June 2017
Volume 58, Issue 7
Open Access
Retina  |   June 2017
Axial Length Variation Impacts on Superficial Retinal Vessel Density and Foveal Avascular Zone Area Measurements Using Optical Coherence Tomography Angiography
Author Affiliations & Notes
  • Danuta M. Sampson
    Centre for Ophthalmology and Visual Science, The University of Western Australia, Perth, Western Australia, Australia
    Lions Eye Institute, Perth, Western Australia, Australia
  • Peijun Gong
    Optical+Biomedical Engineering Laboratory, School of Electrical, Electronic & Computer Engineering, The University of Western Australia, Perth, Western Australia, Australia
  • Di An
    College of Optical Science and Engineering, Zhejiang University, Hangzhou, China
  • Moreno Menghini
    Department of Ophthalmology, Sir Charles Gairdner Hospital, Perth, Western Australia, Australia
  • Alex Hansen
    Centre for Ophthalmology and Visual Science, The University of Western Australia, Perth, Western Australia, Australia
  • David A. Mackey
    Centre for Ophthalmology and Visual Science, The University of Western Australia, Perth, Western Australia, Australia
    Lions Eye Institute, Perth, Western Australia, Australia
  • David D. Sampson
    Optical+Biomedical Engineering Laboratory, School of Electrical, Electronic & Computer Engineering, The University of Western Australia, Perth, Western Australia, Australia
    Centre for Microscopy, Characterisation & Analysis, The University of Western Australia, Perth, Western Australia, Australia
  • Fred K. Chen
    Centre for Ophthalmology and Visual Science, The University of Western Australia, Perth, Western Australia, Australia
    Lions Eye Institute, Perth, Western Australia, Australia
    Department of Ophthalmology, Royal Perth Hospital, Perth, Western Australia, Australia
  • Correspondence: Fred K. Chen, Centre for Ophthalmology and Visual Science, The University of Western Australia, 2 Verdun Street, Perth, Western Australia 6009, Australia; fred.chen@lei.org.au
Investigative Ophthalmology & Visual Science June 2017, Vol.58, 3065-3072. doi:https://doi.org/10.1167/iovs.17-21551
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Danuta M. Sampson, Peijun Gong, Di An, Moreno Menghini, Alex Hansen, David A. Mackey, David D. Sampson, Fred K. Chen; Axial Length Variation Impacts on Superficial Retinal Vessel Density and Foveal Avascular Zone Area Measurements Using Optical Coherence Tomography Angiography. Invest. Ophthalmol. Vis. Sci. 2017;58(7):3065-3072. https://doi.org/10.1167/iovs.17-21551.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: To evaluate the impact of image magnification correction on superficial retinal vessel density (SRVD) and foveal avascular zone area (FAZA) measurements using optical coherence tomography angiography (OCTA).

Methods: Participants with healthy retinas were recruited for ocular biometry, refraction, and RTVue XR Avanti OCTA imaging with the 3 × 3-mm protocol. The foveal and parafoveal SRVD and FAZA were quantified with custom software before and after correction for magnification error using the Littman and the modified Bennett formulae. Relative changes between corrected and uncorrected SRVD and FAZA were calculated.

Results: Forty subjects were enrolled and the median (range) age of the participants was 30 (18–74) years. The mean (range) spherical equivalent refractive error was −1.65 (−8.00 to +4.88) diopters and mean (range) axial length was 24.42 mm (21.27–28.85). Images from 13 eyes were excluded due to poor image quality leaving 67 for analysis. Relative changes in foveal and parafoveal SRVD and FAZA after correction ranged from −20% to +10%, −3% to +2%, and −20% to +51%, respectively. Image size correction in measurements of foveal SRVD and FAZA was greater than 5% in 51% and 74% of eyes, respectively. In contrast, 100% of eyes had less than 5% correction in measurements of parafoveal SRVD.

Conclusions: Ocular biometry should be performed with OCTA to correct image magnification error induced by axial length variation. We advise caution when interpreting interocular and interindividual comparisons of SRVD and FAZA derived from OCTA without image size correction.

