We have shown that the riboflavin TPF signal is attenuated with increasing depth in corneal tissue and demonstrated a correction method designed to increase measurement accuracy for ex vivo comparisons of riboflavin absorption for different soak protocols.
There are a number of potential causes of increased TPF signal attenuation with depth, including absorption and scattering of both excitation and emission radiation and aberration of the excitation beam focus. Fluorescence photons emitted at wavelengths corresponding to the overlap between the absorption and emission spectra (
Fig. 1) can be reabsorbed. This “inner-filter” effect is more pronounced the higher the concentration of fluorescence photons in the sample. As such, when fluorescence microscopy is used to determine fluorophore concentration, fluorescence intensity will only be proportional to the concentration at less-concentrated solutions. We initially observed this effect at the highest (0.1%) riboflavin concentration, with an increasingly attenuated signal when imaging through calibration drops on a well slide. By shifting the microscope detector range beyond the riboflavin absorption spectrum (>525 nm), we minimized this phenomenon, obtaining near-flat TPF signals with increasing depth through the drop (
Fig. 3B).
Previous studies using TPF microscopy to quantify corneal riboflavin absorption ignore the inner-filter effect. Cui et al.
10 measured both fluorescein and riboflavin absorption in feline corneas, and while the effects of reabsorption of the fluorescence emitted from fluorescein were mitigated by a bandpass filter of appropriate wavelength, no such detection filter was used for the riboflavin measurements. By way of a signal attenuation correction factor, Kampik et al.
12 used the intensity loss in collagen fiber second-harmonic generation (SHG) signals taken simultaneously during the same z-scan. However, the SHG detector wavelength channel used (420–460 nm), and to a lesser extent that used for riboflavin (505–555 nm), overlap with the absorption spectrum of riboflavin (300–525 nm;
Fig. 1). This would be expected to cause significant reabsorption of the emitting SHG light, compromising its use a reference to correct TPF signal attenuation.
Our calibration method is based on two assumptions: first, that the concentration within the riboflavin reservoir above the cornea remains constant throughout the entire soak; and second, that in a fully soaked cornea, the riboflavin concentration within the cornea equals that of the reservoir. In practice, however, these assumptions may not be strictly correct.
Figure 4B reveals that the TPF signal within the reservoir does decrease with time, albeit by only a few percent. With a finite (small) volume, the reservoir is potentially susceptible to photobleaching. We investigated this by repeating the above
z-stack protocol through the same custom chamber reservoir of riboflavin placed on a slide (i.e., no cornea). Over 50 minutes we observed near-identical TPF signal plots over a depth of 1200 μm, with no evidence of photobleaching. We have additionally investigated the possibility of photobleaching within the corneal tissue, by performing a final
z-stack at the end of the full 60 minute protocol in a new adjacent axial location (i.e., unirradiated tissue). We observed near identical TPF signal plots over a depth of 800 μm, with no evidence of photobleaching. Fluorescence excitation restricted to the focal plane is a key advantage of multi-photon microscopy, explaining the lack of photobleaching demonstrated. An alternative explanation could be that the reduction in TPF signal with time is due to loss of fluorescent riboflavin molecules as they diffuse into the cornea. However, as illustrated
Figure 4B, this decrease over 50 minutes was small, with a mean of 2.5% (±1.1) for all imaged eyes. For future work, this small effect could be reduced by increasing the volume of the reservoir. The assumption that the concentration in a fully soaked cornea is the same as that in the reservoir also appears to be not quite correct. We consistently observed a step-like decrease in TPF signal as the imaged plane transitioned from the fluid-filled reservoir into the cornea (
Fig. 4B), measured to be 6.1% (±1.6) after a 50-minute soak. This small loss of signal may result from optical aberrations encountered as the laser light crosses the curved surface of the cornea or it may reflect a true difference in concentration between the reservoir and the fully soaked cornea.
The percentage transmission of light of wavelengths between 600 nm and 1000 nm has been previously measured in early postmortem eyes between 95% and 98%.
15 Scattering of light may still occur in an otherwise healthy optical system, becoming more prevalent in disease (for example cataract, corneal edema or corneal scarring). Edema is likely to be particularly relevant in the porcine corneas used in this current study which, although optically clear to the naked eye, were regularly over 800 μm in central corneal thickness (pre-riboflavin soak), compared with published in vivo measurements of 666 ± 68 μm.
16 This is presumed to be due to early endothelial failure within the first 24 hours after harvest and is a limitation of this animal model.
There is a lack of consensus regarding the exact nature and location of CXL-induced collagen cross-links.
17,18 Two-photon microscopy of corneal collagen has been used in an in vivo rabbit model to quantify corneal collagen cross-links after treatment.
19 In this study, the investigators identified a well-demarcated change in both autofluorescence and SHG signal down to a depth of 250 μm. This correlates well with a demarcation line visible by optical coherence tomography (OCT) in human patients, typically around 250 to 300 μm after epithelium-off CXL.
20,21 At this depth (300 μm), following a standard 30-minute soak, we measured a mean stromal concentration of 0.086%. It is, however, not possible to make any absolute correlation between the concentrations reported in our study (using an HPMC-based riboflavin preparation) and demarcation line tissue changes observed in vivo on OCT (using a dextran-based riboflavin preparation). We attempted to apply our methodology to determine concentration depth profiles using 0.1% riboflavin in 20% dextran (500 kDa) as used in the original CXL study,
9 in both well-slide and soaked corneas, but were unable to achieve any consistent TPF signals. Dextran has previously been shown to strongly absorb near-infrared light
22 and absorption of the TPF excitation beam may have been raising the temperature of dextran and affecting its optical properties. This is likely to have been the cause of variations in the TPF signal levels we measured. TPF microscopy may therefore not be suitable to investigate corneal riboflavin absorption for dextran-based preparations. We note that these are being replaced by HPMC-based preparations which appear to have faster tissue diffusion rates
23 as well as causing less corneal thinning during treatment.
24
In conclusion, we present evidence that TPF microscopy of corneas in an ex vivo animal model is affected by significant signal attenuation when imaging at depth. To account for this, we have presented a new time-lapse measurement approach carried out during the riboflavin soak in which each cornea acts as its own internal reference. Correction for TPF signal attenuation should enhance the accuracy of experimental riboflavin absorption measurements used to guide the development of clinical protocols for CXL.