control studies. In fact, it is well known that the latter can yield more optimistic estimates of diagnostic performance due to the sharp separation of the measurements of diseased patients from healthy control subjects. 17 18  
Because the ETDRS definition of CSME applies to all types of patients with diabetic retinopathy, we did not exclude specific subtypes of patients in this review. Finally, we did not take into account the role of the biomicroscopic finding of a thickened hyaloid associated with diabetic macular edema on the diagnostic performance of OCT. 
Sensitivity analyses were planned with the exclusion of studies of lower methodological quality and the exclusion of case–control studies. 
Data Analyses
Descriptive statistics and graphs were produced using MetaDiSc, software presented in an open-source, peer-reviewed, methodologic journal 19 which can be downloaded free at http://www.hrc.es/investigacion/metadisc_en.htm. Forest plots of sensitivity, specificity, positive likely ratios, and negative likely ratios were presented for descriptive purposes (see the Glossary). To pool diagnostic odds ratios (DORs, see the Glossary) we planned to use fixed effects if three or fewer articles were found and random effects if more than three were found. Heterogeneity was assessed using the χ2 test. Sources of DOR heterogeneity were investigated using multilevel logistic regression as suggested by Siadaty et al. 20  
The primary analysis in this review was the pooling of pairs of sensitivity and specificity in a random-effects bivariate model, as proposed by Reitsma et al., 21 obtaining a heterogeneous summary receiver operating characteristic (ROC) curve. The bivariate model uses a random effects approach in the estimation of summary estimates of sensitivity and specificity and their corresponding 95% CIs. Because the bivariate model estimates the strength and the shape of the correlation between sensitivity and specificity, we can either draw a 95% confidence ellipse around their mean values or draw a 95% prediction ellipse for individual values of sensitivity and specificity. 21 This model overcomes the statistical shortcomings of Moses’ summary ROC approach 22 and may be a simpler alternative to hierarchical ROC models, 23 which cannot be fit with many common statistical packages. 
Finally, we present the results of a clinical decision-making model 24 that helps the clinician develop a rational, quantitative approach to the use of OCT for diagnosing CSME. 
Analyses were conducted with commercial statistical software (Stata ver. 9.2; StataCorp, College Station, TX). 
Results
Search of Studies and Study Flow
The electronic search strategy is displayed in Table 1 . From this search, 29 articles were obtained in full. Of these, 14 were excluded 25 26 27 28 29 30 31 32 33 34 35 36 37 38 for various reasons presented in Table 2
Characteristics of Included Studies
The 15 included studies 9 12 39 40 41 42 43 44 45 46 47 48 49 50 51 are shown and described in Table 3 . These studies enrolled 1297 patients (2044 eyes). Twelve of 15 studies evaluated 1449 eyes, 627 with CSME and 822 without CSME. There were only two eyes with CSME of 131 in Strom et al., 48 which was a reproducibility study with no sensitivity and specificity data for diagnosing CSME. Two more studies 50 51 analyzed 45 and 59 eyes, respectively, without providing cross-tabulation data on OCT performance. 
Only six studies 9 12 39 40 41 44 reported a cross tabulation of positive/negative OCT results and presence/absence of CSME. The data presented by Ozdek et al. 43 could not be used, since they used diabetic macular edema and not CSME as a definition of the target disease. 
The type of gold standard varied and included stereophotography, contact or noncontact biomicroscopy, and a combination of these methods (Table 3)
Criteria Used to Define Positive versus Negative OCT Results
Data based on different cutoffs of retinal thickness at the central point were extracted for four studies: 300 μm in Brown, 44 250 μm in Hee et al. 12 and Gaucher et al., 41 and 265 μm in Browning et al. 9 Goebel and Franke 40 adopted a cutoff of 230 μm of retinal thickness in the central retinal subfield, using the automated macular thickness map protocol built into the OCT software. All of these studies presented data on the central type of CSME—that is, CDME. Sadda et al. 39 used the Macular Grid 5 scanning protocol and a complex diagnostic algorithm that accounted for a separate recognition of CDME and the noncentral type of CSME. 
Hee et al. 12 Brown et al. 44 and Gaucher et al. 41 reported data for more than one cutoff value. We extracted data based on the cutoff points mentioned, since they provided a good number of eyes in each cell of the cross-tabulation. 
Methodological Quality of Included Studies: QUADAS Checklist
The methodological quality of the included studies is shown in Table 4 , in which the quality issue of each QUADAS item is presented. It can be seen that items related to sample recruitment and representativeness, inclusion criteria, patient flow, index and reference test execution were of good quality for all or most studies. However, it was unclear in most studies if the gold standard (stereophotography or biomicroscopy) was interpreted independent of OCT (index test) and vice versa, with three studies stating that OCT interpretation was unmasked to gold standard results. We think that it is unlikely that this is an important source of bias for OCT interpretation, given that OCT provides an objective quantitative estimate of retinal thickness and the assessment procedure is sufficiently standardized. However, the gold-standard evaluation, which is a subjective one, could have been affected by OCT results if the examiners were unmasked. More important, the report and handling of uninterpretable test results, such as low-quality OCT examination with unreliable thickness estimate, was unclear or poorly reported in most studies. 
Methodological Quality of Included Studies: Additional Items
Additional information concerning study quality that is not included in the QUADAS checklist is shown in Table 3 . Although most studies were prospective, in six studies the authors did not declare whether all consecutive and eligible patients seen within the beginning and end of the study were enrolled. Since studies were conducted at referral centers, it is unclear how this could have influenced the characteristics of the sample which is already subject to selection due to referral. 
Another issue was a mixed design in seven studies, which enrolled both patients with diabetic retinopathy (with or without CSME) and healthy control subjects. This is expected to enhance the accuracy of OCT, but it does not correspond to its use in clinical practice, during which ophthalmologists want to rule in or rule out the presence of CSME in cases of persons with diabetic retinopathy who are at risk of having this macular complication. As already stated, research has shown that the risk of overestimating the performance of the index test is high in case–control studies. 17 18 The study by Gaucher et al. 41 was classified as having a case–control design, since patients were enrolled as two groups with either presence or absence of CSME, the latter including a majority of diabetic subjects with no retinopathy. 
We considered an additional statistical issue that is specific to ophthalmology (i.e., how the correlation between eyes of the same patient is treated). All but one study (Goebel and Kretzchmar-Gross 45 ) included both eyes of some patients, but no study presented estimates that took into account the within-patient correlation. 
Finally, sample size calculations were not reported in any study, a well-known issue in diagnostic research. 52  
Results of Analyses
A cross tabulation of dichotomous results with the index and reference test could be obtained for six studies. Raw data and additional study features are available in Table 5
Individual Study Results.
Individual estimates and plots of sensitivity, specificity, positive and negative likelihood ratios, and the DOR are displayed in Figure 1for each study. 
Gaucher et al. 41 yielded a maximum sensitivity (100%) and a nearly maximum specificity (96%), which is typical of case–control studies, often providing a biased estimate of diagnostic performance. 17 18  
DORs, Meta-analysis, and Assessment of Heterogeneity.
We initially assessed DOR heterogeneity among the six studies providing data for the meta-analysis. Heterogeneity was high if all six studies were included, since the χ2 value was 49.5%, which is close to the threshold of 50% and is suggestive of statistical heterogeneity, 53 even when the probability only approaches statistical significance (such as in this case; P = 0.078), due to the small number of included studies. 
After exclusion of the case–control study by Gaucher et al., 41 as explained in the Methods section, there was no DOR heterogeneity (χ2 = 0.0%). Therefore, we excluded this study from further analyses. 
The pooled DOR estimate using random effects was 27.1 (95% CI: 17.4–42.2), which is consistent with a good performance of OCT. Of note, the results would not have changed had we incorporated data from Ozdek et al. 43 in the meta-analysis, despite their using diabetic macular edema and not CSME for disease definition, since its estimates of sensitivity and specificity were identical with those in Hee et al. 12  
Because only five studies provided data, and no heterogeneity was found, we could not evaluate sources of heterogeneity consistently. We compareed only two studies 12 39 which used the OCT 2000 with three studies using the Stratus OCT. 9 40 45 The same two studies also included healthy control subjects. Their diagnostic performance did not differ from that of the other three studies in a multilevel logistic regression model. 24  
There was also no power to investigate the asymmetry of the summary ROC curve (i.e., the fact that the performance of OCT—the DOR—may be better at high levels of specificity rather than at high levels of sensitivity). 54 55  
Assessment of Threshold Effect of Retinal Thickness Cutoff Values.
A threshold effect was suggested by the typical shoulder arm displayed by the studies in the ROC plane (Fig. 2)and supported by the good correlation (0.61) of the logit-transformed sensitivity and specificity, 21 although it did not reach statistical significance (P = 0.285), due to the small number of studies included. 
In four studies using central retinal thickness, the estimate was located as expected from a threshold effect due to the cutoff value in the ROC plane. In fact, Brown et al. 44 yielded the highest specificity with a thickness threshold of 300 μm and Goebel and Franke 40 obtained the lowest specificity using a cutoff of 230 μm. 
Sadda et al. 39 did not use thickness and yielded the highest DOR among all by adopting a complex algorithm accounting for the noncentral type of CSME. 
Summary ROC Curve.
Figure 2presents the summary ROC curve based on a bivariate model. 21 If there is no heterogeneity between the studies, such as in our meta-analysis, the best summary estimate of test performance will be a single point on the summary ROC curve (the operating point). 54 The expected operating point on the summary ROC corresponded to a sensitivity of 0.79 (95% CI: 0.71–0.86), a specificity of 0.88 (95% CI: 0.80–0.93), a positive likelihood ratio of 6.5 (95% CI: 4.0–10.7) and a negative likelihood ratio of 0.24 (95% CI: 0.17–0.32). The pooled DOR was 27.7 (95% CI: 17.0–45.3) with this method. 
A Clinical Decision Model: Applying the Results to Clinical Practice
A decision model can help clinicians place the diagnostic performance of a test in a realistic scenario by integrating its diagnostic power in a question about treatment. 24 The current clinical question would be whether a positive or negative OCT result changes the decision to use or not to use photocoagulation on a patient in whom CDME is suspected. 
In a diagnostic clinical decision model, we first estimate the treatment threshold 24 based on the available literature (i.e., the level of suspicion or probability of CDME, above which the benefits of applying focal or grid photocoagulation outweigh the harm). We computed the treatment threshold by using data from the Early Treatment of Diabetic Retinopathy Study (ETDRS), 7 in which grid/focal laser photocoagulation reduced the risk of losing 3 or more lines of visual acuity from approximately 30% to 15% at 3 years. This 15% difference represents the benefit. An estimate of the cost (i.e., the probability of harm when applying treatment to patients, including those incorrectly diagnosed) was not available in the ETDRS or other studies. We arbitrarily used the value of 5% to weigh the loss of the perimacular visual field, the patient’s discomfort, and the rare risk of accidental foveal photocoagulation or occurrence of choroidal neovascularization. This method would result in a cost/benefit ratio of 0.33 and a treatment threshold at 25% suspicion of CDME, which may be clinically reasonable for such a safe treatment. 
