ARVO Annual Meeting Abstract  |   September 2016
Assessing the optical quality of commercially available intraocular lenses by means of modulation transfer function and straylight
Author Affiliations & Notes
  • ELENI PAPADATOU
    University of Valencia, Valencia, Spain
  • Grzegorz Labuz
    Rotterdam Ophthalmic Institute, Rotterdam, Netherlands
  • Thomas J T P Van Den Berg
    Rotterdam Ophthalmic Institute, Rotterdam, Netherlands
  • José-Juan Esteve-Taboada
    University of Valencia, Valencia, Spain
  • David Madrid-Costa
    University of Valencia, Valencia, Spain
  • Norberto Lopez-Gil
    University of Murcia, Murcia, Spain
  • Robert Montés-Micó
    University of Valencia, Valencia, Spain
  • Footnotes
    Commercial Relationships   ELENI PAPADATOU, None; Grzegorz Labuz, None; Thomas Van Den Berg, Royal Netherlands Academy of Arts and Sciences (P); José-Juan Esteve-Taboada, None; David Madrid-Costa, None; Norberto Lopez-Gil, None; Robert Montés-Micó, None
  • Footnotes
    Support  This study was supported by the AGEYE project (608049), the Marie Curie Initial Training Network program (FP7-PEOPLE-2013-ITN), granted by the European Commission, Brussels, Belgium.
Investigative Ophthalmology & Visual Science September 2016, Vol.57, 3115. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      ELENI PAPADATOU, Grzegorz Labuz, Thomas J T P Van Den Berg, José-Juan Esteve-Taboada, David Madrid-Costa, Norberto Lopez-Gil, Robert Montés-Micó; Assessing the optical quality of commercially available intraocular lenses by means of modulation transfer function and straylight. Invest. Ophthalmol. Vis. Sci. 2016;57(12):3115.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose : In vitro evaluation of intraocular lenses (IOLs) is used for evaluation of optical quality. Diffractive IOLs, designed to concentrate light at different vergences, can generate visual artifacts due to diffractive effects. This work aimed at evaluating the optical performance of five commercially available diffractive multifocal IOLs compared to one monofocal. The through focus modulation transfer function (MTF) at 50 cycles per mm (c/mm) and the straylight (expressed by the straylight parameter, s) were used to assess optical quality.

Methods : The multifocal IOLs studied were the AT LISA 809M (Zeiss), the AT LISA 839MP (Zeiss), the FineVision Micro F (PhysIOL), the Technis Symphony (Abbott) and the Acrysof Restor +3 D (Alcon). The monofocal IOL was the Acrysof IQ (Alcon). All IOLs were fresh, free of glistenings or other defects. Through focus MTF from +1 D to -5 D vergence, at two pupil diameters (3mm and 4.5 mm), was obtained with the PMTF instrument (Lambda-X). Straylight of the IOLs was recorded using the C-Quant device (Oculus) with modification proposed by Labuz et al., (Biomed Opt Express, 2015) that allows for in vitro IOL straylight assessment.

Results : For the 0 D vergence Acrysof IQ yielded the highest MTF value (0.78 at 3mm) comparing to the multifocal IOLs (0.37 to 0.51 at 0 D vergence at 3 mm). The through focus MTF of the IOLs decreased for 4.5 mm pupil size. The highest decay was found for the Acrysof IQ (26% at 0 D vergence). According to their addition powers, multifocal IOLs showed one or two extra peaks located at different vergences (from -1.5 D to -4.5 D). The MTF values of those peaks decreased for 4.5 mm by 0.5% to 14%. All multifocal IOLs showed low s values ranging from 0.50±0.13 deg2/sr to 1.40±0.65 deg2/sr. The monofocal IOL had an s value of 0.63±0.17 deg2/sr.

Conclusions : The monofocal IOL showed the highest MTF value (at 0 D vergence), but the multifocal IOLs were found to be affected less by pupil size at all vergences. As expected, multifocal IOLs are better at some other vergence. In the present study, straylight from the IOLs was rather low. The multifocal IOLs had values similar to the monofocal IOL, thus diffractive effects per se of these IOLs are not expected to elevate postoperative straylight. Ex vivo study of explanted IOLs is needed.