Abstract
purpose. To check whether myopic LASIK induces changes on the posterior corneal surface.
methods. A Scheimpflug system (Pentacam; Oculus, GmbH, Wetzlar, Germany) was used to measure preoperative and postoperative posterior corneal topography in 27 eyes (of 14 subjects) that had undergone standard myopic LASIK surgery (attempted corrections between −1.25 and −8.50 D) and on 18 nonoperated eyes (9 subjects). A hybrid porcine-plastic eye model was developed to validate the measurement technique. Longitudinal displacement of the posterior corneal apex and changes of the apical radius of curvature and asphericity were computed.
results. Measurements on a hybrid model eye of known posterior corneal geometry showed that the measured posterior corneal radius of curvature was minimally affected by the geometry of the anterior surface. The measurements on patients showed that, on average, the only relevant (though clinically unimportant) change in radius of curvature and asphericity occurred the first day after surgery (ΔR = −28 ± 34 μm and ΔQ = −0.06 ± 0.06). No statistically significant change was observed afterward. The change in radius was more pronounced in the vertical direction than in the horizontal direction. On average, there was no significant displacement of the posterior corneal apex. Individual changes over time did not show a systematic trend across patients, and control subjects experienced changes of the same order of magnitude.
conclusions. The Pentacam Scheimpflug system can be used reliably to assess changes in the posterior corneal radius of curvature after LASIK. There is no evidence of surgically induced changes in the corneal posterior surface beyond 1 week after surgery.
Corneal laser in situ keratomileusis (LASIK) intends to change the shape of the anterior surface of the cornea, thus changing its refractive power.
1 The technique successfully eliminates refractive errors, but high-order aberrations are generally induced by the procedure.
2 3 The most recent ablation algorithms (wavefront-guided or wavefront-optimized) have been shown to induce fewer high-order aberrations than standard algorithms or even to reduce some high-order aberrations, though discrepancies from the expected optical outcomes are still observed.
4 5 Refractive outcomes are also subject to some variability because of regression,
6 hydration,
7 and other factors that lead to empiric adjustments of the ablation nomogram.
8 9 Although the surgical procedure does not directly affect the posterior surface of the cornea and a large proportion of the optical changes can be explained at the level of the anterior corneal surface,
10 a modification in the posterior corneal surface may be expected because the surgery weakens the cornea. Because corneal biomechanical properties are largely unknown, it is not possible to predict theoretically the extent of this potential deformation.
11
There are some open clinical questions that might be explained by changes on the posterior surface of the cornea. It has been reported that the change in manifest refraction after LASIK differs from the measured change in corneal power, estimated using the standard keratometric index.
12 Discrepancies may arise from the change in ratio of the anterior/posterior corneal curvature with LASIK (and, therefore, the effective keratometric index),
12 13 the change in the corneal effective index of refraction caused by epithelial hyperplasia,
14 or the change in posterior corneal curvature with LASIK. Moreover, theoretical calculations of postoperative corneal shape after subtraction of standard ablation profiles differ dramatically from clinical outcomes.
15 16 17 18 Although it has been demonstrated that, to a large extent, this effect can be explained by physical laws,
19 20 given that it is also encountered when plastic corneas are ablated,
21 results are likely affected by the assumption of a mechanically inert cornea.
22 Empiric corrections of the ablation algorithms can compensate for losses in laser efficiency
21 and perhaps for systematic deviations found in the average population, but they cannot provide individual adjustments because the reason for the intersubject variability in achieved correction is unknown.
11 Greater understanding of the biomechanical processes that take place in the cornea after surgery would help to improve the predictability and stability of achieved corrections.
Some previous experimental studies have reported significant changes in the posterior surface of the cornea after LASIK.
23 24 25 26 27 However, these results have been largely contested because they may be the result of artifacts caused by the optical distortion produced by the anterior surface of the cornea.
28 29 Recently a Scheimpflug imaging—based commercial device that corrects for this distortion, Pentacam (Oculus GmbH, Wetzlar, Germany), has become available. The first studies with this device have reported no significant changes in the posterior surface of the cornea after LASIK and photorefractive keratectomy.
30 31 32 Although some studies have reported high repeatability of the Pentacam instrument in normal
33 and post-LASIK eyes,
34 the accuracy of this system to measure the posterior cornea had not been validated.