We then computed the range of pretest probability within which, using likelihood ratios, a positive or a negative test result can change the treatment decision by, respectively, increasing the probability above the treatment threshold or decreasing it below it. The reader can refer to classic evidence-based medicine books to obtain graphical methods or simple formulas for this purpose. 56 57 Using the likelihood ratios derived from our meta-analysis, the range of usefulness of OCT for diagnosing CDME would be from 5% to 58%, which covers levels of low to medium pretest probability or suspicion of disease. The fact that this range mostly coincides with the prevalence of CDME—the pretest probability—in the included studies is reassuring both for the possibility of generalizing the results and for the fact that OCT was indeed used by those authors for patients with this level of diagnostic uncertainty. Indeed, it would be incorrect to use OCT when deciding on treatment if very little uncertainty exists, such as when CDME is almost definitely present or absent with biomicroscopy. 
Discussion
OCT is increasingly used in clinical practice to study central retinal thickness and morphology in numerous diseases. It has been shown to be a reproducible and objective test in several studies, 10 58 59 and its performance improved when the high-resolution Stratus OCT version was made commercially available in 2002 (Stratus OCT [OCT3]; Carl Zeiss Meditec, Dublin, CA). 
We found a relatively large number of studies (n = 15) that assessed the performance of OCT for diagnosing CSME in diabetic patients, comparing it with well-established gold standard tests such as stereoscopic fundus photography or fundus biomicroscopy. However, few studies (n = 6) presented sufficient data to extract sensitivity and specificity estimates based on one or more cutoffs of OCT retinal thickness at the central point or central area or using other OCT criteria. In three studies, mean retinal thickness and its SD were presented, and in three more studies these values were given for subgroups of retinopathy severity or healthy control subjects. Such studies cannot be used for meta-analysis. 
The results of this systematic review suggest that OCT is a useful tool for diagnosing suspect CDME. The central pooled estimate of our meta-analysis yielded positive and negative likelihood ratios of 6.5 and 0.24, respectively. These values are close to the corresponding values of 5 and 0.2 which were believed to be convincing evidence, but they do not reach the levels of 0.1 and 10 which were considered strong evidence. 54 60  
Decision Criteria to Use with OCT
We found that the five studies used to obtain the summary ROC curve provided results that suggested a threshold effect. Unfortunately, not all studies used the same OCT diagnostic criteria, nor did they present data for more than one retinal thickness cutoff. Furthermore, a threshold effect can also be due to patients’ characteristics, rather than to the cutoff value of macular thickness in each study (implicit threshold effect). 54  
Values of 250 to 265 μm were adopted for three of five studies in our meta-analysis and a thickness of 250 μm has also been used as an OCT inclusion criterion in most studies on CSME by the DRCR.net. 13 In our review, the study adopting a cutoff of 300 μm obtained the highest specificity. Evidence-Based Medicine 57 advises the use of the acronyms SpPIn and SnNOut to remind readers that a valid test should be calibrated at high specificity (positive result) when the purpose is to rule In disease and at high sensitivity (negative result) to rule out disease. 55 56 57 We suggest that, in cases with clinical uncertainty about the existence of CDME, macular photocoagulation should not be used if central retinal thickness is below 250 μm, whereas a value of 300 or more strongly indicates the need for treatment. The pretest probability of CDME should always be taken into account (e.g., one should not delay photocoagulation if a thickness slightly inferior to 250 μm is found in a patient with very high clinical suspicion of CSME), especially if the noncentral type has been identified. In fact, it may be thought that using central retinal thickness results in a poor performance in detecting the second ETDRS definition of CSME, which includes a large patch of retinal edema that does not involve the central part of the macula. The algorithm used by Sadda et al. 39 was of interest and deserves further study. The other studies in our meta-analysis did not report data on OCT performance for subgroups of types of CSME, since they were based on data on the central type only (i.e., CDME). In the discussion, the authors of one study 44 observed that it is not known whether results would have been similar if macular edema not involving the fovea had been evaluated. They also stated that other studies 11 45 demonstrate a strong correlation between extrafoveal and foveal thickness measurements, suggesting that results would be similar. 
Validity and Reproducibility of OCT and the Gold Standard
Validity and reproducibility are key issues in therapeutic studies, as in diagnostic studies. A study is valid insofar as the results represent an unbiased estimate of the underlying truth. Reproducibility is the ability of a measure to yield the same result when reapplied to stable patients. 54 In diagnostic accuracy studies, an imperfect gold standard may limit the performance of the index test. Fundus biomicroscopy may be less reliable when performed by ophthalmologists with diverse training, compared with the high reproducibility of OCT. 58 59 The coefficient of reproducibility of OCT in eyes with diabetic macular edema was shown to be less than 10% with the fast macular thickness mapping protocol of Stratus OCT. 61 Clinicians should consider the reliability indexes provided by the Fast Retinal Thickness Map of the Stratus OCT software and average data from semiautomatic measurements in cases with low reliability. The strength of the signal and the SD displayed in the OCT map can help for this purpose. 
Risk of Bias in This Review
A few issues regarding the quality of the studies included in this review may be of concern. Although the methodological quality of the eligible studies was good overall for most QUADAS items, the masking of index and reference test was not reported in most studies. Uninterpretable results and withdrawals were also poorly or not reported (four of five studies with data used for meta-analysis). Nonetheless, the number of such patients was negligible, and so this is not expected to represent an important source of bias. 
The fact that some studies included healthy control subjects was also of concern (two of five studies with data used for meta-analysis). Their inclusion is expected to enhance the accuracy of OCT, but it does not reproduce its clinical use, in which ophthalmologists want to rule in or rule out CSME in patients with diabetic retinopathy who are at risk of having this macular complication. 
A specific quality issue not covered by the QUADAS checklist is the lack of statistical adjustment for the correlation between the patient’s eyes in all studies. This deficit is expected to yield too optimistic estimates of the standard errors (i.e., too narrow). To the best of our knowledge, methods for obtaining sensitivity and specificity values or ROC curves that take into account correlated data are available, 62 but how these should be treated by a systematic reviewer has not been standardized in specific guidelines. 
Implications for Practice
This review suggests that OCT can be used to diagnose CSME, particularly its central type or CDME, and decide on laser photocoagulation in patients with intermediate suspicion of disease. The strength of our conclusion is limited by the fact that data could be extracted from only a fraction of the published literature due to limitations in the reporting of the included papers. Furthermore, the precision of our estimate is probably worse than that reported due to the failure to take into the account correlation between eyes of the same patient in all the included studies. 
Implications for Research
Several suggestions for further research can be drawn from this review. Most of them refer to the quality of reporting and highlight the need to disseminate the STARD (Standards for Reporting of Diagnostic Accuracy) initiative 63 among authors of diagnostic studies in ophthalmology. 
The inclusion criteria should be made explicit after framing the clinical question of interest. Including patients with any level of diabetic macular edema was common among the included studies and seems a sensible choice when OCT is used as the last objective diagnostic assessment before deciding on the indication to administer photocoagulation. The inclusion of healthy control subjects or diabetics without retinopathy should be avoided. A case–control design should also be avoided, as previously suggested. 17 18  
The availability of visual acuity and other clinical data to graders of fundus photographs should be disclosed by the authors. 64 Although it may be intuitively reasonable to allow clinical information in a diagnostic study, in fact this information was shown to either enhance or impair the interpretation of a test. 64  
The reporting of results should include sensitivity and specificity data for more than one cutoff value of OCT retinal thickness, and an ROC curve should be plotted. Furthermore, we suggest that a cross tabulation of the data also be displayed with 25-μm intervals for thicknesses between 200 and 350 μm. The diagnostic performance of OCT for detecting the central and the noncentral types of CSME should be reported separately in subgroup analyses. More complex but less common OCT decision algorithms should also be reported separately and compared to the performance of the thickness criterion. 
Finally, the correlation between eyes of the same patient should be taken into account in estimates of sensitivity, specificity, and DOR. However, this process will pose particular problems to systematic reviewers, which could be solved by providing them with individual patient data or making datasets available as additional files on the journal’s Web site. 
Our review has summarized the performance of OCT in cross-sectional diagnostic accuracy studies adopting the current, although evolving, methodology of diagnostic systematic reviews with the purpose of investigating whether OCT is as valid as stereophotography and biomicroscopy for detecting CSME. However, such diagnostic accuracy studies are only one of the steps in the evaluation of diagnostic technologies, which was categorized into six levels from studies that address technical feasibility (level 1) to those that address societal impact (level 6). 65 After diagnostic accuracy has been established with respect to traditional gold standards, longitudinal studies of outcome impact are needed to enable incorporation of a new diagnostic test into prognostic and therapeutic clinical pathways. Similar studies by the DRCR.net will help assess whether OCT can replace fundus biomicroscopy as a new objective and quantitative gold standard for diagnosing diabetic macular edema. 13  
Appendix 1
Glossary
Term definitions are as follows 54 56 :
  •  
    Sensitivity: proportion of diseased patients who have positive test results (true positives/total diseased);
  •  
    Specificity: proportion of nondiseased persons who have negative test results (true negatives/total nondiseased);
  •  
    Positive predictive value (PPV): proportion with positive test results who actually have the disease (true positives/total positive);
  •  
    Negative predictive value (NPV): proportion with negative test results who actually do not have the disease (true negatives/total negative);
  •  
    Positive likelihood ratio (LR+): likelihood that a person with disease would have a positive test result compared with a person without disease (true positives/total diseased)/(false positives/total nondiseased) = sensitivity/(1 – specificity);
  •  
    Negative likelihood ratio (LR−): likelihood of a person with disease having a negative test result compared with a person without disease (false positives/total diseased)/(true negatives/total nondiseased) = (1 − sensitivity)/specificity;
  •  
    Diagnostic odds ratio (DOR): odds of positivity among diseased persons divided by the odds of positivity among nondiseased persons; positive likelihood ratio/negative likelihood ratio;
  •  
    SnNout: when a sign/test/symptom has a high sensitivity (Sn), a negative result (N) can help rule out (out) the diagnosis;
  •  
    SpPin: when a sign/test/symptom has a high specificity (Sp), a positive result (P) can help to rule in (in) the diagnosis;
  •  
    Receiver Operating Characteristic (ROC) curve: a figure depicting the power of a diagnostic test. The ROC curve presents the test’s true-positive rate (i.e., sensitivity) on the horizontal axis and the false-positive rate (i.e., 1 – specificity) on the vertical axis for different cutoffs dividing a positive from a negative test. An ROC curve for a perfect test has an area under the curve = 1.0, whereas a test that performs no better than by chance has an area under the curve of only 0.5. A summary ROC curve is obtained by pooling the sensitivity/specificity pairs of each study.
 
