Effect of Repeated Applanation Tonometry on the Accuracy of Intraocular Pressure Measurements
Author Affiliations & Notes
  • M. Ehrenberg
    Ophthalmology, Rabin Medical Center, Petah Tikwa, Israel
    Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • D. Gaton
    Ophthalmology, Rabin Medical Center, Petah Tikwa, Israel
    Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • M. Lusky
    Ophthalmology, Rabin Medical Center, Petah Tikwa, Israel
    Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • G. Dotan
    Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • D. Weinberger
    Ophthalmology, Rabin Medical Center, Petah Tikwa, Israel
    Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • M. Snir
    Ophthalmology, Rabin Medical Center, Petah Tikwa, Israel
    Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel
  • Footnotes
    Commercial Relationships  M. Ehrenberg, None; D. Gaton, None; M. Lusky, None; G. Dotan, None; D. Weinberger, None; M. Snir, None.
  • Footnotes
    Support  None.
Investigative Ophthalmology & Visual Science April 2010, Vol.51, 554. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      M. Ehrenberg, D. Gaton, M. Lusky, G. Dotan, D. Weinberger, M. Snir; Effect of Repeated Applanation Tonometry on the Accuracy of Intraocular Pressure Measurements. Invest. Ophthalmol. Vis. Sci. 2010;51(13):554.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : To evaluate the stability of successive applanation tonometry measurements in glaucomatous eyes.

Methods: : A prospective, comparative, randomized, and evaluator-masked study was conducted in a tertiary medical center. Sixty-seven patients with glaucoma attending our glaucoma clinic underwent four successive intraocular pressure (IOP) measurements with Goldmann-type applanation tonometry. Findings were compared to 70 patients scheduled for cataract surgery, similar in age and sex. The results were statistically analyzed with repeated measures analysis.

Results: : In the glaucoma group, the difference between the first and second IOP measurements was statistically significant (15.94 vs 14.9 mmHg, p< 0.0001), as was the decrease in IOP from the first measurement to each of the other successive measurements. No significant change in IOP was noted in the control group (mean, 13.7 mmHg).

Conclusions: : Patients with glaucoma show a decrease in IOP on repeated applanation tonometry measurements. By contrast, in individuals without glaucoma, no significant decrease in IOP on repeated applanation tonometry measurements was found.

Keywords: intraocular pressure • outflow: trabecular meshwork 
×
×