In this study, we validate and use the Pentacam topographer to study changes in the posterior corneal surface after myopic LASIK, comparing them to physiological changes observed in control subjects.
We conducted standard LASIK surgery using a narrow-beam, flying-spot excimer laser (Chiron Technolas 217-C equipped with the Plano Scan program; Bausch & Lomb Surgical, Rochester, NY). This laser has an emission wavelength of 193 nm, a fixed pulse repetition rate of 50 Hz, and a radiance exposure of 120 mJ/cm2. The procedure was assisted by an eye tracker. Flaps were performed with suction rings of 8.5-mm (3 eyes) or 9.5-mm (24 eyes) diameter. We used a microkeratome (Hansatome; Bausch & Lomb) programmed for 160-μm (22 eyes) or 180-μm (5 eyes) depth. Photoablation was applied to an optical zone of diameter ranging from 5.2 to 7 mm (6.6 ± 0.6 mm). The LASIK procedures were conducted at the Instituto de Oftalmobiología Aplicada, Universidad de Valladolid, Spain.
Preoperative central pachymetry (measured by Pentacam) ranged from 510 to 643 μm (572 ± 33 μm). Residual bed thickness 1 day after surgery (computed as central pachymetry 1 day after surgery minus programmed flap depth) ranged between 271 μm and 421 μm (329 ± 41 μm).
Hybrid Porcine-Plastic Eye Model for Experimental Validation of the Scheimpflug Systems
Validation of the accuracy of the measurements of the posterior corneal surface geometry was performed in vitro using a hybrid porcine/plastic eye model. Corneal tissue was used to achieve similar intracorneal scattering in the images, which appears to be critical to achieve good edge detection, and a similar index of refraction, which is a parameter in the reconstruction algorithms.
The hybrid porcine/plastic eye models consisted of excised porcine corneas mounted on a 12-mm-diameter plastic piston finished on a spherical surface.
Figure 1shows a schematic diagram of the model and mount. A strip of sclera was left around the cornea. Corneal samples were fixed on a custom-designed support, using an annular metallic ring that pressed the scleral strip. The piston (hydrated with hyaluronic acid) was slid inside the supporting piece until the corneal button was fit on the spherical surface. Because of scattering properties, Pentacam detects the back surface of the cornea and not the underlying plastic support. This mount was designed to avoid bubbles, folds, and creases of the corneal tissue by achieving a smooth corneal back surface conformed to the plastic surface. The amount of stress produced by the piston affected the anterior corneal surface geometry. Enucleated eyes were obtained in a local slaughterhouse, and the procedures were performed within 4 hours of death.
Several plastic surfaces (acrylic material) of known radii of curvature (7.47, 7.93, and 8.75 mm) were used as support for the posterior corneas in this validation study. The plastic surfaces were polished in a precision optics lathe. Radii of curvature of the sphere were measured using a microscope noncontact profilometer (PLμ; Sensofar, Barcelona, Spain) that also reported negligible deviations from the spherical shape.
Given that accurate reconstruction of the posterior corneal surface relies on accurate measurements of the anterior corneal surface and optical and geometrical distortion reconstruction method, validation of the anterior corneal surface was also performed using high-quality optical glass surfaces (optical calipers with nominal radii of 9.65, 8.00 and 6.15 mm).
Glass spherical surfaces acting as anterior corneas were measured on the Pentacam system. Ten measurements were conducted on each surface. One of the plastic spherical surfaces used in the hybrid porcine/plastic eye model (7.93 mm) was also measured directly (acting as the anterior surface).
Twelve porcine corneas were used in the validation study (5 porcine corneas were mounted on 7.47-mm and 8.75-mm spherical corneas, and 2 were mounted on the 7.93-mm spherical cornea). Each hybrid eye was measured up to five times on the Pentacam (total of 36 measurements) using procedures identical to those used in the glass/plastic eyes and patients. The piston pressure was slightly incremented between every two measurements. Ten measurements (including the whole set of measurements from one porcine cornea) that showed evidence of corneal opacification, edema, or bubble formation either from visual inspection or on the Pentacam were discarded.
