Effects of Intraocular Lens Opacification on Light Scatter, Stray Light, and Overall Optical Quality/Performance
Author Affiliations & Notes
  • Liliana Werner
    John A. Moran Eye Center University of Utah, Salt Lake City, Utah, United States
  • John C. Stover
    The Scatter Works, Inc., Tucson, Arizona, United States
  • Jim Schwiegerling
    College of Optical Sciences, University of Arizona, Tucson, Arizona, United States
  • Kamal K. Das
    Alcon Laboratories, Inc., Fort Worth, Texas, United States
  • Correspondence: Liliana Werner, John A. Moran Eye Center, University of Utah, 65 Mario Capecchi Drive, Salt Lake City, UT 84132, USA; [email protected]
Investigative Ophthalmology & Visual Science June 2016, Vol.57, 3239-3247. doi:https://doi.org/10.1167/iovs.16-19514
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Liliana Werner, John C. Stover, Jim Schwiegerling, Kamal K. Das; Effects of Intraocular Lens Opacification on Light Scatter, Stray Light, and Overall Optical Quality/Performance. Invest. Ophthalmol. Vis. Sci. 2016;57(7):3239-3247. https://doi.org/10.1167/iovs.16-19514.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: To evaluate light scatter and stray light in intraocular lenses (IOLs) explanted because of postoperative opacification (13 calcified hydrophilic acrylic, 1 calcified silicone, and 4 polymethylmethacrylate [PMMA] lenses with snowflake degeneration), as well as effect of opacification on other optical quality/performance indicators, in comparison with controls.

Methods: The Complete Angle Scatter Instrument (CASI) scatterometer was used to measure the forward light scattering (FLS) of the IOLs, and the stray light values at various angles were calculated from the measured FLS. Modulation transfer function (MTF) was obtained with an optical bench, and a Badal optometer was used to obtain letter chart images through the lenses. Back light scatter and light transmittance were also measured.

Results: Average stray light values (Log (s)) at a scattered angle of 10° were 1.79 ± 0.37 for hydrophilic acrylic IOLs (controls 0.36 ± 0.05), 1.53 for the silicone lens (control 0.41), and 1.62 ± 0.46 for PMMA IOLs (control 0.25). Stray light was significantly higher for explanted opacified lenses (N = 18) in comparison with controls (N = 7; two-tail P < 0.001 at 10°). Modulation transfer function and Badal image contrast were drastically reduced in lenses with calcification and snowflake degeneration.

Conclusions: Different studies described the impact of stray light in human vision, with serious hindrance above 1.47 Log (s). Lenses explanted from patients because of clinically significant opacification are associated with a considerable increase in light scatter and stray light, as well as with a decline of other optical quality/performance indicators.

