Subclinical Keratoconus Detection by Multiparametric Corneal Biomechanical Characterization with the Ocular Response Analyzer
Author Affiliations & Notes
  • Jeremias G. Galletti
    Ophthalmology, Hospital de Clinicas, University of Buenos Aires, Buenos Aires, Argentina
    ECOS (Clinical Ocular Studies) Laboratory, Buenos Aires, Argentina
  • Marianella Delrivo
    ECOS (Clinical Ocular Studies) Laboratory, Buenos Aires, Argentina
  • Tomas Pförtner
    ECOS (Clinical Ocular Studies) Laboratory, Buenos Aires, Argentina
  • Fernando Fuentes Bonthoux
    Ophthalmology, Hospital de Clinicas, University of Buenos Aires, Buenos Aires, Argentina
    ECOS (Clinical Ocular Studies) Laboratory, Buenos Aires, Argentina
  • Footnotes
    Commercial Relationships  Jeremias G. Galletti, None; Marianella Delrivo, None; Tomas Pförtner, None; Fernando Fuentes Bonthoux, None
  • Footnotes
    Support  None
Investigative Ophthalmology & Visual Science March 2012, Vol.53, 1117. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Jeremias G. Galletti, Marianella Delrivo, Tomas Pförtner, Fernando Fuentes Bonthoux; Subclinical Keratoconus Detection by Multiparametric Corneal Biomechanical Characterization with the Ocular Response Analyzer. Invest. Ophthalmol. Vis. Sci. 2012;53(14):1117.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : To compare corneal waveform descriptors of the Ocular Response Analyzer (ORA) in healthy and subclinical keratoconus eyes to assess their diagnostic capacity.

Methods: : 95 eyes from 95 healthy subjects in group 1 and 70 fellow eyes from 70 unilateral manifest keratoconus patients were evaluated with ORA, corneal topography, aberrometry and anterior segment optical coherence tomography. Corneal hysteresis (CH), corneal resistance factor (CRF) and other 41 parameters were obtained from the corneal waveform curves and compared between groups. Linear regression was used to correct for the effect of central corneal thickness (CCT) on correlated parameters and diagnostic performance was assessed by receiver-operating characteristic curve analysis.

Results: : 29 of the 41 additional biomechanical parameters, along with corneal hysteresis (CH) and corneal resistance factor (CRF), showed statistically meaningful differences between the groups (p<0.05, Student’s t test). Six parameters, including CH and CRF, were significantly correlated with CCT in both groups. The parameters CRF, TimeIn, H2, H21 and Dive2 had the best performances (area under the curve, AUC), and only CRF’s was improved after correcting for CCT. Combinations of these parameters had better diagnostic yields (error rate < 20%) than when considered individually, the highest being the combination of the 10 best parameters (AUC=0.82). These results applied also to eyes with normal higher-order 5-mm anterior corneal surface aberrations (<0.344 µm, taken from mean+1 SD of group 1), outperforming corneal topography.

Conclusions: : Multiparametric biomechanical analysis with the ORA can detect a higher proportion of eyes with subclinical keratoconus than corneal topography, with a better diagnostic yield than CH and CRF when considered individually.

Keywords: keratoconus • refractive surgery: complications • cornea: clinical science 
×
×