Quantitative assessment of superficial retinal vessel density (SRVD) and foveal avascular zone area (FAZA) serves as a potential biomarker for macular ischemia in diabetic retinopathy and other retinal vascular diseases.1 Therefore, it is essential that normal variations in SRVD and FAZA in the healthy population are established definitively. Normative ranges of vessel densities27 and FAZA1,8 measured with optical coherence tomography (OCT) angiography (OCTA) have recently been published. However, current commercial OCTA devices do not correct density or area measurements for changes in image size magnification due to axial length variation. 
It is well known that the true size of the fundus area imaged is dependent on the combined magnification of the camera and the eye.9 Various methods have been introduced to correct for the magnification of the eye.10,11 These include methods based on one or more of a combination of variables, including refractive error,11 keratometry,1113 axial length,10,12,14 anterior chamber depth,10 and crystalline lens thickness.10 These methods have been compared by Garway-Heath et al.,9 who concluded that the simplest approach of using only axial length to correct for magnification error was accurate and produced very similar results to the more detailed calculations that incorporated keratometry, refractive error, and lens thickness.15 Given most eye clinics have access to devices that measure axial length and that incorporating axial length has been shown to increase the accuracy of optic nerve head size and retinal layer thickness evaluated by OCT,1623 we hypothesize that OCTA image magnification correction using axial length measurement is feasible and necessary for accurate determination of OCTA-derived variables. 
In this study, we sought to investigate the effect of image magnification on quantitative characterization of the retinal vasculature. Specifically, we evaluated the effect of axial length on SRVD and FAZA measurements using OCTA in individuals with healthy eyes. 
Methods
This study was approved by The University of Western Australia Human Ethics Research Office (RA/4/1/8570) and was conducted in compliance with the tenets of the Declaration of Helsinki. All patients gave informed consent prior to participation in the study and did not receive financial compensation. 
Study Participants
Individuals with no history of retinal disease were eligible for the study. Exclusion criteria were: best-corrected visual acuity of worse than 20/20; evidence of macular or retinal disease; and presence of media opacity, such as cataracts, corneal, or vitreous opacities. All participants underwent ophthalmic examination, ocular biometry (IOL Master; Carl Zeiss Meditec, Inc., Dublin, CA, USA), autorefraction (Ark1, Auto Ref/Keratometer; Nidek, Gamagori, Japan), and OCTA imaging in both eyes (RTVue XR Avanti; Optovue, Inc., Fremont, CA, USA). All biometric measurements and OCTA imaging were performed by trained operators. 
Optical Coherence Tomography Angiography Imaging Protocol
The instrument used in this study operates at 70,000 A-scans per second. According to the manufacturer, RTVue XR Avanti has an axial resolution of 5 μm and the lateral resolution, defined as full-width at half maximum beam diameter at the retinal plane, is approximately 22 μm. The instrument uses a superluminescent diode light source that has a central wavelength of 840 nm. 
Successive OCTA images were taken using a scanning procedure designed to reduce motion artifact. Two volumes scans were acquired sequentially using orthogonal raster scan patterns (denoted, respectively, x-fast and y-fast) over an assumed 3 × 3-mm retinal region centered on the fovea. For the x-fast scan pattern, each B-scan comprised 304 A-scans (along the x-axis) and two consecutive B-scans were captured at each of the 304 locations along the y-axis of the cube. With a scan rate of 209 frames per second, a total of 608 B-scans (304 y-positions × 2 B-scans per position) were acquired in 3 seconds. The flow signal is detected by examining the consecutive B-scans of the same area for differences (decorrelation) using the so-called split-spectrum amplitude decorrelation angiography algorithm. This process allows generation of the nominally 3 × 3-mm x-fast en face OCTA images. The procedure to obtain y-fast en face OCTA images is defined analogously, except that B-scans were acquired parallel to the y-axis. Finally, the Motion Correction Technology (MCT) of the AngioVue software (Optovue, Inc.) automatically aligned and merged the OCTA x-fast and y-fast volume scans to reduce motion artifacts and produce an OCTA en face projection image by automatic segmentation of the superficial retinal capillary plexus and applying maximum intensity projection mapping in the axial direction. 
Optical Coherence Tomography Angiography Fundus Image Magnification
The nominally 3 × 3-mm retinal scan area is derived from calculation of the linear distance on the surface of the retina subtended by a fixed field of view of approximately 10° × 10° using a defined axial length (for the RTVue XR Avanti system, this length is 23.82 mm) (Optovue, Inc., personal communication, 2017). However, the true physical linear dimension of the image (Dt) varies proportionally with the axial length of the eye being imaged (Fig. 1). Therefore, we used the Littman and the modified Bennett formulae to calculate the image size; thereby, determining the relationship between measurements with and without correction.10 
Figure 1
 
Schematic diagrams of the scanning mechanism in an OCTA system for varying axial lengths. The incident beam is shown for three scan paths (yellow, brown, and green; β0 is the incident angle at the cornea and γ0 is angle subtended on the retina); yellow and green paths represent the maximum deflections and the brown path is on axis. Refraction from the scan lens (not shown), cornea, and lens focus each light path on the retina. (A) Default axial length (ALeye0) for which the linear size of the measured retinal diameter (Dm) corresponds to the true retinal diameter (Dt). (B) Left: for shorter axial length (ALeye1), Dt1 < Dm; and right: for longer axial length (ALeye2), Dt2 > Dm.
Figure 1
 
