Determining The Stability Of Implanted Iols With A Dynamic Purkinje-meter System
Author Affiliations & Notes
  • Juan Tabernero
    Laboratorio de Optica, Universidad de Murcia, Murcia, Spain
  • Nino Hirnschall
    Hanusch Hospital, Vienna, Austria
  • Sophie Tatzreiter
    Hanusch Hospital, Vienna, Austria
  • Oliver Findl
    Hanusch Hospital, Vienna, Austria
  • Pablo Artal
    Laboratorio de Optica, Universidad de Murcia, Murcia, Spain
  • Footnotes
    Commercial Relationships  Juan Tabernero, JT (P); Nino Hirnschall, None; Sophie Tatzreiter, None; Oliver Findl, None; Pablo Artal, PA (P)
  • Footnotes
    Support  Supported by "Ministerio de Educación y Ciencia", Spain (grants nº FIS2010-14926 and CSD2007-00013) and "Fundación Séneca", Murcia, Spain (grant 04524/GERM/06).
Investigative Ophthalmology & Visual Science March 2012, Vol.53, 1373. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Juan Tabernero, Nino Hirnschall, Sophie Tatzreiter, Oliver Findl, Pablo Artal; Determining The Stability Of Implanted Iols With A Dynamic Purkinje-meter System. Invest. Ophthalmol. Vis. Sci. 2012;53(14):1373.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : Different intraocular lens designs may provide more or less stability after surgery. To improve IOL centration and stability outcomes, an instrument to measure both parameters is required. Along this objective, we developed a new instrument to quantify IOL misalignments and stability from the dynamic recording of Purkinje images.

Methods: : The prototype is based on a previous version of a system recording Purkinje images (Tabernero et al., Optics Express 2007), but includes significant improvements and modifications. The system allows video recording at 50 Hz and the capture during saccadic eye movements (9º amplitude) induced by two flickering fixation LEDs placed in a central and a peripheral position with respect to the objective-camera axis. Flickering frequency can be switched to 1, 0.5 or 0.33 Hz. In order to investigate directional IOL stability, the peripheral positions of the flickering LEDs included four possible orientations (nasal, temporal, inferior and superior). When an IOL suffers from wobbling, a quick oscillation of Purkinje images PIII and PIV (from anterior and posterior IOL surfaces) can be observed immediately after a change of gaze. Stability is assessed by fitting the relative movement of PIII and/or PIV with respect to PI (reflection from the anterior corneal surface) to the solutions of a classical dumped harmonic oscillator system. The wobbling effect for each IOL is characterized in terms of the "dumping ratio" values resulting from the fitting analysis. Dumping ratios ≥ 1 means that wobbling is not present (the ideal situation in this case) and null values means that the system oscillates without dumping.

Results: : Purkinje images were recorded in five pseudophakic patients (implanted with clinically stable monofocal IOLs) during a series of six saccadic movements (amplitude 9º; frequency 1 Hz). The damping ratio values ranged from 0.07 to 0.30 (average 0.13; standard deviation 0.05), showing a measurable wobbling effect in each case. The IOLs returned to a stationary point in a time that ranged from 1/10 to 1/2 seconds. The analysis revealed that the origin of the wobbling came from tilt oscillations (up to 1 degree). The IOL was also slightly decentered (0.1 mm) as a consequence of the saccadic movements, but in this case it returned to its original position with nearly no oscillations.

Conclusions: : A new instrument to measure IOL misalignments and stability has been developed and successfully tested in patients. It may be useful for a wide range of future studies involving IOL stability.