Cataract surgery is the most prevalent operation in the United States, with more than 3 million procedures performed in 2006 alone.1 A variety of new artificial intraocular lens (IOL) designs, manufactured from different biomaterials are continuously being made available to cataract surgeons to replace the refractive power of the surgically-removed opacified crystalline lens. The biomaterials used in IOL manufacture are basically divided into two main classes: acrylic (rigid and foldable) and silicone (foldable). Although relatively rare, explantation of an IOL may be required due to problems such as error in IOL power calculation, postoperative inflammation, and postoperative opacification of the IOL, among others.2 
Postoperative calcification represents the most important indication for explantation of hydrophilic acrylic IOLs.27 Silicone lenses were also shown to calcify in eyes with asteroid hyalosis.810 The condition named snowflake degeneration of rigid polymethylmethacrylate (PMMA) lenses still leads to explantation of a significant number of these IOLs, although it is usually observed in three-piece lenses manufactured in the 80s and early 90s.11,12 Calcification and snowflake degeneration are leading causes of clinically significant IOL opacification.3 Clinical studies on opacified IOLs demonstrated that they may be more associated with complaints of glare and “foggy” or “cloudy” vision, while visual acuity is not drastically reduced.13,14 The quality of vision loss in affected patients appears to be predominantly related to increased forward light scattering and subsequently retinal stray light, which can give rise to a variety of subjective complaints, including glare in various illuminating conditions, halos around bright lights, color and contrast loss, as well as hazy vision.15 These symptoms may prompt IOL explantation even in the absence of measurable decrease in visual acuity.1315 
In previous studies, we evaluated light scattering and light transmission in IOLs explanted because of clinically significant postoperative opacification.16,17 Very high levels of light scattering and a potential for decrease in light transmittance were found. However, only back scattered light was measured with a Scheimpflug camera (dispersion of light reflected out of the eye that can be seen by an external observer). More recently, a technique developed for measuring the forward light scattering (light toward the retina) of IOLs in vitro was described.18 Through specific mathematical relationships, forward light scattering can be converted to a more clinically relevant metric to determine the potential impact of optic opacification on visual performance, the stray light.18 Therefore, the aim of the current study was to provide a complete characterization of the effects of IOL opacification on the visual function, by measuring forward light scattering and stray light, as well as other optical quality/performance indicators. 
Materials and Methods
Intraocular lenses explanted because of optic opacification associated with significant decrease in the visual function, with symptoms of glare and cloudy vision, and sent to the John A. Moran Eye Center at the University of Utah (Salt Lake City, Utah, USA) for evaluation were used in this study. All lenses underwent gross and light microscopy. In order to remove surface crystals corresponding to dry ophthalmic viscosurgical device (OVD) and/or balanced salt solution (BSS) used during the explantation procedure, the IOLs were rinsed with distilled water and allowed to dry at room temperature before re-evaluation under gross and light microscopy. Because the IOLs needed to be fixated inside special supports for the various measurements, only IOLs explanted without major deformation of the haptic components could be used. The following groups of IOLs were included: 
  1.  
    Single-piece or three-piece hydrophilic acrylic lenses, explanted because of calcification (N = 13);
  2.  
    Single-piece plate silicone lens, explanted because of calcification in an eye with asteroid hyalosis (N = 1); and
  3.  
    Three-piece PMMA lenses, explanted because of snowflake degeneration (N = 4).
Never implanted hydrophilic acrylic, silicone plate, and PMMA lenses were used as controls (5 hydrophilic acrylic lenses, 1 silicone, and 1 PMMA lens). Table 1 provides information on the models of explanted and control IOLs used in this study. Regarding the explanted calcified hydrophilic acrylic lenses, as well as the control hydrophilic acrylic lenses, the water content of their materials is as follows: 25% (Ioflex; Mediphacos, Belo Horizonte, Brazil, and Aqua Sense; Aaren Scientific, Ontario, CA, USA), 26% (PolyLens; Tekia, Irvine, CA, USA, and Centerflex; Rayner, Hove, UK), 18% (Hydroview; Bausch & Lomb, Rochester, NY, USA), and 20% (MemoryLens; CibaVision, Duluth, GA, USA). None of the hydrophilic acrylic lenses included has a hydrophobic coating/surface modification.3,16,17 
Table 1
 