Schematic diagrams of the scanning mechanism in an OCTA system for varying axial lengths. The incident beam is shown for three scan paths (yellow, brown, and green; β0 is the incident angle at the cornea and γ0 is angle subtended on the retina); yellow and green paths represent the maximum deflections and the brown path is on axis. Refraction from the scan lens (not shown), cornea, and lens focus each light path on the retina. (A) Default axial length (ALeye0) for which the linear size of the measured retinal diameter (Dm) corresponds to the true retinal diameter (Dt). (B) Left: for shorter axial length (ALeye1), Dt1 < Dm; and right: for longer axial length (ALeye2), Dt2 > Dm.
According to the Littman formula, the relationship between the measured OCTA image diameter, Display Formula\(\def\upalpha{\unicode[Times]{x3B1}}\)\(\def\upbeta{\unicode[Times]{x3B2}}\)\(\def\upgamma{\unicode[Times]{x3B3}}\)\(\def\updelta{\unicode[Times]{x3B4}}\)\(\def\upvarepsilon{\unicode[Times]{x3B5}}\)\(\def\upzeta{\unicode[Times]{x3B6}}\)\(\def\upeta{\unicode[Times]{x3B7}}\)\(\def\uptheta{\unicode[Times]{x3B8}}\)\(\def\upiota{\unicode[Times]{x3B9}}\)\(\def\upkappa{\unicode[Times]{x3BA}}\)\(\def\uplambda{\unicode[Times]{x3BB}}\)\(\def\upmu{\unicode[Times]{x3BC}}\)\(\def\upnu{\unicode[Times]{x3BD}}\)\(\def\upxi{\unicode[Times]{x3BE}}\)\(\def\upomicron{\unicode[Times]{x3BF}}\)\(\def\uppi{\unicode[Times]{x3C0}}\)\(\def\uprho{\unicode[Times]{x3C1}}\)\(\def\upsigma{\unicodeTimes]{x3C3}}\)\(\def\uptau{\unicode[Times]{x3C4}}\)\(\def\upupsilon{\unicode[Times]{x3C5}}\)\(\def\upphi{\unicode[Times]{x3C6}}\)\(\def\upchi{\unicode[Times]{x3C7}}\)\(\def\uppsy{\unicode[Times]{x3C8}}\)\(\def\upomega{\unicode[Times]{x3C9}}\)\({D_m}\), and the true diameter on the fundus, Display Formula\({D_t}\), can be expressed as: Display Formula\({D_t} = p \cdot q \cdot {D_m},\) where Display Formula\(p \cdot q\) is the overall image magnification factor; p is that of the imaging system and q is that of the eye.10 The factor q can be determined from the Bennett formula: Display Formula\(q = 0.01306 \cdot (AL - 1.82)\), where 1.82 is a constant related to the distance between the corneal apex and the second principal plane and AL is the axial length.10 The factor p, omitting any effect arising from image distortion, can be readily calculated from the Bennet formula if the axial length at which Dt = Dm is known (i.e., 23.82 mm here). When Dt = Dm, then Display Formula\(p = 1/q\) and, therefore, Display Formula\(p = 1/\left[ {0.01306\left( {23.82 - 1.82} \right)} \right] = 3.48.\) Thus, the magnification factor of the image should be corrected by: Display Formula\({D_t}^2/{D_m}^2 = 0.002066{(AL - 1.82)^2}\).  
Validation of Custom-Written Vessel Segmentation
The highest resolution en face OCTA image exported from AngioVue software is 72 dpi (Fig. 2A). This image is first resized to 300 dpi (Corel Photo-Paint X6; Corel, Ottawa, Ontario, Canada) by using nearest-neighbor interpolation prior to automated analysis with our custom-written algorithm in MATLAB (vR2012a; The MathWorks, Inc., Natick, MA, USA). Resizing and interpolation is necessary so that our custom software can accurately identify the pixel intensity threshold used by the AngioVue Analytics software, since the raw decorrelation value for each pixel is not accessible. Vessel density is defined as the percentage of the number of pixels with intensity greater than a fixed threshold to the total number of pixels in the region of interest and expressed in units of percent pixels. The threshold value for each nominally 3 × 3-mm image is set individually by benchmarking against the values obtained from AngioVue Analytics software (version 2016.1.0.26). The optimal threshold for each image is chosen when the sum of the absolute differences in vessel densities (SRVD from our custom software minus SRVD from AngioVue software), across all nine square zones, reaches a minimum (Fig. 2B). This procedure is repeated so that an image-specific threshold is determined and used for calculating our custom-derived vessel density from each OCTA image. 
Figure 2
 
Diagram showing methodology for quantification of SRVD using real data from an eye with axial length shorter than the ideal axial length used by the OCTA device. The en face OCTA image of retinal vasculature is resized to 300 dpi (A). The image is divided into nine squares that are used to find the optimal histogram threshold for the whole image by benchmarking against the AngioVue data for each image (B). The vessel density is measured before and after correction for magnification error in the foveal (yellow, i) and parafoveal (pale blue, ii) regions, as defined in the text (C). Cropped foveal and parafoveal regions are shown before and after correction (D). FAZA is measured before correction and the area values are scaled by the magnification factor to obtain FAZA for the corrected images (E).
Figure 2
 
