August 2016
Volume 57, Issue 10
Open Access
Cornea  |   August 2016
Intraocular Pressure Changes and Relationship With Corneal Biomechanics After SMILE and FS-LASIK
Author Affiliations & Notes
  • Hua Li
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Yan Wang
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Rui Dou
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Pinghui Wei
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Jiamei Zhang
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Wei Zhao
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Liuyang Li
    Tianjin Eye Hospital & Eye Institute Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Medical University, Tianjin, China
  • Correspondence: Yan Wang, Tianjin Eye Hospital & Eye Institute, No. 4 Gansu Road, Heping District, Tianjin 300020, China; wangyan7143@vip.sina.com
Investigative Ophthalmology & Visual Science August 2016, Vol.57, 4180-4186. doi:https://doi.org/10.1167/iovs.16-19615
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Hua Li, Yan Wang, Rui Dou, Pinghui Wei, Jiamei Zhang, Wei Zhao, Liuyang Li; Intraocular Pressure Changes and Relationship With Corneal Biomechanics After SMILE and FS-LASIK. Invest. Ophthalmol. Vis. Sci. 2016;57(10):4180-4186. https://doi.org/10.1167/iovs.16-19615.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: The purpose of this article was to evaluate intraocular pressure (IOP) changes and investigate the relationship with corneal biomechanics after small-incision lenticule extraction (SMILE) and femtosecond laser-assisted laser in situ keratomileusis (FS-LASIK).

Methods: A total of 193 eyes of 193 patients who underwent SMILE and FS-LASIK procedures were included in this retrospective study. Data were collected preoperatively and postoperatively, including Goldmann-correlated IOP (IOPg), corneal-compensated IOP (IOPcc), corneal hysteresis (CH), and corneal resistance factor (CRF) by ocular response analyzer, noncontact intraocular pressure (IOPNCT) by noncontact tonometer, and Ehlers, Shah, Dresden, Kohlhaas, Orssengo/Pye by the Pentacam corrected system. Changes in both groups and differences between groups were evaluated. Multiple linear regression models were constructed to explore factors influencing IOP changes.

Results: In SMILE, the IOPg, IOPcc, IOPNCT, and Kohlhaas decreased significantly at 1 month postoperatively (P < 0.01), whereas with the Ehlers formula they significantly increased (P < 0.01). IOPs decreased at 3 and 6 months compared with all preoperative values except Ehlers values (P < 0.01), but there was no significant difference between 3 and 6 months (P > 0.05). In FS-LASIK, the IOPg, IOPcc, and IOPNCT decreased significantly at 1 month (P < 0.01), whereas in the Ehlers and Shah formulas they significantly increased (P < 0.01). Compared with preoperative values, the IOPs decreased at 3 and 6 months except in the Ehlers and Shah formulas (P < 0.01). Only IOPg and IOPcc differed between 3 and 6 months (P < 0.05). The Ehlers and Shah formulas were closer to the preoperative IOP for both groups, with variation approximately 1 mm Hg at 6 months postoperatively. Preoperative IOP, postoperative corneal resistance factor, corneal hysteresis, and flat keratometry were enrolled into the regression equations.

Conclusions: IOP underestimation after SMILE or FS-LASIK was related to corneal biomechanics as well as preoperative IOP and flat keratometry. IOP after SMILE seem to remain more stable. Accordingly, the Ehlers and Shah formulas were closer to the preoperative IOP. It may be useful to estimate future IOP with the best-fit models after surgery.

