All subjects were identically examined by color Doppler imaging
(CDI) in supine position, shielded from the monitor. The measurements
were performed by an experienced laboratory technician. Blood flow
velocities in the ophthalmic and central retinal arteries were measured
in each pair of eyes seven times under standardized conditions of light
and darkness
(Fig. 1) . The right eye was always examined before the left. The subject kept
the eye not being examined open, when the room was in light. The
initial recording was done in standardized light (120 lux in the
position of the eyes;
Fig. 1 ). Then the subject looked with both eyes
into a 55-W lamp for 5 minutes from a distance of 20 cm (1350 lux in
the position of the eyes), after which the right and the left eyes were
examined. Consequently, the left eye was exposed to light somewhat
longer than the right eye before examination
(Fig. 1) . Then the monitor
and the control panel were covered with red plastic film (Lee Filters,
027 medium red, BS 3944: Part 1, 1992; Light transmission: 0% [600
nm], >50% [650 nm], >80% [>700 nm]). All leaks of light were
eradicated. The examination room was in complete darkness except for
the red light emitted from the monitor and control panel. The subject
kept both eyes closed during the period of darkness. Measurements in
the right and left eyes were done after 5, 15, and 25 minutes of
darkness. The light in the room was then turned on, the red filter was
removed, and the sixth measurement of flow velocities in the right and
left eyes was performed. Finally, the subject with both eyes open was
once again exposed to the light from the 55-W lamp for 5 minutes,
whereafter the right and the left eyes were reexamined.