Diagram showing methodology for quantification of SRVD using real data from an eye with axial length shorter than the ideal axial length used by the OCTA device. The en face OCTA image of retinal vasculature is resized to 300 dpi (A). The image is divided into nine squares that are used to find the optimal histogram threshold for the whole image by benchmarking against the AngioVue data for each image (B). The vessel density is measured before and after correction for magnification error in the foveal (yellow, i) and parafoveal (pale blue, ii) regions, as defined in the text (C). Cropped foveal and parafoveal regions are shown before and after correction (D). FAZA is measured before correction and the area values are scaled by the magnification factor to obtain FAZA for the corrected images (E).
Superficial Retinal Vessel Density and Foveal Avascular Zone Area Quantification
Following validation of the custom segmentation software, SRVD was measured in two regions: the foveal region, defined as a circular area with 0.5-mm radius centered on the foveal center (Fig. 2C yellow region labeled i), and the parafoveal region, defined as the annular region between 0.5- and 1-mm radius from the foveal center (Fig. 2C, pale blue region labeled ii). The foveal avascular zone (FAZ) was delineated in uncorrected images by image segmentation using the Chan-Vese approach (Fig. 2E).24 A contour within the FAZ defined by four points, manually selected by the user, initiated the search routine. FAZA was calculated as the total number of black pixels enclosed by the segmentation contour and expressed in millimeters squared. This procedure was repeated for each image. Finally, the area values were scaled by the magnification correction factor to obtain the corrected FAZA. 
The impact of axial length on deviation of true SRVD or FAZA from uncorrected values was examined by plotting their relative changes against axial length. The relative change in SRVD was defined as: 100 × (SRVD after correction − SRVD before correction) / SRVD before correction. The relative change in FAZA was defined analogously as a percentage of uncorrected FAZA. Image magnification correction factor was defined as: image area after correction for magnification error divided by image area before correction. Bland-Altman analysis was also performed to determine limits of agreement between corrected and uncorrected SRVD and FAZA, as described in detail below. 
Statistical Analysis
Agreement between the results from custom-written and AngioVue Analytics software and the SRVD and FAZA measured before and after correction for magnification error were evaluated using Bland-Altman plots and limits of agreement.2527 First, a scatter plot of absolute differences in values between the two methods against their means was plotted to confirm that there was no relationship between error and magnitude. If this condition was met, then the bias (mean difference), standard deviation of the differences (SD of diff.) and confidence limits for the bias (i.e., the limits of agreement, LA, defined as: mean difference ± 1.96 SD of the differences) were calculated. The bias and LA are displayed in Figures 3 and 4 as solid and dashed lines, respectively. The LA represents the range of values for the differences between the methods that can be expected 95% of the time. Paired t-tests were used to examine the significance of the bias. The t-statistic (tstat) is given by Display FormulaImage not available, where Display FormulaImage not available is the mean difference, and Display FormulaImage not available is the standard error of the mean difference calculated under the null hypothesis. A P value of 0.05 or less was considered statistically significant.  
Figure 3
 
Bland-Altman plots illustrating the agreement between the custom-designed and AngioVue software for the measurement of blood vessel density for all nine regions (I–IX) as defined in Figure 2.
Figure 3
 
Bland-Altman plots illustrating the agreement between the custom-designed and AngioVue software for the measurement of blood vessel density for all nine regions (I–IX) as defined in Figure 2.
Figure 4
 
Bland-Altman plots illustrating the agreement before and after correction for the measurement of SRVD in foveal (A) and parafoveal (B) regions and FAZA (C). tstat and P value are, respectively, −2.0 and 0.05 for SRVD in the foveal region, −1.1 and 0.28 for SRVD in the parafoveal region, and 1.8 and 0.08 for FAZA. The Littman and the modified Bennett formula were applied to correct image size for a magnification error.
Figure 4
 
Bland-Altman plots illustrating the agreement before and after correction for the measurement of SRVD in foveal (A) and parafoveal (B) regions and FAZA (C). tstat and P value are, respectively, −2.0 and 0.05 for SRVD in the foveal region, −1.1 and 0.28 for SRVD in the parafoveal region, and 1.8 and 0.08 for FAZA. The Littman and the modified Bennett formula were applied to correct image size for a magnification error.
Results
Demographics
Forty individuals were enrolled in this study (24 women, 16 men). The median (range) age of the participants was 30 (18–74) years. Nineteen subjects were under 30 years, eight between 30 and 40, four between 40 and 50, and nine individuals older than 50 years. Of the subjects, 30 were Caucasian and 10 were Asian. Sixty-seven eyes from 40 volunteers produced adequate OCTA image quality for analysis. Images from the remaining 13 eyes were excluded from the study due to poor image quality. Among the 67 eyes, only 62 images had an identifiable FAZ for measurement of FAZA. The mean (SD, range) spherical equivalent refraction was −1.65 (2.54, −8.00 to +4.88) diopters (D). The mean (SD, range) axial length was 24.42 (1.59, 21.27–28.85) mm. 
Validation of Custom-Written Vessel Segmentation Software
There was good overall agreement between our custom software and the AngioVue Analytics software for the measurement of retinal vessel density for all subjects and in all nine regions (Fig. 3) defined in Figure 2B. The mean (SD) difference and the lower and upper bounds of the LAs in density between our method and the commercial software are presented in Figure 3 and the Table. A slight bias was noted in squares V and VII but the LAs showed close agreement overall (Fig. 3; Table). 
Table
 
Bland-Altman Analysis and Paired t-test Showing the Agreement Between Custom-Written Software and the AngioVue Analytics Software for the Measurement of Retinal Vessel Density
Table
 
Bland-Altman Analysis and Paired t-test Showing the Agreement Between Custom-Written Software and the AngioVue Analytics Software for the Measurement of Retinal Vessel Density
Image Size Correction Using Littmann and Modified Bennett Formulae
The mean (SD) difference in SRVD before and after correction (vessel density in units of percentage: number of bright pixels/total number of pixels in each image [% pixels]) were −0.6 (2.5) and 0 (0.5) in the fovea (P = 0.05) and parafovea (P = 0.28 Figs. 4A–B), respectively. The lower and upper bounds of the LAs in SRVD before and after correction were −5 and +4, and −1 and +1 in the fovea and parafovea, respectively. The mean (SD) difference in FAZA before and after correction (in units of mm2 in each case) was 0 (0.03) and the respective lower and upper bounds of LAs were −0.06 and +0.06 (P = 0.08, Fig. 4C). 
Relative change in foveal and parafoveal SRVD and FAZA, before and after correction for magnification error at various axial lengths and refractive errors, are shown in Figure 5. While the relative change in parafoveal SRVD (% of the uncorrected value) was minimal (−3 to +2), the relative change in foveal SRVD (% of the uncorrected value) ranged from −20 to +10. The relative change in FAZA (% of uncorrected value) ranged from −20 to +51 after image size correction. 
Figure 5
 