Corneal refractive surgery, such as femtosecond laser-assisted laser in situ keratomileusis (FS-LASIK) and small-incision lenticule extraction (SMILE), has emerged as having good efficacy, predictability, safety, and stability for surgical correction of low, moderate, and high myopia.13 However, the central corneal thickness (CCT), the corneal curvature, and corneal biomechanics change after corneal refractive surgery, which may affect intraocular pressure (IOP) measurements.4 
Meanwhile, myopia is a risk factor for open-angle glaucoma,5 and the routine application of steroids postoperatively will increase the risk of elevated IOP after corneal refractive surgery, even causing steroid-induced glaucoma.6 In addition, elevated IOP is recognized as the most important risk factor for progressive glaucomatous damage.7 Therefore, the accurate evaluation of IOP is a great concern for clinicians. 
Previous studies have showed IOP changes after some refractive surgeries, including photorefractive keratectomy,8,9 laser-assisted subepithelial keratectomy,10,11 laser in situ keratomileusis (LASIK),12 epipolis laser in situ keratomileusis,13 and FS-LASIK.14 Up to now, there has been no study regarding IOP change after SMILE, especially a comparison of IOPs between SMILE and FS-LASIK. We therefore decided to conduct a new retrospective study to evaluate IOP changes and the relationship with corneal biomechanics after both surgeries. 
Methods
Patients
The institutional review board at Tianjin Eye Hospital, Tianjin Medical University, approved this study protocol, which adhered to the tenets of the Declaration of Helsinki. All patients provided informed consent for surgery. 
This was a retrospective study performed at Tianjin Eye Hospital, and patients with myopia or myopic astigmatism who underwent SMILE or FS-LASIK were enrolled. Inclusion criteria were (1) patients older than 18 years; (2) absence of corneal, ocular, or systemic diseases; (3) refractive diopter maintained stable the past 2 years (change of ± 0.50 diopter); (4) patients who wore soft contact lens were off contact lenses for at least 2 weeks and off rigid contact lenses for at least 4 weeks; (5) preoperative noncontact intraocular pressure (IOPNCT) less than 21 mm Hg; (6) no family history of glaucoma; (7) postoperative IOPNCT increase less than 7 mm Hg11,15; (8) and no use of IOP-lowering medications before or after refractive surgery. From a total of 193 eyes, 97 eyes of 97 patients underwent SMILE and 96 eyes of 96 patients underwent FS-LASIK. 
Measures
Patients underwent preoperative measurements of uncorrected distance visual acuity; corrected distance visual acuity; manifest and cycloplegic refractions; IOPNCT, Goldmann-correlated IOP (IOPg), and corneal-compensated IOP (IOPcc) measured by an Ocular Response Analyzer (ORA), corneal topography, and Pentacam-corrected IOPs by the Scheimpflug tomography system (Pentacam; Oculus GmbH, Wetzlar, Germany). Patients were followed up at 1, 3, and 6 months after surgery. 
Ocular Response Analyzer
The ORA (Reichert, Inc., Depew, NY, USA) was used to measure corneal biomechanics and IOPs. Measured parameters are produced by the dynamic bidirectional applanator based on the two pressure measurements at applanation, P1 in the inward direction and P2 in the outward direction. The four parameters included (1) corneal hysteresis (CH), with CH = a[P1-P2]16 representing the ability of the cornea to absorb or dampen the applied force; (2) corneal resistance factor (CRF), with CRF = a[P1-0.7P2] + d described as the whole corneal resistance cumulative effect; (3) IOPg, with IOPg = a[(P1-P2)/2] + c, which were highly correlated with IOPg17; (4) IOPcc, with IOPcc = b[P2-0.43P1] + e, which compensated for corneal properties in estimating IOP, producing a more accurate value in softer eyes and patients after refractive surgery,18 where a, b, c, d, and e are calibration and regression constants. For each studied eye, three records were reserved with the signal score higher than 3.5,19 and the average values were calculated. 
Noncontact Tonometer
The noncontact tonometer (TX-F; Canon, Tokyo, Japan) was used to measure IOP avoiding contact and thus reducing the risk of infection. Some research suggested that IOPNCT had good repeatability and precision.20 This tonometer records three readings and then calculates the average value automatically. 
Pentacam Scheimpflug Images
