March 2012
Volume 53, Issue 14
Free
ARVO Annual Meeting Abstract  |   March 2012
Eliminating Infusion Bubbles And Uncontrolled Reflux By The Constellation Vitrectomy System
Author Affiliations & Notes
  • Stephen R. Russell
    Ophthalmology, Univ of Iowa Hospitals & Clinics, Iowa City, Iowa
  • Vinit B. Mahajan
    Ophthalmology, Univ of Iowa Hospitals & Clinics, Iowa City, Iowa
  • H. Culver Boldt
    Ophthalmology, Univ of Iowa Hospitals & Clinics, Iowa City, Iowa
  • James C. Folk
    Ophthalmology, Univ of Iowa Hospitals & Clinics, Iowa City, Iowa
  • Elliott H. Sohn
    Ophthalmology, Univ of Iowa Hospitals & Clinics, Iowa City, Iowa
  • Christine N. Kay
    Ophthalmology, Univ of Florida, Gainsville, Florida
  • Ryan M. Tarantola
    Ophthalmology, Saint Louis University School of Medicine, Saint Louis, Missouri
  • Footnotes
    Commercial Relationships  Stephen R. Russell, None; Vinit B. Mahajan, None; H. Culver Boldt, Alcon (C); James C. Folk, None; Elliott H. Sohn, None; Christine N. Kay, None; Ryan M. Tarantola, None
  • Footnotes
    Support  Research to Prevent Blindness
Investigative Ophthalmology & Visual Science March 2012, Vol.53, 2599. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Stephen R. Russell, Vinit B. Mahajan, H. Culver Boldt, James C. Folk, Elliott H. Sohn, Christine N. Kay, Ryan M. Tarantola; Eliminating Infusion Bubbles And Uncontrolled Reflux By The Constellation Vitrectomy System. Invest. Ophthalmol. Vis. Sci. 2012;53(14):2599.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : To identify the sources of and eliminate infusion bubbles and uncontrolled reflux from the Alcon Constellation® Vitrectomy system.

Methods: : Surgical and analytical videos were evaluated to identify the source of intraoperative bubbles, which localized to the duckbill valve (DV). Intraoperatively we modified the infusion tubing and its control by removing the DV. The DV was re-purposed as a one-way valve to block reflux originating from the vitrectomy console.

Results: : Infusion bubbles at the DV developed with each transitory tubing pressure drop upon opening of the infusion clamp. Modification of the infusion and aspiration tubing resulted in elimination of infusion bubbles and console-originated reflux in 20 eyes. Since instituting these modifications, Infusion bubbles developed in only 1 eye in which the scrub nurse forgot to make modification to the setup. Placement of the DV to block reflux eliminated uncontrolled and purposeful console-originated reflux.

Conclusions: : Intraoperative modification of Constellation® tubing may eliminate two potentially harmful problems until manufacturer correction is instituted. Since our modified connections represent off-label connectivity, the manufacturer cannot contact potentially affected surgeons or suggest temporary alternative connectivity improvements.