Explanted and Control Intraocular Lenses Included in the Study
Table 1
 
Explanted and Control Intraocular Lenses Included in the Study
The explanted and control silicone lenses were represented by the Chiroflex C10UB (Chiron Vision, Emeryville, CA, USA). Two of the explanted PMMA lenses were represented by model UVL-304-01 (Ioptex, Irwindale, CA, USA). The control PMMA lens was model 809A (Pharmacia, Uppsala, Sweden). 
Forward Light Scattering
Measurements were performed with the Complete Angle Scattering Instrument (CASI) scatterometer. This instrument uses a laser source of 633 nm to illuminate BSS-immersed IOLs within a cylindrical wet cell. The wet cell was placed at the center of rotation of a goniometer arm. A detector measured light scattered per unit solid angle from the IOL as a function of angle. The measurements provided a profile of the scatter distribution for regions outside of the directly transmitted beam represented as the bidirectional scattered distribution function (BSDF). Scattering results were then converted to a more clinically relevant metric to determine the potential impact of optic opacification on visual performance (stray light values). Details on the measurement of forward scattering by using this device, as well as the conversion of measured values into stray light values have been recently described.18 
Modulation Transfer Function and Badal Images
Modulation transfer is the ability of a lens or optical system to transfer an object's contrast to its image. Modulation transfer function (MTF) of the lenses was obtained in the hydrated state using an optical bench and a model eye. The model eye imaged an illuminated slit and the MTF was proportional to the modulus of the Fourier transform of the slit image. The MTF values were obtained using 3- and 5-mm pupils. Badal images were also obtained using a letter chart object with the same pupil sizes.19 
Back Light Scattering and Light Transmittance
In order to confirm results from previous studies, back scattering and light transmittance were measured.16,17 A customized three-piece dark eye model with a PMMA cornea was used to hold the IOLs under immersion in BSS. Care was taken to prevent the presence of air bubbles inside of the eye model during loading and assembly. The BSS-filled model containing the IOL was then placed in front of a Nidek (Gamagori, Japan) EAS-1000 Scheimpflug camera (cornea facing the device), and the room lights were turned off. A cross-sectional image of the IOL inside of the model was then obtained (settings: flash levels 200 W and 50 W; slit length 10 mm; meridian angle 0) and analyzed using the densitometry peak function. Light scattering values were expressed in computer compatible tape (CCT). This is a measure of brightness or intensity of reflected (scattered) light.16,17 According to the operator's manual, the raw pixel density level of the device ranges from 0 to 255 without background correction. 
Measurements of light transmittance were performed with a Lambda 35 UV/Vis spectrophotometer, equipped with a Lab Sphere RSA-PE-20 integrating sphere with a 50-mm diameter (Perkin-Elmer, Waltham, MA, USA) operated in a single beam configuration. Each IOL was fitted to a plastic custom insert with a 5.0-mm diameter aperture for the optic (designed to hold a 6.0-mm diameter optic). The insert containing the IOL was then mounted into a standard rectangular quartz cuvette filled with BSS. Care was taken to prevent the presence of air bubbles inside the cuvette. The assembly was then placed directly in front of the opening of the integrating sphere so that the anterior surface of the IOL was facing the light source. Prior to the measurements, a background correction was performed with the empty inserter immersed in BSS inside the quartz cuvette. Background transmittance spectra were checked to ensure that 100 ± 0.5% transmittance was achieved. Intraocular lens spectra were then collected at room temperature with the following settings: wavelength range 300 to 850 nm; slit width 2 nm; scan speed 120 nm/min; data interval 1 nm. At least two measurements were recorded for each IOL and averaged; background transmittance was checked every other sample to ensure that it did not shift during measurements. Results were expressed as percent light transmittance (%T) in the visible light spectrum (400–700 nm).16,17 
Results
Figure 1 shows representative gross photographs of the lenses explanted because of clinically significant optic opacification. Gross and light microscopy of the explanted hydrophilic acrylic lenses showed calcified deposits, which were predominantly located within the optic substance (N = 9) or on the surface/subsurface of the lenses (N = 4). The single-piece, plate silicone lens with large positioning holes exhibited a dense central area of calcified deposits located on the posterior optic surface only. Gross and light microscopy of the explanted PMMA lenses confirmed the presence of typical intraoptic snowflake lesions. They were usually confined to the most central part of the optic, leaving a clear peripheral edge devoid of lesions. 
Figure 1
 
Gross (macroscopic) photographs of a representative lens from each of the three groups. (A) Calcified single-piece hydrophilic acrylic lens. Diffuse calcification was found predominantly within the substance of the lens optic and haptics. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. The calcification was localized to the center of the posterior optic surface. (C) Three-piece PMMA lens with snowflake degeneration. Snowflake lesions were present within the optic of the lens, sparing the optic periphery.
Figure 1
 
Gross (macroscopic) photographs of a representative lens from each of the three groups. (A) Calcified single-piece hydrophilic acrylic lens. Diffuse calcification was found predominantly within the substance of the lens optic and haptics. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. The calcification was localized to the center of the posterior optic surface. (C) Three-piece PMMA lens with snowflake degeneration. Snowflake lesions were present within the optic of the lens, sparing the optic periphery.
Forward Light Scattering
The plots in Figures 2 through 4 display the forward light scattering of representative lenses from the three different groups. Details on the stray light values, calculated from the measured forward light scattering values are presented in Table 2. Forward light scattering and stray light values from lenses explanted because of calcification and snowflake degeneration were considerably higher than the values for the corresponding controls. Comparison of stray light values between explanted opacified lenses (N = 18) and controls (N = 7) was statistically significant for all angles analyzed (two-tail P < 0.001 at 10°, t-test; 95% confidence interval for the mean difference 1.08–1.68). 
Figure 2
 
Graphs comparing forward light scattering from four representative explanted hydrophilic acrylic lenses exhibiting calcium deposits predominantly within the optic substance, and two control lenses of the same design. PSF, point spread function.
Figure 2
 
Graphs comparing forward light scattering from four representative explanted hydrophilic acrylic lenses exhibiting calcium deposits predominantly within the optic substance, and two control lenses of the same design. PSF, point spread function.
Figure 3
 
Graphs comparing forward light scattering from one explanted calcified silicone lens, and a control lens of the same design.
Figure 3
 
Graphs comparing forward light scattering from one explanted calcified silicone lens, and a control lens of the same design.
Figure 4
 