The rotating Scheimpflug camera identified 25,000 true elevation points to measure corneal thickness and curvature. It also had five IOP correction formulas to correct IOP. This applied for measurements with the applanation tonometer as well as the NCT. When IOPNCT data were input into Pentacam, Ehlers, Shah, Dresden, Kohlhaas, and Orssengo/Pye formulas, corrected IOPs were generated. The following descriptions explains the calculation of IOP change for the correction formula: (1) Ehlers: IOP change = 0.071 × (545 − corneal thicknessmeasured)21; (2) Shah: IOP change = 0.050 × (550 − corneal thicknessmeasured)22; (3) Dresden: IOP change = 0.040 × (550 − corneal thicknessmeasured)23; 4) Kohlhaas: IOP change = (540 − corneal thicknessmeasured)/71 + (43 − K)/2.7 + 0.7524; (5) Orssengo/Pye: IOP change = IOPmeasured/K.25 These formulas can correct the effects of corneal thickness and corneal curvature on IOP. 
Surgical Procedures
All surgical procedures were performed by the same surgeon (Y.W.). After routine irrigation of the conjunctival sac and periocular sterilization, topical anesthesia was applied with two to three drops of oxybuprocaine hydrochloride (benoxyl; Santen, Inc., Osaka, Japan) twice before surgery, 5 minutes apart. 
SMILE Procedure
SMILE was performed using only the femtosecond laser platform (Visumax; Carl Zeiss Meditec AG, Jena, Germany), with typical pulse energy of approximately 140 to 150 nJ and a pulse repetition rate of 500 kHz. Four sequential cleavages were created to make an intrastromal lenticule and a small incision. The laser scanning order was, first, the posterior surface of the refractive lenticule, then the lenticule border, then the anterior surface of the lenticule and, finally, the side-cutting of the 3-mm width incision at the 12:00 location of the cornea. The lenticule diameter was 6.5 mm, and the side-cut angle was 90°. The corneal cap thickness was 110 μm, and its diameter was 7.5 mm. A basement of 10 to 15 μm above the lenticule was added to remove the lenticule successfully. 
FS-LASIK Procedure
In the FS-LASIK group, flap creation was performed using the same femtosecond laser with pulse energy at 165 to 175 nJ. Flap diameter was 8.0 mm, and the thickness was 100 to 110 μm. The flap hinge was positioned nasally. After lifting the flap, ablation of the stromal bed was performed by an excimer laser system (Allegretto; WaveLight Laser Technologie AG, Erlangen, Germany). Finally, the corneal flap was repositioned. 
Patients of both groups were medicated with topical 0.5% levofloxacin (Cravit; Santen, Inc.) eye drops four times daily for 2 days, and 0.1% fluorometholone (FML; Flumetholon; Santen, Inc.) was applied four times daily for 2 weeks and then tapered down to once every 2 weeks. 
Statistical Analysis
Statistical analysis was performed using SPSS (version 20.0; IBM Corp., Chicago, IL, USA). In this study, changes were expressed by the Δ symbol and calculated as preoperative value minus 6-month postoperative value; the thickness of the tissue removed was expressed as lenticule thickness (LT) in the SMILE group and as ablation depth (AD) in the FS-LASIK group. Furthermore, because SMILE uses a unique corneal cutting algorithm, the LT was based on the algebraic sum of spherical power and astigmatic power. In the FS-LASIK group, the spherical equivalent was calculated. 
The normality of all data samples were checked with the Kolmogorov–Smirnov test. The means of all of the IOPs at different examination points within a group were compared by repeated-measures analysis of variance, and the further pairwise comparison used least-significant difference. The Greenhouse Geisser correction was applied to degrees of freedom if sphericity was violated. The comparison of IOP changes between two groups were performed using independent t-tests. The Pearson correlation coefficient (r) was used to evaluate the correlations between variables. Multiple linear regressions were used to explore factors influencing IOP changes in both groups. P values less than 0.05 were considered statistically significant. 
Results
Table 1 shows the preoperative baseline characteristics of patients by group. In the SMILE and FS-LASIK groups, 97 and 96 eyes completed postoperative examinations at 1 and 3 months, respectively; 44 and 38 of these eyes completed the 6-month follow-up. There were no significant differences between preoperative baseline characteristics. 
Table 1
 