Table 1.
 
Search Strategy Using Medline and Embase
Table 1.
 
Search Strategy Using Medline and Embase
No. Query Results
Medline (PubMed 1966–September 2006)
 1 “Tomography, Optical Coherence” [MeSH] 1,464
 2 “Optical coherence tomography” 2,413
 3 OCT Field: Title/Abstract 5,337
 4 1 OR 2 OR 3 6,497
 5 “Macular edema*” Field: Title/Abstract 2,523
 6 “Macular oedema*” Field: Title/Abstract 574
 7 “Macular Edema, Cystoid” [MeSH] 1,689
 8 5 OR 6 OR 7 3,483
 9 4 AND 8 409
 10 “Diabetes Mellitus” [MeSH] 201,736
 11 Diabete* OR Diabetic* Field: Title/Abstract 231,923
 12 10 OR 11 271,713
 13 9 AND 12 193
Embase (embase.com 2002–October 2006)
 1 ‘Optical coherence tomography’/exp AND [2002–2006]/py 2,389
 2 oct AND [2002–2006]/py 3,800
 3 1 OR 2 AND [2002–2006]/py 5,035
 4 ‘Macular edema’ AND [2002–2006]/py 1,204
 5 ‘Macular oedema’ AND [2002–2006]/py 238
 6 Retina macula cystoid edema, cystoid’/exp AND [2002–2006]/py 715
 7 4 OR 5 OR 6 AND [2002–2006]/py 1,665
 8 3 AND 7 AND [2002–2006]/py 403
 9 ‘Diabetes mellitus’/exp AND [2002–2006]/py 88,702
 10 Diabete* OR diabetic* AND [2002–2006]/py 107,084
 11 9 OR 10 AND [2002–2006]/py 107,460
 12 8 AND 11 AND [2002–2006]/py 187
Table 2.
 
Characteristics of Excluded Studies
Table 2.
 
Characteristics of Excluded Studies
Study Reasons for Exclusion
Pierre-Kahn et al. 25 No gold standard
Browning and Fraser 26 No gold standard
Ciardella et al. 30 Only diseased subjects included (all patients with CSME)
Panozzo et al. 28 No gold standard
Kang et al. 27 Only diseased subjects included (all patients with CSME)
Browning 31 No gold standard
Moreira et al. 32 Unclear gold standard (undefined indirect ophthalmoscopy)
Rivellese et al. 35 No gold standard
Schaudig et al. 29 No gold standard
Brancato 38 No gold standard
Schaudig et al. 36 Review article
Pal et al. 37 Qualitative study on various retinal disease
Massin et al. 33 Patients had diabetic retinopathy but not CSME
Massin et al. 34 Small reproducibility study
Table 3.
 
Characteristics of Included Studies
Table 3.
 
Characteristics of Included Studies
Study ID Design (CSME Used to Define Diseased Eyes) Disease Severity of Analyzed Patients (Number of Eyes) Type of OCT Type of Reference Test Patients (Eyes) Analyzed Patients/Eyes Eligible but Not Analyzed
Goebel and Franke 40 Prospective cohort and healthy controls Various level of DR (124) and healthy controls (13) OCT 2000 Stereophotography 82 (124) 13 eyes excluded for poor photographic quality
Sadda et al. 39 Retrospective cohort Patients with DME Stratus OCT Biomicroscopic examination (not defined) and stereophotography 40 (71) None
Gaucher et al. 41 Retrospective case-control 35 diabetic patients with DME and 35 diabetic patients without DME (27 with no DR) Stratus OCT Contact and noncontact biomicroscopy 70 (98) None
Hussain et al. 42 Prospective cohort Various level of DR (83 NPDR, 22 PDR) Stratus OCT Biomicroscopic examination and fundus photography (not defined) 55 (105) Five eyes excluded for vitreous hemorrhage (4) and epiretinal membrane (1)
Ozdek et al. 43 Retrospective cohort Various level of DR (148 NPDR, 47 PDR) Stratus OCT Noncontact biomicroscopy 110 (195) Unclear if all consecutive and eligible patients were included
Brown et al. 44 Prospective cohort Various level of DR Stratus OCT Contact lens fundus biomicroscopy 95 (172) Ten patients declined participation two did not perform OCT
Browning et al. 9 Prospective cohort Patients with DME Stratus OCT Noncontact fundus biomicroscopy 80 (143) Unclear if all consecutive and eligible patients were included
Goebel and Kretzchmar-Gross 45 Prospective cohort and healthy controls Various levels of DR (136) and healthy controls (30) OCT 2000 Contact and noncontact fundus biomicroscopy 166 (166) Fourteen patients due to improper fixation during OCT
Lattanzio et al. 46 Prospective cohort and healthy controls Various levels of DR (116 NPDR, 54 PDR), and diabetes without DR (46) OCT 2000 Slit-lamp biomicroscopy and stereophotography 137 (216) None
Sanchez-Tocino et al. 47 Prospective cohort Various levels of DR (54 NPDR, 21 PDR, 28 DR and CSME), diabetes without DR (45) and healthy controls (44) OCT 2000 Biomicroscopy with slit-lamp and fundus lens 111 (192) Unclear if consecutive patients, no statement on inclusion of all eligible patients
Strom et al. 48 Prospective cohort and healthy controls 47 Patients with DME (80 without CSME, 4 CSME) and 25 healthy controls (33 eyes) OCT 2000 Stereophotography 71 (117) Twelve eyes excluded for various reasons
Yang et al. 49 Prospective cohort and healthy controls Patients with DME (19 CSME, 14 no CSME) OCT 2000 Stereophotography and noncontact biomicroscopy 17 (33) None
Yamamoto et al. 50 Prospective cohort and healthy controls Patients with DME (45, not all with CSME) and healthy controls (21) OCT 2000 Slit lamp biomicroscopy 46 (66) Unclear if all consecutive and eligible patients were included
Otani et al. 51 Prospective cohort and healthy controls Patients with DME (59, not all with CSME) and healthy controls (10) OCT 2000 Biomicroscopy 52 (69) Unclear if all consecutive and eligible patients were included
Hee et al. 12 Cohort and healthy controls (unclear if prospective) Various level of DR (148 NPDR, 34 PDR), diabetes without DR (55), healthy controls (73) OCT 2000 Noncontact biomicroscopy 189 (310) Unclear if all consecutive and eligible patients were included
Table 4.
 
Methodological Quality of Included Studies
Table 4.
 
Methodological Quality of Included Studies
Table 5.
 
Included Studies Data
Table 5.
 