Graphs comparing forward light scattering from four PMMA lenses explanted because of snowflake degeneration, and a control PMMA lens.
Figure 4
 
Graphs comparing forward light scattering from four PMMA lenses explanted because of snowflake degeneration, and a control PMMA lens.
Table 2
 
Stray Light Values (Log (s)) for Explanted Hydrophilic Acrylic, Silicone, and PMMA Lenses Calculated From Forward Light Scattering Measurements Obtained at Various Angles.
Table 2
 
Stray Light Values (Log (s)) for Explanted Hydrophilic Acrylic, Silicone, and PMMA Lenses Calculated From Forward Light Scattering Measurements Obtained at Various Angles.
Modulation Transfer Function and Badal Images
Modulation transfer function and Badal image contrast were found to be reduced for all explanted lenses with calcification and snowflake degeneration. Figure 5 shows the MTF graphs of a representative lens from the calcified hydrophilic acrylic group with corresponding control. Figure 6 shows Badal images obtained from a representative PMMA lens with snowflake degeneration and corresponding control. 
Figure 5
 
Modulation transfer function of a hydrophilic acrylic lens explanted because of calcification, as well as a control lens of the same design, measured at 3.0- and 5.0-mm pupils.
Figure 5
 
Modulation transfer function of a hydrophilic acrylic lens explanted because of calcification, as well as a control lens of the same design, measured at 3.0- and 5.0-mm pupils.
Figure 6
 
Badal images of a PMMA lens explanted because of snowflake degeneration (A, C), as well as a control PMMA lens (B, D), measured at 3.0- (A, B) and 5.0- (C, D) mm pupils.
Figure 6
 
Badal images of a PMMA lens explanted because of snowflake degeneration (A, C), as well as a control PMMA lens (B, D), measured at 3.0- (A, B) and 5.0- (C, D) mm pupils.
Back Scattered Light
Figure 7 shows Scheimpflug photographs with densitometry measurements of representative explanted lenses from each of the three groups. Eight of 18 explanted opacified lenses (44%) had saturated peak density measurements, as values were greater than or equal to 200 CCT at both the highest (200 W) and lowest (50 W) flash settings. In contrast, 100% of the control samples were less than or equal to 9 CCT at the highest (200 W) flash setting. The percent incidence of explanted opacified lenses with CCT greater than or equal to 200 was compared with the controls using Fisher's exact test. Two-tailed P value was less than 0.001 at 200 W. 
Figure 7
 
Scheimplug photographs with densitometry analyses of a representative lens from each of the three groups (same lenses as in Fig. 1). (A) Calcified single-piece hydrophilic acrylic lens; backward light scattering was measured as 222 CCT. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. Backward scattering = 223 CCT. (C) Three-piece PMMA lens with snowflake degeneration. Backward scattering = 225 CCT (image obtained with a flash level of 50 W).
Figure 7
 
