Ocular Sensor Reader Systems for Treatment and Management of Glaucoma
Author Affiliations & Notes
  • M. A. Tarbell
    Visual and Autonomous Exploration Systems Research Laboratory, California Institute of Technology, Pasadena, California
  • W. Fink
    Visual and Autonomous Exploration Systems Research Laboratory, California Institute of Technology, Pasadena, California
    Doheny Eye Institute, University of Southern California, Los Angeles, California
  • Footnotes
    Commercial Relationships  M.A. Tarbell, None; W. Fink, Bausch & Lomb, F; Caltech, P.
  • Footnotes
    Support  Bausch & Lomb Research Funding
Investigative Ophthalmology & Visual Science May 2008, Vol.49, 4227. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      M. A. Tarbell, W. Fink; Ocular Sensor Reader Systems for Treatment and Management of Glaucoma. Invest. Ophthalmol. Vis. Sci. 2008;49(13):4227.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: : To develop user-friendly, non-invasive external readout systems to obtain (optically) information from, e.g., the anterior chamber of the eye such as but not limited to indicated intraocular pressure (IOP) from an implanted IOP sensor, fill-level of implanted drug delivery systems, status monitoring of glaucoma relieve valves/shunts, and other physiological conditions.

Methods: : Two designs have been devised and prototyped: (1) an indirect digital camera-based reader, and (2) a direct opto-mechanical reader. The digital reader comprises an off-the-shelf consumer digital camera (5 MP) that has been modified, i.e., equipped with custom adapter and custom lens to shorten the focal length for close-up imaging. Focus is maintained with the built-in autofocus. The opto-mechanical reader is a U-shaped binocular, using only optical components, i.e., lenses and prisms/mirrors to reroute the optical pathway. The user monitors with one eye the anterior chamber of the respective other eye, and vice versa. The focus of the optical lens system and the width between the two eyepieces of the opto-mechanical reader are user-adjustable. Both the digital and the opto-mechanical reader are equipped with battery-powered, intensity-adjustable illumination lights (i.e., LEDs: visible, IR, and UV possible). Both systems are further equipped with eye cushions.

Results: : The digital reader images the eye at about 20-50x magnification. For example, it takes a snapshot of the iris, which may carry an implanted sensor (e.g., MEMS-based IOP sensor attached to the iris) and displays the image on the built-in LCD viewer. With the optical zoom engaged the reader currently achieves a resolution of about 5µm per pixel and with the additional digital zoom a resolution of about 1.2µm per pixel across a 2560 x 1920 pixel image. These performance parameters are camera-dependent, i.e., they are linearly coupled to the horizontal and vertical camera resolution at hand. The opto-mechanical reader achieves currently a 20-50x magnification.

Conclusions: : Both the digital and the opto-mechanical reader are technically straightforward. The digital reader utilizes off-the-shelf consumer digital cameras and hence may provide a low-cost digital microscope capability. The advantage of the opto-mechanical reader is that it does not require any type of camera, image processing, or other electronic parts. Both reader designs may enable users to image or self-examine their eyes without the help of others. They may assist the treatment and management of glaucoma.

Keywords: microscopy: light/fluorescence/immunohistochemistry • anterior chamber • intraocular pressure 
×
×