Abstract
Purpose:
To examine and evaluate ocular biomechanical metrics and additionally derived corneal and orbital components using a noncontact Scheimpflug-based tonometer (CorVis ST) in a population of healthy eyes.
Methods:
A total of 152 eyes of 152 participants were examined by slit-lamp biomicroscopy, corneal tomography, and the CorVis ST (CST). This determined the distribution of outputs from the CST, such as deformation amplitude (DA), and additionally derived parameters, including maximum corneal deformation (MCD), maximum orbital deformation (MOD), and corneal energy dissipation (CED).
Results:
The mean age of participants was 35.88 ± 13.8 years. Deformation amplitude significantly correlated with age (r = 0.24, P = 0.002) but not sex or ethnicity (P > 0.05). Multiple linear regression analysis revealed significant correlations between DA and age (r = 0.19, P = 0.006) and DA and IOP (r = −0.59, P < 0.001). Age correlated with MCD (r = 0.20, P = 0.01), MOD (r = 0.18, P = 0.03), and CED (r = 0.39, P < 0.001). Males had a lower MOD than females (0.24 vs. 0.26 mm, respectively, P = 0.01); however, there were no differences in MCD or CED between sexes (P > 0.05). There were no significant differences between ethnicities for MCD, MOD, and CED (P > 0.05). Multiple linear regression analysis revealed significant correlations between MCD and IOP (r = −0.65, P < 0.001), CED and age (r = 0.41, P < 0.001), CED and IOP (r = 0.28, P = 0.001), and between CED and central corneal thickness (CCT) (r = −0.36, P < 0.001).
Conclusions:
The isolation of the corneal component (MCD) should be used when analyzing deformation characteristics in diseases that only affect the cornea. This study establishes a baseline for a population of healthy eyes. Future publications will identify differences in MCD, MOD, and CED between healthy and diseased populations.
The cornea is a viscoelastic structure that exhibits classic biomechanical properties such as elasticity, creep, stress relaxation, and hysteresis.
1 These mechanical properties stem from the constituent components of the cornea and therefore are governed by factors such as the extracellular matrix and the organization of collagen fibers.
2
The assessment of corneal biomechanical properties has a number of potential clinical applications such as improving the accuracy of IOP measurements,
3,4 the diagnosis of keratoconus (KC),
5 assessing the biomechanical effects of corneal collagen crosslinking (CXL),
6 and assessing the risk of post-LASIK ectasia.
7,8
Recently, the CorVis ST (CST; Oculus, Wetzlar, Germany) was introduced as a clinical tool for the assessment of corneal biomechanical properties in vivo. The CST is a noncontact tonometer that emits a collimated air-pulse (3.05-mm diameter) that is coupled with a high-speed Scheimpflug camera (4330 frames/sec, aligned at 45 degrees to the cornea). The air-pulse causes the cornea to deform inward into a concave configuration before returning to its initial shape. The real-time video analysis of the deformation profile allows for a number of unique measurements to be made, including deformation amplitude (DA).
9 This is the total displacement of the corneal apex from its initial position to the point of highest concavity. Previously, studies have described DA as a solely corneal reaction.
10–12 However, the in vivo response to an air-pulse is not purely corneal.
13 It includes an indirect and collateral globe displacement secondary to the transfer of energy to the orbital tissue and extraocular muscles.
13 Unfortunately, the inbuilt CST software (v6.07r24) does not provide the isolated pure corneal and orbital components. However, separated components may be derived through external analysis of the CST videos.
14
Application of a cyclic load to viscoelastic materials results in a loss of energy per cycle due to the inherent viscosity in the system.
15 By calculating the area within the loading and unloading curves of the stress-strain graph, the dissipated energy can be determined quantitatively.
16 Ishii et al.
17 previously described the area within the graph produced as “elastic hysteresis.” They utilized a noncontact tonometer in conjunction with a high-speed camera angled at 90° to the cornea and plotted the change in corneal apex position (strain) against the amount of force applied onto the cornea by the air-pulse (stress). However, the camera could not image the cornea past the applanation point.
17
This problem is overcome by using the videos provided by the CST, in which the corneal apex position during the whole deformation process can be tracked (including past the applanation point). This is then refined to exclude the orbital deformation, and correlated to the force applied by the air-pulse. We have termed the area within the resulting graph produced as corneal energy dissipation (CED).
The aim of this study included the determination of the distribution of CST parameters, separated components (corneal and orbital), and CED in a population of healthy eyes. It also describes the repeatability of the separated components and hence establishes baseline data. Finally, the statistically significant and nonsignificant correlations for age, IOP, maximum keratometry (KMAX), and central corneal thickness (CCT) with the additional parameters were examined. This is to aid in attempting to control for parameters that are potential confounders when comparing between healthy and diseased populations.