Data of anterior and posterior radius of curvature (from the best-fitting sphere or mean horizontal and vertical radii of curvature), asphericity Q (defined as p-1), and astigmatism were obtained from the Pentacam software. Anterior and posterior corneal elevation maps were also used in custom-developed fitting routines (Matlab; MathWorks).
We collected topographies of both eyes of each patient before and several times after surgery. Thirteen eyes were measured at three different points after surgery—1.3 ± 0.5, 10.5 ± 3.2, and 33.4 ± 5.5 days—that we will refer as 1 day, 1 week, and 1 month after surgery. Another eye was measured postoperatively only 1 day after surgery, six eyes were measured 1 week after surgery, two eyes were measured 1 week and 1 month after surgery, and five eyes were measured 1 month after surgery. Average measurement times for the whole group of eyes were 1.3, 10.8, and 33.4 days. In control subjects, we acquired several consecutive topographies of both eyes separated by 1 day and 1 week. These data were used as baseline for normal short-term and midterm changes in the posterior corneal shape. Time of the measurement across subjects varied between 9:00 am and 4:00 pm, but for each subject measurements were conducted around the same time of day (within 3.4 ± 1.4 hours). We acquired three to six valid topographies per session. One of the control eyes was measured 23 consecutive times to study instrument repeatability.
Surface Fitting.
Significance of Changes.
Difference Maps and Corneal Elevation.
Validation of the Anterior Surface.
Validation of the Posterior Surface.
Table 2shows the nominal and estimated posterior radii of curvature in the hybrid porcine/plastic eye model. Each datum is the average of the different eyes with similar posterior plastic cornea. Correlation is good between nominal and estimated radius, but bias (underestimation) is higher than for the anterior cornea (150–400 μm). Despite the low accuracy (the absolute values are not recovered exactly), the repeatability is good (below 100 μm) considering that these are individual measurements (nonaveraged) of hybrid eyes whose anterior surface varies across measurements.
Correlations between anterior and posterior corneal surfaces were performed to assess that the optical distortion correction was successfully applied. Statistically significant correlations between the posterior (fixed) and the anterior (variable, as dependent from the specimen and pressure exerted by the piston) corneal surfaces would be indicative of artifacts in the correction. This is particularly critical in the present study, when the same eyes were evaluated before and after a change in the anterior cornea.
Our hybrid porcine/plastic model corneas showed a wide range of anterior radii (from 8.0 to 9.6 mm) for the different corneas. A z-test showed no correlation between anterior and posterior corneal radius for any of the eyes (P > 0.6), indicating that the anterior surface did not affect the measurement of the posterior surface. In addition, the anterior corneas typically showed significant amounts of astigmatism (1.9 ± 1.0 D) with preferential orientations in the vertical and horizontal directions, whereas the estimated astigmatism for the posterior surface was much lower (0.31 ± 0.15 D), with no preferential orientation. These retrieved values of astigmatism of the posterior corneal surfaces conformed to spherical plastic surfaces and were similar to those found in the calibrations of spherical model eyes (anterior surfaces) in the previous section. There was no correlation between the astigmatism of the anterior and posterior corneal surfaces (P = 0.57 for all eyes). Finally, there was no correlation between the anterior and posterior corneal asphericities (P = 0.28 for all the eyes). However, the posterior corneal asphericity differed from zero and was variable (−0.33 ± 0.44).
We tested correlations between posterior corneal changes in radius of curvature, asphericity, or central elevation with the attempted correction, change of pachymetry, preoperative central pachymetry, postoperative central pachymetry, or postoperative residual bed thickness. We tested all these correlations separately for changes 1 day, 1 week, and 1 month after surgery. Of those correlations, only the correlation between attempted correction and posterior corneal apex shift 1 week after surgery showed a statistically significant correlation (r = 0.665; P = 0.0005). We found near-significant correlations (P = 0.04–0.051) only between attempted correction and posterior corneal apex shift 1 day after surgery, and pachymetry, posterior apex shift, and radius of the posterior surface 1 week after surgery. These correlations disappear if eyes with attempted corrections greater than −6 D are removed from the calculation. There were no statistically significant correlations 1 month after surgery.
We have presented a validation of the Pentacam Scheimpflug system to assess the posterior corneal surface geometry and measurements of changes in the posterior corneal surface in patients who underwent myopic LASIK refractive surgery and in control subjects.