Included Studies Data
Study Eyes with CSME (Prevalence, %) Eyes without CSME (n) Sensitivity TP/(TP + FN) Specificity TN/(TN + FP) OCT Thickness Threshold (μm) Mean CRT/CRA (SD)
CSME No CSME
Goebel and Franke 40 * 64 (57.7) 47, ‡ 82.8% 53/64 76.6% 36/47 230 μm (central subfield) NA NA
Sadda et al. 39 35 (55.6) 28 88.6% 31/35 85.7% 24/28 NA 317 (77) 251 (47)
Gaucher et al. 41 49 (51.5) 46 100% 0/49 95.7% 2/46 250 μm (central point) 451.2 (104.8) 191.5 (29.0)
Hussain et al. 42 50 (47.6) 55 NA NA NA NA NA
Ozdek et al. 43 112 (57.4) 83 NA NA NA (data available for CME only) NA NA
Brown et al. 44 , † , ‡ 33 (19.2) 139 66.7% 22/33 95.7% 133/139 300 μm (central point) NA NA
Browning et al. 9 50 (35.0) 93 84.0% 42/50 81.7% 76/93 265 μm (central point) NA NA
Goebel and Kretzchmar-Gross 45 116 (85.3) 20 NA NA NA 333 (131) NA
Lattanzio et al. 46 NA NA NA NA NA NA NA
Sanchez-Tocino et al. 47 28 (13.9) 174 NA NA NA 371 (180) NA
Strom et al. 48 14 (42.4) 19 NA NA Spongelike changes used, not thickness 255.6 (138.9) 174.6 (38.2)
Yang et al. 49 131 CSME definition not used for reporting data (only 2 CSME eyes)
Yamamoto et al. 50 45 CSME definition not used for reporting data
Otani et al. 51 59 CSME definition not used for reporting data
Hee et al. 12 , ‡ , § 76 (39.8) 115, § 73.7% 56/76 91.3% 105/115 250 μm CTP NA NA
Figure 1.
 
Individual estimates of sensitivity, specificity, positive and negative likelihood ratios, and DORs for the six studies for which the data were complete. For Gaucher et al., 41 the presence of a cell with a null value was treated adding 0.5 to that cell to compute all accuracy parameters.
Figure 1.
 
Individual estimates of sensitivity, specificity, positive and negative likelihood ratios, and DORs for the six studies for which the data were complete. For Gaucher et al., 41 the presence of a cell with a null value was treated adding 0.5 to that cell to compute all accuracy parameters.
Figure 2.
 
Summary ROC curve pooling the results of five studies included in the meta-analysis, which are shown as gray circles proportional to the sample size. The square symbol in the center of the ellipses is the expected operating point on the summary ROC. The inner ellipse (long-dashed line) is the 95% confidence region of the pooled estimate. The outer ellipse (short-dashed line) is the prediction region of the pairs of sensitivity and specificity on the heterogeneous summary ROC curve, which shows the typical shoulder arm morphology. The summary ROC curve is not presented in full because only the portion within the area of the studies should be considered, and no inference can be drawn outside the data of the meta-analysis.
Figure 2.
 