Baseline Characteristics of the Eyes by Group
Table 1
 
Baseline Characteristics of the Eyes by Group
Change in the IOP Measurement Values and Corneal Parameters Over Time
Table 2 shows IOPs, CH, CRF, CCT, flat keratometry (Kf), steep keratometry (Ks), and mean keratometry (Km) changes after SMILE and FS-LASIK. In SMILE, the IOPg, IOPcc, IOPNCT, and Kohlhaas decreased significantly at 1 month postoperatively (P < 0.01), whereas in the Ehlers formula, they significantly increased (P < 0.01). All of the IOPs showed severe decreases at the 3- and 6-month follow-ups except for those in Ehlers (P < 0.01), but there were no significant differences between 3 and 6 months (P > 0.05). CH, CRF, CCT, Kf, Ks, and Km decreased significantly at 1, 3, and 6 months when compared with preoperative values (P < 0.01). CH and CRF showed no significant differences between 1, 3, and 6 months (P > 0.05). 
Table 2
 
The IOP Measurement Values and Relevant Parameter Variations Pre- and Post-SMILE and FS-LASIK
Table 2
 
The IOP Measurement Values and Relevant Parameter Variations Pre- and Post-SMILE and FS-LASIK
In FS-LASIK, the IOPg, IOPcc, and IOPNCT decreased significantly at 1 month (P < 0.01), whereas in the Ehlers and Shah formulas they significantly increased (P < 0.01). All of the IOPs decreased sharply at the 3- and 6-month follow-ups except in the Ehlers and Shah formulas (P < 0.01) when compared with preoperative values. Only IOPg and IOPcc differed between 3 and 6 months (P < 0.05). CH, CRF, CCT, Kf, Ks, and Km decreased significantly at 1, 3, and 6 months when compared with preoperative values (P < 0.01). CH and CRF showed no significant differences between 1 and 3 months (P > 0.05; Table 2). 
Between-Group Comparisons
Table 3 shows IOP variation per removed or ablated tissue (ΔIOP/LT or AD) between two groups, only ΔIOPg/AD and ΔIOPcc/AD in FS-LASIK were higher than ΔIOPg/LT and ΔIOPcc/LT in SMILE at the 6-month follow-up (P < 0.05). The Figure shows the ΔIOP at 1, 3, and 6 months postoperatively, and only ΔIOPg in FS-LASIK was higher than SMILE at the 6-month follow-up (P < 0.05). Similarly, ΔCRF, ΔCH, ΔKf, ΔKs, and ΔKm per removed or ablated tissue in FS-LASIK were higher than in SMILE (0.05 ± 0.02 vs. 0.04 ± 0.01, 0.03 ± 0.02 vs. 0.02 ± 0.01, 0.05 ± 0.02 vs. 0.04 ± 0.01, 0.06 ± 0.02 vs. 0.04 ± 0.01, 0.05 ± 0.02 vs. 0.04 ± 0.01; P = 0.001, 0.009, 0.001, <0.001, <0.001; respectively). 
Table 3
 
The ΔIOP per Removed or Ablated Tissue at 6 Months Between SMILE and FS-LASIK
Table 3
 
The ΔIOP per Removed or Ablated Tissue at 6 Months Between SMILE and FS-LASIK
Figure
 
The comparisons of ΔIOPg (A), ΔIOPcc (B), ΔEhlers (C), and ΔShah (D) between the SMILE and FS-LASIK groups and their changes over time (pre- and postoperative 1 month [1M], pre- and postoperative 3 months [3M], and pre- and postoperative 6 months [6M]). The standard error (SE) is shown by the error bar (*P < 0.05).
Figure
 
The comparisons of ΔIOPg (A), ΔIOPcc (B), ΔEhlers (C), and ΔShah (D) between the SMILE and FS-LASIK groups and their changes over time (pre- and postoperative 1 month [1M], pre- and postoperative 3 months [3M], and pre- and postoperative 6 months [6M]). The standard error (SE) is shown by the error bar (*P < 0.05).
Correlation Analysis
Table 4 shows the correlations between the ΔIOPg, ΔIOPcc, ΔIOPNCT, and the pre-IOP, ΔCCT, ΔCRF, ΔCH, ΔKm, respectively. The ΔIOP in the FS-LASIK group had more influencing factors and correlated with ΔCH and ΔCRF. In SMILE, the ΔIOPg, ΔIOPcc, and ΔIOPNCT correlated with the pre-IOP (r = 0.532, 0.478, 0.512, respectively, P < 0.01). Although in FS-LASIK, the ΔIOPg, ΔIOPcc, and ΔIOPNCT correlated with the pre-IOP (r = 0.780, 0.664, 0.614, respectively, P < 0.01). 
Table 4
 
The Correlation Analysis Between ΔIOPg, ΔIOPcc, ΔIOPNCT, and the Pre-IOP, ΔCCT, ΔCRF, ΔCH, ΔKm, Respectively
Table 4
 
The Correlation Analysis Between ΔIOPg, ΔIOPcc, ΔIOPNCT, and the Pre-IOP, ΔCCT, ΔCRF, ΔCH, ΔKm, Respectively
Regression Analysis of ΔIOPNCT and Influencing Factors
Table 2 shows that IOPNCT was stable at 3 months after SMILE and FS-LASIK, so patients' completed 3-month follow-ups were analyzed. Considering that IOPNCT was common in clinical examinations, a multiple linear regression analysis of ΔIOPNCT was created to analyze the relationships between possible preoperative influencing factors (pre-IOP, pre-CCT, Km, Kf, Ks, CRF, CH, and spherical power and astigmatic power/spherical equivalent) and intraoperative parameters (LT or AD) as well postoperative influencing factors (post-CCT, Km, Kf, Ks, CRF, CH) and variations after surgery (ΔCCT, ΔKm, ΔKf, ΔKs, ΔCRF, ΔCH). 
Multiple linear regression models were constructed to explore the factors influencing ΔIOPNCT in both groups. Preoperative IOP and postoperative CRF, CH, and Kf were enrolled into regression equations in both groups. Table 5 shows the definite regression coefficients between ΔIOPNCT and the influencing factors. It is possible to obtain the following two regression equations to evaluate ΔIOPNCT: after SMILE,  after FS-LASIK,  The actual postoperative IOP may be evaluated by the measured IOP plus the ΔIOP.  
Table 5
 