Plots of the relative change in SRVD in the foveal (A), and parafoveal (B), regions and FAZA (C), before and after correction for magnification error. The arrow represents the default axial length of 23.82 mm for the RTVue instrument. The dashed horizontal lines indicate the 5% error margins in vessel density. The fine dashed lines are nonlinear least-squares fits to the data. Plots of the relative change in SRVD in the foveal (D), and parafoveal (E), regions and FAZA (F), before and after correction for magnification error against spherical equivalent.
Figure 5
 
Plots of the relative change in SRVD in the foveal (A), and parafoveal (B), regions and FAZA (C), before and after correction for magnification error. The arrow represents the default axial length of 23.82 mm for the RTVue instrument. The dashed horizontal lines indicate the 5% error margins in vessel density. The fine dashed lines are nonlinear least-squares fits to the data. Plots of the relative change in SRVD in the foveal (D), and parafoveal (E), regions and FAZA (F), before and after correction for magnification error against spherical equivalent.
The relative change in foveal SRVD was less than 5% (relative to uncorrected values) for 33 of 67 eyes (49%), with axial lengths ranging from 22.28 to 24.71 mm (Fig. 5A). This contrasts with parafoveal SRVD (Fig. 5B), which exhibited changes of less than 5% in 67 eyes (100%) with axial lengths ranging from 21.27 to 28.85 mm. Only 16 of 62 eyes (26%), with axial lengths ranging from 23.29 to 24.33 mm, had less than 5% change in FAZA after correction (Fig. 5C). There was no specific range of refractive error that predicted less than 5% change in foveal and parafoveal SRVD (Figs. 5D, 5E), as well as in FAZA (Fig. 5E). 
Discussion
We found that, after correction, relative changes in foveal and parafoveal SRVD and FAZA ranged from −20% to +10%, −3% to +2%, and −20% to +51%, respectively. Image size correction was greater than 5% in 51% and 74% of eyes, respectively, in measurements of foveal SRVD and FAZA. By contrast, 100% of eyes exhibited less than 5% correction of parafoveal SRVD. 
The differential effects of image magnification on foveal versus parafoveal SRVD can be explained as follows: the foveal region (central 1-mm circle) incorporates the FAZ (black pixels) surrounded by dense capillaries (white pixels) at the edge of the image. A small change in the image boundary location, as a result of magnification correction, will affect bright pixel density disproportionately; thus, having greater impact on foveal SRVD (Fig. 2D). In contrast, the parafoveal region has a relatively uniform vascular network in healthy eyes; therefore, small changes in its boundary from image size correction are less likely to induce significant change in overall density. If we excluded the FAZ from the calculation of vessel density in the foveal region, the error in foveal SRVD would be smaller and may even be comparable to that of parafoveal SRVD. However, our main purpose is to illustrate the extent of the impact of image magnification on readily accessible parameters, such as foveal and parafoveal SRVD without the need to exclude an arbitrarily defined avascular zone, as this may introduce another source of variability to the definition of normative intervals in SRVD. Furthermore, we believe it is necessary to include the FAZ in characterization of the foveal SRVD because pathologic retinal capillary dropout can alter the size of FAZA and also lead to the development of new regions of avascular zones in the parafovea. Such arbitrary exclusion of avascular regions from the measurement of foveal and parafoveal SRVD may result in reduced sensitivity of OCTA in detecting retinal vascular disease. In addition, the impact of image size correction on parafoveal SRVD may be greater in diseased eyes, where significant ischemia (black pixels) may occur in the parafoveal region simulating the pattern seen in the healthy fovea. 
The impact of correction for magnification error on the interpretation of OCT findings in a variety of retinal disease conditions has been previously shown by others. The effect of axial length on peripapillary retinal nerve fiber layer (RNFL) thickness in myopic, hyperopic, and emmetropic eyes has been investigated using different OCT instruments. For example, Kang et al.18 and Savini et al.22 used the Cirrus HD OCT instrument to evaluate the influence of axial length on RNFL measurements in adult eyes. In other studies, the RTVue OCT was employed to evaluate the same relationship in a group of adults28 and, separately, in a group of children.23 Those studies showed that RNFL thickness was significantly different between myopic, hyperopic, and emmetropic eyes in adults and children; however, these differences disappeared after correction for magnification error using the modified Bennett formula.10 Similarly, measurement of the optic disc size has an important bearing on the evaluation of optic nerve diseases and anomalies. Leung et al.20 used a Stratus OCT instrument to evaluate associations between optic disc area and axial length/spherical equivalent. The relationship between the measured size of the OCT image and the actual fundus size imaged was also described using Littmann's