Scheimplug photographs with densitometry analyses of a representative lens from each of the three groups (same lenses as in Fig. 1). (A) Calcified single-piece hydrophilic acrylic lens; backward light scattering was measured as 222 CCT. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. Backward scattering = 223 CCT. (C) Three-piece PMMA lens with snowflake degeneration. Backward scattering = 225 CCT (image obtained with a flash level of 50 W).
Light Transmittance
The average light transmittance (T) in the visible light spectrum for all 13 calcified hydrophilic acrylic lenses was 87.38 ± 4.85%. The mean value obtained for the control hydrophilic acrylic lenses was 97.87 ± 0.70%. The %T for the silicone plate lens explanted because of calcification in an eye with asteroid hyalosis was 94.6%; it was 97.7% for the never implanted control silicone lens. The %T for the PMMA lenses exhibiting snowflake degeneration was 88.9 ± 3.7%. The value obtained for the never implanted control PMMA lens was 98.8%. 
Discussion
Calcification (hydrophilic acrylic and silicone IOLs), and snowflake degeneration (PMMA IOLs) usually lead to progressive opacification of the IOL optic, resulting in significant decrease of the visual function requiring explantation/exchange of the affected lens.3 Calcification of hydrophilic acrylic lenses appears to be a multifactorial problem, and factors related to IOL manufacture, IOL packaging, surgical techniques and adjuvants, as well as patient metabolic conditions, among others may be implicated. As the exact combination of factors and sequence of events ultimately leading to calcification of the lenses is still unknown, continuous research on this complication is warranted.3 The majority of the studies on calcified hydrophilic acrylic lenses describe explantation during the second postoperative year, or earlier. In some cases, the deposits causing the opacification are found on the optical surface/subsurface of the lenses, while they are predominantly found within the optic substance in others.47 In cases of calcification of silicone lenses in eyes with asteroid hyalosis, studies demonstrated deposits only on the posterior optic surface of the lenses. The deposits could be partially removed with Nd:YAG laser, but there was a reaccumulation after the procedure because the asteroid bodies are rich in calcium/phosphate.810 Snowflake degeneration is a slowly progressive condition; it is the result of PMMA degradation caused by long-term ultraviolet light exposure, not calcium deposition. The degree of optic opacification of the majority of the PMMA lenses included in this study may take 10 to 20 years to develop. The intraoptic spherical lesions observed in these cases are thought to correspond to sites of degenerated PMMA material.11,12 
We have already performed two preliminary studies evaluating back scattered light and light transmittance in IOLs explanted because of clinically significant optic opacification.16,17 In both studies, the same Scheimpflug and spectrophotometer devices described in our current study were used. The first study included various designs of hydrophilic acrylic lenses with different patterns of calcification, as well as a silicone lens explanted because of calcification on the posterior optic surface in an eye with asteroid hyalosis, and PMMA lenses with snowflake degeneration.16 Back scattered light in that series was extremely high, with 26 of 31 explanted lenses measured between 205 and 227 CCT. On average, the percentage of light transmittance in that same study was reduced by 8.43% in hydrophilic acrylic IOLs with calcification, 3% in the calcified silicone lens, and 9.53% in the PMMA IOLs with snowflake degeneration.16 In the second study, seven explanted calcified hydrophilic acrylic lenses and eight control lenses of the same design were studied.17 Back light scattering was 219.71 ± 2.62 CCT for explanted IOLs, and 4.75 ± 2.50 CCT for controls IOLs. The average light transmittance in the visible light spectrum was 75.94% to 87.25% for explanted IOLs, and 97.54% to 98.97% for control IOLs.17 
The fact that the earlier publications included only the measurements of back light scattering was a major limitation of those studies.16,17 The forward light scattering measurements and subsequently the stray light would provide a meaningful correlation with possible clinical symptoms related to glare disability and stray light hindrance.1315 Therefore, in the current study forward light scatter, stray light, and other optical quality/performance indicators were also assessed for lenses explanted because of clinically significant opacification. Both the back and forward light scattering were significantly increased in the explanted lenses, which resulted in the reduction of light transmission, MTF, and Badal image contrast for the opacified lenses. 