The validation experiments using a hybrid porcine/plastic eye model indicate that the posterior corneal radius can be obtained with high repeatability and that the reconstructed posterior surface is not affected by changes or asymmetries in the anterior cornea. Posterior corneal radius seems to be slightly underestimated and asphericity overestimated in comparison with the nominal values of the posterior cornea in the model eye used for calibration. Although studies on the repeatability of this system had been published (particularly of the anterior radius of curvature and pachymetry
33 34 ), this is, to our knowledge, the first time the accuracy (for both anterior and posterior surfaces) has been studied.
Our results for anterior surface mean preoperative radius (
R ant = 7740 ± 230 μm) and asphericity (
Q ant = −0.11 ± 0.13) in our group of patients agree well with those reported by Dubbelman et al.,
36 who used a custom-adapted Scheimpflug imaging system (
R ant = 7870 ± 270 μm;
Q ant = −0.18 ± 0.18).
1 There is also good agreement between our mean preoperative posterior corneal radius of curvature (
R back = 6440 ± 250 μm) and that of Dubbelman et al.
36 (
R back = 6400 ± 280 μm). However, our mean preoperative posterior corneal asphericity was significantly higher (
Q back = 0.18 ± 0.21 in our study and
Q back = −0.38 ± 0.27 in Dubbelman et al.
36 for 6- and 7-mm fitted areas, respectively). Differences might have arisen from differences in the refractive state and age of both populations because our results from the validation using a hybrid model eye indicate an underestimation rather than an overestimation of the posterior corneal asphericity.
We have not found evidence of a systematic influence of LASIK on the posterior surface of the cornea on average. We detected a change in the radius of curvature and asphericity of the posterior corneal surface the first day after surgery, but afterward this change disappeared. The cause of this change is unclear to us. The fact that changes disappeared in days to weeks suggests the action of biological processes within that time. Potential factors affecting posterior corneal shape temporarily may include hydration, keratocyte activity, stress produced by the suction ring of the microkeratome, or medication. Furthermore, we observed a remarkable similarity between both eyes of each subject, suggesting a physiological mechanism acting bilaterally in the same way because both corneas of the same patient are expected to have similar biomechanical properties and to follow similar biological processes. Bilateral similarity also tended to occur in left and right eyes of the control subjects.
We have found that changes in the posterior radius of curvature occur primarily in the vertical direction in patients after LASIK. There are at least three possible causes for this effect. First, it is possible that surgery affects corneal stability more strongly in the vertical direction possibly because of an asymmetrical ablation for correction of astigmatism, but we did not find any statistically significant correlation between corrected astigmatism and radius change. Even with symmetric ablation corneal stability could be more strongly affected in the vertical direction because of the direction of the flap, which is cut in the vertical direction, leaving a superior hinge. Second, meridian-independent surgery may cause a greater change in the vertical meridian if there is a greater mechanical stress of the cornea in that direction. This higher mechanical stress may be caused by the eyelid, which presses on the superior part of the cornea and has been shown to modify the corneal geometry and to have an impact on corneal aberrations.
37 Third, intralamellar cohesive strength, studied in human eye bank corneas, has been shown to be lower in the vertical than in the horizontal meridian, suggesting that even a symmetric force applied to the cornea may result in an asymmetric corneal deformation.
38 In previous studies,
34 the vertical posterior radius of curvature was found to be the least repeatable of the parameters measured by Pentacam, in repeated measurements of patients after LASIK.
Control subjects also experienced statistically significant changes of the same order of magnitude as found in patients, though the reported differences between vertical and horizontal meridians were unique to patients (see
Fig. 7D ; see also
Fig. 6B ). These differences between patients and control subjects were indicative of some surgical effect on the posterior corneal surface, though the average magnitude of the changes observed in patients was similar to that observed in control subjects. This suggested that most of the changes observed in patients were normal, perhaps because of changes in intraocular pressure. In fact, preliminary results of inflation experiments on porcine eyes show that the radius of curvature of the posterior corneal surface changes approximately 30 μm/mm Hg (Pérez-Escudero A, et al.
IOVS 2008;49:ARVO Abstract 664). The changes in posterior radius of curvature that we report in the present study (up to 120 μm) are consistent with changes of intraocular pressure of the order of 5 mm Hg, which is the average change in intraocular pressure throughout the day.
