Cup-shaped Ag-AgCl electrodes were fixed with collodion in the
following positions: active electrode in Oz, reference electrode in
Fpz, and ground in the left arm. The interelectrode resistance was kept
below 3 kΩ. The bioelectric signal was amplified (gain 20,000),
filtered (band-pass 1–100 Hz), and averaged (200 events free from
artifacts were averaged for every trial) by BM 6000. The analysis time
was 250 msec. The transient VEP response is characterized by a number
of waves with three subsequent peaks of negative, positive, and
negative polarity, respectively. In normal subjects these peaks have
the following latencies: 75, 100, and 145 msec (N75, P100, N145).
In the recording session, simultaneous PERGs and VEPs were recorded at
least twice, and the resultant waveforms were superimposed to check the
repeatability of the results. We accepted PERG and VEP signals with a
signal-to-noise ratio higher than 2. The noise was measured by
recording the bioelectric signals (200 averaged events) while the
monitor was screened by a cardboard. Noise less than 0.1 μV (mean
0.086 μV) was observed in all subjects tested. For all PERGs and VEPs
the peak latency and the peak amplitude of each of the averaged waves
were measured directly on the display by means of a pair of cursors.