Bleaching difference reflectance spectroscopy. (
A–
C) Data of an albino Balb/c mouse; (
D–
F) data of a pigmented C57Bl/6J mouse. (
A,
D) Reflectance difference spectra (
Equation 2) collected during serial scanning of small retinal regions as in
Figure 2, but with two differences: The light source entering the pupil was a broadband spectrum (
Fig. 1D); the spectrum of the light reflected from the scanned area was captured and analyzed by spectrometer 1 (
Fig. 1B). The quantity plotted is the power of the light reflected from the eye with the power of the initial scan subtracted at each wavelength, and normalized overall by the saturated maximum at 500 nm that is reached after 50 to 60 scans. (For clarity, only a sample of the difference spectra is shown: For Figure 4A, the plotted spectra are from scans 3, 5, 7, 9, 11, 13, 17, 23, and 28 and the average of scans 41 to 50; for the intrinsically noisier data of the C57Bl/6 mouse of Figure 4D, the plotted spectra are averages of scans 2 and 3; 4 and 5; 6 and 7; 8 to 10; 11 to 15; 16 to 20; 21 to 30; 31 to 40; and 41 to 50. Note that the same ordinate scale applies to all parts of the figure. (
B,
E) The normalized difference values in the two spectral bands (mid-wavelength, 500 to 510 nm,
green symbols; long-wavelength, 610 to 630 nm,
red symbols; see
gray bars in [
A,
D]) are plotted as a function of the scan number. The long-wavelength data (
red symbols) were fitted with smooth curves to extract the trends over the scan series. The mid-wavelength data were adjusted as expressed in
Equation 10, and the resultant data fitted with the serial bleaching model (
smooth trace through green symbols;
Equation 7; cf.
Fig. 2). (
C,
F) Replotting of the mid-wavelength data in (
B,
E), respectively, as a function of the cumulative energy density to illustrate the determination of photosensitivity. The energy density in these plots is presented in rhodopsin-equivalent photons μm
−2, calculated by integrating the measured power density spectrum (
Fig. 1D) converted to photon units against the normalized mouse rhodopsin absorption spectrum template (
Fig. 1C). The adjustment of the mid-wavelength reflectance data with the long-wavelength trend had negligible effect on the quality of fit and parameters extracted from the albino mouse's data, but improved the quality of the fit and caused the value of log
10(
Qe) estimated to be 0.12 log
10 units lower than when the fitting of the pigmented mouse's data was performed without the adjustment.