Summary ROC curve pooling the results of five studies included in the meta-analysis, which are shown as gray circles proportional to the sample size. The square symbol in the center of the ellipses is the expected operating point on the summary ROC. The inner ellipse (long-dashed line) is the 95% confidence region of the pooled estimate. The outer ellipse (short-dashed line) is the prediction region of the pairs of sensitivity and specificity on the heterogeneous summary ROC curve, which shows the typical shoulder arm morphology. The summary ROC curve is not presented in full because only the portion within the area of the studies should be considered, and no inference can be drawn outside the data of the meta-analysis.
The authors thank the following colleagues: Vanna Pistotti (Trial Search Coordinator at the Italian Cochrane Centre, Mario Negri Institute, Milan, Italy) for suggestions about the search strategy; David Gaucher (Service d’Ophthalmologie de l’hôpital Lariboisière, Paris, France) for providing additional data on his study; Piergiorgio Duca and Giovanni Casazza (Department of Medical Statistics and Biometry, University of Milan, Italy) for comments on the paper; Roger Harbord (MRC Health Services Research Collaboration and Department of Social Medicine, University of Bristol, UK) for providing the nonofficial Stata Command fitting the bivariate model metandi; and Sarah Harvey for editing the manuscript. 
World Health Organisation. The World Health Report 2002. ;http://www.who.int/whr/en
KingH, AubertRE, HermanWH. Global burden of diabetes, 1995–2025: prevalence, numerical estimates, and projections. Diabetes Care. 1998;21:1414–1431. [CrossRef] [PubMed]
AielloLP, GardnerTW, KingGL, BlankenshipG, CavalleranoJD, FerrisFL, III. Diabetic retinopathy. Diabetes Care. 1998;21:143–156. [CrossRef] [PubMed]
National Institute of Diabetes and Digestive and Kidney Disease. Diabetes in America. 1995; 2nd ed.National Institutes of Health, National Institute of Diabetes and Digestive and Kidney Disease Bethesda, MD.
The Diabetic Retinopathy Study Research Group. Indications for photocoagulation treatment of diabetic retinopathy. DRS Report Number 14. Int Ophthalmol Clin. 1987;27:239–253. [CrossRef] [PubMed]
JavittJC, CannerJK, SommerA. Cost effectiveness of current approaches to the control of retinopathy in type I diabetics. Ophthalmology. 1989;96:255–264. [CrossRef] [PubMed]
Early Treatment Diabetic Retinopathy Study Research Group. Photocoagulation treatment for Diabetic Macular Edema. ETDRS Report Number 1. Arch Ophthalmology. 1985;103:1796–1806. [CrossRef]
KinyounJ, BartonF, FisherM, HubbardL, AielloL, FerrisF. Detection of diabetic macular edema: ETDRS report number 5. Ophthalmology. 1989;96:746–751. [CrossRef] [PubMed]
BrowningDJ, McOwenMD, BowenRM, Jr, O’MarahTL. Comparison of the clinical diagnosis of diabetic macular edema with diagnosis by optical coherence tomography. Ophthalmology. 2004;111:712–715. [CrossRef] [PubMed]
KoozekananiD, RobertsC, KatzSE, HerderickEE. Intersession repeatability of macular thickness measurements with the Humphrey 2000 OCT. Invest Ophthalmol Vis Sci. 2000;41:1486–1491. [PubMed]
HeeMR, PuliafitoCA, WongC, et al. Quantitative assessment of macular edema with optical coherence tomography. Arch Ophthalmol. 1995;113:1019–1029. [CrossRef] [PubMed]
HeeMR, PuliafitoCA, DukerJS, et al. Topography of diabetic macular edema with optical coherence tomography. Ophthalmology. 1998;105:360–370. [CrossRef] [PubMed]
Diabetic Retinopathy Clinical Research Network Web site. http://public.drcr.net/ Accessed on February 20, 2007
LeeflangMMG, ScholtenRJPM, RutjesAWS, ReitsmaJB, BossuytPMM. Use of methodological search filters to identify diagnostic accuracy studies can lead to the omission of relevant studies. J Clin Epidemiol. 2006;59:234–240. [CrossRef] [PubMed]
WhitingP, RutjesAWS, ReitsmaJB, BossuytPMM, KleijnenJ. The development of QUADAS: a tool for the quality assessment of studies of diagnostic accuracy included in systematic reviews. BMC Med Res Method. 2003;3:25. [CrossRef]
WhitingPF, WeswoodME, RutjesAWS, ReitsmaJB, BossuytPNM, KleijnenJ. Evaluation of QUADAS, a tool for the quality assessment diagnostic accuracy studies. BMC Med Res Method. 2006;6:9. [CrossRef]
WhitingP, RutjesAW, ReitsmaJB, GlasAS, BossuytPM, KleijnenJ. Sources of variation and bias in studies of diagnostic accuracy: a systematic review. Ann Intern Med. 2004;140:189–202. [CrossRef] [PubMed]
LijmerJG, MolBW, HeisterkampS, et al. Empirical evidence of design-related bias in studies of diagnostic tests. JAMA. 1999;282:1061–1066. [CrossRef] [PubMed]
ZamoraJ, AbrairaV, MurielA, KhanKS, CoomarasamyS. Meta-DiSc: a software for meta-analysis of test accuracy data. BMC Med Res Method. 2006;6:31. [CrossRef]
SiadatyMS, PhilbrickbJT, HeimcSW, SchectmanbJM. Repeated-measures modeling improved comparison of diagnostic tests in meta-analysis of dependent studies. J Clin Epidemiol. 2004;57:698–711. [CrossRef] [PubMed]
ReitsmaJB, GlasAS, RutjesAWS, ScholtenRJPM, BossuytPM, ZwindermanAH. Bivariate analysis of sensitivity and specificity produces informative summary measures in diagnostic reviews. J Clin Epidemiol. 2005;58:982–990. [CrossRef] [PubMed]
MosesLE, ShapiroD, LittenbergB. Combining independent studies of a diagnostic test into a summary ROC curve: data-analystic approaches and some additional considerations. Stat Med. 1993;12:1293–1316. [CrossRef] [PubMed]
MacaskillP. Empirical Bayes estimates generated in a hierarchical summary ROC analysis agreed closely with those of a full Bayesian analysis. J Clin Epidemiol. 2004;57:925–932. [CrossRef] [PubMed]
PaukerSG, KassirerJP. The threshold approach to clinical decision making. N Engl J Med. 1980;302:1109–1117. [CrossRef] [PubMed]
Pierre-KahnV, TadayoniR, HaouchineB, MassinP, GaudricA. Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurement. Br J Ophthalmol. 2005;89:1581–1585. [CrossRef] [PubMed]
BrowningDJ, FraserCM. Regional patterns of sight-threatening diabetic macular edema. Am J Ophthalmol. 2005;140:117–124. [PubMed]
KangSW, ParkCY, HamDI. The correlation between fluorescein angiographic and optical coherence tomographic features in clinically significant diabetic macular edema. Am J Ophthalmol. 2004;137:313–322. [CrossRef] [PubMed]
PanozzoG, ParoliniB, GussonE, et al. Diabetic macular edema: an OCT-based classification. Semin Ophthalmol. 2004;19:13–20. [CrossRef] [PubMed]
SchaudigU, ScholzF, LercheRC, RichardG. Optical coherence tomography for macular edema: classification, quantitative assessment, and rational usage in the clinical practice (in German). Ophthalmologe. 2004;101:785–793. [PubMed]
