April 2010
Volume 51, Issue 13
Free
ARVO Annual Meeting Abstract  |   April 2010
Image Quality Analysis for an Aspheric Toric Apodized Diffractive Intraocular Lens Using Modulation Transfer Function Testing
Author Affiliations & Notes
  • J. Schwiegerling
    Ophthalmology, University of Arizona, Tucson, Arizona
  • Footnotes
    Commercial Relationships  J. Schwiegerling, Alcon Research, F.
  • Footnotes
    Support  None.
Investigative Ophthalmology & Visual Science April 2010, Vol.51, 5727. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      J. Schwiegerling; Image Quality Analysis for an Aspheric Toric Apodized Diffractive Intraocular Lens Using Modulation Transfer Function Testing. Invest. Ophthalmol. Vis. Sci. 2010;51(13):5727.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : To compare the optical image quality of AcrySof IQ ReSTOR +3.0 D and a new aspheric toric apodized diffractive intraocular lens using the modulation transfer function (MTF).

Methods: : Best focus MTF at 100 lp/mm which corresponds to Snellen visual acuity of 20/20 was evaluated using the 0.2 micron Spherical Aberration (SA) Modified ISO model eye at distance focus for Acrysof IQ ReSTOR +3.0 D and aspheric toric apodized diffractive intraocular lenses. Data was collected on the Optikos Video MTF measurement system with 3 mm and 5 mm apertures to simulate both photopic and mesopic pupil sizes. Ten lenses for each model were evaluated. Through-focus MTF testing was performed in the ISO Model Eye with a 3 mm aperture.

Results: : At 100 lp/mm, the mean distance MTF values at the 3 mm and 5 mm apertures were 31% and 21% for AcrySof IQ ReSTOR +3.0 D model and the mean distance MTF values at the 3 mm and 5 mm aperture were 34% and 25% for the aspheric toric apodized diffractive lens model. The through-focus MTF results showed two distinct peaks at distance and near foci in both models with separations correspond to the designed add powers.

Conclusions: : MTF testing at 3 mm and 5 mm apertures for the aspheric toric apodized diffractive IOL model showed good image quality that was comparable to that of IQ ReSTOR +3.0 D model.

Keywords: intraocular lens • astigmatism • presbyopia 
×
×