formula.20 The main finding of that study was that the optic disc area increases with axial length and myopic refraction. For each millimeter increase in axial length, the optic disc area increased by 0.095 mm2.20 Huang et al.14 also studied the relationship between RNFL thickness, optic disc size, and image magnification using an approach based on axial length only. In that study, after accounting for the effect of magnification, RNFL thickness was no longer dependent on axial length. Our observation is in agreement with previously described reports that axial length affects the assessment of retinal vascular biomarkers and should be taken into account in their quantification. Furthermore, we report here the range of axial lengths (22.28–24.71 mm) that occurred in 49% of the eyes that result in a less than 5% change in foveal SRVD following image size correction. Theoretically, axial lengths in the range 23.29 to 24.33 mm would be predicted to correspond to less than 5% change in FAZA, but only 26% of the eyes in our cohort had axial lengths within this range. Therefore, 51% and 74% of eyes may have significant error (>5%) in foveal SRVD and FAZA, respectively, if image size correction is not applied. 
There are several limitations in the present study. First, our sample size is small and accordingly our range of axial lengths is smaller than expected in the general population. Further studies with a larger number of healthy subjects, encompassing a wider range of axial lengths, age, and spherical equivalents, are necessary to fully evaluate vessel density as a biomarker of retinal health. Second, we only examined healthy subjects with no clinical evidence of retinal disease based on indirect fundoscopy. The relative error in SRVD measurement may be greater in patients with capillary dropouts arising from diseases, such as diabetic retinopathy, radiation retinopathy, macular telangiectasia, and retinal vein occlusion. Third, we did not examine the impact of magnification on deep retinal vessel density (DRVD). Quantitative analysis of the deep capillary plexus is not reliable using the AngioVue software because of projection artifacts cause the FAZ to be not as clearly defined in the deep capillary plexus as in the superficial capillary plexus. For these reasons, we recommend further studies on DRVD once projection artifacts can be removed in commercial instruments.29,30 
Further work is required to design a method for vessel density calculation that is independent of the AngioVue software and to optimize FAZA segmentation. There were no cases in our study population where the FAZ was not continuous (i.e., divided by bridging vessels into separate zones). However, this phenomenon is not uncommon in diseased eyes where discrete avascular regions can be separated by capillaries requiring manual adjustment. An important related issue, which did not arise in our study, is the need for segmentation and calculation of not only the FAZA but also of total avascular zone areas in the macula. Hwang et al.31 demonstrated that objective quantification of total capillary nonperfusion can be a better biomarker of retinal vascular disease than FAZA, because of FAZA's relatively high variability in healthy individuals. Thus, an accurate and automated method for measurement of both FAZA and total avascular area would be advantageous. 
Conclusions
We have demonstrated the impact of image magnification due to axial length variation on SRVD and FAZA measurement in OCTA. Large errors in these parameters can arise in the absence of image size correction. Our results show correction of foveal SRVD is required in 51% of eyes and of FAZA in 74% of eyes. We advise caution when interpreting interocular and interindividual comparisons of SRVD and FAZA derived from OCTA without image size correction. 
Acknowledgments
The authors thank Avenell Chew and Yi Chen for assistance with data collection. 
Supported in part by research grants from the Ophthalmic Research Institute of Australia (DMS, FKC; Surry Hills, NSW, Australia); Global Ophthalmology Award Program Bayer (FKC; Leverkusen, Germany); National Health & Medical Research Council (Early Career Fellowship; APP1054712, FKC; Canberra, ACT, Australia); Retina Australia (FKC; Perth, WA, Australia); donations from the Miocevich, Constantine, and Saleeba families (FKC; Perth, WA, Australia) and Health Department of Western Australia (FKC; Perth, WA, Australia). 
Disclosure: D.M. Sampson, None; P. Gong, None; D. An, None; M. Menghini, None; A. Hansen, None; D.A. Mackey, None; D.D. Sampson, None; F.K. Chen, None 
References
Kim AY, Chu Z, Shahidzadeh A, Wang RK, Puliafito CA, Kashani AH. Quantifying microvascular density and morphology in diabetic retinopathy using spectral-domain optical coherence tomography angiography. Invest Ophthalmol Vis Sci. 2016; 57: 362–370.
Lupidi M, Coscas F, Cagini C, et al. Automated quantitative analysis of retinal microvasculature in normal eyes on optical coherence tomography angiography. Am J Ophthalmol. 2016; 169: 9–23.
Sridhar J, Shahlaee A, Rahimy E, et al. Optical coherence tomography angiography and en face optical coherence tomography features of paracentral acute middle maculopathy. Am J Ophthalmol. 2015; 160: 1259–1268.