All participants were recruited from staff, relatives of staff, and relatives of patients at the Greenlane Clinical Centre, Auckland District Health Board, New Zealand. Exclusion criteria were previous ocular trauma/surgery or infections, contact lens wear, or any systemic or ocular disease that may affect the cornea.
For each participant, demographic data, such as age, sex, and ethnicity, were collected. Only the right eye was examined for each participant unless it fell under one of the exclusion criteria listed above, in which case, the left eye was examined.
Informed consent was obtained from all participants and the study design adhered to the tenets of the Declaration of Helsinki. The study protocol was approved by the University of Auckland human ethics committee (reference number 010853).
Statistical analysis was performed using SPSS software version 22.0 (IBM Corporation, Chicago, IL, USA) and Prism software version 6.01 (GraphPad, La Jolla, CA, USA). The normality of the data was tested using the D'Agostino and Pearson omnibus normality test.
Descriptive statistical analysis was performed on baseline characteristics. A Wilcoxon matched-pairs signed rank test was used for differences between DA, MCD, and MOD. Two-tailed Spearman's correlation coefficient was used to measure statistical dependence between nonparametric variables. Significant differences between sexes were investigated using the Mann-Whitney U test. A Kruskal-Wallis test was conducted to explore the differences in variables between ethnicities. Newly derived parameters (MCD, MOD, and CED) were tested for repeatability using the Bland-Altman methodology. The mean within-subject SD (Sw) was determined. Precision (Pr) and repeatability were calculated using the formulas 1.96 × Sw and 2.77 × Sw, respectively. The mean of within-subject means was divided by Sw to derive the coefficient of variation (CV). Intraclass correlation (ICC) was calculated by using ANOVA. An ICC of more than 0.70 and a CV less than 20% were considered highly repeatable; an ICC of more than 0.60 and CV less than 20% were considered fairly repeatable.
Multiple linear regression analysis was performed to determine if age, IOP, CCT, and K
MAX are significant predictors of DA, MCD, MOD, or CED. Only highly repeatable and reliable CST parameters have been included in the multiple linear regression analysis.
5,19 Central corneal thickness and K
MAX were obtained using the Pentacam HR (v6.07r29).
P values of 0.05 or less were considered to be statistically significant.
A total of 152 healthy subjects were recruited with a mean age of 35.88 ± 13.8 years (range, 13–74 years). There was a larger proportion of females (62%, n = 95) than males (38%, n = 57). Most participants were of European ethnicity (53.8%, n = 81), followed by Indian subcontinent (18.6%, n = 29), Southeast Asian (11.5%, n = 18), Pacific Island Nations (5.1%, n = 8), Middle Eastern (4.5%, n = 7), Maori (4.5%, n = 6), and African descent (1.9%, n = 3).
Values for all CST parameters are presented in
Table 1.
Table 1 The Mean, SD, and Range of Values for Each Parameter Measured by the CorVis ST in Healthy Eyes (n = 152)
Table 1 The Mean, SD, and Range of Values for Each Parameter Measured by the CorVis ST in Healthy Eyes (n = 152)
The mean DA was 1.07 ± 0.10 mm. There was a significant correlation between DA and age (
r = 0.24,
P = 0.002;
Fig. 2); however, there were no significant differences in DA between sexes (
P = 0.63) or ethnicities (
P = 0.20). Multiple linear regression analysis revealed that DA was significantly positively correlated with age (
r = 0.19,
P = 0.006) and negatively correlated to IOP (
r = −0.59,
P < 0.001).
The in vivo eyeball reaction to an air-pulse includes contributions from the cornea, orbital tissue, and extraocular muscles.
13 Koprowski
14 described methods of analyzing CST videos to derive supplementary parameters accounting for these components. These additional methods of analyzing CST videos have not been investigated in healthy corneas. We determined the distribution of CST parameters, and separated components and the correlation of these parameters with age, sex, ethnicity, corneal thickness, and corneal curvature in a population with healthy corneas.
Previous studies have considered DA as an exclusively corneal phenomenon
10–12; however, this study provides evidence of the distribution and significant difference between DA (provided by the device) and the isolated corneal (MCD) and orbital (MOD) components in a population of healthy corneas. Isolation of the orbital component should be used when analyzing deformation characteristics in diseases that only affect the cornea. Furthermore, future publications need to consider these additional parameters, as they could identify whether the corneal or the orbital component contributes to the differences in overall DA.
The aging cornea is known to change its stromal microstructure. The cornea undergoes natural glycation-induced crosslinks causing an increased cross-sectional area within collagen fibers and fibrillar molecules.
20,21 Corresponding biomechanical increases in corneal tissue stiffness have been observed through ex vivo inflation testing.
22–24 However, biomechanical changes with age using in vivo testing methods are still debated.
There are two commercially available devices that claim to measure biomechanical properties of the cornea in vivo, the Ocular Response Analyzer (ORA; Reichert, Inc., Depew, NY, USA) and the CST, although direct comparisons between these instruments are not possible, as the pressure profiles of the ORA vary between measurements, whereas the CST has a consistent pressure profile in all measurements.
