To confirm this qualitative assessment, the values of
Neq and efficiency that were derived from the LAM were plotted as a function of
Et0 in
Figures 5 and
6, respectively.
Figure 5 shows log
Neq versus log
Et0 for the 10 patients and 10 control subjects. The upper
x-axis gives the log
CS0 equivalents of the log
Et0 values. The vertical and horizontal gray regions demarcate the normal range of log
Et0 and log
Neq , respectively. Two patients had log
Et0 values (and log
CS0 values) that were within the range of normal. The remaining eight patients had values of log
Et0 that were elevated by as much as 0.92 log units (approximately a factor of 8) above the upper limit of normal, corresponding to reductions of
CS0 by as much as approximately a factor of 3. The values of log
Neq , were above the range of normal for all 10 patients, and there was a statistically significant difference in log
Neq between the patients and controls (
t = −6.32,
P < 0.05). Log
Neq and log
Et0 were correlated significantly for the RP patients (
r = −0.80,
P < 0.05), but not for the control subjects (
r = 0.34,
P = 0.34). This is consistent with the observation that the individual functions for the patients with RP shown in
Figure 4 were shifted primarily upward and rightward by equal amounts, whereas this was not the case for the control subjects (
Fig. 3). There was also a statistically significant correlation between log
Neq and visual field diameter (
r = −0.69,
P < 0.05), such that patients with more restricted fields, representing greater disease progression, had higher values of log
Neq .