Different studies demonstrated that stray light increases with age in the healthy eye.2025 Under 40 years of age, a normal stray light value (Log (s)) is 0.9 (s = 8). This value increases to Log (s) = 1.2 (s = 16) at age 65, which corresponds to double the amount of stray light in a 40-year-old eye. On average, stray light remains relatively unchanged at up to 40 years of age. It doubles in noncataractous eyes by the age of 65 years and triples by the age of 77 years as compared with the young eye. Therefore, the older crystalline lens is an important source of stray light, even if it is clear. Any disturbances of the optic media (e.g., cataracts) will further increase the stray light with age.2025 Phacoemulsification with IOL implantation has the potential to revert the increase in stray light related to cataracts, as well as age. A prospective observational case series objectively compared the mean level of intraocular stray light 6 months postoperatively in patients implanted with a multifocal or a monofocal posterior chamber IOL.26 The study found that cataract surgery lowered the mean level of intraocular stray light compared with the level in age-matched noncataractous subjects.26 Also, the amount of light scatter from different IOLs measured in vitro was typically lower than that found in healthy donor crystalline lenses of various ages.27 
van den Berg et al.2025 described the impact of stray light in human vision, with serious hindrance found above 1.47 Log (s). This corresponds to approximately four times the amount of stray light in a 40-year-old eye (s = 30 for 1.47 Log (s) versus s = 8 for 0.9 Log (s)). A 4-fold increase in stray light can be considered a serious handicap, as one can imagine that knowing the hindrance one experiences already before the age of 40 years from a low sun or other against-the-light viewing conditions.2025 Considering the values shown in Table 2, practically all stray light values for the explanted lenses were greater than or equal to 1.47 Log (s). Interestingly, in another study performed on cadaver-eye removed hydrophobic acrylic (AcrySof; Alcon Laboratories, Fort Worth, TX, USA) lenses with subsurface nanoglistenings (small water molecules present within the lens subsurface), none of the lenses exhibited stray light values greater than or equal to 1.47 Log (s).28 Also laboratory aged AcrySof lenses demonstrated stray light values much lower than the stray light hindrance.18 However, further studies are necessary to assess the effects of glistenings (microvacuoles filled with water present within the IOL substance) on stray light based on the measurements of forward light scattering. 
To the best of our knowledge, there is only one study with couple of lenses in the literature assessing stray light measurements in explanted opacified IOLs. van der Meulen et al.29 measured the amount of forward light scattered by two hydrophilic acrylic lenses explanted because of calcification. They used an in vitro set up for angles from 1.7° to 22°, which had been previously used to evaluate the scattered light of donor lenses and spectacle lenses.2024 At 7.5°, stray light values of the opacified lenses were Log (s) 1.8 (s = 63) and Log (s) 2.9 (s = 794), respectively.29 These represent approximately an 8-fold and a 100-fold increase in stray light compared with young normal values, which are expected to lead to major reduction in quality of vision. We obtained stray light values at broader light scattering angles based on the measurements of forward light scattering. In our study, as an example, mean stray light values for opacified hydrophilic acrylic, silicone, and PMMA lenses at 10° were 1.79 ± 0.37, 1.53, and 1.62 ± 0.46 Log (s), respectively. These represent an increase in stray light of approximately 8-fold (s = 62 for 1.79 Log (s)), 4.5-fold (s = 35 for 1.53 Log (s)), and 5.5-fold (s = 43 for 1.62 Log (s)), respectively, in comparison with the young eye (s = 8 for 0.9 Log (s)). These values would also be expected to reduce the quality of vision as these are above the stray light hindrance or threshold. This can also be related to the significant reduction of the MTF and quality of the Badal images obtained from the explanted lenses, in comparison with controls. 
One of the lenses in the study by van der Meulen et al.29 exhibited a degree of stray light largely above the values found in our study (2.9 vs. 1.79 ± 0.37 for calcified hydrophilic acrylic lenses). This is likely a result of different in vitro set up for the measurements, as well as differences in the severity/extent of IOL calcification. In our study, we measured for the first time stray light at a broad light scattering angles in a calcified silicone lens, and in PMMA lenses with snowflake degeneration. Other methods to evaluate light scattering/stray light have been described in the literature.30 In this study, we used the CASI scatterometer to evaluate explanted opacified lenses. This is the same method we have recently used to evaluate hydrophobic acrylic IOLs with subsurface nanoglistenings removed from cadaver eyes.28 
In summary, we provided a complete characterization of the effects of IOL opacification on the visual function, by measuring forward light scattering and subsequently obtaining stray light, as well as other optical quality/performance indicators in a large series of lenses explanted because of calcification and snowflake degeneration. The levels of stray light found with the explanted lenses would be expected to cause a serious negative impact on the visual function, even if visual acuity was not significantly reduced. 
Acknowledgments
The study was presented in part at the ARVO annual meeting, May 2015, Denver, Colorado, USA. 
Supported in part by unrestricted grants from Research to Prevent Blindness, Inc. (New York, NY, USA) and Alcon Laboratories (Fort Worth, TX, USA) to the Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah. 
