In this work, we compared the optical properties of the most commonly used methods to create a stable air–cornea interface for in vivo ocular imaging in rodents. The optical properties of the system formed by the eye, its air–cornea interface, and the group of lenses L3 when present (
Fig. 1), dictate the maximum resolution that can be achieved when imaging the retina. Of particular importance is the amount of scattering and the wavefront aberrations present.
When the ability to reduce scattering was compared, the CS performed much better than any other condition. The CL was second in performance, showing a particularly large variability.
Figure 4 (bottom panel) shows, however, that during the early experiments the amount of scattering present with the CL was similar to that found with the CS. The most likely explanation is that we used a new CS for each experiment but rinsed and reused the expensive custom-made CL. Therefore, despite careful cleaning, scratches, and deposits may have accumulated on the CL, leading to an increase in scattering over time. The use of a new CL after a few measurements or more thorough cleaning procedures would very likely lead to much better results in terms of scattering.
Quantification of forward light-scattering based on HS spot images is restricted to the small angle scattering (<1° from the center of the PSF). The use of a specially designed double-pass system instead could have slightly increased the angle covered.
10 To truly quantify the large angle domain or straylight (1° to 100° from the center of the PSF) a separate measurement with a different device would have been necessary, which was not possible in our experimental setting. Nevertheless, our measurement setup was sufficiently sensitive to demonstrate differences in scattering between conditions. The systematic increase of the CAHM value (
Fig. 4) in the CL case discussed above strongly indicates that the method indeed quantifies forward-scattering derived from the optical media. Given the purpose of the study, to determine the method of best protecting the cornea and maximizing resolution in retinal imaging, a full characterization of wide-angle scattering in the rat eye was beyond the scope of this work.
In our particular experimental conditions with the amount of defocus and all other aberrations measured, any changes in shape of the HS spots induced by aberrations were theoretically and experimentally verified to be negligible. A theoretical analysis of our system showed that for the effects of the first-pass PSF to be noticeable on the HS spots we should have at least 5.3 D of defocus, which is approximately 20 times larger than the average defocus values we recorded. The same applies for other higher order aberrations such as coma or trefoil. Numerical simulations of the HS spots using the actual wavefront measurements (including defocus and astigmatism) showed all HS spots to be airy disks of the same size. Finally, during our pilot data collection before these experiments (Diaz-Santana L, et al. IOVS 2010;51:ARVO E-Abstract 2322), we verified experimentally that the size of the spots was invariant, independent of the amount of defocus present (within the range expected). Hence, any changes observed in the HS spots in the data presented in this article were due to scattering.
In the presence of very large aberrations, or larger HS lenslet apertures, aberrations may affect the HS spot shape. In this case a deconvolution can be used to disentangle the scattering information from the aberrations
33,34 ; however, one must be careful, as the wavefront information recorded under these conditions may be biased, because the accuracy with which the positions of the HS spots is determined can be severely affected by the change in shape of the HS spots. It is generally accepted that if the shape of the spots is affected by aberrations, the HS operates outside of its dynamic range, and the data should not be used. We ensured that our system was operating within its dynamic range in all data sets.
The two independent measurements of the naked eye (NK1 and NK2), the second of which served as an internal control, showed a reduction in scattering between the first and the second set of measurements with the time between the two measurements being approximately 15 to 20 minutes. Although NK1 showed a decrease in scattering as the experiment progressed, NK2 stayed almost constant (
Fig. 4). The last few eyes measured show very similar levels of scattering in both cases. We think that the period between the onset of anesthesia and the time when we started recording the data may have decreased as aligning the eye with the system became more routine. This may have reduced alterations to the corneal epithelium that occurred despite the use of lubricant eye drops.
Differences in scattering between the different interfaces could be due to changes in the optical properties of other optical elements in the eye, as a consequence of the anesthetic used. The specific optical properties and surface roughness of the CL, CS, and cornea, together with the scattering properties of the lubricant eye drops used to couple them to the cornea, are very likely to be the determinant factors in how much scattering is present. Direct measurement of these parameters, however, was outside the scope of this work.
The use of MO appears to have slightly reduced scattering, but less than the CL or the CS. We cannot provide a reasonable explanation for the slight but systematic increase in scattering as data collection progressed. The measurements needed to be collected very quickly, as breakup of the oil surface occurred faster than the normal tear film breakup. This phenomenon could be observed in the live image of the HS spots, appearing as shadows moving across the screen. Compared with the CL and CS, applying the oil to the cornea is much easier. Thus, for very short imaging sessions without aiming for highest resolution (e.g., cell counts), this method may be suitable. Alternatively, the use of highly viscous lubricant eye drops which are specially made to increase the breakup time of the tear film could in fact be superior to MO for quick image-acquisition procedures.
The wavefront aberrations of the eye were influenced in very different ways by the different air–cornea improvements. The MO had no effect compared to the naked eye. In contrast, the CL significantly reduced the wavefront aberrations, whereas the CS significantly increased them: In both cases, the effect was clearer mainly in aberrations of third or lower order. A possible explanation for the aberration increase with the CS may be that, analogous to the human eye, the internal optics may compensate for corneal aberrations.
