The repeatability of visual changes measured with tests of visual acuity and contrast sensitivity
Author Affiliations & Notes
  • Manonmani Murugappan
    College of Osteopathic Medicine, Nova Southeastern University, Fort Lauderdale, Florida, United States
  • Andrea M Janoff
    College of Optometry , Nova Southeastern University, Tampa, Florida, United States
  • Luis Andres Lesmes
    Adaptive Sensory Technology, California, United States
  • Emma Flor
    College of Optometry , Nova Southeastern University, Tampa, Florida, United States
  • Maryn JaNet Barnes
    College of Optometry , Nova Southeastern University, Tampa, Florida, United States
  • Ava K Bittner
    College of Optometry , Nova Southeastern University, Tampa, Florida, United States
    Ophthalmology, University of California, Los Angeles, Los Angeles, California, United States
  • Footnotes
    Commercial Relationships   Manonmani Murugappan, Adaptive Sensory Technology (F), Adaptive Sensory Technology (R); Andrea Janoff, Adaptive Sensory Technology (F); Luis Lesmes, Adaptive Sensory Technology (I), Adaptive Sensory Technology (E), Adaptive Sensory Technology (P); Emma Flor, Adaptive Sensory Technology (F), Adaptive Sensory Technology (R); Maryn Barnes, Adaptive Sensory Technology (F); Ava Bittner, Adaptive Sensory Technology (F)
  • Footnotes
    Support  an investigator initiated award from Adaptive Sensory Technology
Investigative Ophthalmology & Visual Science July 2019, Vol.60, 5904. doi:
  • Views
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Manonmani Murugappan, Andrea M Janoff, Luis Andres Lesmes, Emma Flor, Maryn JaNet Barnes, Ava K Bittner; The repeatability of visual changes measured with tests of visual acuity and contrast sensitivity. Invest. Ophthalmol. Vis. Sci. 2019;60(9):5904.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose : The recent applications of active learning algorithms to testing visual acuity (VA) and contrast sensitivity function (CSF) – quantitative VA (qVA) and quantitative CSF (qCSF) – reflect an attempt to develop tools with higher stimulus resolution, better test precision, and improved detection of vision changes related to intervention or ocular disease progression. We compared the test-retest repeatability of the qVA and qCSF tests to the standard ETDRS VA and Pelli-Robson CS charts.

Methods : At two visits about one week apart, the same test battery was repeated, involving two measures of distance VA (ETDRS trans-illuminated chart and qVA) and two tests of distance CS (Pelli-Robson and qCSF) in a total of 50 eyes in 25 normally-sighted, pre-presbyopic adults without ocular disease. Subjects performed all tests with daily disposable Acuvue or Alcon contact lenses with distance-only and multifocal correction to introduce some visual degradation. Between-visit repeatability was determined with 95% coefficients of repeatability (CR).

Results : For the two visual conditions, 95% CRs for distance-only and multifocal correction were 0.18 and 0.18 log units for ETDRS VA, 0.12 and 0.16 log units for qVA, 0.20 and 0.21 logCS for Pelli-Robson, and 0.23-0.25 and 0.24-0.29 logCS for qCSF area under the log curve (AUC) or at 1.5, 3 and 6 cpd, respectively. The magnitude of vision loss with the multifocal lens was not significantly different between the two visits for each of the four tests (all p>0.05). Cohen’s d effect size reflects both the magnitude of visual change and test repeatability, which was 1.16 and 1.61 for ETDRS VA and qVA, respectively, 0.77 for Pelli-Robson CS, 1.20 for qCSF AUC, and 0.31, 0.75, and 1.11 for qCSF at 1.5, 3 and 6 cpd, respectively.

Conclusions : As part of central visual function test validation and selection, it is important to determine and consider both the test repeatability and magnitude of visual changes of interest that are documented with each test. Improving Cohen’s d effect size for detected visual changes has the potential to reduce sample sizes in clinical trials.