All ERG waveforms consistently showed a- and b-wave amplitudes and OPs on the rise of the b-wave at luminances greater than −0.54 log cd·s/m
2 (
Fig. 3A). Like the pSTRs, amplitudes generally were greater with higher luminance levels (
P = 0.0011) and during elevated IOP (
Figs. 3A–C). Specifically, a-wave amplitudes during loop wear (−228.4 ± 29.7 μV) were significantly greater in magnitude than those during the preloop (−31.4 ± 9.9 μV;
P < 0.0001) and postloop (−49.0 ± 12.6 μV;
P < 0.0001) conditions (
Fig. 3B). Similarly, b-wave amplitudes during loop wear (747.9 ± 74.5 μV) were significantly greater than those before (123.8 ± 29.0 μV;
P < 0.0001) and following (200.0 ± 56.0 μV;
P < 0.0001) loop wear (
Fig. 3C). In the treated eyes, there were no differences between the pre- and postloop conditions in the a-wave amplitudes (
P = 0.4602), but b-waves amplitudes during postloop condition were greater than those before loop wear (compare 200.0 ± 56.0 vs. 123.8 ± 29.0 μV, respectively;
P = 0.0081). The enhanced a-wave and b-wave amplitudes in the treated eyes were significantly greater than those in the control eyes (
P < 0.0001 for both;
Figs. 3B,
3C). A- and b-wave amplitudes in control eyes did not differ as a function of contralateral loop wear (
P = 1.000 for both ERG components). A-wave implicit times also were longer during loop wear (31.6 ± 2.4 ms) than those during preloop (27.8 ± 2.6 ms;
P < 0.0001) and postloop (27.8 ± 1.8 ms;
P < 0.0001) conditions (
Fig. 3B, inset). Similarly, b-wave implicit times were also longer during loop wear (78.9 ± 5.6 ms) than those during pre- (68.9 ± 5.6 ms;
P < 0.0001) and postloop (73.5 ± 5.2 ms,
P = 0.0223) conditions (
Fig. 3C, inset). There were no differences between pre- and postloop a-wave implicit times (
P = 1.000), however, b-wave implicit times for postloop conditions were longer than those for preloop (73.5 ± 5.2 vs. 68.9 ± 5.6 ms respectively;
P = 0.0247) conditions. Implicit times during loop wear for a- and b-waves were greater than control eyes (31.6 ± 2.4 vs. 28.5 ± 2.0 ms,
P = 0.0008 for a-waves; 78.9 ± 5.2 vs. 72.5 ± 3.8 ms,
P = 0.0054 for b-waves; insets of 3B and 3C). For both a- and b-wave implicit times, control eyes did not change for all comparisons (
P > 0.1275). Oscillatory potential amplitudes were greater in the treated eyes during loop wear (
Fig. 4A) and the root mean square values the OPs during IOP elevation (33.1 ± 10.1 μV) increased significantly compared to pre- (7.0 ± 1.3 μV;
P < 0.0001) and postloop (8.4 ± 2.9 μV;
P < 0.0001) wear conditions (
Fig. 4B). For treated eyes, pre- and post-RMS values did not significantly differ (
P = 1.000). Root mean square values did not change for control eyes for all loop conditions (
P = 1.000).