The Regression Coefficients of Influencing Factors Enrolled Into the Equation
Table 5
 
The Regression Coefficients of Influencing Factors Enrolled Into the Equation
Discussion
IOP measurement values decreased after both SMILE and FS-LASIK, which was consistent with previous research results.814,19 In this study, corneal biomechanical method ORA combined with NCT were used to investigate the relationship between corneal biomechanics and IOP change. Both tonometers estimated IOP by directing a pulse of air at the cornea, causing it to flatten and allowing an electro-optical collimation detector to evaluate corneal curvature. Both SMILE and FS-LASIK surgeries used a myopic ablation profile to remove tissue from the cornea to correct refractive error. After the surgery, CCT, corneal curvature, and corneal biomechanics decreased significantly,26 and the corneal structure also changed. These corneal changes altered postsurgical measurement of IOP.24 The corneal biomechanic parameters from ORA could present CH and CRF, with CH representing the viscoelastic property of the cornea and CRF the corneal elasticity with a strong positive correlation for CCT. The decrease of CH and CRF then indicates the decrease of cornea stiffness. When the IOP was measured, the cornea was more likely to be flattened, so these values were lower and underestimated the real IOP. Some studies suggested that there were no clinical symptoms of low IOP after surgery, based on animal experiments but also on clinical observation.27 This IOP reduction was an illusion and what really changed were the influencing factors of IOP measurement. 
In this study, IOP measurement values remained stable at 3 months after SMILE, with CH and CRF also restored at 1 month. Moreover, CCT remained stable at 3 months and corneal curvature kept changing. Although some studies showed that the corneal curvature and CCT were related to IOP reduction,15,28 Liu and Roberts29 reported that differences in corneal biomechanics between individuals may contribute more than corneal thickness or curvature to IOP measurement errors, which was consistent with this study. After surgery, patients needed routine application of 0.1% FML to prevent inflammation and refractive regression for 2 months. In a comparative study, 6 weeks of FML treatment resulted in an IOP increase greater than 5 mm Hg in 8.3% of cases, although topical FML was thought to have less effect on IOP than other steroid agents.30 All of these possible influencing factors, especially corneal biomechanics, regained stability, and after stopping FML medication, IOP measurement values were stable again. Regarding IOP changes in FS-LASIK, most IOP measurement values remained stable at 3 months except IOPg and IOPcc. Similarly, CCT, Kf, Ks, and Km also did not regain its stability at 3 months except in corneal biomechanics. When compared with SMILE, FS-LASIK induced more keratocyte apoptosis, proliferation, and inflammation,31 so IOP values after FS-LASIK seemed less stable, which may be a result of corneal structure variation. 
In both groups, the low IOP readings after surgery resulted in a delayed diagnosis of glaucoma or recognition of ocular hypertensive patients, so it was very important to measure postoperative IOP or the change of IOP in an accurate way. Interestingly, the decrease in postoperative 6-month IOP in the current study was significant when measured using NCT (approximately 5 mm Hg) and IOPg (approximately 6 mm Hg) when compared with IOPcc (approximately 3 mm Hg) and Pentacam (approximately 2 mm Hg). This finding is supported by the hypothesis that IOPcc is less influenced by corneal properties, such as CCT, and does not decrease significantly following LASIK and epipolis laser in situ keratomileusis.13 Furthermore, among all of the ΔIOP after SMILE, ΔShah was smallest (0.83 mm Hg), followed by ΔEhlers (−0.87 mm Hg). After FS-LASIK, ΔShah was the smallest (0.35 mm Hg) and ΔEhlers second (−1.43 mm Hg). The comparison between ΔEhlers and ΔShah showed no significant difference. Accordingly, the Ehlers and Shah formulas were more close to preoperative IOP for both surgeries, with IOP variation approximately 1 mm Hg. Previous studies have found that IOPcc and Pentacam-Ehlers were consistent before and after LASIK13 because only Pentacam–Ehlers was available in that version of the machine. The corrected IOP can eliminate the influence of CCT. In this study, the Ehlers significantly increased at 6 months postoperatively, which may be related to the corrected coefficients and formula. It needs further investigation. 
A comparison of ΔIOP per removed or ablated tissue showed that ΔIOPg and ΔIOPcc per removed tissue in SMILE was lower than with FS-LASIK. ΔIOPg had a moderate correlation with ΔCRF (Table 3). Reasons for this change may be the better preservation of corneal integrity and corneal biomechanics after SMILE.26 