CiardellaAP, KlancnikJ, SchiffW, BarileG, LangtonK, ChangS. Intravitreal triamcinolone for the treatment of refractory diabetic macular oedema with hard exudates: an optical coherence tomography study. Br J Ophthalmol. 2004;88:1131–1136. [CrossRef] [PubMed]
BrowningDJ. Potential pitfalls from variable optical coherence tomograph displays in managing diabetic macular edema. Am J Ophthalmol. 2003;136:555–557. [CrossRef] [PubMed]
MoreiraRO, TrujilloFR, MeirellesRM, EllingerVC, ZaguryL. Use of optical coherence tomography (OCT) and indirect ophthalmoscopy in the diagnosis of macular edema in diabetic patients. Int Ophthalmol. 2001;24:331–336. [CrossRef] [PubMed]
MassinP, ErginayA, HaouchineB, MehidiAB, PaquesM, GaudricA. Retinal thickness in healthy and diabetic subjects measured using optical coherence tomography mapping software. Eur J Ophthalmol. 2002;12:102–108. [PubMed]
MassinP, VicautE, HaouchineB, ErginayA, PaquesM, GaudricA. Reproducibility of retinal mapping using optical coherence tomography. Arch Ophthalmol. 2001;119:1135–1142. [CrossRef] [PubMed]
RivelleseM, GeorgeA, SulkesD, ReichelE, PuliafitoC. Optical coherence tomography after laser photocoagulation for clinically significant macular edema. Ophthalmic Surg Lasers. 2000;31:192–197. [PubMed]
SchaudigUH, GlaefkeC, ScholzF, RichardG. Optical coherence tomography for retinal thickness measurement in diabetic patients without clinically significant macular edema. Ophthalmic Surg Lasers. 2000;31:182–186. [PubMed]
PalE, GivortG, LarocheA, BaralePO, LimonS, UllernM. Macular imaging with optical coherence tomography (in French). J Fr Ophtalmol. 1998;21:484–494. [PubMed]
BrancatoR. Optical coherence tomography (OCT) in macular edema. Doc Ophthalmol. 1999;97:337–339. [CrossRef] [PubMed]
SaddaSR, TanO, WalshAC, SchumanJS, VarmaR, HuangD. Automated detection of clinically significant macular edema by grid scanning optical coherence tomography. Ophthalmology. 2006;113:1187. [PubMed]
GoebelW, FrankeR. Retinal thickness in diabetic retinopathy: comparison of optical coherence tomography, the retinal thickness analyzer, and fundus photography. Retina. 2006;26:49–57. [CrossRef] [PubMed]
GaucherD, TadayoniR, ErginayA, HaouchineB, GaudricA, MassinP. Optical coherence tomography assessment of the vitreoretinal relationship in diabetic macular edema. Am J Ophthalmol. 2005;139:807–813. [CrossRef] [PubMed]
HussainA, HussainN, NuthetiR. Comparison of mean macular thickness using optical coherence tomography and visual acuity in diabetic retinopathy. Clin Experiment Ophthalmol. 2005;33:240–245. [CrossRef] [PubMed]
OzdekSC, ErdincMA, GurelikG, AydinB, BahceciU, HasanreisogluB. Optical coherence tomographic assessment of diabetic macular edema: comparison with fluorescein angiographic and clinical findings. Ophthalmologica. 2005;219:86–92. [CrossRef] [PubMed]
BrownJC, SolomonSD, BresslerSB, SchachatAP, DiBernardoC, BresslerNM. Detection of diabetic foveal edema: contact lens biomicroscopy compared with optical coherence tomography. Arch Ophthalmol. 2004;122:330–335. [CrossRef] [PubMed]
GoebelW, Kretzchmar-GrossT. Retinal thickness in diabetic retinopathy: a study using optical coherence tomography (OCT). Retina. 2002;22:759–767. [CrossRef] [PubMed]
LattanzioR, BrancatoR, PierroL, et al. Macular thickness measured by optical coherence tomography (OCT) in diabetic patients. Eur J Ophthalmol. 2002;12:482–487. [PubMed]
Sanchez-TocinoH, Alvarez-VidalA, MaldonadoMJ, Moreno-MontanesJ, Garcia-LayanaA. Retinal thickness study with optical coherence tomography in patients with diabetes. Invest Ophthalmol Vis Sci. 2002;43:1588–1594. [PubMed]
StromC, SanderB, LarsenN, LarsenM, Lund-AndersenH. Diabetic macular edema assessed with optical coherence tomography and stereo fundus photography. Invest Ophthalmol Vis Sci. 2002;43:241–245. [PubMed]
YangCS, ChengCY, LeeFL, HsuWM, LiuJH. Quantitative assessment of retinal thickness in diabetic patients with and without clinically significant macular edema using optical coherence tomography. Acta Ophthalmol Scand. 2001;79:266–270. [CrossRef] [PubMed]
YamamotoS, YamamotoT, HayashiM, TakeuchiS. Morphological and functional analyses of diabetic macular edema by optical coherence tomography and multifocal electroretinograms. Graefes Arch Clin Exp Ophthalmol. 2001;239:96–101. [CrossRef] [PubMed]
OtaniT, KishiS, MaruyamaY. Patterns of diabetic macular edema with optical coherence tomography. Am J Ophthalmol. 1999;127:688–693. [CrossRef] [PubMed]
BachmannML, PuhanMA, ter RietG, BossuytPM. Sample sizes of studies on diagnostic accuracy: literature survey. BMJ. 2006;332:1127–1129. [CrossRef] [PubMed]
FletcherJ. What is heterogeneity and is it important?. BMJ. 2007;334:94–96. [CrossRef] [PubMed]
DinnesJ, DeeksJ, KirbyJ, RoderickP. A methodological review of how heterogeneity has been examined in systematic reviews of diagnostic test accuracy. Health Technol Assess. 2005;9(12)1–113.
PewsnerD, BattagliaM, MinderC, MarxA, BucherHC, EggerM. Ruling a diagnosis in or out with “SpPIn” and “SnNOut”: a note of caution. BMJ. 2004;329:209–213. [CrossRef] [PubMed]
GuyattGH RennieD eds. Users’ Guides to the Medical Literature. A Manual for Evidence-Based Clinical Practice. 2002;AMA Press Chicago.
StrausES, RichradsonWS, GlasziouP, HaynesRB. Diagnosis and screening. Evidence-Based Medicine. How to Practice and Teach EBM. 2005;67–100.Elsevier Amsterdam.
BaumannM, GentileRC, LiebmannJM, et al. Reproducibility of retinal thickness measurements in normal eyes using optical coherence tomography. Ophthalmic Surg Lasers. 1998;29:280–285. [PubMed]
MuscatS, ParksS, KempE, KeatingD. Repeatability and reproducibility of macular thickness measurements with the Humphrey OCT system. Invest Ophthalmol Vis Sci. 2002;43:490–495. [PubMed]
JaeschkeR, GuyattG, SackettDL. Users guides to the medical literature. VI. How to use an article about a diagnostic test. B: What are the results and will they help me in caring for my patients. JAMA. 1994;271:703–707. [CrossRef] [PubMed]
PolitoA, Del BorrelloM, IsolaM, GemellaN, BandelloF. Repeatability and reproducibility of fast macular thickness mapping with stratus optical coherence tomography. Arch Ophthalmol. 2005;123:1330–1337. [CrossRef] [PubMed]
ZhouX, ObuchowskiNA, McClishDK. Analysis of correlated ROC data. Statistical Methods in Diagnostic Medicine. 2002;274–306.John Wiley & Sons Inc New York.
BossuytPM, ReitsmaJB, BrunsDE, et al. Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. BMJ. 2003;326:41–44. [CrossRef] [PubMed]
IrwigL, MacaskillP, WalterSD, HoussamiN. New methods give better estimates of changes in diagnostic accuracy when prior information is provided. J Clin Epidemiol. 2006;59:299–307. [CrossRef] [PubMed]
TatsioniA, ZarinDA, AronsonN, et al. Challenges in Systematic Reviews of Diagnostic Technologies. Ann Intern Med. 2005;142:1048–1055. [CrossRef] [PubMed]
Figure 1.
 
