An optical oxygen sensor and probe (Oxylab pO2 optode; Oxford Optronix, Oxford, UK) was used as previously described, to measure oxygen levels in several locations within the eye.
8 17 The diameter of the probe is 200 to 220 μm (∼25 gauge). Although this probe does not have the same spatial resolution as the microelectrodes used in some studies,
5 20 it has the advantage that it does not consume oxygen and is most sensitive at low oxygen levels. To measure oxygen partial pressures in rabbits under circumstances that were as close as possible to physiological, we avoided the use of drugs that might affect ocular blood flow. The pupils were not dilated. To avoid damaging the rabbit lens, which occupies a larger proportion of the globe than in the human eye, we used two sclerotomies. The first was 2.5 mm and the second 6.5 mm posterior to the corneal limbus. These locations are labeled A and B, respectively, in
Figure 1 . The first sclerotomy was made with the tip of a 23-gauge needle at position A, the probe was inserted through the sclerotomy and gently moved into the posterior and anterior chambers. Because of the flexibility of the scleral wall in rabbits, the sclera and cortical vitreous sealed around the shaft of the probe, and there was no leakage of fluid or obvious loss of intraocular pressure. Measurements were first taken in the posterior chamber close to the pars plana of the ciliary body (
Fig. 1 ; PC1), in the posterior chamber close to the pars plicata of the ciliary body (PC2), in the posterior chamber under the middle of the iris (PC3,) and in the posterior chamber at the edge of the pupil (PC4). The blue light from the tip of the optode was visible under the iris of the albino animals, assisting the positioning of the probe at locations PC3 and PC4. The probe was then positioned in the anterior chamber at the surface of the lens (AC lens), in the center of the anterior chamber (AC center), at the posterior surface of the cornea (AC cornea), and, finally at the anterior chamber angle (AC angle). The probe was then withdrawn to the posterior chamber, and a measurement was made at the equator of the lens (A LEQ). Oxygen measurements were recorded for 2 to 4 minutes at each position. To minimize the distortion of aqueous humor flow, the probe was moved from point to point, slowly and gently, under a surgical microscope. Oxygen readings were repeatable in the same animal by returning to the same point, signifying that the flow of aqueous humor and oxygen metabolism were not disturbed by the measurement procedure. When the oxygen measurements were completed in the anterior of the eye, the probe was removed. The wound site was sealed by a small plug of cortical vitreous, and no aqueous humor leaked out of the eye. A second, posterior sclerotomy was made 6.5 mm from the limbus and the probe was inserted. The partial pressure of oxygen was measured, in sequence, at the central posterior surface of lens (VC lens), in the center of the vitreous chamber (VC center), in the vitreous body just anterior to the central retina (VC retina), and near the equatorial retina on the opposite side of the eye (VC eq). The probe was then withdrawn to measure the oxygen level just posterior to the lens equator (B LEQ). Because bright light can alter the values obtained with the optode, only normal laboratory illumination was used during measurements. We confirmed that this amount of light did not alter the readings by checking oxygen levels with the lights on or off. In the low light that reached the interior of the eye, the blue excitation light at the tip of the probe permitted accurate positioning for each of the oxygen measurements. The mean and SEM of P
o 2 are reported for each location. IOP decreased by an average of 11% after the second sclerotomy and was at all times within the normal range for rabbits.