A strong positive linear correlation was found between ΔIOP and preoperative IOP, which was consistent with Schallhorn et al.,32 being the strongest predictor of postoperative IOP. In the FS-LASIK group, ΔIOP had more influencing factors and correlated with CH and CRF. Regarding ΔCCT, the correlation coefficient was smaller than the corneal biomechanic parameter CRF. Similarly, Liu et al.29 reported that corneal biomechanics may contribute more than corneal thickness or curvature to IOP change, with controversy regarding how much ΔIOP can be explained by the ΔCCT.33 
From the available data, it was possible to construct models for IOP change after SMILE and FS-LASIK. These models cannot evaluate all of the variance after refractive surgery. The best-fit model for SMILE and FS-LASIK can explain 50.5% and 55.2% variance, respectively. In both models, preoperative IOP was still the main predictor of ΔIOP. Furthermore, postoperative CH, CRF, and Kf also enrolled in the equation, which was a new finding and a reminder of the important role of corneal biomechanics in IOP changes. In addition, NCT was a simple and safe device for routine IOP measurements. Previous data have showed that NCT can produce accurate IOP that is comparable to a Goldmann tonometer,34,35 which was the gold standard, allowing these models to be useful in evaluating future IOP changes. The calculation of ΔIOP needs preoperative IOP and postoperative corneal biomechanics, which may need more comprehensive examinations. 
Previous studies have constructed statistical models to predict postoperative IOP after LASIK. Cheng et al.4 used NCT to measure IOP after LASIK in 123 eyes and gave a model involving postoperative CCT, postoperative corneal curvature, and AD. That study did not consider the corneal biomechanics, which were the main influencing factors. Schallhorn et al.32 also used NCT to measure IOP after LASIK in a large sample investigation, and the best-fit model, including manifest refraction spherical equivalent and preoperative IOP, can explain 45% variance. In that study, although the large sample was important, it only considered CCT, corneal curvature, and manifest refraction spherical equivalent while ignoring corneal biomechanics. The models in this study may be valuable when evaluating IOP after SMILE if the influence of corneal biomechanics is considered. Although new influencing factors have been added to the model to evaluate the real IOP, this study needs more samples for further study. Other IOP measurement devices need be investigated in future, such as the Goldmann tonometer, which was a gold standard for IOP measurements of a normal cornea. Some studies have showed that the Goldmann tonometer was no more accurate and could lower the IOP after PRK (photorefractive keratectomy),15 LASEK,10 and LASEK (laser-assisted subepithelial keratectomy),36 because it can be affected by the corneal thickness. However, the Goldmann tonometer measurement still needs to be investigated in the further research. 
In summary, IOP measurement underestimation after SMILE and FS-LASIK was related to corneal biomechanics as well as to preoperative IOP and Kf. IOP after SMILE seemed to remain more stable than FS-LASIK. Accordingly, the Ehlers and Shah formulas were closer to the preoperative values, with IOP variation approximately 1 mm Hg at 6 months postoperatively. The best-fit models for IOP change may be useful to evaluate the long-term postoperative IOP after SMILE and FS-LASIK. Further investigations on accurate IOP measurements are required. 
Acknowledgments
Supported by the National Natural Science Foundation of China (Grant 81470658) and Tianjin Research Program of Application Foundation and Advanced Technology (14JCZDJC35900). The authors alone are responsible for the content and writing of the paper. 
Disclosure: H. Li, None; Y. Wang, None; R. Dou, None; P. Wei, None; J. Zhang, None; W. Zhao, None; L. Li, None 
References
Lee JK, Chuck RS, Park CY Femtosecond laser refractive surgery: small-incision lenticule extraction vs. femtosecond laser-assisted LASIK. Curr Opin Ophthalmol. 2015 ; 26: 260–264.
Vestergaard A, Ivarsen AR, Asp S, et al. Small-incision lenticule extraction for moderate to high myopia: predictability, safety, and patient satisfaction. J Cataract Refract Surg. 2012 ; 38: 2003–2010.
Reinstein DZ, Carp GI, Archer TJ, et al. Outcomes of small incision lenticule extraction (SMILE) in low myopia. J Refract Surg. 2014 ; 30: 812–818.
Cheng AC, Fan D, Tang E, et al. Effect of corneal curvature and corneal thickness on the assessment of intraocular pressure using noncontact tonometry in patients after myopic LASIK surgery. Cornea. 2006 ; 25: 26–28.
Marcus MW, de Vries MM, Junoy Montolio FG, et al. Myopia as a risk factor for open-angle glaucoma: a systematic review and meta-analysis. Ophthalmology. 2011 ; 118: 1989–1994.
Tomi T, Nicula C, Nicula D. Postoperative complications of refractive surgery Lasik technique. Oftalmologia. 2003 ; 58: 60–63.
