CCT measurements have wide diagnostic applications. Regarding glaucoma diagnosis, it has been reported that thick or thin corneal pachymetry is clinically correlated with intraocular pressure (IOP) readings.
35,36 In addition, Gordon et al.
4 reported that each 40-μm reduction in CCT is associated with a relative risk of 1.71 for development of primary open-angle glaucoma. Furthermore, progression of visual field loss in open-angle glaucoma was significantly associated with thinner CCT.
37 Moreover, Doughty and Zaman
38 published a meta-analysis study that found that a 10% difference in CCT results in a 3.4 mm Hg difference in IOP. Likewise, when assessing candidates for refractive surgery, reliable CCT measurements are also demanded because it has been reported that a preoperative CCT thinner than 500 μm is a relative contraindication for LASIK
39 and a postoperative residual stroma thinner than 250 μm or a total CCT thinner than 400 μm, should be avoided when planning excimer laser ablations.
40,41 Currently, ultrasonic pachymetry remains the gold standard for measuring CCT; however, most authors agree
15–21 that it has several disadvantages, such as risk of infection or epithelial damage, patient discomfort, and/or probe misplacement. Therefore, consistent noncontact pachymetry methods to measure CCT would be highly beneficial. The reliability of ultrasonic pachymetry is high, its repeatability coefficient (2.77 × S
w) when assessing intraobserver variability has been reported recently to be about 5 μm, after averaging five consecutive measurements.
42 Therefore, a noncontact technique that replaces ultrasonic pachymetry must provide similar variability values. Axial resolution using SD-OCT has been reported to be limited up to 2.1 μm in laboratory studies.
43 Thus, OCT may be a reliable alternative to ultrasonic pachymetry. Cirrus HD-OCT is a commercially available SD-OCT device that provides a reported axial resolution of 5 μm
23; thus, it is worth assessing its reliability when measuring the CCT.