To overcome the drawback of low efficacy in accelerated CXL as predicted by the S-formula, a RF CCM is proposed as follows. In the conventional Dresden protocol, extra RF drops were instilled during the UV exposure (with a frequency Fdrop = 5–10), which reduced the effective dose from 5.4 J/cm
2 to approximately 4.0 J/cm
2, based on our calculations.
2,3 For an optimal protocol (for fast and efficient CXL in the anterior stroma), I propose Fdrop = 1 to 4 to compensate the fast RF depletion in the anterior stroma, especially in high intensity (>18 mW/cm
2). In contrast to the conventional Dresden protocol, which keeps the RF in a saturated condition during the UV exposure, CCM proposes to turn off the UV light after each of the extra RF drops applied to the stroma and waiting for a period approximately 1.0 to 2.0 minutes to allow enough RF diffusion (with a diffusion depth D > 150 μm) before it is turned on again. In the above proposed CCM, my theory predicts comparable efficacy (for the same dose) for intensity of 1.5 to 45 mW/cm
2, based on a combined efficacy formula defined as c-Ceff = 1− exp [−(S1 + S2 + … Sj)], with j = Fdrop, and Fdrop is given by the integer portion of square root of [I
0/3], that is, Fdrop = (1, 1, 2, 3, 3, 4), for I
0 = (1.5, 3, 9, 18, 30, 45) mW/cm
2 and exposure time t = (30, 30, 10, 5, 3, 2) minutes. The above CCM proposes that higher intensity requires larger Fdrop (or more RF resupply) to compensate the faster bleaching effect in the anterior stroma (100–250 μm), which is re-treated by Fdrop times, and the waiting period (with UV off) after each RF drops secures enough diffusing depth (D > 150 μm). Numerical simulation of c-Ceff (to be shown elsewhere) under the new CCM protocol shows a stronger correlation with the measured data of Wernli et al.
3 than the simple protocol (with Fdrop = 0) or Dresden protocol (with Fdrop > 5). The
Figure shows an example of c-Ceff for various UV intensities of 3 to 45 mW/cm
2, with Fdrop = (1, 2, 3, 3, 4), where all cases have efficacy above the threshold value 80% within the anterior stroma crosslinked depth range of 0 to 200 μm. In comparison, curves in the range 400 to 500 μm are associated with the situation of Fdrop = 0, which shows low efficacy < 80% for high intensity with I
0 > 18 mW/ cm
2, where diffusion depth D = 500, 200, 150, and 150 μm were used for Fdrop = 1, 2, 3, and 4. The cutoff (or maximum) intensity with c-Ceff < 80% (within 0–200 μm stroma) predicted by the S-formula is approximately 45 to 55 mW/cm
2, consistent with the clinical data of Wernli et al.
4 The above formulas also demonstrate that not only crosslink depth (z* > 150 μm) but also crosslink strength (S* > 1.6, or c-Ceff > 0.8) is required in order to achieve high crosslinked stroma volume which is proportional to (z*S*), as also suggested by Bekesi et al.
1