19,35 As the CS nulls the refraction of the cornea, the internal aberrations of the eye may be uncovered, leading to an increase in total aberrations compared with the naked eye. Furthermore, lens L3, added to compensate for the loss of optical power, has an impact on the resulting wavefront's reaching the HS sensor. Improving the optical quality of L3—for example, by using a microscope objective—may decrease the aberrations measured with the CS. However, this benefit may be negligible, as the combination of lenses used in L3 was already designed to minimize aberrations in our setup. Using two microscope objectives, a Badal system could be built in an approach similar to that proposed by Artal et al.
36 Then, the focusing correction for the loss of optical power in the CS case would already be included, and addition of L3 would not be necessary. A different strategy could be to design L3 in a way that it introduces aberrations complementary to the wavefront aberrations derived from the inner eye optics and which can rotate/translate together with the rat eye ensuring an optimal coupling at all times. This approach would result in fewer total aberrations and could therefore be attractive for imaging purposes.
For practical high-resolution imaging, not only is the amplitude of the wavefront of great importance but also its variability, both as a function of eccentricity and between individuals. Large changes in the wavefront aberration amplitude from eye to eye, or even within one session when changing the imaging field, will challenge every imaging system, whether correcting methods such as AO are used or not. In other words, the less these aberrations vary, the less demanding it will be to design a system that can obtain high-resolution images over a larger field of view in a large proportion of eyes. Thus, it is an important feature of the CL that it reduces not only the aberration amplitude but also its variability. The CS, in contrast, greatly increases the aberration variability compared with all other interfaces. Previously reported work on ray tracing simulations of the rat eye, and its coupling to the CL or CS, showed that the wavefront aberrations increased rapidly with eccentricity when using a CS, but much less with the CL or naked eye. (Diaz-Santana L, et al. IOVS 2010;51:ARVO E-Abstract 2322). On the basis of these results, we speculate that the reason for the larger variability observed with the CS was due to misalignment of the optical axis during our experiments inducing larger wavefront aberrations, whereas similar misalignments with the CL did not translate into larger wavefront errors. That is, the CL appears to be more tolerant to eccentric imaging. Imaging eccentric to the optical axis often occurs in live imaging, not by accident but deliberately when attempting to reach more peripheral areas of the retina. Therefore, the CL may have a clear advantage over the other methods, especially the CS, as it reduces the aberration amplitude, increases the predictability of the wavefront aberrations and, consequently, the image quality. Further experimental work is necessary to confirm that it is indeed more tolerant to eccentric imaging. The disadvantage in scattering reduction may be overcome by using a new or thoroughly cleaned lens after a few imaging sessions.
To illustrate the practical implications of both aberrations and scattering on image quality, we include as part of this discussion numerical simulations of image formation under some of the different conditions measured. They are limited by the assumptions made to estimate the effects of scattering and by the mathematical formalism used (Fraunhofer approximation). They should not be treated as strict predictions, but as an aid to understand our results.
Details of how the simulations were performed are presented in Appendix A. The original image (
Fig. 7A) used for these simulations shows a fluorescent retinal ganglion cell in a flat-mounted retina from a YFPH transgenic mouse. Each pixel in the image measures 0.5 μm.
Figure 7D shows the effects of diffraction alone. An ideal rat eye without either aberrations or scattering would produce such an image. The blurring observed is due to the wave nature of light alone. In this case, the experimental width of the SH spots provided by the manufacturer gave an area CAHM = 7.5.
The two center panels (
Figs. 7B,
7E) show the effect of the aberrations measured with the CL in eye 3, while the rightmost panels (
Figs. 7C,
7F) show the effect of aberrations in this same eye when using the CS. In the top panels (
Figs. 7B,
7C) the minimum amount of scattering measured in all the experiments (CAHM = 19.11) was used, while in the bottom panels (
Figs. 7E,
7F), we used the maximum amount of scattering measured (CAHM = 30.28). When comparing the images, it becomes apparent that aberrations have a dominant role. In the CS case, the effect of scattering is practically invisible compared with the degradation introduced by aberrations alone. In the case of the CL, the two values of scattering used result in a very small difference in contrast. Scattering, however, still prevents us from reaching the optimal quality of a diffraction-limited image, even in the absence of aberrations.
For practical high-resolution in vivo imaging experiments, obtaining good-quality images in a reliable manner is of fundamental importance. Considering this, we conclude from our study that using a CL is the best way to create a stable, optically improved air–cornea interface.
Supported by Imagine Eyes, which provided equipment and software loans.
Disclosure:
C. van Oterendorp, Imagine Eyes, France (F);
L. Diaz-Santana, Imagine Eyes, France (F);
N. Bull, None;
J. Biermann, None;
J.F. Jordan, None;
W.A. Lagrèze, None;
K.R. Martin, None