The mean CCTs, which were measured during the first visit using the UHR-OCT, UL-OCT, and RTVue devices with an automatic segmentation algorithm applied to each measurement, were not different from each other (
Table 2). The mean CCT measurements derived from the RTVue and Visante pachymetry maps for both groups were thicker than those measurements obtained from the UHR-OCT, UL-OCT, and RTVue devices in conjunction with an automatic segmentation algorithm applied to each measurement (
Tables 2,
3). The coefficient of repeatability of the healthy subjects ranged from 2.01 to 4.88 μm with an ICC that was greater than 0.997 for the CCT measurements from all of the OCT devices (
Table 2). The coefficient of reproducibility was less than 6.0 μm with an ICC greater than 0.996 (
Table 3). For the ET measurements from the UHR-OCT, UL-OCT, and RTVue devices, the coefficient of repeatability was less than 2.2 μm with an ICC greater than 0.890 (
Table 4), and the coefficient of reproducibility ranged from 2.8 to 3.9 μm with an ICC greater than 0.74 (
Table 5). For the LASIK patients, the coefficient of repeatability ranged from 3.5 to 7.8 μm with an ICC greater than 0.992 for the CCT measurements (
Table 2), and the coefficient of reproducibility was less than 5.6 μm with an ICC greater than 0.992 (
Table 3). The coefficient of repeatability for the ET measurements was less than 4.8 μm with an ICC greater than 0.84 (
Table 4), and the coefficient of reproducibility ranged from 2.7 to 3.6 μm with an ICC greater than 0.86 (
Table 5). In both groups, there were no significant differences between the various scans that were obtained either during the same visit or during different visits (all
P > 0.05).
Figure 5 shows the relationship between the coefficient of repeatability and the axial resolution of the OCT device when used to measure both the CCT and ET. In both groups, a higher axial resolution correlated with a lower coefficient of repeatability.