Abstract
Purpose:
To evaluate the repeatability and reproducibility of retinal blood flow (RBF) measurements in humans by using new auto-alignment and measurement software in a commercially available Doppler optical coherence tomography (DOCT) system.
Methods:
The DOCT flowmeter assessed the intrasession repeatability and the intersession and interobserver reproducibility of the RBF measurements. For intrasession repeatability, the coefficients of variation (CVs) of five repeated RBF measurements were calculated at the retinal arteries and veins in 20 normal eyes of 20 healthy volunteers. For intersession reproducibility, two sets of three measurements obtained by one observer on 2 different days were compared. For interobserver reproducibility, two sets of three measurements obtained by two observers on the same day were compared. Intraclass correlation coefficients (ICCs) also were used to evaluate the repeatability and reproducibility. The relevance of the DOCT flowmeter and laser Doppler velocimetry (LDV) also was assessed.
Results:
Regarding intrasession repeatability, the ICC of the RBF exceeded 0.90 in arterioles and venules (ICC: 0.994 and 0.970, respectively). The CVs of the RBF in the arterioles and venules were 6.0% ± 3.4% and 8.8% ± 5.1%, respectively. The intersession and interobserver RBF values had high reproducibility in the arterioles (ICC: 0.980 and 0.993, respectively) and venules (ICC: 0.982 and 0.986, respectively). The RBF measured with the DOCT flowmeter was correlated strongly with LDV in the arterioles (r = 0.76; P < 0.001).
Conclusions:
The DOCT flowmeter had good reproducibility in the arterioles and venules and precisely measured the RBF as compared to the LDV in the arterioles.
Evaluation of the ocular circulation is highly important for gaining an understanding of the physiologic and pathologic features in a number of retinal diseases.
1 Although many devices are used to measure retinal ocular blood flow—that is, the blue field entoptic technique, microsphere method, laser Doppler flowmeter, laser Doppler velocimetry (LDV), oxygen measurements, retinal vessel analyzer, color Doppler imaging, and laser speckle technique
2—we have shown that LDV is a reliable, noninvasive, and clinically useful tool for evaluating the retinal circulation in humans, based on the absolute vessel diameter and blood velocity.
3,4 Considering these advantages, we have measured the retinal blood flow (RBF) in several retinal diseases such as diabetic retinopathy (DR) in type 1 diabetes mellitus (DM),
5 DR in type 2 DM,
6 branch retinal vein occlusion,
7 and age-related macular degeneration.
8 However, the RBF measurements obtained with the LDV method have a number of disadvantages, that is, the required RBF should be considered as the parabolic flow distribution to calculate the absolute values of blood velocity,
9 fine alignment of the laser beam, and maintenance of strict eye fixation for measurements. Therefore, it is difficult to measure the RBF at the bifurcation or the margin of the optic nerve head because of uncertainty regarding the parabolic flow distribution. LDV also is limited to use in cooperative patients.
A novel velocimetry technique using optical coherence tomography (OCT) technology, namely, Doppler OCT (DOCT), has been developed recently.
10–12 OCT can detect not only morphologic images but also a Doppler shift of reflected light, which provides information about flow and movement.
13,14 Although the technique facilitates measurements of the absolute RBF values in humans, with good repeatability,
10–12 those DOCT methods have some shortcomings. In such methods, it is possible to measure the retinal venous blood flow velocity but not the arterial velocity. Another research group
15 has reported the results with the swept-source–based OCT flowmeter in which both the arterial and venous total flow are assessed by using their prototype OCT. We have previously developed a DOCT instrument with novel software, referred to as a segmental-scanning method,
16 which enables simultaneous measurement of the RBF in the retinal arterioles and adjacent venules during one cardiac cycle. We have previously reported the accuracy of the measurements in in vitro glass capillaries. We then evaluated the reproducibility of the blood velocity measurements in the retinal arterioles and venules in anesthetized cats. Recently, a new DOCT instrument has been developed that can be used in a clinical setting, namely, the DOCT flowmeter (Yoshida A, et al.
IOVS 2016;57:ARVO E-Abstract 5922).
The goal of the current study was to determine the repeatability and reproducibility of RBF measurements with the DOCT flowmeter in healthy subjects and the relevance of the arteriolar data when the DOCT flowmeter was compared with the LDV.
A total of 20 healthy volunteers (7 men, 13 women; age range, 22–26 years) with no systemic or ophthalmic disease with mild refractive errors (1.00 and −3.75 diopters of spherical equivalent) participated in this study. This study adhered to the guidelines approved by the ethics committee of our institution and the tenets of the Declaration of Helsinki. Each subject provided written informed consent before enrollment after receiving a complete explanation of the study design and protocol.
