Until recently, one criterion used to determine keratoconus progression was a change of 1 D in
K-max with the Pentacam device. This was based on the limits of reproducibility (0.8 D) established in 100 healthy eyes, as reported by McAlinden et al.
4 A report on the limits of reproducibility in 32 keratoconus eyes, demonstrated a clear difference in repeatability of K-max (2.3 D).
1 In this study by Hashemi et al.,
1 three images were taken by two observers. The tolerance can be calculated on the reproducibility limits of a single image taken by each observer (
TR1), average of pairs of images taken by each observer (
TR2), and average of triplets of images taken by each observer (
TR3). The tolerance index for reproducibility of
K-max,
TR1K-max, is Log
n(2.3/0.8) = 1.1, the
TR1 cutoff is 0.24 (N1 = 100, N2 = 32, see
Table 1). Because in this example
TR1K-max > TR cutoff, this summarizes the statistically significant difference. The traditional reproducibility limits are given in the column labeled “
R” in
Table 2, and there is a scale attached (mm
3, D, deg, or μm). In keratoconus, up to 20 parameters are monitored by the clinician, yet no comparative interpretation can be made without background knowledge on normative limits. Tolerance (column labeled
TR1) provides an easier way to note where additional care is required, as values outside normal limits (greater than the TR cutoff) are evident. Those significantly better than normal are marked in italic, those that are significantly worse are marked in bold, those similar to normal are in plain font (
Table 2, TR values). The
K-max is clearly outside normal limits, but so also is pachymetry at thinnest corneal thickness (TCT) and Apex, corneal power in diopters (KPD), anterior chamber (AC) volume, and AC depth, with
K-max and pachymetry measures being the most affected.