The Advanced Glaucoma Intervention Study. The relationship between control of intraocular pressure and visual field deterioration. Am J Ophthalmol. 2000 ; 130: 429–440.
Chatterjee A, Shah S, Bessant DA, et al. Reduction in intraocular pressure after excimer laser photorefractive keratectomy correlation with pretreatment myopia. Ophthalmology. 1997 ; 104: 355–359.
Sadigh AL, Fouladi RF, Hashemi H, et al. A comparison between Goldmann applanation tonometry and dynamic contour tonometry after photorefractive keratectomy. Graefes Arch Clin Exp Ophthalmol. 2013 ; 251: 603–608.
Johannesson G, Hallberg P, Eklund A, et al. Change in intraocular pressure measurement 2 years after myopic laser-assisted subepithelial keratectomy. J Cataract Refract Surg. 2012 ; 38: 1637–1642.
Han KE, Kim H, Kim NR, et al. Comparison of intraocular pressures after myopic laser-assisted subepithelial keratectomy: Tonometry-pachymetry, Goldmann applanation tonometry, dynamic contour tonometry, and noncontact tonometry. J Cataract Refract Surg. 2013 ; 39: 888–897.
Kaufmann C, Bachmann LM, Thiel MA. Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci. 2003 ; 44: 3790–3794.
Shousha SM, Abo Steit MA, Hosny MH, et al. Comparison of different intraocular pressure measurement techniques in normal eyes, post surface and post lamellar refractive surgery. Clin Ophthalmol. 2013 ; 7: 71–79.
Shemesh G, Soiberman U, Kurtz S. . Intraocular pressure measurements with Goldmann applanation tonometry and dynamic contour tonometry in eyes after IntraLASIK or LASEK. Clin Ophthalmol. 2012; 6: 1967–1970.
Garzozi HJ, Chung HS, Lang Y, et al. Intraocular pressure and photorefractive keratectomy: a comparison of three different tonometers. Cornea. 2001 ; 20: 33–36.
Roberts CJ. Concepts and misconceptions in corneal biomechanics. J Refract Surg. 2014 ; 40: 862–869.
Lam A, Chen D, Chiu R, et al. Comparison of IOP measurements between ORA and GAT in normal Chinese. Optom Vis Sci. 2007 ; 84: 909–914.
Pepose JS, Feigenbaum SK, Qazi MA, et al. Changes in corneal biomechancis and intraocular pressure following LASIK using static, dynamic and noncontact tonometry. Am J Ophthalmol. 2007 ; 143: 39–47.
Dou R, Wang Y, Xu L, et al. Comparison of corneal biomechanical characteristics after surface ablation refractive surgery and novel lamellar refractive surgery. Cornea. 2015 ; 34: 1441–1446.
Yilmaz I, Altan C, Aygit ED, et al. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL. Clin Ophthalmol. 2014 ; 8: 1069–1074.
Ehlers N, Bramsen T, Sperling S. Applanation tomometry and central corneal thickness. Acta Ophthalmol. 1975 ; 53: 34–43.
Shah S, Chatterjee A, Mathai M, et al. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology. 1999 ; 106: 2154–2160.
Kohlhaas M, Boehm AG, Spoerl E, et al. Effect of central corneal thickness, corneal curvature, and axial length on applanation tomometry. Arch Ophthalmol. 2006 ; 124: 471–476.
Kohlhaas M, Spoerl E, Boehm AG, et al. A correction formula for the real intraocular pressure after LASIK for the correction of myopic astigmatism. J Refract Surg. 2006 ; 22: 263–267.
Orssengo GJ, Pye DC. Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol. 1999 ; 61: 551–572.
Wu D, Wang Y, Zhang L, et al. Corneal biomechanical effects: small-incision lenticule extraction versus femtosecond laser-assisted laser in situ keratomileusis. J Cataract Refract Surg. 2014 ; 40: 954–962.
Siganos DS, Papastergion GI, Moedas C. Assessment of the pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg. 2004 ; 30: 746–751.
Cacho I, Sanchez-Naves J, Batres L, et al. Comparison of Intraocular pressure before and after laser in situ keratomileusis refractive surgery measured with Perkins tonometry noncontact tonometry, and transpalpebral tonometry. J Ophthalmol. 2015 ; 2015: 683895.
Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg. 2005 ; 31: 146–155.
Armaly MF Statistical attributes of the steroid hypertensive response in the clinically normal eye: the demonstration at three levels of response. Invest Ophthalmol. l965 ; 4: 187–197.
Dong Z, Zhou X, Wu J, et al. Small incision lenticule extraction (SMILE) and femtosecond laser LASIK: comparison of corneal wound healing and inflammation. Br J Ophthalmol. 2014 ; 98: 263–269.
Schallhorn JM, Schallhorn SC, Ou Y. Factors that influence intraocular pressure changes after myopic and hyperopic LASIK and photorefractive keratectomy: a large population study. Ophthalmology. 2015 ; 122: 471–479.
Kotecha A. What biomechanical properties of the cornea are relevant for the clinician? Surv Ophthalmol. 2007 ; 52 (suppl 2): S109–S114.
Jorge J, Diaz-Rey JA, Gonzalez-Meijome JM, et al. Clinical performance of the Reichert AT 550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002 ; 22: 560–564.
Cho P, Lui T. Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optom Vis Sci. 1997 ; 74: 51–58.
Kirwan C, O'Keefe M. Measurement of intraocular pressure in LASIK and LASEK patients using the Reichert Ocular Response Analyzer and Goldmann applanation tonometry. J Refract Surg. 2008 ; 24: 366–370.
Figure
 