All volunteers underwent an ophthalmologic examination, including review of the medical history, slit-lamp biomicroscopy, intraocular pressure (IOP) measurement, and funduscopic examination using a 90-diopter lens. The axial length of each eye was measured with an A-mode ultrasound system (IOLMaster 500; Carl Zeiss Meditec, Jena, Germany). The mean arterial blood pressure (MABP) and heart rate (HR) were measured with an electronic sphygmomanometer (EP-88Si; Colin, Tokyo, Japan). The IOP was monitored by applanation tonometry (Haag Streit, Bern, Switzerland). The ocular perfusion pressure was calculated as 2/3MABP − IOP. The pupil was dilated with one drop of 0.4% tropicamide (Mydrin M; Santen Pharmaceuticals, Osaka, Japan). The subjects were asked to abstain from drinking coffee and alcohol and smoking for 24 hours before the examination. Each subject rested for 10 minutes in a quiet room before the examination began. After the pupils were dilated, the subjects were required to rest in the sitting position for at least 5 minutes in a quiet dimly lit room in which the temperature was maintained at 25°C, after which their BP and HR were measured in the left arm. The RBF measurements were assessed after the BP, HR, and IOP measurements. The relatively large arteries (>80 μm and sufficiently far from the bifurcations) chosen for measurement had relatively straight segments one disc diameter away from the optic disc.
To assess intrasession repeatability, 20 eyes of 20 subjects were recruited. Measurement of the right or left eye was determined by a random number table. One experienced observer (TY, observer 1) performed the DOCT flowmeter measurements for each subject. The RBF measurements of the arterioles and venules in either the superior or inferior temporal sites were repeated five times by using the follow-up mode in the software. The observer was masked to the results of the retinal circulatory parameters. The coefficients of variation (CVs) for the retinal circulatory parameters, retinal vessel D, maximal and averaged retinal blood velocity during one cardiac cycle (Vmaximum and Vaverage), and RBF were determined.
The intersession reproducibility of the DOCT flowmeter measurements was evaluated. The same observer performed three measurements on 2 different days at each of the 20 arterioles and venules (days 1 and 2). Both measurements were performed in the morning under fasting conditions. The observer was masked to the results of the retinal circulatory parameters. The correlations for the different days of the D, Vaverage, Vmaximum, and RBF were determined.
The interobserver reproducibility of the DOCT flowmeter measurements was evaluated. Observer 1 and another inexperienced observer (TO, observer 2) measured the same sites of the same 20 eyes three times on the same day. A minimum of 10 minutes was allowed between measurements. The observers were masked to the results of the retinal circulatory parameters. The correlation between the two observers' measurements was determined and the differences in the D, Vaverage, Vmaximum, and RBF were determined.
In the last part of this study, the correlations between the DOCT flowmeter and LDV were evaluated in the arterioles in the same 20 subjects. After an experienced observer (Y-SS, observer 3) performed the LDV measurements, observer 1 performed the DOCT flowmeter measurements at the same vessel sites of the LDV on the same day as those measured by experienced observer 4 (TT). The observers were masked to the results of the retinal circulatory parameters. A minimum of 10 minutes was allowed between the LDV and DOCT flowmeter measurements. The correlation of the measurements between the two different methods was determined in the RBF.
Table 2 summarizes the results of intrasession repeatability of the arterioles and venules measured by DOCT flowmeter (
n = 20). In the arterioles, the Ds ranged from 80.9 to 123.4 μm (mean, 94.5 μm). The
Vaverages ranged from 9.2 to 21.3 mm/s (mean, 15.5 mm/s). The
Vmaximums ranged from 18.1 to 44.4 mm/s (mean, 30.7 mm/s). The RBFs ranged from 3.2 to 14.2 μL/min (mean, 7.8 μL/min). The ICCs of the D,
Vaverage,
Vmaximum, and RBF exceeded 0.9 (0.977, 0.957, 0.943, and 0.994, respectively). The CVs for the D,
Vaverage,
Vmaximum, and RBF of the measured arterioles were 3.4% ± 2.1%, 8.8% ± 6.2%, 10.8% ± 3.9%, and 6.0% ± 3.4%, respectively.
Table 2 Results of Intrasession Repeatability of Arterioles and Venules Measured on DOCT Flowmeter
Table 2 Results of Intrasession Repeatability of Arterioles and Venules Measured on DOCT Flowmeter
In the venules, the Ds ranged from 98.9 to 158.6 μm (mean, 118.7 μm). The Vaverages ranged from 5.6 to 21.2 mm/s (mean, 13.1 mm/s). The Vmaximums ranged from 7.2 to 26.3 mm/s (mean, 16.1 mm/s). The RBFs ranged from 4.9 to 19.4 μL/min (mean, 9.8 μL/min). The ICCs of the D, Vaverage, Vmaximum, and RBF exceeded 0.9 (0.973, 0.977, 0.982, and 0.944, respectively). The CVs for the D, Vaverage, Vmaximum, and RBF were 4.2% ± 2.7%, 8.8% ± 3.3%, 8.5% ± 3.6%, and 8.8% ± 5.1%, respectively.
