For details regarding the reliability and stimulus characteristics for all of the variables, see the extended treatment in Hammond et al.
20 The apparatus and procedure used to measure macular pigment optical density (MPOD) have also been described in detail previously.
20,21 This method of measuring macular pigment has been extensively validated.
2 A macular densitometer (Macular Metrics, Inc., Rehoboth, MA, USA) was employed. The protocol was the same as that previously published.
20,22 A schematic of the optical setup of the apparatus is given in Wooten et al.
21 (see
Fig. 3 therein). Lutein and Z in the serum were measured using reverse-phase high performance liquid chromatography (HPLC), and that measurement is also detailed in Hammond et al.
20
Three efficacy parameters were measured (PR, GD, and CC), and all three parameters were assessed on the same apparatus, modified for each parameter. The apparatus was a Maxwellian view optical system providing up to three channels. A schematic of the setup of the apparatus is given in Wooten et al.
21 (see
Fig. 1 therein). Consistent Maxwellian delivery of the stimulus to the eye was maintained by a dental impression bite bar and forehead rest assembly. A monitor with an infrared camera was used to monitor pupil position and ensure that subjects were both aligned and viewing the stimuli during intense exposures.
Based on the assumption that the mechanisms for accumulating MP evolved as a result of outdoor activities, we strove to design the stimuli in a manner that was as ecologically valid as possible. Thus, as shown in
Figure 1 of Hammond et al.,
20 we used a xenon source that closely matched the spectral distribution of sunlight (an even better match can be made to the solar spectrum provided in
Fig. 1 of Hammond et al.
23). For the GD measurement, subjects adjusted the intensity of an annulus (centered at 11.5° eccentricity) until it veiled a 1° circular grating (4 cyc/deg) target. Photostress recovery was assessed by exposing the subjects to a bright (5.5 log trolands) circular disc (5°) of xenon light for 5 seconds. Based on calculations by Margrain et al.,
24 this bleaches approximately 50% of the central photopigment. Recovery was measured as the time needed to see a dim flashing target (200-ms cycle) that begins shuttering after the cessation of the photostress. Chromatic contrast was determined by having subjects adjust a circular blue (460 nm) surround until it caused the disappearance of a circular (1°) yellow (570 nm) grating target. The entire series of visual measurements took approximately 1 hour to complete.
Fasting samples (10 mL whole blood) were collected on the morning of a study visit by a licensed phlebotomist for quantification of carotenoids, retinol, and tocopherol. The samples, collected in 10-mL lithium heparin-coated vacutainers (BD, Franklin Lakes, NJ, USA) were separated by centrifugation at 1500
g for 20 minutes at 4°C. After separation, the plasma was distributed into 1.5-mL light-protected Eppendorf vials, 1 mL per vial. The samples were stored at −80°C until express shipment (in insulated boxes on dry ice) to DSM Nutritional Products, Inc., in Kaiseraugst, Switzerland, for analysis. Blood samples were prepared and analyzed for L and Z determination as described by Hartmann et al.
25
Extensive calibration was conducted in this study, since the intervention lasted an entire year and it was critical that the stimuli were stable, effectively, over several years. Hence, both radiometric and photometric calibrations were regularly performed. Prior to each experimental sitting, a dedicated radiometer was used to ensure that total light output remained constant (S370 Optometer; UDT Instruments, Hawthorne, CA, USA). The neutral-density wedge was calibrated using a second radiometer (model 370; Graseby Optronics, Orlando, FL, USA). Photometric calibrations were done using a telescopic spectral radiometer (model PR650; PhotoResearch, Inc., Chatsworth, CA, USA) with the stimuli projected onto a white reflectance standard calibrated to the instrument. Spatial alignment of the channels was checked every session by increasing the intensity of the light source and checking the precise location of the projected image against a fixed point on a wall, the position of which relative to the equipment never changed (the equipment was bolted to the floor). To assess the daily and long-term laboratory performance of the HPLC plasma analytics, dedicated control plasma was used. This control was composed of pooled human plasma that was characterized internally and then used as a quality control measure to ascertain the daily and long-term repeatability of the HPLC plasma analytics. All of the analytical methods were regularly checked through participation in international ring trials organized by the National Institutes of Standards and Technology (USA) and the Society for Vitamins and Biofactors (France). The control samples were analyzed at least four times a day during the study.