The comparisons of ΔIOPg (A), ΔIOPcc (B), ΔEhlers (C), and ΔShah (D) between the SMILE and FS-LASIK groups and their changes over time (pre- and postoperative 1 month [1M], pre- and postoperative 3 months [3M], and pre- and postoperative 6 months [6M]). The standard error (SE) is shown by the error bar (*P < 0.05).
Figure
 
The comparisons of ΔIOPg (A), ΔIOPcc (B), ΔEhlers (C), and ΔShah (D) between the SMILE and FS-LASIK groups and their changes over time (pre- and postoperative 1 month [1M], pre- and postoperative 3 months [3M], and pre- and postoperative 6 months [6M]). The standard error (SE) is shown by the error bar (*P < 0.05).
Table 1
 
Baseline Characteristics of the Eyes by Group
Table 1
 
Baseline Characteristics of the Eyes by Group
Table 2
 
The IOP Measurement Values and Relevant Parameter Variations Pre- and Post-SMILE and FS-LASIK
Table 2
 
The IOP Measurement Values and Relevant Parameter Variations Pre- and Post-SMILE and FS-LASIK
Table 3
 
The ΔIOP per Removed or Ablated Tissue at 6 Months Between SMILE and FS-LASIK
Table 3
 
The ΔIOP per Removed or Ablated Tissue at 6 Months Between SMILE and FS-LASIK
Table 4
 
The Correlation Analysis Between ΔIOPg, ΔIOPcc, ΔIOPNCT, and the Pre-IOP, ΔCCT, ΔCRF, ΔCH, ΔKm, Respectively
Table 4
 
The Correlation Analysis Between ΔIOPg, ΔIOPcc, ΔIOPNCT, and the Pre-IOP, ΔCCT, ΔCRF, ΔCH, ΔKm, Respectively
Table 5
 
The Regression Coefficients of Influencing Factors Enrolled Into the Equation
Table 5
 
The Regression Coefficients of Influencing Factors Enrolled Into the Equation
×
×

This PDF is available to Subscribers Only

Sign in or purchase a subscription to access this content. ×

You must be signed into an individual account to use this feature.

×