April 2009
Volume 50, Issue 13
Free
ARVO Annual Meeting Abstract  |   April 2009
Comparison of OcuSense and Clifton Nanolitre Osmometers
Author Affiliations & Notes
  • A. Tomlinson
    Department of Vision Sciences, Glasgow Caledonian University, Glasgow, United Kingdom
  • L. McCann
    Department of Vision Sciences, Glasgow Caledonian University, Glasgow, United Kingdom
  • E. I. Pearce
    Department of Vision Sciences, Glasgow Caledonian University, Glasgow, United Kingdom
  • Footnotes
    Commercial Relationships  A. Tomlinson, Allergan, Pfizer, SIFI, F; Pfizer, Ciba Vision, SIFI, C; Birmingham Optical, R; L. McCann, None; E.I. Pearce, Pfizer, R; Allergan, Pfizer, SIFI, F.
  • Footnotes
    Support  None.
Investigative Ophthalmology & Visual Science April 2009, Vol.50, 534. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      A. Tomlinson, L. McCann, E. I. Pearce; Comparison of OcuSense and Clifton Nanolitre Osmometers. Invest. Ophthalmol. Vis. Sci. 2009;50(13):534.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : Tear hyperosmolarity is diagnostic of dry eye disease (DED), yet difficulty in measurement has limited its utility. Development of new instruments allowing clinical application may lead to the adoption of osmolarity as a primary diagnostic test in DED. The purpose of this study was to compare the new OcuSense osmometer (OcuSense Inc., San Diego, CA), based on electrical impedance and ‘lab-on-a-chip’ nanolitre technology, with the freezing point depression Clifton Osmometer (Clifton Technical Physics, Hartford, NY).

Methods: : Thirty-six subjects were recruited, 15 DED (9F/6M; age 41.7±16.9) and 21 controls (12F/9M; 35±12.8). Inclusion criteria for DED were NITBUT <10secs and positive for symptoms. Tear samples were collected from the inferior-temporal conjunctival sac for testing with both osmometers.

Results: : Osmolarity values for controls were 308±6.2mOsm/l (OcuSense) and 310±7.2mOsm/l (Clifton) and for DED 321±16.5 and 323±14.7 and were significantly different. Significant correlation was found between OcuSense and Clifton measurements (r=0.904; p=0.006). A scatter-plot with the regression line (dotted) is shown in the figure. Bland-Altman analysis showed general agreement between techniques; the majority of points fell within the 95% confidence limits. But the actual values differed by less than 1% with the OcuSense being lower. A cut-off value of >316mOsm/l, derived from the distribution of osmolarity values was used to diagnose DED with an effectiveness for the OcuSense of 73% (sensitivity), 90% (specificity) and 85% (PPV) and for the Clifton 73%, 71% and 65%.

Conclusions: : Tear film osmolarity measured with the OcuSense system correlates well with the Clifton Osmometer. The new instrument has the potential to provide clinicians with a readily available, clinically applicable measure, which could become the ‘gold-standard’ in DED.

Keywords: inflammation • cornea: tears/tear film/dry eye 
×
×