Disclosure: L. Werner, AMO (F), Alcon (F, C), Anew Optics (F), Bausch & Lomb (F), ClarVista Medical (F), CORD (F), Hoya (F), LensGen (F), Medicontur (F), Mynosys (F), Omega (F), Powervision (F, C), Zeiss (F); J.C. Stover, P; J. Schwiegerling, None; K.K. Das, Alcon (E) 
References
Rush SW, Gerald AE, Smith JC, Rush JA, Rush RB. Prospective analysis of outcomes and economic factors of same-day bilateral cataract surgery in the United States. J Cataract Refract Surg. 2015; 41: 732–739.
Mamalis N, Brubaker J, Davis D, Espandar L, Werner L. Complications of foldable intraocular lenses requiring explantation or secondary intervention–2007 survey update. J Cataract Refract Surg. 2008; 34: 1584–1591.
Werner L. Causes of intraocular lens opacification or discoloration. J Cataract Refract Surg. 2007; 33: 713–726.
Haymore J, Zaidman G, Werner L, et al. Misdiagnosis of hydrophilic acrylic intraocular lens optic opacification: report of 8 cases with the MemoryLens. Ophthalmology. 2007; 114: 1689–1695.
Werner L, Apple DJ, Escobar-Gomez M, et al. Postoperative deposition of calcium on the surfaces of a hydrogel intraocular lens. Ophthalmology. 2000; 107: 2179–2185.
Werner L, Apple DJ, Kaskaloglu M, Pandey SK. Dense opacification of the optical component of a hydrophilic acrylic intraocular lens: a clinicopathological analysis of 9 explanted lenses. J Cataract Refract Surg. 2001; 27: 1485–1492.
Werner L, Hunter B, Stevens S, Chew JJL, Mamalis N. Role of silicon contamination on calcification of hydrophilic acrylic intraocular lenses. Am J Ophthalmol. 2006; 141: 35–43.
Foot L, Werner L, Gills JP, et al. Surface calcification of silicone plate intraocular lenses in patients with asteroid hyalosis. Am J Ophthalmol. 2004; 137: 979–987.
Werner L, Kollarits CR, Mamalis N, Olson RJ. Surface calcification of a three-piece silicone intraocular lens in a patient with asteroid hyalosis: a clinicopathologic case report. Ophthalmology. 2005; 112: 447–452.
Stringham J, Werner L, Monson B, Theodosis R, Mamalis N. Calcification of different designs of silicone intraocular lenses in eyes with asteroid hyalosis. Ophthalmology. 2010; 117: 1486–1492.
Apple DJ, Peng Q, Arthur SN, et al. Snowflake degeneration of polymethyl methacrylate posterior chamber intraocular lens optic material: a newly described clinical condition caused by unexpected late opacification of polymethyl methacrylate. Ophthalmology. 2002; 109: 1666–1675.
Dahle N, Werner L, Fry L, Mamalis N. Localized central optic snowflake degeneration of a PMMA intraocular lens: clinical report with pathological correlation. Arch Ophthalmol. 2006; 124: 1350–1353.
Blundell MSJ, Mayer EJ, Knox Cartwright NE, Hunt LP, Tole DM, Dick AD. The effect on visual function of Hydroview intraocular lens opacification: a cross-sectional study. Eye. 2010; 24: 1590–1598.
Dagres E, Khan MA, Kyle GM, Clark D. Perioperative complications of intraocular lens exchange in patients with opacified Aqua-Sense lenses. J Cataract Refract Surg. 2004; 30: 2569–2573.
Aslam TM, Haider D, Murray IJ. Principles of disability glare measurement: an ophthalmological perspective. Acta Ophthalmol Scand. 2007; 85: 354–360.
Michelson J, Werner L, Ollerton A, Leishman L, Bodnar Z. Light scattering and light transmittance in intraocular lenses explanted because of optic opacification. J Cataract Refract Surg. 2012; 38: 1476–1485.
Barra D, Werner L, Costa JLP, et al. Light scattering and light transmittance in a series of calcified single-piece hydrophilic acrylic intraocular lenses of the same design. J Cataract Refract Surg. 2014; 40: 121–128.
Das K, Stover J, Schwiegerling J, Karakelle M. Technique for measuring forward light scatter in intraocular lenses. J Cataract Refract Surg. 2013; 39: 770–778.
Choi J, Schwiegerling J. Optical performance measurement and night driving simulation of ReSTOR, ReZoom, and Tecnis multifocal intraocular lenses in a model eye. J Refract Surg. 2008; 24: 218–222.
van den Berg TJ. Analysis of intraocular straylight especially in relation to age. Optom Vis Sci. 1995; 72: 52–59.
van den Berg TJ, Ijspeert JK. Light scattering in donor lenses. Vision Res. 1995; 35: 169–177.
van den Berg TJ. Depth-dependent forward light scattering by donor lenses. Invest Ophthalmol Vis Sci. 1996; 37: 1157–1166.
van den Berg TJ. Light scattering by donor lenses as a function of depth and wavelength. Invest Ophthalmol Vis Sci. 1997; 38: 1321–1332.
van den Berg TJ, Spekreijse H. Light scattering model for donor lenses as a function of depth. Vision Res. 1999; 39: 1437–1445.
van Den Berg TJ, Van Rijn LJ, Michael R, et al. Straylight effects with aging and lens extraction. Am J Ophthalmol. 2007; 144: 358–363.
de Vries NE, Franssen L, Webers CA, et al. Intraocular straylight after implantation of the multifocal AcrySof ReSTOR SA60D3 diffractive intraocular lens. J Cataract Refract Surg. 2008; 34: 957–962.
van der Mooren M, van den Berg TJ, Coppens J, Piers P. Combining in vitro test methods for measuring light scatter in intraocular lenses. Biomed Opt Express. 2011; 2: 505–510.
Werner L, Stover J, Schwiegerling J, Das KK. Light scattering, stray light and optical quality in hydrophobic acrylic intraocular lenses with subsurface nanoglistenings. J Cataract Refract Surg. 2016; 42: 148–156.
van der Meulen IJ, Porooshani H, van den Berg TJ. Light-scattering characteristics of explanted opacified Aquasense intraocular lenses. Br J Ophthalmol. 2009; 93: 830–832.
Söderberg PG, Chen EP, Lindström B. An objective and rapid method for the determination of light dissemination in the lens. Acta Ophthalmol (Copenh). 1990; 68: 44–52.
Figure 1
 