39 The small changes induced by the surgery are superposed to these physiological changes and initiate the subtle correlations mentioned.
All changes we found in the radius of curvature of the posterior corneal surface were smaller than 180 μm (taking into account an interval of confidence of 98.3%). In an average cornea, this change in radius induces a change in the refractive power of the posterior corneal surface below 0.18 D,
40 too small to be clinically relevant. Therefore, the contribution of the posterior corneal surface to shifts from the attempted refraction is minor. Previous studies with a topographer (Orbscan; Bausch & Lomb) reported long-term average changes in posterior radius of curvature up to 400 μm,
24 26 and ectasia measured as forward displacement of the center of the posterior corneal surface up to 40 μm,
23 25 much greater than the changes observed by us (lower than 8 μm on average, including the 98.3% confidence interval;
Fig. 4C ). The discrepancy might have resulted from improper correction of the distortion because of the anterior corneal surface in the topographer (Orbscan; Bausch & Lomb).
28 29 Our results are consistent with recent data obtained with the same device (Pentacam).
30 Along with the findings of this study, more experimental data on corneal biomechanical properties and more accurate models of corneal biomechanics will help to better our understanding of the corneal shape response to LASIK surgery.
Presented in part at the annual meeting of the Association for Research in Vision and Ophthalmology, Fort Lauderdale, Florida, April 2007.
Supported by Project Grants FIS2005–04382 and FIS2008–02065 (Ministerio de Ciencia e Innovación, Spain); Project Grant EURYI-05–102-ES (European Heads of Research Councils-European Science Foundation and Ministerio de Ciencia e Innovación); FPU and FPI Predoctoral Fellowships from the Ministerio de Educación y Ciencia; the Consejo Superior de Investigaciones Científicas JAE Program; and the Unidad Asociada Instituto de Optica-CSIC/Instituto de Oftalmobiología Aplicada-Universidad de Valladolid.
Submitted for publication January 14, 2009; revised March 14, 2009; accepted June 15, 2009.
Disclosure:
A. Pérez-Escudero, None;
C. Dorronsoro, None;
L. Sawides, None;
L. Remón, None;
J. Merayo-Lloves, None;
S. Marcos, None
The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked “
advertisement” in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Corresponding author: Susana Marcos, Instituto de Óptica Daza de Valdés, Consejo Superior de Investigaciones Científicas, Serrano 121, 28006 Madrid, Spain;
[email protected].
Table 1. Validation of Pentacam Measurements for Anterior Radius
Table 1. Validation of Pentacam Measurements for Anterior Radius
Glass Sphere | Nominal R (mm) | Measured R (anterior surface) (mm) | Difference (mm) |
1 | 9.65 | 9.61 ± 0.04 | 0.04 |
2 | 8.00 | 7.97 ± 0.01 | 0.03 |
3 | 6.15 | 6.12 ± 0.02 | 0.03 |
Table 2. Validation of Pentacam Measurements for Posterior Radius
Table 2. Validation of Pentacam Measurements for Posterior Radius
Plastic Sphere | Nominal R (mm) | Measured R (posterior surface) (mm) | Difference (mm) |
1 | 7.47 | 7.32 ± 0.1 | 0.15 |
2 | 7.93 | 7.54 ± 0.05 | 0.39 |
3 | 8.75 | 8.51 ± 0.08 | 0.24 |
Table 3. Average Values for All Patients
Table 3. Average Values for All Patients
| Anterior Corneal Surface | | Posterior Corneal Surface | |
| Radius (μm) | Asphericity | Radius (μm) | Asphericity |
Before surgery | 7700 ± 240 | −0.10 ± 0.12 | 6380 ± 220 | 0.16 ± 0.24 |
1 day after surgery | 8430 ± 430 | 0.42 ± 0.50 | 6420 ± 210 | 0.06 ± 0.28 |
1 week after surgery | 8460 ± 480 | 0.42 ± 0.53 | 6400 ± 180 | 0.19 ± 0.28 |
1 month after surgery | 8520 ± 390 | 0.59 ± 0.53 | 6370 ± 220 | 0.12 ± 0.26 |
The authors thank Samuel Arba Mosquera and José Requejo-Isidro for fruitful discussions and comments and Sara Arganda for help with final formatting of the figures.