Individual estimates of sensitivity, specificity, positive and negative likelihood ratios, and DORs for the six studies for which the data were complete. For Gaucher et al., 41 the presence of a cell with a null value was treated adding 0.5 to that cell to compute all accuracy parameters.
Figure 1.
 
Individual estimates of sensitivity, specificity, positive and negative likelihood ratios, and DORs for the six studies for which the data were complete. For Gaucher et al., 41 the presence of a cell with a null value was treated adding 0.5 to that cell to compute all accuracy parameters.
Figure 2.
 
Summary ROC curve pooling the results of five studies included in the meta-analysis, which are shown as gray circles proportional to the sample size. The square symbol in the center of the ellipses is the expected operating point on the summary ROC. The inner ellipse (long-dashed line) is the 95% confidence region of the pooled estimate. The outer ellipse (short-dashed line) is the prediction region of the pairs of sensitivity and specificity on the heterogeneous summary ROC curve, which shows the typical shoulder arm morphology. The summary ROC curve is not presented in full because only the portion within the area of the studies should be considered, and no inference can be drawn outside the data of the meta-analysis.
Figure 2.
 
Summary ROC curve pooling the results of five studies included in the meta-analysis, which are shown as gray circles proportional to the sample size. The square symbol in the center of the ellipses is the expected operating point on the summary ROC. The inner ellipse (long-dashed line) is the 95% confidence region of the pooled estimate. The outer ellipse (short-dashed line) is the prediction region of the pairs of sensitivity and specificity on the heterogeneous summary ROC curve, which shows the typical shoulder arm morphology. The summary ROC curve is not presented in full because only the portion within the area of the studies should be considered, and no inference can be drawn outside the data of the meta-analysis.
Table 1.
 
Search Strategy Using Medline and Embase
Table 1.
 
Search Strategy Using Medline and Embase
No. Query Results
Medline (PubMed 1966–September 2006)
 1 “Tomography, Optical Coherence” [MeSH] 1,464
 2 “Optical coherence tomography” 2,413
 3 OCT Field: Title/Abstract 5,337
 4 1 OR 2 OR 3 6,497
 5 “Macular edema*” Field: Title/Abstract 2,523
 6 “Macular oedema*” Field: Title/Abstract 574
 7 “Macular Edema, Cystoid” [MeSH] 1,689
 8 5 OR 6 OR 7 3,483
 9 4 AND 8 409
 10 “Diabetes Mellitus” [MeSH] 201,736
 11 Diabete* OR Diabetic* Field: Title/Abstract 231,923
 12 10 OR 11 271,713
 13 9 AND 12 193
Embase (embase.com 2002–October 2006)
 1 ‘Optical coherence tomography’/exp AND [2002–2006]/py 2,389
 2 oct AND [2002–2006]/py 3,800
 3 1 OR 2 AND [2002–2006]/py 5,035
 4 ‘Macular edema’ AND [2002–2006]/py 1,204
 5 ‘Macular oedema’ AND [2002–2006]/py 238
 6 Retina macula cystoid edema, cystoid’/exp AND [2002–2006]/py 715
 7 4 OR 5 OR 6 AND [2002–2006]/py 1,665
 8 3 AND 7 AND [2002–2006]/py 403
 9 ‘Diabetes mellitus’/exp AND [2002–2006]/py 88,702
 10 Diabete* OR diabetic* AND [2002–2006]/py 107,084
 11 9 OR 10 AND [2002–2006]/py 107,460
 12 8 AND 11 AND [2002–2006]/py 187
Table 2.
 
