The results shown in
Figures 1 and
2 show that, for the infrared spectral range of the InGaAs sensor, the contrast of interference fringes is inversely related to wave number and so is an increasing function of wavelength. This supports the proposal that the low contrast of interference fringes from the precorneal tear film at shorter wavelengths
4,12 is related to corneal surface roughness.
The analysis of
Figures 2 and
3 indicates that the Gaussian relation between the amplitude reflectance of the corneal surface and wave number (
Equation 8) is more consistent with experimental results than the exponential model (
Equation 9). Sinha and Tippur
18 showed that the Gaussian function of wave number in
Equation 8 is to be expected from a Gaussian distribution of corneal surface height
h (deviation from the mean height) of the following form:
where
p(
h) is the probability of height (
h),
p0 is a constant, and
σ is the standard deviation of surface height. With this assumption, they show that the amplitude reflectance (
r) from such a rough surface should be given by the following equation:
where
r0 is a constant. Equating this expression with the Gaussian function of
Equation 8 gives the following:
Equation 12 was used to estimate a corneal surface height mean (SD) of 0.129 (0.010) μm between individuals. For the Gaussian distribution of the model, approximately 95% of the surface height may be expected to be within 2 SDs of the mean height or a total range of approximately 0.5 μm. The measurement area of 25 × 33 μm (probably larger because of the effects of diffraction, aberrations, and defocus) is similar to the area of superficial corneal cells.
23 Therefore, this estimate of the roughness of the corneal surface presumably includes both the effects of roughness within individual cells such as microplicae
24–26 and the difference in surface height between cells.
16,17 The above estimate of total range of approximately 0.5 μm is of the same order of magnitude as the height of human microvilli of approximately 0.5 to 0.75 μm from the transmission electron micrographs by Ehlers.
22
Limitations of this estimate of corneal surface roughness should be noted. The anterior (air) surface of the tear film tends to be pulled flat by surface tension and so is probably quite smooth relative to the posterior (corneal) surface. However, variability in lipid thickness
10 may be expected to contribute to the variability of overall tear thickness and so could contribute to the above roughness estimate. In an unpublished study (Powell DR, Chandler HL, King-Smith PE, unpublished observations, 2013) of more than 1400 high-resolution color micrographs of the lipid layer, we found a lipid thickness SD of 0.014 μm over an area of 180-μm diameter (larger than the test area in the present study), which is only 11% of the estimated corneal surface height SD of 0.129 μm. Given that corneal surface roughness and lipid thickness variability are expected to be uncorrelated, the overall variance (square of the standard deviation) in tear film thickness will be given by the sum of corneal surface variance and lipid surface variance, so the contribution from lipid surface variance is expected to be only approximately 1% of the total variance. Another possible complicating factor is that the glycocalyx may have a refractive index that is intermediate between that of the tear film and corneal epithelium and could therefore act somewhat like an antireflection coating, whose effect would vary with wave number; alternatively, it is possible that the glycocalyx may have a higher refractive index than the corneal epithelium, which would tend to increase observed reflection, again in a manner dependent on wave number. Perhaps the most important limitation of the Gaussian roughness analysis is that reflectance spectra in the wavelength range of the silicon detector (562–1030 nm) are often not consistent with this model. For example, the reflectance spectrum of
Figure 2 in the study by King-Smith et al.
4 shows a fairly constant contrast of spectral oscillations over a wide range of wave numbers (approximately 1.1–1.7 μm
−1), which is inconsistent with the strong decay predicted from
Equation 8 and extrapolated from
Figure 2A. The interference at shorter wavelengths (e.g., visible light) is therefore often greater than the prediction of
Equation 8, suggesting a weak contribution from a smoother or more compact component of the corneal surface. In summary, the Gaussian roughness model predicts quite well the characteristics of spectral oscillations in the spectral range of the InGaAs camera (above approximately 900 nm) but is less satisfactory at shorter wavelengths.
Liu and Pflugfelder
27 have shown that corneal surface irregularity (roughness), studied by corneal topography, is increased in aqueous-deficient dry eye. The present study was limited to healthy individuals and concerns roughness on a much smaller scale; it is planned to study whether the surface roughness observed by the methods herein is affected by dry eye disorders.
A final conclusion of the study is that the contrast of interference fringes from the whole thickness of the precorneal tear film is considerably higher in the wavelength range of InGaAs sensors compared with that of silicon-based sensors. Reported images of whole-thickness fringes from the precorneal tear film using silicon-based cameras have been noisy because of their low contrast,
12 so the present study indicates that better images could be obtained by using InGaAs cameras rather than silicon-based cameras.