PallikarisIG, PapatzanakiME, StathiEZ, FrenschockO, GeorgiadisA. Laser in situ keratomileusis. Lasers Surg Med. 1990;10(5)463–468.
[CrossRef] [PubMed]ApplegateRA, HowlandHC. Refractive surgery, optical aberrations and visual performance. J Refract Surg. 1997;13(3)295–299.
[PubMed]Moreno-BarriusoE, Merayo LlovesJ, MarcosS, NavarroR, LlorenteL, BarberoS. Ocular aberrations before and after myopic corneal refractive surgery: LASIK-induced changes measured with laser ray tracing. Invest Ophthalmol Vis Sci. 2001;42:1396–1403.
[PubMed]PadmanabhanP, MrochenM, BasuthkarS, ViswanathanD, JosephR. Wavefront-guided versus wavefront-optimized laser in situ keratomileusis: contralateral comparative study. J Cataract Refract Surg. 2008;34(3)389–397.
[CrossRef] [PubMed]KimA, ChuckRS. Wavefront-guided customized corneal ablation. Curr Opin Ophthalmol. 2008;19(4)314–320.
[CrossRef] [PubMed]ChayetAS, AssilKK, MontesM, Espinosa-LaganaM, CastellanosA, TsiouliasG. Regression and it mechanisms after laser in situ keratomileusis in moderate and high myopia. Ophthalmology. 1998;105(7)1194–1199.
[CrossRef] [PubMed]FelthamMH, StapletonF. The effect of water content on the 193 nm excimer laser ablation. Clin Experiment Ophthalmol. 2002;30(2)99–103.
[CrossRef] [PubMed]MrochenM, HafeziF, IseliHP, LöfflerJ, SeilerT. Nomograms for the improvement of refractive outcomes. Ophthalmolge. 2006;103(4)331–338.
[CrossRef] DitzenK, HandzelA, PiegerS. Laser in situ keratomileusis nomogram development. J Refract Surg. 1999;15(2 suppl)S197–S201.
[PubMed]MarcosS, BarberoS, LlorenteL, Merayo-LlovesJ. Optical response to LASIK surgery for myopia from total and corneal aberration measurements. Invest Ophthalmol Vis Sci. 2001;42(13)3349–3356.
[PubMed]DuppsWJ, WilsonSE. Biomechanics and wound healing in the cornea. Exp Eye Res. 2006;83:709–720.
[CrossRef] [PubMed]TangM, LiY, AvilaM, HuangD. Measuring total corneal power before and after laser in situ keratomileusis with high-speed optical coherence tomography. J Cataract Refract Surg. 2006;32:1843–1850.
[CrossRef] [PubMed]JaradeE, Abi NaderF, TabbaraK. Intraocular lens power calculation following LASIK: determination of the new effective index of refraction. J Refract Surg. 2006;22:75–80.
[PubMed]SpadeaL, FascianiR, NecozioneS, BalestrazziE. Role of the corneal epithelium in refractive changes following laser in situ keratomileusis for high myopia. J Refract Surg. 2000;16:133–139.
[PubMed]CanoD, BarberoS, MarcosS. Comparison of real and computer-simulated outcomes of LASIK refractive surgery. J Opt Soc Am A. 2004;21(6)926–936.
[CrossRef] MarcosS, CanoD, BarberoS. Increase in corneal asphericity after standard laser in situ keratomileusis for myopia is not inherent to the Munnerlyn algorithm. J Refract Surg. 2003;19(5)S592–S596.
[PubMed]GatinelD, Hoang-XuanT, AzarDT. Determination of corneal asphericity after myopia surgery with the excimer laser: a mathematical model. Invest Ophthalmol Vis Sci. 2001;42(8)1736–1742.
[PubMed]JiménezJR, AneraRG, Jiménez del BarcoL. Equation for corneal asphericity after corneal refractive surgery. J Refract Surg. 2003;19(1)65–69.
[PubMed]MrochenM, SeilerT. Influence of corneal curvature on calculation of ablation patterns used in photorefractive laser surgery. J Refract Surg. 2001;17:584–587.
JiménezJR, AneraRG, Jiménez del BarcoL, HitaE. Effect on laser-ablation algorithms of reflection losses and nonnormal incidence on the anterior cornea. Appl Phys Lett. 2002;81(8)1521–1523.