The current study showed the good repeatability of the intrasession measurements and the reproducibility of the intersession and interobserver measurements of the RBF, using the newly developed DOCT flowmeter in healthy subjects. This study also found that absolute RBF values measured with the DOCT flowmeter were similar to the RBF values of the same arterioles measured by LDV. The RBF measurements using the DOCT flowmeter and LDV also were correlated significantly, indicating that the performance of the DOCT flowmeter is equivalent to that of LDV (
Fig. 4). In addition, in our previous LDV study,
4 the CVs of the RBF for intrasession repeatability reported are 13.8% in the arterioles and 12.7% in the venules. Regarding the current DOCT flowmeter, the CVs of the RBF were 6.0% in the arterioles and 8.8% in the venules (
Table 2). The Bland-Altman plots showed that the deviations of the arterioles and venules in the RBF values were good in the plots of the intersession and interobserver measurements (
Figs. 2,
3). These results also indicated that the DOCT flowmeter provides an accurate representation of the RBF.
Until the advent of the DOCT technique, LDV was the only commercially available instrument that measured the absolute value of the RBF in the arterioles and venules. Werkmeister et al.
20 showed for the first time that the DOCT (dual-beam bidirectional Doppler Fourier-domain OCT) measurements of the human venules are well correlated with those obtained with LDV (LDV-5000; Oculix, Inc., Arbaz, Switzerland). No studies have investigated the correlation in the human retinal arterioles. Therefore, the current findings showed for the first time that the DOCT flowmeter measurements of the human arterioles are well correlated with those of the LDV (
Fig. 4). No commercially available instrument has used the DOCT technique yet. Although the DOCT flowmeter is an experimental prototype, we are in the process of developing a commercially available product.
The LDV method can measure only the maximal blood velocity. In this technique, the blood flow in the retinal vessels should be considered as laminar flow to calculate the absolute values of blood velocity.
9 It is difficult to measure the bifurcation or the margin of the optic nerve head where the flow is considered to be nonlaminar in those areas. In addition, since the LDV technique requires strict focusing on the target vessel, it must be performed by a well-trained examiner. Therefore, there are some difficult situations when measuring the velocity with LDV. However, since the DOCT technique can detect not only the maximal blood velocity but also the entire blood velocity, it does not require parabolic flow definition. There is no restriction on the measurement vessel site. Our DOCT flowmeter system required appropriate detection of the Doppler angle (<86° in target vessels).
16 It is difficult to manually find the proper Doppler angle. We overcame the angle problem by using an automatic setting of the proper Doppler angle using a combination of OCT intensity and phase images.
16 The DOCT flowmeter does not require any specialized technique for observation. After the simple preparations to set the head on the instrument (as with the standard OCT measurements) are completed, the observer selects the target vessels. The RBF is measured automatically. The simplicity of this method makes it suitable for screening large populations and for clinical use. These are the major advantages of the DOCT flowmeter.
Another advantage of the DOCT flowmeter is the ability to analyze the blood flow pattern, that is, determination of the degree of laminar flow. Laminar flow is a type of blood flow in which the concentric layers of blood move in parallel through the vessel. The maximal velocity occurs in the center of the vessel and the minimal velocity is at the vessel wall.
21 The laminar flow is associated with shear stress, which is important for regulating blood flow.
22,23 The DOCT flowmeter can visualize the retinal blood velocity profile (RBVP), which is the velocity profile of the red blood cells in the horizontal center of a vessel at a given moment (
Fig. 1C). Recently, we first showed an abnormal retinal arterial RBVP during the diastolic phase in a patient with Takayasu's arteritis and aortic insufficiency (Tani T, et al.
IOVS 2015;56:ARVO E-Abstract 5892). These results indicate that the aortic valve function is related to the retinal arterial circulation. Therefore, not only absolute RBF but also RBVP analyses are expected to elucidate retinal circulatory disturbances.
The DOCT flowmeter has some limitations. First, if the Doppler angle exceeds 86° at the measured vessels, even though the incident beam was shifted automatically, the RBF was not considered reliable in the DOCT flowmeter measurement. Second, in some clinical cases, that is, when a vessel is close to the optic disc or in patients with small pupils, it might be difficult to find the proper Doppler angle. Third, the number of vessels detected in a single scan is limited to a couple of vessels when using the 1-mm scan length. Even when the scan length was expanded to 4 mm, not all vessels could be covered in one scan. If the total blood flow measurement is required, multiple DOCT scans are necessary to cover all the vessels.
Finally, scanning also might be difficult when there is an opacity in the eye, that is, a vitreous hemorrhage or hard cortical cataract.
In conclusion, the current study confirmed the reproducibility of intraobserver, intersession, and interobserver measurements in a newly developed DOCT flowmeter and that the DOCT flowmeter can precisely measure the RBF, compared with the LDV. These data suggest that the DOCT flowmeter is acceptable for measuring the RBF measurements in humans. Thus, the DOCT flowmeter has potential for clinical use and screening RBF changes in a large population.
The authors thank Lynda Charters for manuscript review and also thank Jun Sakai and Shunsuke Nakamura for technical support.
Supported by a Grant-in-Aid for Scientific Research from the Ministry of Education, Culture, Sports, Science, and Technology in Japan (15K20242 to TT and 16K20296 to TY).
Disclosure: T. Tani, None; Y.-S. Song, None; T. Yoshioka, None; T. Omae, None; A. Ishibazawa, None; M. Akiba, None; A. Yoshida, None