Agemy SA, Scripsema NK, Shah CM, et al. Retinal vascular perfusion density mapping using optical coherence tomography angiography in normals and diabetic retinopathy patients. Retina. 2015; 35: 2353–2363.
Gadde SGK, Anegondi N, Bhanushali D, et al. Quantification of vessel density in retinal optical coherence tomography angiography images using local fractal dimension. Invest Ophthalmol Vis Sci. 2016; 57: 246–252.
Yu J, Jiang C, Wang X, et al. Macular perfusion in healthy Chinese: an optical coherence tomography angiogram study. Invest Ophthalmol Vis Sci. 2015; 56: 3212–3217.
Al-Sheikh M, Tepelus TC, Nazikyan T, Sadda SR. Repeatability of automated vessel density measurements using optical coherence tomography angiography. Br J Ophthalmol. 2017; 101: 449–452.
Coscas F, Sellam A, Glacet-Bernard A, et al. Normative data for vascular density in superficial and deep capillary plexuses of healthy adults assessed by optical coherence tomography angiography. Invest Ophthalmol Vis Sci. 2016; 57: 211–223.
Garway-Heath DF, Rudnicka AR, Lowe T, et al. Measurement of optic disc size: equivalence of methods to correct for ocular magnification. Br J Ophthalmol. 1998; 82: 643–649.
Bennett AG, Rudnicka AR, Edgar DF. Improvements on Littmann's method of determining the size of retinal features by fundus photography. Graefes Arch Clin Exp Ophthalmol. 1994; 232: 361–367.
Bengtsson B, Krakau CE. Some essential optical features of the Zeiss fundus camera. Acta Ophthalmol. 1977; 55: 123–131.
Bengtsson B, Krakau CE. Correction of optic disc measurements on fundus photographs. Graefes Arch Clin Exp Ophthalmol. 1992; 230: 24–28.
Littmann H. Determination of the true size of an object on the fundus of the living eye. By H. Littmann from the original article, “Zur Bestimmung der wahren Grosse eines Objektes auf dem Hintergrund des lebenden Auges,” which originally appeared in Klinisches Monats. Optom Vis Sci. 1992; 69: 717–720.
Huang D, Chopra V, Lu AT-H, Tan O, Francis B, Varma R. Does optic nerve head size variation affect circumpapillary retinal nerve fiber layer thickness measurement by optical coherence tomography? Invest Ophthalmol Vis Sci. 2012; 53: 4990–4997.
Langenbucher A, Seitz B, Viestenz A. Computerised calculation scheme for ocular magnification with the Zeiss telecentric fundus camera. Ophthalmic Physiol Opt. 2003; 23: 449–455.
Bayraktar S, Bayraktar Z, Yilmaz OF. Influence of scan radius correction for ocular magnification and relationship between scan radius with retinal nerve fiber layer thickness measured by optical coherence tomography. J Glaucoma. 2001; 10: 163–169.
Rauscher F, Navneet S, Feuer W, Budenz D. Myopia affects retinal nerve fiber layer measurements as determined by optical coherence tomography. J Glaucoma. 2009; 18: 501–505.
Kang SH, Hong SW, Im SK, Lee SH, Ahn MD. Effect of myopia on the thickness of the retinal nerve fiber layer measured by cirrus HD optical coherence tomography. Invest Ophthalmol Vis Sci. 2010; 51: 4075–4083.
Nowroozizadeh S, Cirineo N, Amini N, et al. Influence of correction of ocular magnification on spectral-domain OCT retinal nerve fiber layer measurement variability and performance. Invest Ophthalmol Vis Sci. 2014; 55: 3439–3446.
Leung CKS, Cheng ACK, Chong KKL, et al. Optic disc measurements in myopia with optical coherence tomography and confocal scanning laser ophthalmoscopy. Invest Ophthalmol Vis Sci. 2007; 48: 3178–3183.
Hirasawa K, Shoji N, Yoshii Y, Haraguchi S. Determination of axial length requiring adjustment of measured circumpapillary retinal nerve fiber layer thickness for ocular magnification. PLoS One. 2014; 9: 1–5.
Savini G, Barboni P, Parisi V, Carbonelli M. The influence of axial length on retinal nerve fibre layer thickness and optic-disc size measurements by spectral-domain OCT. Br J Ophthalmol. 2012; 96: 57–61.
Aykut V, Öner V, Taş M, Işcan Y, Ağaçhan A. Influence of axial length on peripapillary retinal nerve fiber layer thickness in children: a study by RTVue spectral-domain optical coherence tomography. Curr Eye Res. 2013; 38: 1241–1247.
Wang X, Huang D, Xu H. An efficient local Chan–Vese model for image segmentation. Pattern Recognit. 2010; 43: 603–618.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1: 307–310.
Bunce C. Correlation, agreement, and Bland-Altman analysis: statistical analysis of method comparison studies. Am J Ophthalmol. 2009; 148: 4–6.
Hanneman S. Design, analysis and interpretation of method-compariosn studies. AACN Adv Crit Care. 2008; 19: 223–234.
Oner V, Aykut V, Tas M, Alakus MF, Iscan Y. Effect of refractive status on peripapillary retinal nerve fibre layer thickness: a study by RTVue spectral domain optical coherence tomography. Br J Ophthalmol. 2013; 97: 75–79.
Zhang M, Hwang TS, Campbell P, et al. Projection-resolved optical coherence tomographic angiography. Biomed Opt Express. 2016; 7: 816–828.
Hwang TS, Zhang M, Bhavsar K, et al. Visualization of 3 distinct retinal plexuses by projection-resolved optical coherence tomography angiography in diabetic retinopathy. JAMA Ophthalmol. 2016: 134: 1411–1419.
Hwang TS, Gao SS, Liu L, Lauer AK, et al. Automated quantification of capillary nonperfusion using optical coherence tomography angiography in diabetic retinopathy. JAMA Ophthalmol. 2016: 134: 367–373.
Figure 1
 
