Vitreous Substitute Removal Rates With the Constellation Vitrectomy System
Author Affiliations & Notes
  • H. T. Rahman
    Ophthalmology, Baylor College of Medicine, Houston, Texas
  • A. J. Barkmeier
    Ophthalmology, Baylor College of Medicine, Houston, Texas
  • R. Ray
    Ophthalmology, Baylor College of Medicine, Houston, Texas
  • E. R. Holz
    Ophthalmology, Baylor College of Medicine, Houston, Texas
  • P. E. Carvounis
    Ophthalmology, Baylor College of Medicine, Houston, Texas
  • Footnotes
    Commercial Relationships  H.T. Rahman, None; A.J. Barkmeier, None; R. Ray, None; E.R. Holz, Alcon Laboratories, F; Alcon Laboratories, C; P.E. Carvounis, Alcon Laboratories, C; Alcon Laboratories, F.
  • Footnotes
    Support  None.
Investigative Ophthalmology & Visual Science April 2010, Vol.51, 3595. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      H. T. Rahman, A. J. Barkmeier, R. Ray, E. R. Holz, P. E. Carvounis; Vitreous Substitute Removal Rates With the Constellation Vitrectomy System. Invest. Ophthalmol. Vis. Sci. 2010;51(13):3595.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : To compare the effects of duty cycle modulation and varying cut rates with a recently introduced 20-, 23-, and 25-gauge (ga) high cut rate vitrectomy system.

Methods: : The aspiration rate of 1cc of strained egg white in-vitro was measured with different cut rates (1500, 2500, 5000 cuts per minute (CPM)) and duty cycles (port biased open, 50/50, port biased closed) for the Constellation (Alcon Laboratories, Fort Worth, TX, USA) 20- 23-, and 25-gauge vitrectomy systems.

Results: : Comparable simulated vitreous removal was achieved at higher cut rates with biased open and 50/50 duty cycles within each vitrectomy gauge system (20ga: p = 0.79, 23ga: p = 0.25, 25ga: p = 0.51). However, with the biased closed duty cycle, simulated vitreous removal significantly increases with increasing cut rates within each vitrectomy gauge system (20ga: p = 0.02, 23ga: p = 0.03, 25ga: p = 0.02). At the highest cut rate of 5000 CPM, simulated vitreous removal rates are similar across all three different duty cycles within each vitrectomy gauge system (20ga: p = 0.71, 23ga: p = 0.06, 25ga: p = 0.19). The 20-gauge (p< 0.001) and 23-gauge (p< 0.001) systems achieved considerably faster simulated vitreous removal than 25-gauge at all cut rates and duty cycles.

Conclusions: : Using a fixed, high cut rate of 5000 CPM with proportional vacuum within each vitrectomy gauge system will not considerably compromise vitreous aspiration rates, regardless of duty cycle.

Keywords: vitreoretinal surgery • vitreous 
×
×