25 Any comparisons made between these machines must adjust for these differences before drawing conclusions.
Several studies have investigated the relationship between ORA parameters and age.
8,26–29 Foster et al.
26 demonstrated a significant correlation with age for both output parameters (corneal hysteresis [CH] and corneal resistance factor [CRF]) in a large study involving 4184 eyes; however, several factors were not incorporated into a multivariate analysis, including IOP, CCT, and keratometry. Other studies have reported similar observations but with weaker correlations.
8,27,28 Conversely, Kirwan, et al.
29 showed no correlation between CH and age in a subset of children.
In the current study, there was a significant correlation between age and DA, and the correlation remained after isolating the corneal component. This is in agreement with Leung et al.,
11 who demonstrated a statistically significant, weak relationship (
r = 0.002) between age and DA in a group of normal, glaucoma, and glaucoma-suspect eyes. Tian et al.
12 also described a significant positive correlation (
r = 0.43) between age and DA. Conversely, others have discovered no significant correlations between age and DA.
5,30
An increase in DA with age is counterintuitive to ex vivo studies that exhibit increased stiffness with age
22–24; however, the deformation response to an air-pulse has been claimed to be largely influenced by corneal interlamellar sliding resistance.
31 These resistance forces would be influenced by the content and distribution of the proteoglycan matrix interspersed between collagen fibrils and collagen lamellae. Age is associated with an increase in cross-sectional collagen fibril and fibrillar molecular area but with decreased interfibrillar spacing.
20,21,32 This indicates reduced proteoglycan substance in the aged cornea, which may subsequently cause decreased interlamellar sliding resistance. Furthermore, the nonenzymatic crosslinks shown to occur with age may not affect these resistance forces.
20
Loss or atrophy of orbital fat is a natural aging process.
33 Observations in this study imply a more compliant orbit with age, as MOD increased with age (
r = 0.18,
P = 0.03). This is in agreement with a previous study by Frueh,
34 which demonstrated that the force required to displace the orbit posteriorly a fixed distance using a contact lens was lower in the older age group (>45) when compared with the younger age group (>45). Furthermore, our data revealed a statistically significant (
P = 0.01) difference in MOD between sexes. The magnitude of this difference (0.02 mm) is not clinically significant.
Hysteresis is a crucial parameter for understanding corneal biomechanical behavior. The physical property describes the difference between the loading and unloading curves.
16 The behavior hails from the viscous component dissipating energy within the viscoelastic system.
25 Analysis of corneal hysteresis could reveal the viscous damping effect in the complex viscoelastic cornea.
24 The area between the loading and unloading curves for a deformation response to an air-pulse has been previously described by Ishii et al.
17 as “elastic hysteresis.” However, the cornea was not imaged past the applanation point in the deformation process and the orbital response to the air-pulse was not isolated.
17
In the current study, CED incorporates the corneal apex position past the applanation point and considers only the pure corneal response. A significant correlation between age and CED was observed in this dataset. However, ex vivo inflation studies were not able to demonstrate statistically significant differences in hysteresis with age.
24 In this study, the positive correlation identified between age and CED is in contrast to the observations of Ishii et al.,
17 who reported a statistically significant decrease in hysteresis with age.
Hysteresis is not only dependent on the viscosity of the material but the nature of the pressure applied onto the cornea. Previously published in vivo and ex vivo studies used different instruments that use alternate pressure characteristics. Thus, direct comparisons in hysteresis values between these studies are not possible.
A potential limitation of this study is that the true corneal apex position may have been displaced slightly during the deformation process or, alternatively, the air-pulse profile may not have been directly applied at the center of the cornea.
As shown in
Figure 5, normal measurements can be taken where the apex is tilted away from the center of the air-pulse profile. This could impact the deformation analysis.
6 The true apex cannot be followed unless a method of tracking the apex is provided or the air-pulse is applied correctly in every measurement. Lago et al.
35 proposed a methodology for an in vivo estimation of the biomechanical behavior. The biggest contributor to dissimilarity between simulated deformed corneas and real deformed corneas were in cases in which the positions of the air-pulse were not aimed directly at the center.
35 Thus, misaligned measurements can cause significant changes to deformation characteristics.
Additional research is needed to predict the magnitude of change in deformation characteristics corresponding to variations in tilt. This would first require the development of a reliable and repeatable method of deriving the amount of tilt in individual videos.
A three-dimensional imaging technique would allow a thorough understanding of the energy dissipated in a viscoelastic structure; however, the CST provides two-dimensional images. Additionally, CED only incorporates information over a single dimension. This reduction in complexity may have contributed to the results observed in this study.
Future studies will develop methods of accounting for this tilt and analyze the impact of these additional components between normal and diseased populations.