Schematic diagrams of the scanning mechanism in an OCTA system for varying axial lengths. The incident beam is shown for three scan paths (yellow, brown, and green; β0 is the incident angle at the cornea and γ0 is angle subtended on the retina); yellow and green paths represent the maximum deflections and the brown path is on axis. Refraction from the scan lens (not shown), cornea, and lens focus each light path on the retina. (A) Default axial length (ALeye0) for which the linear size of the measured retinal diameter (Dm) corresponds to the true retinal diameter (Dt). (B) Left: for shorter axial length (ALeye1), Dt1 < Dm; and right: for longer axial length (ALeye2), Dt2 > Dm.
Figure 1
 
Schematic diagrams of the scanning mechanism in an OCTA system for varying axial lengths. The incident beam is shown for three scan paths (yellow, brown, and green; β0 is the incident angle at the cornea and γ0 is angle subtended on the retina); yellow and green paths represent the maximum deflections and the brown path is on axis. Refraction from the scan lens (not shown), cornea, and lens focus each light path on the retina. (A) Default axial length (ALeye0) for which the linear size of the measured retinal diameter (Dm) corresponds to the true retinal diameter (Dt). (B) Left: for shorter axial length (ALeye1), Dt1 < Dm; and right: for longer axial length (ALeye2), Dt2 > Dm.
Figure 2
 
Diagram showing methodology for quantification of SRVD using real data from an eye with axial length shorter than the ideal axial length used by the OCTA device. The en face OCTA image of retinal vasculature is resized to 300 dpi (A). The image is divided into nine squares that are used to find the optimal histogram threshold for the whole image by benchmarking against the AngioVue data for each image (B). The vessel density is measured before and after correction for magnification error in the foveal (yellow, i) and parafoveal (pale blue, ii) regions, as defined in the text (C). Cropped foveal and parafoveal regions are shown before and after correction (D). FAZA is measured before correction and the area values are scaled by the magnification factor to obtain FAZA for the corrected images (E).
Figure 2
 
Diagram showing methodology for quantification of SRVD using real data from an eye with axial length shorter than the ideal axial length used by the OCTA device. The en face OCTA image of retinal vasculature is resized to 300 dpi (A). The image is divided into nine squares that are used to find the optimal histogram threshold for the whole image by benchmarking against the AngioVue data for each image (B). The vessel density is measured before and after correction for magnification error in the foveal (yellow, i) and parafoveal (pale blue, ii) regions, as defined in the text (C). Cropped foveal and parafoveal regions are shown before and after correction (D). FAZA is measured before correction and the area values are scaled by the magnification factor to obtain FAZA for the corrected images (E).
Figure 3
 
Bland-Altman plots illustrating the agreement between the custom-designed and AngioVue software for the measurement of blood vessel density for all nine regions (I–IX) as defined in Figure 2.
Figure 3
 
Bland-Altman plots illustrating the agreement between the custom-designed and AngioVue software for the measurement of blood vessel density for all nine regions (I–IX) as defined in Figure 2.
Figure 4
 
Bland-Altman plots illustrating the agreement before and after correction for the measurement of SRVD in foveal (A) and parafoveal (B) regions and FAZA (C). tstat and P value are, respectively, −2.0 and 0.05 for SRVD in the foveal region, −1.1 and 0.28 for SRVD in the parafoveal region, and 1.8 and 0.08 for FAZA. The Littman and the modified Bennett formula were applied to correct image size for a magnification error.
Figure 4
 
Bland-Altman plots illustrating the agreement before and after correction for the measurement of SRVD in foveal (A) and parafoveal (B) regions and FAZA (C). tstat and P value are, respectively, −2.0 and 0.05 for SRVD in the foveal region, −1.1 and 0.28 for SRVD in the parafoveal region, and 1.8 and 0.08 for FAZA. The Littman and the modified Bennett formula were applied to correct image size for a magnification error.
Figure 5
 
Plots of the relative change in SRVD in the foveal (A), and parafoveal (B), regions and FAZA (C), before and after correction for magnification error. The arrow represents the default axial length of 23.82 mm for the RTVue instrument. The dashed horizontal lines indicate the 5% error margins in vessel density. The fine dashed lines are nonlinear least-squares fits to the data. Plots of the relative change in SRVD in the foveal (D), and parafoveal (E), regions and FAZA (F), before and after correction for magnification error against spherical equivalent.
Figure 5
 
Plots of the relative change in SRVD in the foveal (A), and parafoveal (B), regions and FAZA (C), before and after correction for magnification error. The arrow represents the default axial length of 23.82 mm for the RTVue instrument. The dashed horizontal lines indicate the 5% error margins in vessel density. The fine dashed lines are nonlinear least-squares fits to the data. Plots of the relative change in SRVD in the foveal (D), and parafoveal (E), regions and FAZA (F), before and after correction for magnification error against spherical equivalent.
Table
 
Bland-Altman Analysis and Paired t-test Showing the Agreement Between Custom-Written Software and the AngioVue Analytics Software for the Measurement of Retinal Vessel Density
Table
 
Bland-Altman Analysis and Paired t-test Showing the Agreement Between Custom-Written Software and the AngioVue Analytics Software for the Measurement of Retinal Vessel Density
×
×