Gross (macroscopic) photographs of a representative lens from each of the three groups. (A) Calcified single-piece hydrophilic acrylic lens. Diffuse calcification was found predominantly within the substance of the lens optic and haptics. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. The calcification was localized to the center of the posterior optic surface. (C) Three-piece PMMA lens with snowflake degeneration. Snowflake lesions were present within the optic of the lens, sparing the optic periphery.
Figure 1
 
Gross (macroscopic) photographs of a representative lens from each of the three groups. (A) Calcified single-piece hydrophilic acrylic lens. Diffuse calcification was found predominantly within the substance of the lens optic and haptics. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. The calcification was localized to the center of the posterior optic surface. (C) Three-piece PMMA lens with snowflake degeneration. Snowflake lesions were present within the optic of the lens, sparing the optic periphery.
Figure 2
 
Graphs comparing forward light scattering from four representative explanted hydrophilic acrylic lenses exhibiting calcium deposits predominantly within the optic substance, and two control lenses of the same design. PSF, point spread function.
Figure 2
 
Graphs comparing forward light scattering from four representative explanted hydrophilic acrylic lenses exhibiting calcium deposits predominantly within the optic substance, and two control lenses of the same design. PSF, point spread function.
Figure 3
 
Graphs comparing forward light scattering from one explanted calcified silicone lens, and a control lens of the same design.
Figure 3
 
Graphs comparing forward light scattering from one explanted calcified silicone lens, and a control lens of the same design.
Figure 4
 
Graphs comparing forward light scattering from four PMMA lenses explanted because of snowflake degeneration, and a control PMMA lens.
Figure 4
 
Graphs comparing forward light scattering from four PMMA lenses explanted because of snowflake degeneration, and a control PMMA lens.
Figure 5
 
Modulation transfer function of a hydrophilic acrylic lens explanted because of calcification, as well as a control lens of the same design, measured at 3.0- and 5.0-mm pupils.
Figure 5
 
Modulation transfer function of a hydrophilic acrylic lens explanted because of calcification, as well as a control lens of the same design, measured at 3.0- and 5.0-mm pupils.
Figure 6
 
Badal images of a PMMA lens explanted because of snowflake degeneration (A, C), as well as a control PMMA lens (B, D), measured at 3.0- (A, B) and 5.0- (C, D) mm pupils.
Figure 6
 
Badal images of a PMMA lens explanted because of snowflake degeneration (A, C), as well as a control PMMA lens (B, D), measured at 3.0- (A, B) and 5.0- (C, D) mm pupils.
Figure 7
 
Scheimplug photographs with densitometry analyses of a representative lens from each of the three groups (same lenses as in Fig. 1). (A) Calcified single-piece hydrophilic acrylic lens; backward light scattering was measured as 222 CCT. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. Backward scattering = 223 CCT. (C) Three-piece PMMA lens with snowflake degeneration. Backward scattering = 225 CCT (image obtained with a flash level of 50 W).
Figure 7
 
Scheimplug photographs with densitometry analyses of a representative lens from each of the three groups (same lenses as in Fig. 1). (A) Calcified single-piece hydrophilic acrylic lens; backward light scattering was measured as 222 CCT. (B) Calcified single-piece plate silicone lens, removed from an eye with asteroid hyalosis. Backward scattering = 223 CCT. (C) Three-piece PMMA lens with snowflake degeneration. Backward scattering = 225 CCT (image obtained with a flash level of 50 W).
Table 1
 
Explanted and Control Intraocular Lenses Included in the Study
Table 1
 
Explanted and Control Intraocular Lenses Included in the Study
Table 2
 
Stray Light Values (Log (s)) for Explanted Hydrophilic Acrylic, Silicone, and PMMA Lenses Calculated From Forward Light Scattering Measurements Obtained at Various Angles.
Table 2
 
Stray Light Values (Log (s)) for Explanted Hydrophilic Acrylic, Silicone, and PMMA Lenses Calculated From Forward Light Scattering Measurements Obtained at Various Angles.
×
×