Letters to the Editor  |   March 2012
Diurnal Variation of Retinal Thickness Measured by Optical Coherence Tomography in Normal Adults
Author Affiliations & Notes
  • Colin S. H. Tan
    Department of Ophthalmology, Tan Tock Seng Hospital, Singapore;
    the National Healthcare Group Eye Institute, Singapore; and
  • Wei Kiong Ngo
    the National Healthcare Group Eye Institute, Singapore; and
  • Milton C. Chew
    the National Healthcare Group Eye Institute, Singapore; and
  • Kelvin Z. Li
    the National Healthcare Group Eye Institute, Singapore; and
  • Louis W. Lim
    the National Healthcare Group Eye Institute, Singapore; and
  • Srinivas R. Sadda
    the Department of Ophthalmology, University of Southern California Keck School of Medicine, Doheny Eye Institute, Los Angeles, California.
Investigative Ophthalmology & Visual Science March 2012, Vol.53, 1639. doi:https://doi.org/10.1167/iovs.12-9566
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Colin S. H. Tan, Wei Kiong Ngo, Milton C. Chew, Kelvin Z. Li, Louis W. Lim, Srinivas R. Sadda; Diurnal Variation of Retinal Thickness Measured by Optical Coherence Tomography in Normal Adults. Invest. Ophthalmol. Vis. Sci. 2012;53(3):1639. https://doi.org/10.1167/iovs.12-9566.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
We read with interest the article by Jo et al. 1 in the August 2011 issue describing the diurnal variation of retinal thickness in normal subjects measured by time-domain (TD) and spectral-domain (SD) optical coherence tomography (OCT). The authors concluded that the diurnal variation in the retinal thickness observed with TD-OCT was caused by limitations in the repeatability of the OCT device rather than by actual variation of retinal thickness. 1  
Although this was a well-designed study, one limitation was that the retinal thickness was assessed at only two time points, which provides limited information on the pattern of diurnal change. A difference between two time points is less compelling than if the retinal thickness measurements differ across several additional points during the day. We would like to highlight two other studies in normal subjects 2,3 that provide additional evidence to support the findings of Jo et al. A significant strength of both of the additional studies is that the retinal thickness was measured at five time points during a single day, thus providing more comprehensive and robust data for the analysis of diurnal variation compared with the two time points that were used in the present study. In addition, in both studies, the examination sequence (five OCT scans during the day) was repeated on a separate day to confirm the consistency of the results. 
In a study of 12 normal individuals, 2 sequential SD-OCT scans were performed with the Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany). The authors found no significant variation in the macular thickness over the five time points during the day, with a maximum amplitude (the difference between the maximum and minimum retinal thickness) of 0.9 μm. 2 This result is comparable to the difference of 0.3 μm reported by Jo et al. for central macular thickness measured with the Cirrus OCT (Carl Zeiss Meditec, Dublin, CA). In addition, this study used the tracking function of the Spectralis OCT to reduce the chance of a small displacement of the foveal B-scan position between sequential OCT scans and to ensure that the same point on the retina was measured for all OCT scans. 2 We feel that this is a useful and important advantage, conferred by the tracking function of the Spectralis OCT. 
In another study, Chakraborty et al. 3 used an optical biometer instead of OCT to measure retinal thickness. They also reported no significant diurnal variation in retinal thickness (mean amplitude, 8 μm) over the 2 days of examination. 
Another interesting point that merits discussion is that, in Jo et al., in the sectors that showed a significant difference in retinal thickness on TD-OCT, the mean difference ranged from 2.5 to 3.5 μm. Although this difference may be statistically significant, it is unlikely to be of clinical relevance. In addition, the coefficient of repeatability of Stratus OCT retinal thickness in normal individuals has been reported to be approximately 17 μm, 4,5 which is much greater than the difference between the two time points in this study. 
Jo et al. also cited earlier studies that have reported diurnal variation in retinal thickness in patients with macular diseases. 6 8 In these studies, various versions of TD-OCT were used to perform the OCT scans. Although it is certainly possible that the diurnal variation in retinal thickness may be due to the underlying disease process, it is also possible that at least part of it is due to motion artifact resulting from poor fixation as a result of the underlying disease process. 5 It would be interesting to know whether a similar degree of diurnal change in retinal thickness in diseased eyes would be observed with tracking-capable SD-OCT devices. 
In summary, in confirming this finding using SD-OCT with tracking, we agree with Jo et al. that there does not appear to be significant diurnal variation in retinal thickness in normal individuals. 
References
Jo Y-J Heo D-W Shin Y-I Kim J-Y . Diurnal variation of retina thickness measured with time domain and spectral domain optical coherence tomography in healthy subjects. Invest Ophthalmol Vis Sci. 2011;52:6497–6500. [CrossRef] [PubMed]
Tan CS Ouyang Y Ruiz H Sadda SR . Diurnal variation of choroidal thickness in normal, healthy subjects measured by spectral domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2012;53:261–266. [CrossRef] [PubMed]
Chakraborty R Read SA Collins MJ . Diurnal variations in axial length, choroidal thickness, intraocular pressure, and ocular biometrics. Invest Ophthalmol Vis Sci. 2011;52:5121–5129. [CrossRef] [PubMed]
Leung CK Cheung CY Weinreb RN . Comparison of macular thickness measurements between time domain and spectral domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2008;49:4893–4897. [CrossRef] [PubMed]
Forooghian F Cukras C Meyerle CB Chew EY Wong WT . Evaluation of time domain and spectral domain optical coherence tomography in the measurement of diabetic macular edema. Invest Ophthalmol Vis Sci. 2008;49:4290–4296. [CrossRef] [PubMed]
Diabetic Retinopathy Clinical Research Network. Diurnal variation in retinal thickening measurement by optical coherence tomography in center-involved diabetic macular edema. Arch Ophthalmol. 2006;124:1701–1707. [CrossRef] [PubMed]
Frank RN Schulz L Abe K . Temporal variation in diabetic macular edema measured by optical coherence tomography. Ophthalmology. 2004;111:211–217. [CrossRef] [PubMed]
Gupta B Grewal J Adewoyin T . Diurnal variation of macular oedema in CRVO: prospective study. Graefes Arch Clin Exp Ophthalmol. 2009;247:593–596. [CrossRef] [PubMed]
×
×

This PDF is available to Subscribers Only

Sign in or purchase a subscription to access this content. ×

You must be signed into an individual account to use this feature.

×