Characteristics of Excluded Studies
Table 2.
 
Characteristics of Excluded Studies
Study Reasons for Exclusion
Pierre-Kahn et al. 25 No gold standard
Browning and Fraser 26 No gold standard
Ciardella et al. 30 Only diseased subjects included (all patients with CSME)
Panozzo et al. 28 No gold standard
Kang et al. 27 Only diseased subjects included (all patients with CSME)
Browning 31 No gold standard
Moreira et al. 32 Unclear gold standard (undefined indirect ophthalmoscopy)
Rivellese et al. 35 No gold standard
Schaudig et al. 29 No gold standard
Brancato 38 No gold standard
Schaudig et al. 36 Review article
Pal et al. 37 Qualitative study on various retinal disease
Massin et al. 33 Patients had diabetic retinopathy but not CSME
Massin et al. 34 Small reproducibility study
Table 3.
 
Characteristics of Included Studies
Table 3.
 
Characteristics of Included Studies
Study ID Design (CSME Used to Define Diseased Eyes) Disease Severity of Analyzed Patients (Number of Eyes) Type of OCT Type of Reference Test Patients (Eyes) Analyzed Patients/Eyes Eligible but Not Analyzed
Goebel and Franke 40 Prospective cohort and healthy controls Various level of DR (124) and healthy controls (13) OCT 2000 Stereophotography 82 (124) 13 eyes excluded for poor photographic quality
Sadda et al. 39 Retrospective cohort Patients with DME Stratus OCT Biomicroscopic examination (not defined) and stereophotography 40 (71) None
Gaucher et al. 41 Retrospective case-control 35 diabetic patients with DME and 35 diabetic patients without DME (27 with no DR) Stratus OCT Contact and noncontact biomicroscopy 70 (98) None
Hussain et al. 42 Prospective cohort Various level of DR (83 NPDR, 22 PDR) Stratus OCT Biomicroscopic examination and fundus photography (not defined) 55 (105) Five eyes excluded for vitreous hemorrhage (4) and epiretinal membrane (1)
Ozdek et al. 43 Retrospective cohort Various level of DR (148 NPDR, 47 PDR) Stratus OCT Noncontact biomicroscopy 110 (195) Unclear if all consecutive and eligible patients were included
Brown et al. 44 Prospective cohort Various level of DR Stratus OCT Contact lens fundus biomicroscopy 95 (172) Ten patients declined participation two did not perform OCT
Browning et al. 9 Prospective cohort Patients with DME Stratus OCT Noncontact fundus biomicroscopy 80 (143) Unclear if all consecutive and eligible patients were included
Goebel and Kretzchmar-Gross 45 Prospective cohort and healthy controls Various levels of DR (136) and healthy controls (30) OCT 2000 Contact and noncontact fundus biomicroscopy 166 (166) Fourteen patients due to improper fixation during OCT
Lattanzio et al. 46 Prospective cohort and healthy controls Various levels of DR (116 NPDR, 54 PDR), and diabetes without DR (46) OCT 2000 Slit-lamp biomicroscopy and stereophotography 137 (216) None
Sanchez-Tocino et al. 47 Prospective cohort Various levels of DR (54 NPDR, 21 PDR, 28 DR and CSME), diabetes without DR (45) and healthy controls (44) OCT 2000 Biomicroscopy with slit-lamp and fundus lens 111 (192) Unclear if consecutive patients, no statement on inclusion of all eligible patients
Strom et al. 48 Prospective cohort and healthy controls 47 Patients with DME (80 without CSME, 4 CSME) and 25 healthy controls (33 eyes) OCT 2000 Stereophotography 71 (117) Twelve eyes excluded for various reasons
Yang et al. 49 Prospective cohort and healthy controls Patients with DME (19 CSME, 14 no CSME) OCT 2000 Stereophotography and noncontact biomicroscopy 17 (33) None
Yamamoto et al. 50 Prospective cohort and healthy controls Patients with DME (45, not all with CSME) and healthy controls (21) OCT 2000 Slit lamp biomicroscopy 46 (66) Unclear if all consecutive and eligible patients were included
Otani et al. 51 Prospective cohort and healthy controls Patients with DME (59, not all with CSME) and healthy controls (10) OCT 2000 Biomicroscopy 52 (69) Unclear if all consecutive and eligible patients were included
Hee et al. 12 Cohort and healthy controls (unclear if prospective) Various level of DR (148 NPDR, 34 PDR), diabetes without DR (55), healthy controls (73) OCT 2000 Noncontact biomicroscopy 189 (310) Unclear if all consecutive and eligible patients were included
Table 4.
 
Methodological Quality of Included Studies
Table 4.
 
Methodological Quality of Included Studies
Table 5.
 
Included Studies Data
Table 5.
 
Included Studies Data
Study Eyes with CSME (Prevalence, %) Eyes without CSME (n) Sensitivity TP/(TP + FN) Specificity TN/(TN + FP) OCT Thickness Threshold (μm) Mean CRT/CRA (SD)
CSME No CSME
Goebel and Franke 40 * 64 (57.7) 47, ‡ 82.8% 53/64 76.6% 36/47 230 μm (central subfield) NA NA
Sadda et al. 39 35 (55.6) 28 88.6% 31/35 85.7% 24/28 NA 317 (77) 251 (47)
Gaucher et al. 41 49 (51.5) 46 100% 0/49 95.7% 2/46 250 μm (central point) 451.2 (104.8) 191.5 (29.0)
Hussain et al. 42 50 (47.6) 55 NA NA NA NA NA
Ozdek et al. 43 112 (57.4) 83 NA NA NA (data available for CME only) NA NA
Brown et al. 44 , † , ‡ 33 (19.2) 139 66.7% 22/33 95.7% 133/139 300 μm (central point) NA NA
Browning et al. 9 50 (35.0) 93 84.0% 42/50 81.7% 76/93 265 μm (central point) NA NA
Goebel and Kretzchmar-Gross 45 116 (85.3) 20 NA NA NA 333 (131) NA
Lattanzio et al. 46 NA NA NA NA NA NA NA
Sanchez-Tocino et al. 47 28 (13.9) 174 NA NA NA 371 (180) NA
Strom et al. 48 14 (42.4) 19 NA NA Spongelike changes used, not thickness 255.6 (138.9) 174.6 (38.2)
Yang et al. 49 131 CSME definition not used for reporting data (only 2 CSME eyes)
Yamamoto et al. 50 45 CSME definition not used for reporting data
Otani et al. 51 59 CSME definition not used for reporting data
Hee et al. 12 , ‡ , § 76 (39.8) 115, § 73.7% 56/76 91.3% 105/115 250 μm CTP NA NA
×
×

This PDF is available to Subscribers Only

Sign in or purchase a subscription to access this content. ×

You must be signed into an individual account to use this feature.

×