[CrossRef] DorronsoroC, CanoD, Merayo-LlovesJ, MarcosS. Experiments on PMMA models to predict the impact of corneal refractive surgery on corneal shape. Opt Exp. 2006;14:6142–6156.
[CrossRef] MunnerlynCR, KoonsSJ, MarshallJ. Photorefractive keratectomy: a technique for laser refractive surgery. J Cataract Refract Surg. 1988;14:46–52.
[CrossRef] [PubMed]WangZ, ChenJ, YangB. Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness. Ophthalmology. 1999;106:406–410.
[CrossRef] [PubMed]SeitzB, TorresF, LangenbucherA, BehrensA, SuárezE. Posterior corneal curvature changes after myopic laser in situ keratomileusis. Ophthalmology. 2001;108:666–672.
[CrossRef] [PubMed]BaekTM, LeeKH, KagayaF, TomidokoroA, AmanoS, OshikaT. Factors affecting the forward shift of posterior corneal surface after laser in situ keratomileusis. Ophthalmology. 2001;108:317–320.
[CrossRef] [PubMed]TwaMD, RobertsC, MahmoudAM, ChangJS. Response of the posterior corneal surface to laser in situ keratomileusis for myopia. J Cataract Refract Surg. 2005;31:61–71.
[CrossRef] [PubMed]GrzybowskiDM, RobertsCJ, MahmoudAM, ChangJS. Model for nonectatic increase in posterior corneal elevation after ablative procedures. J Cataract Refract Surg. 2005;31:72–81.
[CrossRef] [PubMed]UedaT, NawaY, MasudaK, IshibashiH, HaraY, UozatoH. Posterior corneal surface changes after hyperopic laser in situ keratomileusis. J Cataract Refract Surg. 2005;31:2084–2087.
[CrossRef] [PubMed]SeitzBS, TorresF, LangenbucherA, et al. Posterior corneal curvature changes after myopic laser in situ keratomileusis (discussion). Ophthalmology. 2001;108:666–673.
[CrossRef] [PubMed]CiolinoJB, BelinMW. Changes in the posterior cornea after laser in situ keratomileusis and photorefractive keratectomy. J Cataract Refract Surg. 2006;32:1426–1431.
[CrossRef] [PubMed]CiolinoJ, KhachikianS, CorteseM, BelinM. Long-term stability of the posterior cornea after laser in situ keratomileusis. J Cataract Refract Surg. 2007;33:1366–1370.
[CrossRef] [PubMed]MatsudaJ, HiedaO, KinoshitaS. Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery. Jpn J Ophthalmol. 2008;52:245–249.
[CrossRef] [PubMed]LacknerB, SchmidingerG, PiehS, FunovicsM, SkorpikC. Repeatability and reproducibility of central corneal thickness measurement with Pentacam. Orbscan Ultrasound Opt Vis Sci. 2005;82:892–899.
[CrossRef] JainR, DilrajG, GrewalSPS. Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis. Ind J Ophthalmol. 2007;55:341–347.
[CrossRef] RosalesP, MarcosS. Pentacam Scheimpflug quantitative imaging of the crystalline lens and intraocular lens. J Refract Surg. 2009;25:421–428.
[CrossRef] [PubMed]DubbelmanM, WeeberHA, van der HeijdeRGL, Völker-DiebenHJ. Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography. Acta Ophthalmol Scand. 2002;80:379–383.
[CrossRef] [PubMed]BuehrenT, CollinsMJ, CarneyL. Corneal aberrations and reading. Optom Vis Sci. 2003;80(2)159–166.
[CrossRef] [PubMed]SmolekMK. Interlamellar cohesive strength in the vertical meridian of human eye bank corneas. Invest Ophthalmol Vis Sci. 1993;34(10)2962–2969.
[PubMed]DavidR, ZangwillL, BriscoeD, DaganM, YagevR, YassurY. Diurnal intraocular pressure variations: an analysis of 690 diurnal curves. Br J Ophthalmol. 1992;76:280–283.
[CrossRef] [PubMed]BarberoS. Refractive power of a multilayer rotationally symmetric model of the human cornea and tear film. J Opt Soc Am A. 2006;23(7)1578–1585.
[CrossRef]