Comparison of electroretinography measurements between tabletop and handheld stimulators in healthy subjects
Author Affiliations & Notes
  • BUMGI KIM
    SANGGYE PAIK HOSPITAL, Seoul, Nowon, Korea (the Democratic People's Republic of)
  • Jae Suk Kim
    SANGGYE PAIK HOSPITAL, Seoul, Nowon, Korea (the Democratic People's Republic of)
  • Won Hyuk Oh
    SANGGYE PAIK HOSPITAL, Seoul, Nowon, Korea (the Democratic People's Republic of)
  • Jae Yong Park
    SANGGYE PAIK HOSPITAL, Seoul, Nowon, Korea (the Democratic People's Republic of)
  • Miryoung Song
    SANGGYE PAIK HOSPITAL, Seoul, Nowon, Korea (the Democratic People's Republic of)
  • Je Hyung Hwang
    SANGGYE PAIK HOSPITAL, Seoul, Nowon, Korea (the Democratic People's Republic of)
  • Footnotes
    Commercial Relationships   BUMGI KIM, None; Jae Suk Kim, None; Won Hyuk Oh, None; Jae Yong Park, None; Miryoung Song, None; Je Hyung Hwang, None
  • Footnotes
    Support  no support
Investigative Ophthalmology & Visual Science July 2019, Vol.60, 2495. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      BUMGI KIM, Jae Suk Kim, Won Hyuk Oh, Jae Yong Park, Miryoung Song, Je Hyung Hwang; Comparison of electroretinography measurements between tabletop and handheld stimulators in healthy subjects. Invest. Ophthalmol. Vis. Sci. 2019;60(9):2495.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose : To compare full field electroretinography (ffERG) parameters obtained from handheld and tabletop ERG devices in normal subjects

Methods : Twenty volunteers were subjected to ffERG. The responses were recorded with the two stimulators, and the results obtained from the right eyes of the 20 participants were compared

Results : The b-wave in the rod response, a- and b- waves in the maximal combined response, a-wave in the cone response, and the 30-Hz flicker response showed significantly greater amplitudes when recorded with a tabletop stimulator than with the handheld stimulator. The implicit time of response (ITR) in the 30-Hz flicker response was longer when recorded with the handheld stimulator than with the tabletop stimulator. According to the ICC with regard to amplitude, the two stimulators showed moderate-to-high reliability in the measurement of the b-wave in the rod response and a- and b- waves in the maximal combined response. With regard to the ITR, the reliability of measurement of the b-wave in the rod response and a-wave in the maximal combined response was moderate to high.

Conclusions : Relative to a tabletop stimulator, a handheld stimulator resulted in lower ERG amplitude measurements, although there were no significant differences in variability between the two stimulators